Open Access 01.12.2023 | Research

Acceleration of knee magnetic resonance imaging using a combination of compressed sensing and commercially available deep learning reconstruction: a preliminary study

verfasst von: Hiroyuki Akai, Koichiro Yasaka, Haruto Sugawara, Taku Tajima, Masaru Kamitani, Toshihiro Furuta, Masaaki Akahane, Naoki Yoshioka, Kuni Ohtomo, Osamu Abe, Shigeru Kiryu

Erschienen in: BMC Medical Imaging | Ausgabe 1/2023

Abstract

Purpose

To evaluate whether deep learning reconstruction (DLR) accelerates the acquisition of 1.5-T magnetic resonance imaging (MRI) knee data without image deterioration.

Materials and methods

Twenty-one healthy volunteers underwent MRI of the right knee on a 1.5-T MRI scanner. Proton-density-weighted images with one or four numbers of signal averages (NSAs) were obtained via compressed sensing, and DLR was applied to the images with 1 NSA to obtain 1NSA-DLR images. The 1NSA-DLR and 4NSA images were compared objectively (by deriving the signal-to-noise ratios of the lateral and the medial menisci and the contrast-to-noise ratios of the lateral and the medial menisci and articular cartilages) and subjectively (in terms of the visibility of the anterior cruciate ligament, the medial collateral ligament, the medial and lateral menisci, and bone) and in terms of image noise, artifacts, and overall diagnostic acceptability. The paired t-test and Wilcoxon signed-rank test were used for statistical analyses.

Results

The 1NSA-DLR images were obtained within 100 s. The signal-to-noise ratios (lateral: 3.27 ± 0.30 vs. 1.90 ± 0.13, medial: 2.71 ± 0.24 vs. 1.80 ± 0.15, both p < 0.001) and contrast-to-noise ratios (lateral: 2.61 ± 0.51 vs. 2.18 ± 0.58, medial 2.19 ± 0.32 vs. 1.97 ± 0.36, both p < 0.001) were significantly higher for 1NSA-DLR than 4NSA images. Subjectively, all anatomical structures (except bone) were significantly clearer on the 1NSA-DLR than on the 4NSA images. Also, in the former images, the noise was lower, and the overall diagnostic acceptability was higher.

Conclusion

Compared with the 4NSA images, the 1NSA-DLR images exhibited less noise, higher overall image quality, and allowed more precise visualization of the menisci and ligaments.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
ACL
Anterior cruciate ligament
CNR
Contrast-to-noise ratio
DL
Deep learning
DLR
Deep learning reconstruction
LM
Lateral meniscus
MCL
Medial collateral ligament
MM
Medial meniscus
MRI
Magnetic resonance imaging
NSA
Number of signal averages
ROI
Region of interest
SD
Standard deviation
SNR
Signal-to-noise ratio

Introduction

Magnetic resonance imaging (MRI) plays a central role in knee evaluation and is commonly used in various clinical settings [13]. Acute or chronic knee pain is the principal indication for knee MRI [4, 5]. Patients in pain find it difficult to remain motionless during MRI; thus, the acceleration of the MRI scan is one of the key elements in successful knee MRI evaluation.
For the acceleration of knee MRI, various image acquisition techniques have been applied. Parallel imaging was first employed to this end [6, 7], but the disadvantages included a reduced SNR, noise enhancement, aliasing, and reconstruction artifacts [8]. Several advanced parallel imaging sequences as GeneRalized Autocalibrating Partially Parallel Acquisition (GRAPPA) and Controlled Aliasing in Parallel Imaging Results in Higher Acceleration (CAIPIRINHA) have also been applied to knee MRI overcoming some of these disadvantages [9]. Compressed sensing allows rapid MRI because it reconstructs highly undersampled k-space data and also has been successfully applied to knee MRI [10]. However, the disadvantages include image blurring and long post-processing times [11]. Finally, multislice (or multiband) imaging (the simultaneous acquisition of multiple slices) was introduced [12]. These techniques can be used in combination; recently, Del Grande et al. achieved a fourfold-accelerated 5-minute knee MRI protocol including five sequences by the combination of multislice imaging and parallel imaging [13].
Over the last decade, deep learning (DL) has found many applications (for example, in image processing, speech recognition, and natural language processing [14]). Radiologists have employed DL for image segmentation [15, 16] and lesional evaluation [17, 18]. Recently, deep learning has been applied to image reconstruction, and DL-based reconstruction (DLR) of images effectively denoised the images without compromising contrast [1921].
We hypothesized that we could accelerate the MRI scan by using DLR without deteriorating the image quality. For routine 1.5T knee MRI, usually number of signal averages (NSA) of 2 or 3 is applied [22, 23]. Therefore, in the present study, we accelerated MRI four-fold, applied DLR, and compared objective and subjective image features with those of conventional images.

Materials and methods

The research ethics committee of our institution (International University of Health and Welfare Chiba District Ethics Review Committee) approved the study (approval no. 20-Nr-059), and all subjects provided written informed consent. Twenty-one healthy volunteers (17 men and 4 women; mean age ± standard deviation [SD] 44.7 ± 10.9 years) were enrolled. The inclusion criteria were (1) age over 20, (2) no medical history of the knee (e.g., surgery, intra-articular injection), and (3) no apparent knee pain.

MRI examination

All volunteers underwent 1.5-T MRI (Vantage Orian, Canon Medical Systems Corporation) using a 16-channel knee coil; proton-density-weighted images of the right knee were obtained in the coronal plane using the following parameters (repetition time 2,000 ms; echo time 33 ms; NSA 1 or 4; echo train length 8; flip angle 90°; pixel bandwidth 217 Hz; field of view 160 mm; acquisition matrix 512 × 512; slice thickness 1.5 mm; spacing between slices 2 mm; and slice number 18). Compressed sensing with parallel imaging (Compressed SPEEDER: Canon Medical Systems Corporation) was employed in this study. This technique first processes the image domain parallel imaging with random undersampling (k space undersampling rate = 64.1%) in phase encoding, followed by compressed sensing implemented with wavelet transform to remove the artifacts. The scan times were 100 s for a NSA of 1 (1NSA) and 390 s for a NSA of 4 (4NSA). The 1NSA images were subjected to DLR, yielding 1NSA-DLR images. DLR was implemented using the Advanced Intelligent Clear IQ Engine (Canon Medical Systems Corporation) [19]. Briefly, this convolutional neural network-based technique uses 7 × 7 discrete cosine transform to divide the image data into a zero-frequency component and other high-frequency components at the feature extraction layer. The former component follows a separate collateral path to maintain the image contrast, and the latter components are processed to 22 subsequent feature conversion layers for denoising. This DLR technique has also been shown to improve the image quality of cervical spine MRI [24] and diffusion-weighted whole-body imaging with background body signal suppression [25].

Objective evaluation of image quality

For the objective evaluation of image noise, SNRs of the medial meniscus (MM) and lateral meniscus (LM) were calculated. As we used a compressed sensing technique, the SNR could not be derived using the background SD. Thus, we calculated the SNR of the meniscus by employing the SD of the meniscus per se [26]. We chose meniscus as target since (1) meniscus is the one of the main structures in the knee MRI evaluation, and (2) a structure with low signal intensity will clearly show the effect of denoising. First, we manually defined a regions of interest (ROIs) that included the entire meniscus in the slices of the middle body showing the smallest meniscus area (Fig. 1A). The SNR was calculated by dividing the mean signal intensity of the ROI by the SD of the ROI.
For objective evaluation of contrast, the contrast-to-noise ratios (CNRs) of the LM and the articular cartilage of the lateral femoral condyle and the CNRs of MM and the articular cartilage of the medial femoral condyle were calculated. We defined ROIs that covered the entire areas of cartilage in the slices with the most prominent image slice in the posterior part (Fig. 1B). This tissue-specific CNR was [26]:
$${\text{CNR}}=({\text{Mean}}_{{\rm cartilage}} - {\text{Mean}}_{{\rm meniscus}})/\sqrt({\text{SD}}_{{\rm cartilage}}^{2} + {\text{SD}}_{{\rm meniscus}}^{2}).$$
We used ImageJ software (National Institutes of Health) for quantitative image analysis. All ROIs were placed by a board-certified radiologist with 18 years of experience.

Subjective evaluation of image quality

Three board-certified radiologists (with 17, 12, and 8 years of experience) assessed image quality. The images were presented in a random order to minimize recall bias; the observers were blinded to the acquisition method. All acquired image slices were assigned for the assessment. The visibilities of the anterior cruciate ligament (ACL), medial collateral ligament (MCL), MM, LM, and bone, as well as image noise, artifacts, and overall diagnostic acceptability, were scored using a five-point Likert scale (Table 1).
Table 1
The scale used for subjective image quality analysis
Grade
Visibility
Noise
Artifact
Overall
1
Not visible
Undiagnostic
Undiagnostic
Unacceptable
2
Mostly not visible or blurred
Strong noise but still diagnostic
Strong artifacts but still diagnostic
Average
3
Mostly visible but partially blurred
Acceptable noise
Acceptable artifacts
Fair
4
Subtle heterogeneity or blurring
Minimal noise
Minimal artifacts
Very good
5
Homogeneous internal intensity with sharp edges
No noise
No artifact
Excellent

Statistical analysis

The 1NSA-DLR and 4NSA images were compared. All results are expressed as means ± SDs. The objective noises and contrasts were compared using the paired t-test because the Shapiro–Wilk test confirmed that the data were normally distributed. The subjective image qualities were compared employing the Wilcoxon signed-rank test. A p value < 0.05 was considered to indicate statistical significance. We evaluated interobserver agreement by calculating the Cohen weighted kappa values (quadratic weights); values of 0–0.20 indicate poor agreement, 0.21–0.40 fair, 0.41–0.60 moderate, 0.61–0.80 good, and 0.81–1.00 excellent. All analyses were performed using R software (version 4.0.5; R Foundation for Statistical Computing).

Results

Representative 1NSA-DLR and 4NSA images are shown in Fig. 2.

Objective analysis of image quality

The SNR of the LM was significantly higher in the 1NSA-DLR than in 4NSA images (3.27 ± 0.30 vs. 1.90 ± 0.13, p < 0.001). Similarly, the SNR of the MM was significantly higher in the 1NSA-DLR than in 4NSA images (2.71 ± 0.24 vs. 1.80 ± 0.15, p < 0.001). The CNR between the articular cartilage of the femoral condyle and the meniscus was significantly higher in the 1NSA-DLR than in 4NSA images in both lateral (2.61 ± 0.51 vs. 2.18 ± 0.58, p < 0.001) and medial sides (2.19 ± 0.32 vs. 1.97 ± 0.36, p < 0.001) (Table 2).
Table 2
The results of objective image quality analysis
 
Meniscus
Cartilage
Diff(Cart – Meni)
CNR
p value
Mean
SD
Mean
SD
Mean
Lateral side
 1NSA-DLR
869.7
267.1
3029.3
797.0
2159.6
2.61 ± 0.51
< 0.001
 4NSA
721.0
382.5
2857.7
921.7
2136.7
2.18 ± 0.58
 
Medial side
 1NSA-DLR
1030.8
380.1
3440.6
1032.7
2409.7
2.19 ± 0.32
< 0.001
 4NSA
864.6
484.7
3248.8
1124.8
2384.2
1.97 ± 0.36
 
For meniscus and cartilage, mean and SD of the signal intensity of each structure is described. Diff(Cart – Meni) represents the difference in signal intensity of cartilage and meniscus. SDs are expressed as their mean and CNRs are expressed as the mean ± standard deviation. p values are shown for the comparison of CNR and the p values that indicate significant differences are shown in bold

Subjective analysis of image quality

The results of the subjective analysis are listed in Table 3. All the observers reported that all anatomical structures except bone were better visualized on 1NSA-DLR than 4NSA images (all p < 0.05). Bone visualization was in fact better, and the extent of artifacts was lower in 1NSA-DLR images (all observers), but the differences were not significant (p > 0.05). Noise was lower (p ≤ 0.001) and the overall diagnostic acceptability was higher on the 1NSA-DLR than on 4NSA images (p < 0.01) (all observers). Cohen’s kappa analysis revealed that the extent of interobserver agreement was moderate-to-excellent in terms of structural visibilities (0.45–0.54 for ACL, 0.47–0.59 for MCL, 0.65–0.72 for MM, 0.59–0.68 for LM, and 0.66–0.85 for bone) and fair-to-excellent in terms of noise (0.32–0.61), artifacts (0.84–0.85), and overall image quality (0.60–0.77).
Table 3
The results of subjective image quality analysis
 
ACL
MCL
MM
LM
Bone
Noise
Artifact
Overall
Reader 1
 1NSA-DLR
4.14 ± 0.35
4.90 ± 0.29
4.95 ± 0.21
5.00 ± 0.00
4.00 ± 0.00
3.71 ± 0.45
4.81 ± 0.39
3.76 ± 0.43
 4NSA
3.76 ± 0.43
4.23 ± 0.68
3.95 ± 0.79
4.29 ± 0.88
3.86 ± 0.35
3.05 ± 0.21
4.42 ± 1.14
3.00 ± 0.69
 p value
0.03
0.001
< 0.001
0.005
0.15
< 0.001
0.14
0.003
Reader 2
 1NSA-DLR
4.19 ± 0.39
4.52 ± 0.50
4.48 ± 0.50
4.48 ± 0.50
4.00 ± 0.00
3.95 ± 0.21
5.00 ± 0.00
3.95 ± 0.37
 4NSA
3.71 ± 0.63
3.81 ± 0.59
3.86 ± 0.77
3.95 ± 0.72
3.81 ± 0.39
3.05 ± 0.37
4.52 ± 1.17
2.90 ± 0.61
 p value
0.02
0.001
0.008
0.02
0.07
< 0.001
0.17
< 0.001
Reader 3
 1NSA-DLR
4.19 ± 0.39
4.62 ± 0.49
4.57 ± 0.49
4.52 ± 0.50
4.05 ± 0.21
3.95 ± 0.21
4.86 ± 0.35
3.95 ± 0.21
 4NSA
3.90 ± 0.53
4.14 ± 0.77
3.86 ± 0.94
4.00 ± 0.82
3.90 ± 0.43
3.43 ± 0.58
4.43 ± 1.00
3.29 ± 0.76
 p value
0.04
0.03
0.01
0.02
0.23
0.001
0.10
< 0.001
All variables are expressed as the mean ± standard deviation. p values that indicate significant differences are shown in bold

Discussion

We found that knee MRI could be accelerated four-fold using DLR. Objectively, the 1NSA-DLR images were more uniform and showed higher CNR than the 4NSA images; subjectively, the 1NSA-DLR images revealed all studied structures more clearly than the 4NSA images, regardless of structure size. Although statistical significance was not attained, bone visibility was better and artifacts fewer on 1NSA-DLR images; noise was also significantly lower, imparting better overall image quality.
DL is relatively new, and its clinical applications are few in number. One study employed DLR to comprehensively examine the knee in 5 min without compromising image quality or diagnostic accuracy. Recht et al. performed 406 consecutive knee examinations using a 3-T MRI scanner and found that standard images and deep-learning-based accelerated images were largely equivalent [27]. This was also our experience; DLR accelerated 1.5-T knee MRI four-fold.
One of the advantages of the DLR technique is that this is a post-processing technique, so DLR can be applied to the image obtained by accelerating image acquisition techniques. In the present study, we used a compressed sensing technique in image acquisition. And by combination with DLR, clear coronal proton-density-weighted knee MR images were able to obtain within only 100s. This short acquisition time can reduce motion-related artifacts, and indeed the level of artifacts was scored better for 1NSA-DLR images in the present study. We believe that a combination of accelerating image acquisition techniques and DLR would also be beneficial for patients with knee pain.
Our work had several limitations. First, the number of volunteers was small. Second, all volunteers were recruited from one institution. Selection bias may have been in play. Third, we assessed only proton-density-weighted images because these are optimal for detecting meniscal lesions [28]. DLR is useful for processing T1- and T2-weighted images, fluid-attenuated inversion recovery images, and magnetic resonance cholangiopancreatographic images [19, 29]. Thus, DLR can effectively process other knee MRI sequences as well. Finally, we did not evaluate the disease detectability of DLR. We assume that DLR would facilitate the detection of abnormalities by radiologists, given the high image quality, but further investigations are needed.

Conclusion

In conclusion, DLR significantly improved knee MR image quality. 1NSA-DLR images exhibited less noise, better visualization of menisci and ligaments, and higher overall image quality compared with 4NSA images.

Acknowledgements

Not applicalble.

Declarations

The research ethics committee of our institution (International University of Health and Welfare Chiba District Ethics Review Committee) approved the study (approval no. 20-Nr-059), and all subjects provided written informed consent to participate in the study. All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and national research committee and with the 1964 Helsinki declaration.
Not applicalble.

Competing interests

S.K. disclosed grants from Canon Medical Systems. All other authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Petron DJ, Greis PE, Aoki SK, et al. Use of knee magnetic resonance imaging by primary care physicians in patients aged 40 years and older. Sports Health. 2010;2:385–90.CrossRef Petron DJ, Greis PE, Aoki SK, et al. Use of knee magnetic resonance imaging by primary care physicians in patients aged 40 years and older. Sports Health. 2010;2:385–90.CrossRef
2.
Zurück zum Zitat Pai DR, Strouse PJ. MRI of the pediatric knee. Am J Roentgenol. 2011;196:1019–27.CrossRef Pai DR, Strouse PJ. MRI of the pediatric knee. Am J Roentgenol. 2011;196:1019–27.CrossRef
3.
Zurück zum Zitat Walczak BE, McCulloch PC, Kang RW, Zelazny A, Tedeschi F, Cole BJ. Abnormal findings on knee magnetic resonance imaging in asymptomatic NBA players. J Knee Surg. 2008;21:27–33.CrossRef Walczak BE, McCulloch PC, Kang RW, Zelazny A, Tedeschi F, Cole BJ. Abnormal findings on knee magnetic resonance imaging in asymptomatic NBA players. J Knee Surg. 2008;21:27–33.CrossRef
4.
Zurück zum Zitat Chien A, Weaver JS, Kinne E, Omar I. Magnetic resonance imaging of the knee. Pol J Radiol. 2020;85:e509–31.CrossRef Chien A, Weaver JS, Kinne E, Omar I. Magnetic resonance imaging of the knee. Pol J Radiol. 2020;85:e509–31.CrossRef
5.
Zurück zum Zitat Yusuf E, Kortekaas MC, Watt I, Huizinga TW, Kloppenburg M. Do knee abnormalities visualised on MRI explain knee pain in knee osteoarthritis? A systematic review. Ann Rheum Dis. 2011;70:60–7.CrossRef Yusuf E, Kortekaas MC, Watt I, Huizinga TW, Kloppenburg M. Do knee abnormalities visualised on MRI explain knee pain in knee osteoarthritis? A systematic review. Ann Rheum Dis. 2011;70:60–7.CrossRef
6.
Zurück zum Zitat Kwok WE, Zhong J, You Z, Seo G, Totterman SM. A four-element phased array coil for high resolution and parallel MR imaging of the knee. Magn Reson Imaging. 2003;21:961–7.CrossRef Kwok WE, Zhong J, You Z, Seo G, Totterman SM. A four-element phased array coil for high resolution and parallel MR imaging of the knee. Magn Reson Imaging. 2003;21:961–7.CrossRef
7.
Zurück zum Zitat Zuo J, Li X, Banerjee S, Han E, Majumdar S. Parallel imaging of knee cartilage at 3 Tesla. J Magn Reson Imaging. 2007;26:1001–9.CrossRef Zuo J, Li X, Banerjee S, Han E, Majumdar S. Parallel imaging of knee cartilage at 3 Tesla. J Magn Reson Imaging. 2007;26:1001–9.CrossRef
8.
Zurück zum Zitat Deshmane A, Gulani V, Griswold MA, Seiberlich N. Parallel MR imaging. J Magn Reson Imaging. 2012;36:55–72.CrossRef Deshmane A, Gulani V, Griswold MA, Seiberlich N. Parallel MR imaging. J Magn Reson Imaging. 2012;36:55–72.CrossRef
9.
Zurück zum Zitat Fritz J, Fritz B, Thawait GG, Meyer H, Gilson WD, Raithel E. Three-dimensional CAIPIRINHA SPACE TSE for 5-minute high-resolution MRI of the knee. Invest Radiol. 2016;51:609–17.CrossRef Fritz J, Fritz B, Thawait GG, Meyer H, Gilson WD, Raithel E. Three-dimensional CAIPIRINHA SPACE TSE for 5-minute high-resolution MRI of the knee. Invest Radiol. 2016;51:609–17.CrossRef
10.
Zurück zum Zitat Kijowski R, Rosas H, Samsonov A, King K, Peters R, Liu F. Knee imaging: Rapid three-dimensional fast spin-echo using compressed sensing. J Magn Reson Imaging. 2017;45:1712–22.CrossRef Kijowski R, Rosas H, Samsonov A, King K, Peters R, Liu F. Knee imaging: Rapid three-dimensional fast spin-echo using compressed sensing. J Magn Reson Imaging. 2017;45:1712–22.CrossRef
11.
Zurück zum Zitat Lustig M, Donoho D, Pauly JM, Sparse MRI. The application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58:1182–95.CrossRef Lustig M, Donoho D, Pauly JM, Sparse MRI. The application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58:1182–95.CrossRef
12.
Zurück zum Zitat Barth M, Breuer F, Koopmans PJ, Norris DG, Poser BA. Simultaneous multislice (SMS) imaging techniques. Magn Reson Med. 2016;75:63–81.CrossRef Barth M, Breuer F, Koopmans PJ, Norris DG, Poser BA. Simultaneous multislice (SMS) imaging techniques. Magn Reson Med. 2016;75:63–81.CrossRef
13.
Zurück zum Zitat Del Grande F, Rashidi A, Luna R, et al. Five-minute five-sequence knee MRI using combined simultaneous multislice and parallel imaging acceleration: comparison with 10-Minute parallel imaging knee MRI. Radiology. 2021;299:635–46.CrossRef Del Grande F, Rashidi A, Luna R, et al. Five-minute five-sequence knee MRI using combined simultaneous multislice and parallel imaging acceleration: comparison with 10-Minute parallel imaging knee MRI. Radiology. 2021;299:635–46.CrossRef
14.
Zurück zum Zitat Yasaka K, Akai H, Kunimatsu A, Kiryu S, Abe O. Deep learning with convolutional neural network in radiology. Jpn J Radiol. 2018;36:257–72.CrossRef Yasaka K, Akai H, Kunimatsu A, Kiryu S, Abe O. Deep learning with convolutional neural network in radiology. Jpn J Radiol. 2018;36:257–72.CrossRef
15.
Zurück zum Zitat Weston AD, Korfiatis P, Kline TL, et al. Automated abdominal segmentation of CT scans for body composition analysis using deep learning. Radiology. 2019;290:669–79.CrossRef Weston AD, Korfiatis P, Kline TL, et al. Automated abdominal segmentation of CT scans for body composition analysis using deep learning. Radiology. 2019;290:669–79.CrossRef
16.
Zurück zum Zitat Trebeschi S, van Griethuysen JJM, Lambregts DMJ, et al. Deep learning for fully-automated localization and segmentation of rectal Cancer on multiparametric MR. Sci Rep. 2017;7:5301.CrossRef Trebeschi S, van Griethuysen JJM, Lambregts DMJ, et al. Deep learning for fully-automated localization and segmentation of rectal Cancer on multiparametric MR. Sci Rep. 2017;7:5301.CrossRef
17.
Zurück zum Zitat Yasaka K, Akai H, Abe O, Kiryu S. Deep learning with convolutional neural network for differentiation of liver masses at dynamic contrast-enhanced CT: a preliminary study. Radiology. 2018;286:887–96.CrossRef Yasaka K, Akai H, Abe O, Kiryu S. Deep learning with convolutional neural network for differentiation of liver masses at dynamic contrast-enhanced CT: a preliminary study. Radiology. 2018;286:887–96.CrossRef
18.
Zurück zum Zitat Kiryu S, Yasaka K, Akai H, et al. Deep learning to differentiate parkinsonian disorders separately using single midsagittal MR imaging: a proof of concept study. Eur Radiol. 2019;29:6891–9.CrossRef Kiryu S, Yasaka K, Akai H, et al. Deep learning to differentiate parkinsonian disorders separately using single midsagittal MR imaging: a proof of concept study. Eur Radiol. 2019;29:6891–9.CrossRef
19.
Zurück zum Zitat Kidoh M, Shinoda K, Kitajima M, et al. Deep learning based noise reduction for brain MR Imaging: tests on phantoms and healthy volunteers. Magn Reson Med Sci. 2020;19:195–206.CrossRef Kidoh M, Shinoda K, Kitajima M, et al. Deep learning based noise reduction for brain MR Imaging: tests on phantoms and healthy volunteers. Magn Reson Med Sci. 2020;19:195–206.CrossRef
20.
Zurück zum Zitat Herrmann J, Koerzdoerfer G, Nickel D, et al. Feasibility and implementation of a deep learning MR reconstruction for TSE sequences in musculoskeletal imaging. Diagnostics (Basel). 2021;11:1484.CrossRef Herrmann J, Koerzdoerfer G, Nickel D, et al. Feasibility and implementation of a deep learning MR reconstruction for TSE sequences in musculoskeletal imaging. Diagnostics (Basel). 2021;11:1484.CrossRef
21.
Zurück zum Zitat Naganawa S, Nakamichi R, Ichikawa K, et al. MR imaging of endolymphatic Hydrops: utility of iHYDROPS-Mi2 combined with deep learning reconstruction denoising. Magn Reson Med Sci. 2021;20:272–9.CrossRef Naganawa S, Nakamichi R, Ichikawa K, et al. MR imaging of endolymphatic Hydrops: utility of iHYDROPS-Mi2 combined with deep learning reconstruction denoising. Magn Reson Med Sci. 2021;20:272–9.CrossRef
22.
Zurück zum Zitat Wong S, Steinbach L, Zhao J, Stehling C, Ma CB, Link TM. Comparative study of imaging at 3.0 T versus 1.5 T of the knee. Skeletal Radiol. 2009;38:761–9.CrossRef Wong S, Steinbach L, Zhao J, Stehling C, Ma CB, Link TM. Comparative study of imaging at 3.0 T versus 1.5 T of the knee. Skeletal Radiol. 2009;38:761–9.CrossRef
23.
Zurück zum Zitat Helito CP, Helito PV, Costa HP, et al. MRI evaluation of the anterolateral ligament of the knee: assessment in routine 1.5-T scans. Skeletal Radiol. 2014;43:1421–7.CrossRef Helito CP, Helito PV, Costa HP, et al. MRI evaluation of the anterolateral ligament of the knee: assessment in routine 1.5-T scans. Skeletal Radiol. 2014;43:1421–7.CrossRef
24.
Zurück zum Zitat Yasaka K, Tanishima T, Ohtake Y, et al. Deep learning reconstruction for 1.5 T cervical spine MRI: effect on interobserver agreement in the evaluation of degenerative changes. Eur Radiol. 2022;32:6118–25.CrossRef Yasaka K, Tanishima T, Ohtake Y, et al. Deep learning reconstruction for 1.5 T cervical spine MRI: effect on interobserver agreement in the evaluation of degenerative changes. Eur Radiol. 2022;32:6118–25.CrossRef
25.
Zurück zum Zitat Tajima T, Akai H, Sugawara H, et al. Feasibility of accelerated whole-body diffusion-weighted imaging using a deep learning-based noise-reduction technique in patients with prostate cancer. Magn Reson Imaging. 2022;92:169–79.CrossRef Tajima T, Akai H, Sugawara H, et al. Feasibility of accelerated whole-body diffusion-weighted imaging using a deep learning-based noise-reduction technique in patients with prostate cancer. Magn Reson Imaging. 2022;92:169–79.CrossRef
26.
Zurück zum Zitat Altahawi FF, Blount KJ, Morley NP, Raithel E, Omar IM. Comparing an accelerated 3D fast spin-echo sequence (CS-SPACE) for knee 3-T magnetic resonance imaging with traditional 3D fast spin-echo (SPACE) and routine 2D sequences. Skeletal Radiol. 2017;46:7–15.CrossRef Altahawi FF, Blount KJ, Morley NP, Raithel E, Omar IM. Comparing an accelerated 3D fast spin-echo sequence (CS-SPACE) for knee 3-T magnetic resonance imaging with traditional 3D fast spin-echo (SPACE) and routine 2D sequences. Skeletal Radiol. 2017;46:7–15.CrossRef
27.
Zurück zum Zitat Recht MP, Zbontar J, Sodickson DK, et al. Using Deep Learning to accelerate knee MRI at 3 T: results of an interchangeability study. Am J Roentgenol. 2020;215:1421–9.CrossRef Recht MP, Zbontar J, Sodickson DK, et al. Using Deep Learning to accelerate knee MRI at 3 T: results of an interchangeability study. Am J Roentgenol. 2020;215:1421–9.CrossRef
28.
Zurück zum Zitat Lefevre N, Naouri JF, Herman S, Gerometta A, Klouche S, Bohu Y. A current review of the meniscus imaging: proposition of a useful tool for its radiologic analysis. Radiol Res Pract. 2016;2016:8329296. Lefevre N, Naouri JF, Herman S, Gerometta A, Klouche S, Bohu Y. A current review of the meniscus imaging: proposition of a useful tool for its radiologic analysis. Radiol Res Pract. 2016;2016:8329296.
29.
Zurück zum Zitat Tajima T, Akai H, Sugawara H, et al. Breath-hold 3D magnetic resonance cholangiopancreatography at 1.5T using a deep learning-based noise-reduction approach: comparison with the conventional respiratory-triggered technique. Eur J Radiol. 2021;144:109994.CrossRef Tajima T, Akai H, Sugawara H, et al. Breath-hold 3D magnetic resonance cholangiopancreatography at 1.5T using a deep learning-based noise-reduction approach: comparison with the conventional respiratory-triggered technique. Eur J Radiol. 2021;144:109994.CrossRef
Metadaten
Titel
Acceleration of knee magnetic resonance imaging using a combination of compressed sensing and commercially available deep learning reconstruction: a preliminary study
verfasst von
Hiroyuki Akai
Koichiro Yasaka
Haruto Sugawara
Taku Tajima
Masaru Kamitani
Toshihiro Furuta
Masaaki Akahane
Naoki Yoshioka
Kuni Ohtomo
Osamu Abe
Shigeru Kiryu
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
BMC Medical Imaging / Ausgabe 1/2023
Elektronische ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-023-00962-2

Weitere Artikel der Ausgabe 1/2023

BMC Medical Imaging 1/2023 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.