Skip to main content
Erschienen in: Fluids and Barriers of the CNS 1/2020

Open Access 01.12.2020 | Research

Compressed-sensing accelerated 4D flow MRI of cerebrospinal fluid dynamics

verfasst von: Elena Jaeger, Kristina Sonnabend, Frank Schaarschmidt, David Maintz, Kilian Weiss, Alexander C. Bunck

Erschienen in: Fluids and Barriers of the CNS | Ausgabe 1/2020

Abstract

Background

4D flow magnetic resonance imaging (MRI) of CSF can make an important contribution to the understanding of hydrodynamic changes in various neurological diseases but remains limited in clinical application due to long acquisition times. The aim of this study was to evaluate the accuracy of compressed SENSE accelerated MRI measurements of the spinal CSF flow.

Methods

In 20 healthy subjects 4D flow MRI of the CSF in the cervical spine was acquired using compressed sensitivity encoding [CSE, a combination of compressed sensing and parallel imaging (SENSE) provided by the manufacturer] with acceleration factors between 4 and 10. A conventional scan using SENSE was used as reference. Extracted parameters were peak velocity, absolute net flow, forward flow and backward flow. Bland–Altman analysis was performed to determine the scan-rescan reproducibility and the agreement between SENSE and compressed SENSE. Additionally, a time accumulated flow error was calculated. In one additional subject flow of the spinal canal at the level of the entire spinal cord was assessed.

Results

Averaged acquisition times were 10:21 min (SENSE), 9:31 min (CSE4), 6:25 min (CSE6), 4:53 min (CSE8) and 3:51 min (CSE10). Acquisition of the CSF flow surrounding the entire spinal cord took 14:40 min. Bland–Altman analysis showed good agreement for peak velocity, but slight overestimations for absolute net flow, forward flow and backward flow (< 1 ml/min) in CSE4–8. Results of the accumulated flow error were similar for CSE4 to CSE8.

Conclusion

A quantitative analysis of acceleration factors CSE4–10 showed that CSE with an acceleration factor up to 6 is feasible. This allows a scan time reduction of 40% and enables the acquisition and analysis of the CSF flow dynamics surrounding the entire spinal cord within a clinically acceptable scan time.
Hinweise
Elena Jaeger and Kristina Sonnabend contributed equally to the work.

Supplementary information

Supplementary information accompanies this paper at https://​doi.​org/​10.​1186/​s12987-020-00206-3.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
2D
Two-dimensional
4D
Four-dimensional
AF
Absolute net flow
BF
Backward flow
CI
Confidence interval
CPU
Central processing unit
CS
Compressed sensing
CSE
Compressed SENSE
CSF
Cerebrospinal fluid
ECG
Electrocardiogram
FF
Forward flow
FOV
Field of view
LL
Lower limit
LoA
Limits of agreement
MD
Mean difference
MRI
Magnetic resonance imaging
PCMRI
Phase-contrast magnetic resonance imaging
PV
Peak velocity
REML
Restricted maximum likelihood
ROI
Region-of-interest
SENSE
Sensitivity encoding
SD
Standard deviation
UL
Upper limit

Background

Changes in the cerebrospinal fluid (CSF) flow dynamics have been found to be associated with different neurological diseases such as hydrocephalus [1], Chiari malformation [2, 3], syringomyelia [4] and Alzheimer disease [5]. The quantification of the CSF flow is necessary for a better understanding of the physiology and pathophysiology of CSF dynamics and may assist in diagnosis and treatment guidance of CSF related diseases. At present, two-dimensional phase-contrast magnetic resonance imaging (2D flow MRI) is the most common technique to quantify the CSF flow non-invasively [6]. Since this method is two-dimensional it is not sufficient to fully represent the complex physiological or pathological flow dynamics of the CSF including multidirectional flow occurring in patients with Chiari malformation, inside of arachnoid cysts and syringes as well as during reflux from the third into the lateral ventricles [7, 8]. To represent complex flow patterns adequately a time-resolved three-dimensional MRI measurement (4D flow MRI) method is necessary [9, 10]. However, due to the large amount of data that needs to be recorded, the acquisition of a 4D flow MRI data set is accompanied with long scan times and therefore, rarely integrated into clinical routine scan protocols.
To accelerate 4D flow MRI acquisitions, different methods were proposed by exploiting spatial–temporal correlations or using compressed sensing. However, most work focused on cardio-vascular applications [1115] and no work exists on accelerated 4D flow MRI of the CSF.
The concept of compressed sensing is a promising way to accelerate MRI measurements. It was first proposed in 2006 by Donoho [16] and transferred to MRI by Lustig et al. [17]. Compressed sensing can reduce the required amount of data by random undersampling of the k-space, if an image can be assumed to be “sparse” in a suitable transform domain [18]. Further image acceleration can be accomplished by combining compressed sensing and parallel imaging [18, 19], herein referred to as compressed SENSE (CSE) [20, 21]. Previous studies have shown that CSE can accelerate scan times while obtaining an acceptable quality in aortic [13, 22, 23] and carotid [24] blood flow using acceleration factors between R = 3 and R = 7. In comparison to blood flow, velocities of the CSF are low and therefore more difficult to measure. Contrast of the tissue is lower and no contrast agent can be used making an exact acquisition more challenging. In addition, when measuring the blood flow in the chest or abdomen more coils can be used which improves quality.
The aim of this study was to investigate the applicability of the combination of compressed sensing and parallel imaging to the quantification of CSF flow using 4D flow MRI to improve its clinical applicability. To estimate the maximum feasible image acceleration using CSE, different CSE factors were compared to the standard parallel imaging acceleration using SENSE regarding image quality and different flow parameters, such as peak velocity, forward volume flow, backward volume flow and absolute net flow in healthy subjects. Finally, to further investigate the potential of the CSE acceleration the proposed method was applied to measure the CSF flow surrounding the entire spinal cord (C1-L1) in one healthy subject by the combination of 2 separate acquisitions.

Methods and material

Study population

20 healthy subjects (11 women, 9 men, mean age ± SD: 29.7 ± 13.04 years) with no history of neurological or spinal diseases were recruited. The study was approved by the institutional ethics committee and informed consent was obtained from all subjects prior to the study.

Image acquisition

All examinations were obtained using a clinical 3 T MRI system (Ingenia; Philips Healthcare, Best, The Netherlands). Data were acquired using a spoiled gradient echo sequence with a 16-channel head and neck coil array and the built‐in 8-channel posterior coil array. During the examination, the peripheral heart rate was recorded using a wireless pulse oximeter to allow for retrospective synchronization of the data. Beforehand, a survey scan and a sagittal 3D T2-weighted scan were performed for planning purposes. The imaging volume of the 4D flow acquisition was chosen to cover the cerebral aqueduct and the cervical spine (C1-C7). Each volunteer underwent one scan using SENSE with an acceleration factor of R = 3.75. and four scans using CSE (acceleration factors R = 4, 6, 8 and 10—termed as CSE4, CSE6, CSE8 and CSE10). The CSE sequence is a combination of SENSE and compressed sensing techniques. Both sequences used in this work were provided by the manufacturer as product sequences (Compressed SENSE/ SENSE, Philips Healthcare). The acquired trajectory was cartesian. A pseudorandom undersampling pattern with fully sampling of the center and a randomized pattern of the remaining k-space was used. No sparsity was used in the time domain. A regularized L1-iterative norm in combination with wavelet transform as sparsifying transform was used for reconstruction of the images. In 15 out of the 20 subjects a second scan using SENSE was performed to evaluate the scan-rescan reproducibility. To mitigate the bias from ordering effects the acquisitions were performed in random order, except for the second scan using SENSE, which was always acquired at the end of the session. Repeated scans were acquired without repositioning of the subjects. The MRI protocol took about 60 min depending on the subjects’ heart rate. All imaging parameters are shown in Table 1. The FOV was resized, if necessary.
Table 1
Acquisition parameters
Parameter
Cervical spinal canal
C1-L1
Repetition time (ms)
8.2
8.3
Echo time (ms)
5.1
4.0
Field of view (mm3)
[220–240] × [240–250] × 30 (FH × AP × RL)
277.6 × 277.6 × 30 (FH × AP × RL)
Acquisition voxel size (mm3)
1.2 × 1.2 × 1.2
1.5 × 1.5 × 1.5
Reconstructed voxel size (mm3)
0.6 × 0.6 × 0.6
0.88 × 0.88 × 0.75
Acquired temporal resolution (ms)
65.6
66.4
VENC(cm/s) (all spatial directions)
15
15
Flip angle (deg)
4
4
Reconstructed cardiac phases
15
15
Acceleration factor
3.75,4,6,8,10
6
In one additional healthy subject an acquisition of the cervical and the thoracic spinal canal (C1-L1) was performed by splitting the FOV in two overlapping sagittal stacks using an acceleration factor of R = 6 (CSE6). To reduce the effect of heart rate variability between the two stacks 15 cardiac phases were reconstructed. Scan parameters are also listed in Table 1.
Reconstruction was performed on-line using standard imaging reconstructing hardware (32 GB RAM, Intel Xeon E5-1620 CPU). Concomitant gradient fields were corrected on-line. Eddy current induced background phases were corrected during post processing.

Image analysis

The velocity mapping data were analyzed using the commercially available GTFlow software (version 3.1, Gyrotools LLC, Winterthur, Switzerland). Built-in correction of eddy currents induced background phases were applied and by default, cranially directed velocities were defined as positive. For each volunteer 7 regions of interest (ROI) were defined in the scan using SENSE located at the center of the cervical vertebrae (C1–C7). In each ROI the subarachnoid space of the spinal cord was manually outlined. Contours were drawn at the time point with the most apparent flow and copied to all time points. The defined contours were transferred to the scan using CSE and the repeated scan using SENSE and shifted manually, if necessary. To provide good agreement the spinal canal and the vertebral vessels served as anatomical landmarks. Flow curves were automatically extracted by GTFlow as the sum of the pixels of the contour. Forward volume flow (FF) was calculated as the sum of the cranially directed flow, backward volume flow (BF) was calculated as the sum of the caudally directed flow (see Fig. 1). Additionally, the absolute net flow (AF) was calculated as the average of absolute forward and backward volume flow. Peak velocity (PV) was extracted from the pixel with the maximum velocity of all time points within the ROI.
Data covering the C1-L1 region were processed equally. ROIs were defined at the level of the midportion of each vertebra and additionally at each intervertebral disc, resulting in 36 analyzed ROIs. Both stacks were visually overlapped using Matlab (Matlab 2019a, The MathWorks, Inc., Natick, MA) and flow patterns were visualized by generating pathlines.

Statistical analysis

For statistical analysis, the open source software package R-studio (version 3.6.2) was used. Bland–Altman analyses were performed to assess the agreement and the correlation between the scans using SENSE and the scans using different CSE factors as well as scan-rescan reproducibility between repeated scans using SENSE. Limits of agreement were corrected as recommended for repeated measurements using REML [25].
To additionally assess the significance and magnitude of the potential bias of CSE methods compared to SENSE, pair differences between CSE methods and SENSE were calculated as SENSE-SENSE2 and SENSE-CSE for each position along the spinal cord within each patient. This approach accounts for the dependency of the 7 acquired slices within the subjects. These pair differences were analyzed in a linear mixed effect model, with the mean pair difference as fixed effect and between patient variance as random effect. Based on the fitted models we computed: (1) p values for the test whether mean pair difference is equal to 0 (null hypothesis, no bias compared to SENSE) or differs from 0 (alternative hypothesis, CS methods biased compared to SENSE), and (2) 95% confidence limits for the mean pair difference.
Additionally, the accumulated flow error ER as proposed by Giese et al. [26] was calculated (Eq. 1). This parameter is an approach that is more sensitive to detect possible temporal differences between the flow curves. Since the net CSF flow should be close to zero, instead of dividing by the sum of the net flow it was divided by the sum of the absolute net flow.
$${E}_{R}=\frac{1}{{n}_{S}}\sum_{S=1}^{{n}_{S}}\left(\frac{1}{{n}_{C}}\sum_{C=1}^{{n}_{C}}\left(\frac{\sum_{t=0}^{{t}_{{n}_{T}}}\left|{{Q}_{\text{SENSE}}}_{S,C}\left(t\right)-{{Q}_{\text{CSE R}}}_{S,C}\left(t\right)\right|}{\sum_{t=0}^{{t}_{{n}_{T}}}{{|Q}_{\text{SENSE}}}_{S,C}\left(t\right)|}\right)\right)$$
(1)
\({n}_{S}\) corresponds to the number of subjects, \({n}_{C}\) to the number of contours and \({n}_{T}\) to the number of timepoints. \({{Q}_{\text{SENSE}}}_{S,C}\) and \({{Q}_{\text{CSE R}}}_{S,C}\) delineates the flow through the contour C in the subject S over time in the scan using SENSE and CSE.

Results

Scans were completed in all subjects, none of the scans had to be aborted or repeated. In one subject all slices of the second scan using SENSE and in another subject slices from C3–C7 of the scans using SENSE and CSE6 were excluded due to motion artifacts. Acquisition times averaged over all subjects were 10:21± 1:19 min (SENSE), 9:31 ± 1:12 min (CSE4), 6:25 ± 0:51 min (CSE6), 4:53 ± 0:38 min (CSE8) and 3:51 ± 0:29 min (CSE10). The average reconstruction time of CSE accelerated acquisitions took around 10 min per scan.
Figure 2a exemplary shows phase-contrast, magnitude and pathline images of one subject. Figure 2b shows resulting magnitude and phase-contrast images with the corresponding ROI. A decrease in quality with increasing CSE acceleration factor can be noticed. Time resolved animations of the pathlines are available in the supplementary material (see Additional files 1, 2, 3, 4, 5).
Figure 3 shows the time-accumulated flow error averaged over all subjects. While CSE4, CSE6 and CSE8 show comparable flow errors around 30%, the tendency of an increasing error with increasing acceleration can still be observed. The largest deviation can be seen in CSE10 with a flow error of 53.9 ± 39.1%.
Results of the Bland–Altman analyses are shown in Table 2. For Bland–Altman plots see Fig. 4 and Additional files 6, 7, 8.
Table 2
Results of Bland–Altman analysis for all flow parameters
Flow parameter
Acceleration factor
MD (CI)
SD
UL (CI)
LL (CI)
Absolute mean
Absolute net flow (ml/s)
CSE4
0.31 (0.1; 0.5)
1.19
2.65 (2.3; 3.0)
− 2.02 (− 2.4; − 1.7)
7.94
CSE6
0.82 (0.6; 1.1)
1.42
3.60 (3.2; 4.0)
− 1.96 (− 2.4; − 1.5)
7.45
CSE8
1.13 (0.9; 1.4)
1.4
3.88 (3.5; 4.3)
− 1.61 (− 2.0; − 1.2)
7.12
CSE10
0.82 (0.5;1.2)
2.11
4.95 (4.3; 5.6)
− 3–30 (− 3.9; − 2.7)
7.43
SENSE2
0.08 (− 0.3; 0.4)
1.77
3.55 (3.0; 4.2)
− 3.39 (− 4.0; − 2.8)
7.59
SENSE
    
8.25
Peak velocity (cm/s)
CSE4
0.31 (− 0.2; 0.8)
3.22
6.63 (5.7; 7.6)
− 6.0 (− 5.1; − 6.9)
5.31
CSE6
0.41 (− 0.04;0.9)
2.74
5.78 (5.0; 6.6)
− 4.96 (− 4.2; − 5.8)
5.21
CSE8
0.40 (− 0.05;0.9)
2.71
5.72 (5.0; 6.5)
− 4.91 (− 4.1; − 5.7)
5.20
CSE10
− 0.22 (− 0.7;0.2)
2.84
5.33 (4.5; 6.1)
− 5.79 (− 5.0; − 6–6)
5.85
SENSE2
− 0.32 (− 1.0; 0.3)
3.16
5.88 (4.8; 7.0)
− 6.52 (− 5.4; − 7.7)
5.92
SENSE
    
5.598
Forward flow (ml/s)
CSE4
0.78 (0.5; 1.02)
1.44
3.61 (3.6; 5.6)
− 2.05 (− 1.6; − 2.5)
8.93
CSE6
0.96 (0.7; 1.2)
1.60
4.10 (3.6; 4.6)
− 2.18 (− 1.7; − 2.6)
8.76
CSE8
0.90 (0.6; 1.2)
1.87
4.56 (4.0; 5.1)
− 2.77 (− 2.2; − 3.3)
8.81
CSE10
0.31 (− 0.4; 1)
3.91
7.97 (6.8; 9.1)
− 7.35 (− 6.2; − 8.5)
9.40
SENSE2
− 0.003 (− 0.4; 0.4)
1.91
3.76 (3.1; 4.4)
− 3.76 (− 3.1; − 4.5)
9.23
SENSE
    
9.71
Backward flow (ml/s)
CSE4
0.33 (0.08; 0.6)
1.48
3.23 (2.8; 3.7)
− 2.58 (− 2.1; − 3.0)
9.77
CSE6
0.65 (0.3; 0.97)
1.90
4.37 (3.8; 4.9)
− 3.06 (− 2.5; − 3.6)
9.49
CSE8
0.87 (0.5; 1.2)
2.02
4.83 (4.3; 5.4)
− 3.10 (− 2.5; − 3.7)
9.23
CSE10
0.57 (− 0.08; 1.2)
3.90
8.21 (7.1; 9.3)
− 7.07 (− 5.9; − 8.2)
9.53
SENSE2
− 0.21 (− 0.7; 0.3)
2.29
4.28 (3.5; 5.1)
− 4.71 (− 3.9; − 5.5)
9.97
SENSE
    
10.1
SD standard deviation, CI confidence interval, MD mean difference, UL upper limit, LL lower limit
Upper limit of agreement represents the mean difference + 1.96 standard deviations, lower limit of agreement represents the mean difference − 1.96 standard deviations. All numbers in parentheses indicate the 95% confidence interval
SD is given for difference between CSE and SENSE. Deviations between CSE and SENSE were evaluated in 140 measurements for CSE4, 8 and 10 and 135 for CSE6. Deviations between SENSE 2 and SENSE were evaluated for 93 measurements. Absolute mean values are given for reference
Limits of agreement (LoA) of CSE4, CSE6 and CSE8 were similar for AF, FF and BF with a slight increase with increasing acceleration factors. In CSE10 a notable widening of the confidence intervals can be observed for all parameters. The PV showed relatively larger confidence intervals than other parameters in all acceleration factors with LoAs between 10–12 cm/s, although the mean differences of PV showed only minor deviations in all acceleration factor (CSE4: 0.3 cm/s; CSE6: 0.43 cm/s; CSE8: 0.33 cm/s; CSE10: − 0.223 cm/s). Confidence intervals (CI) of PV included zero in all acceleration factors. The mean differences in AF and BF were smallest in CSE4 (AF: 0.3 ml/s and BF: 0.3 ml/s), increased in CSE6 (AF:0.8 ml/s; BF: 0.6 ml/s) and were largest in CSE8 (1.1 ml/s and BF: 0.9 ml/s).
Table 3 shows the resulting mean pair differences.
Table 3
Statistical interference for the pair differences of CSE vs. SENSE and SENSE2 vs. SENSE
Flow parameter
Acceleration-method
Mean pair difference
Lower CI
Upper CI
p value
Sign
Average net flow (ml/s)
SENSE2
0.08
− 0.71
0.87
0.84
 
CSE4
0.31
− 0.05
0.67
0.10
 
CSE6
0.82
0.37
1.27
0.002
**
CSE8
1.13
0.82
1.44
5.59∙10–7
***
CSE10
0.82
0.15
1.49
0.02
*
Peak velocity (cm/s)
SENSE
− 0.32
− 1.24
0.60
0.49
 
CSE4
0.31
− 0.22
0.85
0.26
 
CSE6
0.41
− 1.18
1.00
0.18
 
CSE8
0.40
− 0.21
1.00
0.20
 
CSE10
− 0.23
− 0.72
0.26
0.36
 
Forward flow (ml/s)
SENSE2
− 0.003
− 0.84
0.84
0.99
 
CSE4
0.78
0.29
1.28
0.01
**
CSE6
0.96
0.42
1.50
0.002
**
CSE8
0.90
0.33
1.46
0.005
**
CSE10
0.31
− 0.47
1.09
0.44
 
Backward flow (ml/s)
SENSE2
− 0.21
− 1.22
0.79
0.67
 
CSE4
0.33
− 0.11
0.77
0.15
 
CSE6
0.65
0.02
1.28
0.05
 
CSE8
0.87
0.32
1.42
0.01
**
CSE10
0.57
− 0.19
1.33
0.15
 
Lower CI lower confidence interval, Upper CI upper confidence interval, p value p value for the test whether mean pair difference is equal to 0 (null hypothesis) or differs from 0 (alternative)
Interference for mean pair difference for SENSE–SENSE2 and SENSE–CSE, based on mixed model fits
Stars indicate that the mean pair difference is significantly different from 0 (* 0.05 > p > 0.01; ** 0.01 > p > 0.001; *** 0.001 > p)
There were no significant differences for PV for all acceleration factors. CSE4, 6 and 8 showed a significant difference for FF of 0.78 ml/s (CSE4), 0.96 ml/s (CSE6) and 0.90 ml/s (CSE8). BF had no significant differences except for CSE8 with an estimated difference of 0.87 ml/s. There was a significant difference for AF for CSE6, 8 and 10, with CSE8 showing the highest difference of 1.13 ml/s (p-value 5.59∙10–7). The significant difference for CSE6 and 10 was 0.82 ml/s with CIs of 0.37–1.27 ml/s (CSE6) and 0.15–1.49 ml/s (CSE10). There is a tendency to underestimate values in all CSE acceleration factors.
The test–retest comparison (see Fig. 4b) showed small differences between the first and the second scan using SENSE in all parameters (mean differences: AF: 0.08 ml/s; PV: − 0.32 ml/s; FF: − 0.003 ml/s; BF: − 0.215 ml/s). CIs included 0 in all parameters. In all parameters the LoA for repeated scans using SENSE were comparable to the LoA in CSE8. The standard deviation (SD) of repeated scans using SENSE acquisitions was higher than in scans using CSE4, CSE6 and CSE8. Mean pair difference showed no significant differences between the first and the second scan using SENSE in all parameters assessed.
Data covering the C1-L1 region were acquired successfully. The acquisition time took 7:20 min per stack and 14:40 min in total. Time-resolved animations of recorded pathlines are available in the supplementary material (see Additional file 9). A screenshot of the pathlines is shown in Fig. 5.

Discussion

To our knowledge, this is the first study evaluating the feasibility of CSE acceleration for 4D flow MRI of CSF flow dynamics. Moreover, by applying an appropriate CSE acceleration, we were able to demonstrate that it is feasible to measure the CSF flow surrounding the entire spinal cord during a single examination with a clinically acceptable scan time.
Peak flow velocities (5.6 ± 2.7 cm/s) were in good agreement with results of previous studies by Pahlavian et al. [27] with a peak velocity of 6.0 ± 1.7 cm/s, by Bunck et al. [9] with values between 3.0 and 4.8 cm/s and by Lindstrøm et al. [28] with values between 1.8 and 4.2 cm/s.
The flow parameters analyzed in this study showed a good test–retest agreement between repeated scans using SENSE. The SD of 3.16 cm/s for PV and around 2 ml/s for BF, AF and FF seem high compared to the CSE data. This can be attributed to both the smaller number of subjects in the test–retest group naturally leading to wider CIs and the physiological fluctuations due to the delay between the acquisitions of the first and the second scan using SENSE. To minimize physiological differences, like respiration or varying heart rate, the order of acquisitions was randomized except for the second SENSE accelerated scan. This may explain an increased SD of the second scan using SENSE compared to the scans using CSE4 and CSE6. Additionally, reduced movement and physiological variations due to the shorter scan time in scans using CSE4 and CSE6 may lead to increased accuracy. This could also serve as a possible explanation why CSE10 showed no significant differences for FF, BF and PV.
Overall Bland–Altman analyses showed good agreement between scans using CSE and SENSE with acceleration factors up to CSE8. The LoA and the SD of the PV are relatively wider than for the rest of the parameters. This finding can most likely be attributed to the internal process of determining this specific value. The PV is defined as the pixel with the highest velocity. Therefore, this parameter can be subject to large fluctuations due to noise. Another reason that could contribute to deviations is the temporal resolution and the temporal smoothing by 4D flow PCMRI [29, 30]. This finding was also observed in the ventricular structures by Stadlbauer et al. [10] and in the cerebral aqueduct by Sartoretti et al. [31]. Although in all parameters the mean difference of CSE10 is comparable to CSE4 and CSE6 and showed no significant difference for FF, BF, and PV, wide confidence intervals, a high accumulated flow error and a substantially decreased image quality led to the conclusion that CSE10 is beyond adequate acceleration factors. SD of the scan-rescan measurement was between 1.77 and 2.29 ml/s. There were no significant differences for CSE4 for BF and AF and for CSE6 for BF. The significant differences that were detected for CSE4 and CSE6 are small (MD < 0.5 ml/s for CSE4, MD < 0.9 ml/s for CSE6). Those deviations are probably clinically not relevant as the alterations in flow velocities such as in stenotic flow are usually well beyond these minor deviations [2]. Small CIs indicate a higher accuracy due to lower variations within different subjects.
Results of the accumulated flow error were broadly in line with results of the Bland–Altman analyses. A flow error around 30% is higher than reported for cardiovascular 4D flow MRI, for which errors between 15 and 20% in aortic flow and an error between 20–30% in the pulmonary arteries have been described [11, 26]. This can most likely be attributed to the CSF flow being more sensitive to disturbances and the smaller flow velocities making stronger gradients necessary [8]. Additionally, CSF velocities are close to zero in a large portion of the cardiac cycle, which increases the measurement error when using a PCMRI sequence with single encoding velocity. Also, outlining of the spinal CSF can be challenging, e.g. by the contamination of adjacent blood vessels [9], especially because we chose to shift the contours manually and not automatically. Previous studies have shown that respiration also effects the CSF flow in both direction and speed [3234]. This might be an additional cause for larger errors between repeated scans. Although CSE6 and CSE8 showed similar results in the Bland–Altman analysis and the accumulated flow error, defining contours in images acquired with CSE8 may be difficult due to the reduced magnitude image quality (see Fig. 2). This could potentially lead to greater deviations than observed in our study, as all contours were outlined in the SENSE accelerated scan and also needs to be considered when choosing the appropriate acceleration factor. Moreover, in contrast to CSE8, mean deviations were less than 10% of absolute values for all tested parameters when applying CSE6 acceleration. Additionally, CSE8 showed a significant difference for AF, FF and BF with a mean difference > 1 ml/s for AF whereas CSE6 only showed a significant difference for AF and FF with mean differences always < 1 ml/s. Compared to SENSE and CSE4, the CSE6 scan achieved a noticeable reduction in scan time. It is possible that the reduced scan time is the reason for the lower accumulated flow error and may improve image quality by reduced motion artifacts. It would also be possible to make use of the time gained in CSE6 by e.g. choosing a higher temporal resolution. Accordingly, we regard CSE6 as the most appropriate acceleration factor.
Although only the cervical spine was assessed in detail in our study, we have demonstrated the feasibility of a measurement covering the entire spinal cord and its surrounding CSF space in two stacks in 14 min. Such an acquisition allows for a more comprehensive visualization of CSF dynamics and may provide important diagnostic information for patients with syringomyelia, Chiari malformation or arachnoid adhesions [3537].
Our study has several limitations. The study assessed only a relatively small number of healthy subjects. However, the results of our study clearly show the feasibility of CSE6 accelerated 4D flow MRI of the CSF dynamics at the cervical spinal canal. With a higher number of subjects, we would expect standard deviations and level of agreements to be even smaller. Since no patient data were recorded, these results are not fully transferable to patients with abnormal, more complex flow dynamics. With CSE6, however, we have chosen a rather conservative acceleration factor, so that its use in everyday clinical practice seems conceivable. Quantitative analysis was only provided for the CSF flow at the cervical spinal canal. However, as the feasibility of an imaging volume covering the entire spinal cord was shown in one subject with sufficient image quality, the results are likely to be applicable to whole spine measurements. A specific challenge of whole spine measurements is the correct cardiac gating of both stacks. This could reduce the accuracy of the correct flow dynamics. A major limitation is the missing respiratory gating. Studies by Spijkerman et al. [38] and Yildiz et al. [39] have shown that respiratory gated 2D flow MRI measurements provide more accurate quantifications of peak velocity, net flow, and stroke volume in 2D acquisitions with significant but small differences between respiratory gated acquisitions in inspiration and expiration. However, respiratory gating prolongs the scan time by about a factor of 2 to 3 and thus leads to extremely increased scan times in 4D flow MRI. Therefore, non-respiratory gated 4D flow MRI acquisitions were acquired to achieve clinical feasible scan times. Future studies may enable further acceleration, making respiratory gated 4D flow MRI acquisitions of the spinal CSF feasible with scan times of less than 10 min and thus enable more accurate quantifications of CSF hydrodynamics.
We have chosen a relatively high parallel imaging factor (SENSE) in order to achieve an acquisition with a similar duration as CSE4. A longer SENSE scan with lower acceleration factors would have exceeded the duration of the entire scan protocol. Additionally, although the image quality is increased by lower acceleration factors, the influence of motion artifacts may be more prevalent with increased scan time.
Finally, further sequence optimizations are still applicable. Long reconstruction times can be further improved by application of dedicated reconstruction algorithms [40] and temporal-spatial redundancy may be exploited leading to increased image acceleration [41].

Conclusions

Our study demonstrates that CSE is a reliable method to accelerate CSF flow measurements. Acceleration factors up to CSE6 seem to be appropriate for clinical use enabling a reduction of acquisition time by 40%. Scan times of about 6 min for the acquisition of a cervical spine imaging volume and about 14 min for a set of two imaging volumes covering the entire spinal cord can be achieved. This facilitates the use of 4D flow MRI for the diagnosis of pathologic CSF dynamics as in Chiari Malformation or syringomyelia.

Supplementary information

Supplementary information accompanies this paper at https://​doi.​org/​10.​1186/​s12987-020-00206-3.

Acknowledgements

Not applicable.
All study participants signed written consent form. The study was approved by the institutional ethics committee.
All subjects gave written consent to participate and to publish.

Competing interests

KW is an employee of Philips Healthcare Germany. The other authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Kelly EJ, Yamada S. Cerebrospinal fluid flow studies and recent advancements. Semin Ultrasound CT MRI. 2016;37(2):92–9. Kelly EJ, Yamada S. Cerebrospinal fluid flow studies and recent advancements. Semin Ultrasound CT MRI. 2016;37(2):92–9.
2.
Zurück zum Zitat Bunck AC, Kroeger JR, Juettner A, Brentrup A, Fiedler B, Crelier GR, et al. Magnetic resonance 4D flow analysis of cerebrospinal fluid dynamics in Chiari I malformation with and without syringomyelia. Eur Radiol. 2012;22(9):1860–70.PubMed Bunck AC, Kroeger JR, Juettner A, Brentrup A, Fiedler B, Crelier GR, et al. Magnetic resonance 4D flow analysis of cerebrospinal fluid dynamics in Chiari I malformation with and without syringomyelia. Eur Radiol. 2012;22(9):1860–70.PubMed
3.
Zurück zum Zitat Haughton VM, Korosec FR, Medow JE, Dolar MT, Iskandar BJ. Peak systolic and diastolic CSF velocity in the foramen magnum in adult patients with Chiari I malformations and in normal control participants. Am J Neuroradiol. 2003;24(2):169–76.PubMedPubMedCentral Haughton VM, Korosec FR, Medow JE, Dolar MT, Iskandar BJ. Peak systolic and diastolic CSF velocity in the foramen magnum in adult patients with Chiari I malformations and in normal control participants. Am J Neuroradiol. 2003;24(2):169–76.PubMedPubMedCentral
4.
Zurück zum Zitat Heiss JD, Patronas N, DeVroom HL, Shawker T, Ennis R, Kammerer W, et al. Elucidating the pathophysiology of syringomyelia. J Neurosurg. 1999;91(4):553–62. PubMed Heiss JD, Patronas N, DeVroom HL, Shawker T, Ennis R, Kammerer W, et al. Elucidating the pathophysiology of syringomyelia. J Neurosurg. 1999;91(4):553–62. PubMed
5.
Zurück zum Zitat Schubert JJ, Veronese M, Marchitelli L, Bodini B, Tonietto M, Stankoff B, et al. Dynamic 11 C-PiB PET shows cerebrospinal fluid flow alterations in Alzheimer’s disease and multiple sclerosis. J Nucl Med. 2019;60(10):1452–60. PubMedPubMedCentral Schubert JJ, Veronese M, Marchitelli L, Bodini B, Tonietto M, Stankoff B, et al. Dynamic 11 C-PiB PET shows cerebrospinal fluid flow alterations in Alzheimer’s disease and multiple sclerosis. J Nucl Med. 2019;60(10):1452–60. PubMedPubMedCentral
6.
Zurück zum Zitat Wagshul ME, Chen JJ, Egnor MR, McCormack EJ, Roche PE. Amplitude and phase of cerebrospinal fluid pulsations: Experimental studies and review of the literature. J Neurosurg. 2006;104(5):810–9.PubMed Wagshul ME, Chen JJ, Egnor MR, McCormack EJ, Roche PE. Amplitude and phase of cerebrospinal fluid pulsations: Experimental studies and review of the literature. J Neurosurg. 2006;104(5):810–9.PubMed
7.
Zurück zum Zitat Yamada S, Tsuchiya K, Bradley WG, Law M, Winkler ML, Borzage MT, et al. Current and emerging MR imaging techniques for the diagnosis and management of CSF flow disorders: a review of phase-contrast and time-spatial labeling inversion pulse. AJNR Am J Neuroradiol. 2015;36(4):623–30. PubMedPubMedCentral Yamada S, Tsuchiya K, Bradley WG, Law M, Winkler ML, Borzage MT, et al. Current and emerging MR imaging techniques for the diagnosis and management of CSF flow disorders: a review of phase-contrast and time-spatial labeling inversion pulse. AJNR Am J Neuroradiol. 2015;36(4):623–30. PubMedPubMedCentral
8.
Zurück zum Zitat Korbecki A, Zimny A, Podgórski P, Sąsiadek M, Bladowska J. Imaging of cerebrospinal fluid flow: fundamentals, techniques, and clinical applications of phase-contrast magnetic resonance imaging. Pol J Radiol. 2019;84:e240–50. PubMedPubMedCentral Korbecki A, Zimny A, Podgórski P, Sąsiadek M, Bladowska J. Imaging of cerebrospinal fluid flow: fundamentals, techniques, and clinical applications of phase-contrast magnetic resonance imaging. Pol J Radiol. 2019;84:e240–50. PubMedPubMedCentral
9.
Zurück zum Zitat Bunck AC, Kröger JR, Jüttner A, Brentrup A, Fiedler B, Schaarschmidt F, et al. Magnetic resonance 4D flow characteristics of cerebrospinal fluid at the craniocervical junction and the cervical spinal canal. Eur Radiol. 2011;21(8):1788–96. PubMed Bunck AC, Kröger JR, Jüttner A, Brentrup A, Fiedler B, Schaarschmidt F, et al. Magnetic resonance 4D flow characteristics of cerebrospinal fluid at the craniocervical junction and the cervical spinal canal. Eur Radiol. 2011;21(8):1788–96. PubMed
10.
Zurück zum Zitat Stadlbauer A, Salomonowitz E, Brenneis C, Ungersbock K, Van Der Riet W, Buchfelder M, et al. Magnetic resonance velocity mapping of 3D cerebrospinal fluid flow dynamics in hydrocephalus: preliminary results. Eur Radiol. 2012;22(1):232–42.PubMed Stadlbauer A, Salomonowitz E, Brenneis C, Ungersbock K, Van Der Riet W, Buchfelder M, et al. Magnetic resonance velocity mapping of 3D cerebrospinal fluid flow dynamics in hydrocephalus: preliminary results. Eur Radiol. 2012;22(1):232–42.PubMed
11.
Zurück zum Zitat Neuhaus E, Weiss K, Bastkowski R, Koopmann J, Maintz D, Giese D. Accelerated aortic 4D flow cardiovascular magnetic resonance using compressed sensing: applicability, validation and clinical integration. J Cardiovasc Magn Reson. 2019;21(1):65.PubMedPubMedCentral Neuhaus E, Weiss K, Bastkowski R, Koopmann J, Maintz D, Giese D. Accelerated aortic 4D flow cardiovascular magnetic resonance using compressed sensing: applicability, validation and clinical integration. J Cardiovasc Magn Reson. 2019;21(1):65.PubMedPubMedCentral
12.
Zurück zum Zitat Carlsson M, Töger J, Kanski M, Bloch KM, Ståhlberg F, Heiberg E, et al. Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: Head to head comparison and validation at 1.5 T and 3 T. J Cardiovasc Magn Reson. 2011;13(1):55.PubMedPubMedCentral Carlsson M, Töger J, Kanski M, Bloch KM, Ståhlberg F, Heiberg E, et al. Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: Head to head comparison and validation at 1.5 T and 3 T. J Cardiovasc Magn Reson. 2011;13(1):55.PubMedPubMedCentral
13.
Zurück zum Zitat Jung B, Stalder AF, Bauer S, Markl M. On the undersampling strategies to accelerate time-resolved 3D imaging using k-t-GRAPPA. Magn Reson Med. 2011;66(4):966–75.PubMed Jung B, Stalder AF, Bauer S, Markl M. On the undersampling strategies to accelerate time-resolved 3D imaging using k-t-GRAPPA. Magn Reson Med. 2011;66(4):966–75.PubMed
14.
Zurück zum Zitat Knobloch V, Boesiger P, Kozerke S. Sparsity transform k-t principal component analysis for accelerating cine three-dimensional flow measurements. Magn Reson Med. 2013;70(1):53–63.PubMed Knobloch V, Boesiger P, Kozerke S. Sparsity transform k-t principal component analysis for accelerating cine three-dimensional flow measurements. Magn Reson Med. 2013;70(1):53–63.PubMed
15.
Zurück zum Zitat Hsiao A, Lustig M, Alley MT, Murphy M, Chan FP, Herfkens RJ, et al. Rapid pediatric cardiac assessment of flow and ventricular volume with compressed sensing parallel imaging volumetric cine phase-contrast MRI. Am J Roentgenol. 2012;198(3):W250–W259259. Hsiao A, Lustig M, Alley MT, Murphy M, Chan FP, Herfkens RJ, et al. Rapid pediatric cardiac assessment of flow and ventricular volume with compressed sensing parallel imaging volumetric cine phase-contrast MRI. Am J Roentgenol. 2012;198(3):W250–W259259.
16.
Zurück zum Zitat Donoho DL. Compressed sensing. IEEE Trans Inf Theory. 2006;52(4):1289–306. Donoho DL. Compressed sensing. IEEE Trans Inf Theory. 2006;52(4):1289–306.
17.
Zurück zum Zitat Lustig M, Donoho DL, Santos JM, Pauly JM. Compressed sensing MRI. IEEE Signal Process Mag. 2008;25(2):72–82. Lustig M, Donoho DL, Santos JM, Pauly JM. Compressed sensing MRI. IEEE Signal Process Mag. 2008;25(2):72–82.
18.
Zurück zum Zitat Feng L, Benkert T, Block KT, Sodickson DK, Otazo R, Chandarana H. Compressed sensing for body MRI. J Magn Reson Imaging. 2017;45:966–87. PubMed Feng L, Benkert T, Block KT, Sodickson DK, Otazo R, Chandarana H. Compressed sensing for body MRI. J Magn Reson Imaging. 2017;45:966–87. PubMed
19.
Zurück zum Zitat Jaspan ON, Fleysher R, Lipton ML. Compressed sensing MRI: a review of the clinical literature. Br J Radiol. 2015;88:20150487. PubMedPubMedCentral Jaspan ON, Fleysher R, Lipton ML. Compressed sensing MRI: a review of the clinical literature. Br J Radiol. 2015;88:20150487. PubMedPubMedCentral
20.
Zurück zum Zitat Otazo R, Kim D, Axel L, Sodickson DK. Combination of compressed sensing and parallel imaging for highly accelerated first-pass cardiac perfusion MRI. Magn Reson Med. 2010;64(3):767–76. PubMedPubMedCentral Otazo R, Kim D, Axel L, Sodickson DK. Combination of compressed sensing and parallel imaging for highly accelerated first-pass cardiac perfusion MRI. Magn Reson Med. 2010;64(3):767–76. PubMedPubMedCentral
21.
Zurück zum Zitat Liu F, Duan Y, Peterson BS, Kangarlu A. Compressed sensing MRI combined with SENSE in partial k-space. Phys Med Biol. 2012;57(21):N391.PubMed Liu F, Duan Y, Peterson BS, Kangarlu A. Compressed sensing MRI combined with SENSE in partial k-space. Phys Med Biol. 2012;57(21):N391.PubMed
22.
Zurück zum Zitat Kwak Y, Nam S, Akçakaya M, Basha TA, Goddu B, Manning WJ, et al. Accelerated aortic flow assessment with compressed sensing with and without use of the sparsity of the complex difference image. Magn Reson Med. 2013;70:851–8. PubMed Kwak Y, Nam S, Akçakaya M, Basha TA, Goddu B, Manning WJ, et al. Accelerated aortic flow assessment with compressed sensing with and without use of the sparsity of the complex difference image. Magn Reson Med. 2013;70:851–8. PubMed
23.
Zurück zum Zitat Ma LE, Markl M, Chow K, Huh H, Forman C, Vali A, et al. Aortic 4D flow MRI in 2 minutes using compressed sensing, respiratory controlled adaptive k-space reordering, and inline reconstruction. Magn Reson Med. 2019;81:3675–90. PubMedPubMedCentral Ma LE, Markl M, Chow K, Huh H, Forman C, Vali A, et al. Aortic 4D flow MRI in 2 minutes using compressed sensing, respiratory controlled adaptive k-space reordering, and inline reconstruction. Magn Reson Med. 2019;81:3675–90. PubMedPubMedCentral
24.
Zurück zum Zitat Tao Y, Rilling G, Davies M, Marshall I. Carotid blood flow measurement accelerated by compressed sensing: validation in healthy volunteers. Magn Reson Imaging. 2013;31:1485–91.PubMed Tao Y, Rilling G, Davies M, Marshall I. Carotid blood flow measurement accelerated by compressed sensing: validation in healthy volunteers. Magn Reson Imaging. 2013;31:1485–91.PubMed
25.
Zurück zum Zitat Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17:571–82. PubMed Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17:571–82. PubMed
26.
Zurück zum Zitat Giese D, Wong J, Greil GF, Buehrer M, Schaeffter T, Kozerke S. Towards highly accelerated cartesian time-resolved 3D flow cardiovascular magnetic resonance in the clinical setting. J Cardiovasc Magn Reson. 2014;16(1):42. PubMedPubMedCentral Giese D, Wong J, Greil GF, Buehrer M, Schaeffter T, Kozerke S. Towards highly accelerated cartesian time-resolved 3D flow cardiovascular magnetic resonance in the clinical setting. J Cardiovasc Magn Reson. 2014;16(1):42. PubMedPubMedCentral
27.
Zurück zum Zitat Pahlavian SH, Bunck AC, Loth F, Tubbs RS, Yiallourou T, Kroeger JR, et al. Characterization of the discrepancies between four-dimensional phase-contrast magnetic resonance imaging and in-silico simulations of cerebrospinal fluid dynamics. J Biomech Eng. 2015;5:051002. Pahlavian SH, Bunck AC, Loth F, Tubbs RS, Yiallourou T, Kroeger JR, et al. Characterization of the discrepancies between four-dimensional phase-contrast magnetic resonance imaging and in-silico simulations of cerebrospinal fluid dynamics. J Biomech Eng. 2015;5:051002.
28.
Zurück zum Zitat Lindstrøm EK, Schreiner J, Ringstad GA, Haughton V, Eide PK, Mardal KA. Comparison of phase-contrast MR and flow simulations for the study of CSF dynamics in the cervical spine. Neuroradiol J. 2018;31:292–8.PubMedPubMedCentral Lindstrøm EK, Schreiner J, Ringstad GA, Haughton V, Eide PK, Mardal KA. Comparison of phase-contrast MR and flow simulations for the study of CSF dynamics in the cervical spine. Neuroradiol J. 2018;31:292–8.PubMedPubMedCentral
29.
Zurück zum Zitat Maier IL, Hofer S, Joseph AA, Merboldt KD, Tan Z, Schregel K, et al. Carotid artery flow as determined by real-time phase-contrast flow MRI and neurovascular ultrasound: a comparative study of healthy subjects. Eur J Radiol. 2018;106:38–45.PubMed Maier IL, Hofer S, Joseph AA, Merboldt KD, Tan Z, Schregel K, et al. Carotid artery flow as determined by real-time phase-contrast flow MRI and neurovascular ultrasound: a comparative study of healthy subjects. Eur J Radiol. 2018;106:38–45.PubMed
30.
Zurück zum Zitat Lin H-Y, Bender JA, Ding Y, Chung Y-C, Hinton AM, Pennell ML, et al. Shared velocity encoding: a method to improve the temporal resolution of phase-contrast velocity measurements. Magn Reson Med. 2012;68:703–10. PubMed Lin H-Y, Bender JA, Ding Y, Chung Y-C, Hinton AM, Pennell ML, et al. Shared velocity encoding: a method to improve the temporal resolution of phase-contrast velocity measurements. Magn Reson Med. 2012;68:703–10. PubMed
31.
Zurück zum Zitat Sartoretti T, Wyss M, Sartoretti E, Reischauer C, Hainc N, Graf N, et al. Sex and age dependencies of aqueductal cerebrospinal fluid dynamics parameters in healthy subjects. Front Aging Neurosci. 2019;11:1999. Sartoretti T, Wyss M, Sartoretti E, Reischauer C, Hainc N, Graf N, et al. Sex and age dependencies of aqueductal cerebrospinal fluid dynamics parameters in healthy subjects. Front Aging Neurosci. 2019;11:1999.
32.
Zurück zum Zitat Dreha-Kulaczewski S, Joseph AA, Merboldt KD, Ludwig HC, Gärtner J, Frahm J. Inspiration is the major regulator of human CSF flow. J Neurosci. 2015;35(6):2485–91. PubMedPubMedCentral Dreha-Kulaczewski S, Joseph AA, Merboldt KD, Ludwig HC, Gärtner J, Frahm J. Inspiration is the major regulator of human CSF flow. J Neurosci. 2015;35(6):2485–91. PubMedPubMedCentral
33.
Zurück zum Zitat Yamada S, Miyazaki M, Yamashita Y, Ouyang C, Yui M, Nakahashi M, et al. Influence of respiration on cerebrospinal fluid movement using magnetic resonance spin labeling. Fluids Barriers CNS. 2013;10(1):36. PubMedPubMedCentral Yamada S, Miyazaki M, Yamashita Y, Ouyang C, Yui M, Nakahashi M, et al. Influence of respiration on cerebrospinal fluid movement using magnetic resonance spin labeling. Fluids Barriers CNS. 2013;10(1):36. PubMedPubMedCentral
34.
Zurück zum Zitat Chen L, Beckett A, Verma A, Feinberg DA. Dynamics of respiratory and cardiac CSF motion revealed with real-time simultaneous multi-slice EPI velocity phase contrast imaging. Neuroimage. 2015;122:281–7.PubMed Chen L, Beckett A, Verma A, Feinberg DA. Dynamics of respiratory and cardiac CSF motion revealed with real-time simultaneous multi-slice EPI velocity phase contrast imaging. Neuroimage. 2015;122:281–7.PubMed
35.
Zurück zum Zitat Milhorat TH, Chou MW, Trinidad EM, Kula RW, Mandell M, Wolpert C, et al. Chiari I malformation redefined: clinical and radiographic findings for 364 symptomatic patients. Neurosurgery. 1999;44(5):1005–177. PubMed Milhorat TH, Chou MW, Trinidad EM, Kula RW, Mandell M, Wolpert C, et al. Chiari I malformation redefined: clinical and radiographic findings for 364 symptomatic patients. Neurosurgery. 1999;44(5):1005–177. PubMed
36.
Zurück zum Zitat Sadique SI, Pandey P, Chaudhuri AK. Cerebrospinal fluid flowmetry in pediatric patients with chiari malformation I with surgical implications. World Neurosurg. 2020;135:e83–e8686.PubMed Sadique SI, Pandey P, Chaudhuri AK. Cerebrospinal fluid flowmetry in pediatric patients with chiari malformation I with surgical implications. World Neurosurg. 2020;135:e83–e8686.PubMed
37.
Zurück zum Zitat Mauer UM, Freude G, Danz B, Kunz U. Cardiac-gated phase-contrast magnetic resonance imaging of cerebrospinal fluid flow in the diagnosis of idiopathic syringomyelia. Neurosurgery. 2008;63:1139–44.PubMed Mauer UM, Freude G, Danz B, Kunz U. Cardiac-gated phase-contrast magnetic resonance imaging of cerebrospinal fluid flow in the diagnosis of idiopathic syringomyelia. Neurosurgery. 2008;63:1139–44.PubMed
38.
Zurück zum Zitat Spijkerman JM, Geurts LJ, Siero JCW, Hendrikse J, Luijten PR, Zwanenburg JJM. Phase contrast MRI measurements of net cerebrospinal fluid flow through the cerebral aqueduct are confounded by respiration. J Magn Reson Imaging. 2019;49:433–44. PubMed Spijkerman JM, Geurts LJ, Siero JCW, Hendrikse J, Luijten PR, Zwanenburg JJM. Phase contrast MRI measurements of net cerebrospinal fluid flow through the cerebral aqueduct are confounded by respiration. J Magn Reson Imaging. 2019;49:433–44. PubMed
39.
Zurück zum Zitat Yildiz S, Thyagaraj S, Jin N, Zhong X, Heidari Pahlavian S, Martin BA, et al. Quantifying the influence of respiration and cardiac pulsations on cerebrospinal fluid dynamics using real-time phase-contrast MRI. J Magn Reson Imaging. 2017;46(2):431–9. PubMed Yildiz S, Thyagaraj S, Jin N, Zhong X, Heidari Pahlavian S, Martin BA, et al. Quantifying the influence of respiration and cardiac pulsations on cerebrospinal fluid dynamics using real-time phase-contrast MRI. J Magn Reson Imaging. 2017;46(2):431–9. PubMed
40.
Zurück zum Zitat Baron CA, Dwork N, Pauly JM, Nishimura DG. Rapid compressed sensing reconstruction of 3D non-cartesian MRI. Magn Reson Med. 2018;79(5):2685–92. PubMed Baron CA, Dwork N, Pauly JM, Nishimura DG. Rapid compressed sensing reconstruction of 3D non-cartesian MRI. Magn Reson Med. 2018;79(5):2685–92. PubMed
41.
Zurück zum Zitat Hutter J, Schmitt P, Saake M, Stübinger A, Grimm R, Forman C, et al. Multi-dimensional flow-preserving compressed sensing (MuFloCoS) for time-resolved velocity-encoded phase contrast MRI. IEEE Trans Med Imaging. 2015;34(2):400–14. PubMed Hutter J, Schmitt P, Saake M, Stübinger A, Grimm R, Forman C, et al. Multi-dimensional flow-preserving compressed sensing (MuFloCoS) for time-resolved velocity-encoded phase contrast MRI. IEEE Trans Med Imaging. 2015;34(2):400–14. PubMed
Metadaten
Titel
Compressed-sensing accelerated 4D flow MRI of cerebrospinal fluid dynamics
verfasst von
Elena Jaeger
Kristina Sonnabend
Frank Schaarschmidt
David Maintz
Kilian Weiss
Alexander C. Bunck
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
Fluids and Barriers of the CNS / Ausgabe 1/2020
Elektronische ISSN: 2045-8118
DOI
https://doi.org/10.1186/s12987-020-00206-3

Weitere Artikel der Ausgabe 1/2020

Fluids and Barriers of the CNS 1/2020 Zur Ausgabe

Leitlinien kompakt für die Neurologie

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Nicht Creutzfeldt Jakob, sondern Abführtee-Vergiftung

29.05.2024 Hyponatriämie Nachrichten

Eine ältere Frau trinkt regelmäßig Sennesblättertee gegen ihre Verstopfung. Der scheint plötzlich gut zu wirken. Auf Durchfall und Erbrechen folgt allerdings eine Hyponatriämie. Nach deren Korrektur kommt es plötzlich zu progredienten Kognitions- und Verhaltensstörungen.

Schutz der Synapsen bei Alzheimer

29.05.2024 Morbus Alzheimer Nachrichten

Mit einem Neurotrophin-Rezeptor-Modulator lässt sich möglicherweise eine bestehende Alzheimerdemenz etwas abschwächen: Erste Phase-2-Daten deuten auf einen verbesserten Synapsenschutz.

Sozialer Aufstieg verringert Demenzgefahr

24.05.2024 Demenz Nachrichten

Ein hohes soziales Niveau ist mit die beste Versicherung gegen eine Demenz. Noch geringer ist das Demenzrisiko für Menschen, die sozial aufsteigen: Sie gewinnen fast zwei demenzfreie Lebensjahre. Umgekehrt steigt die Demenzgefahr beim sozialen Abstieg.

Hirnblutung unter DOAK und VKA ähnlich bedrohlich

17.05.2024 Direkte orale Antikoagulanzien Nachrichten

Kommt es zu einer nichttraumatischen Hirnblutung, spielt es keine große Rolle, ob die Betroffenen zuvor direkt wirksame orale Antikoagulanzien oder Marcumar bekommen haben: Die Prognose ist ähnlich schlecht.

Update Neurologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.