Dual-chamber ICD for left bundle branch area pacing: the cardiac resynchronization and arrhythmia sensing via the left bundle (cross-left) pilot study

verfasst von: Nicolas Clementy, Alexandre Bodin, Vincent Ah-Fat, Dominique Babuty, Arnaud Bisson

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 4/2023

Einloggen, um Zugang zu erhalten

Abstract

Background

Left bundle branch area pacing (LBBAP) has emerged as a promising technique to deliver cardiac resynchronization therapy (CRT). However, safety and efficacy of ventricular arrhythmia sensing via the left bundle in implantable cardioverter-defibrillator (ICD) recipients remain unclear. We sought to evaluate the feasibility of a single LBBAP lead connected to a dual-chamber ICD in patients indicated with a CRT-D implantation.

Methods

The CROSS-LEFT pilot study prospectively included 10 consecutive patients with a reduced ejection fraction and a complete left bundle branch block, indicated with a prophylactic CRT-D. A DF-1 lead was implanted at the right ventricular (RV) apex, and an LBBAP lead through the interventricular septum. Ventricular fibrillation was induced at implantation in both conventional (RV) and left bundle branch area sensing configurations. The latter was the final sensing configuration, and patients were implanted with a dual-chamber DF-1 ICD connected to the atrial lead (RA port), the LBBAP lead (RV IS-1 port), and the defibrillation lead (RV DF-1 port), the IS-1 pin being capped. Atrioventricular delay was optimized to ensure fusion between LBBAP and native conduction from the right bundle. Patients were followed during 6 months.

Results

No difference between both configurations was observed regarding R-wave sensing in sinus rhythm (p = 0.22), ventricular fibrillation median interval detection (p = 1.00), or total induced episode duration (p = 0.78). LBBAP resulted in a significant reduction of median QRS width from 164 to 126 ms (p = 0.002). Median ventricular sensing significantly improved from 9.7 at implantation to 18.8 mV at 6 months (p = 0.01). Median LVEF also significantly improved from 29 to 44% at 6 months (p = 0.002).

Conclusion

Ventricular arrhythmia sensing and defibrillation can be performed via a single LBBAP lead connected to a dual-chamber ICD, and is associated with significant electromechanical reverse remodeling.

Clinical trial registration number

NCT05102227

Graphical abstract

In patients presenting with left bundle branch block and left ventricular systolic dysfunction, a left bundle branch area pacing lead connected to a DF-1 dual-chamber implantable cardioverter-defibrillator provides safe ventricular arrhythmia sensing and efficient electro-mechanical resynchronization.
Literatur
8.
Zurück zum Zitat Chen X, Ye Y, Wang Z, Jin Q, Qiu Z, Wang J, Qin S, Bai J, Wang W, Liang Y, Chen H, Sheng X, Gao F, Zhao X, Fu G, Ellenbogen KA, Su Y, Ge J. Cardiac resynchronization therapy via left bundle branch pacing vs. optimized biventricular pacing with adaptive algorithm in heart failure with left bundle branch block: a prospective, multi-centre, observational study. Europace. 2021. https://doi.org/10.1093/europace/euab249.CrossRefPubMedPubMedCentral Chen X, Ye Y, Wang Z, Jin Q, Qiu Z, Wang J, Qin S, Bai J, Wang W, Liang Y, Chen H, Sheng X, Gao F, Zhao X, Fu G, Ellenbogen KA, Su Y, Ge J. Cardiac resynchronization therapy via left bundle branch pacing vs. optimized biventricular pacing with adaptive algorithm in heart failure with left bundle branch block: a prospective, multi-centre, observational study. Europace. 2021. https://​doi.​org/​10.​1093/​europace/​euab249.CrossRefPubMedPubMedCentral
Metadaten
Titel
Dual-chamber ICD for left bundle branch area pacing: the cardiac resynchronization and arrhythmia sensing via the left bundle (cross-left) pilot study
verfasst von
Nicolas Clementy
Alexandre Bodin
Vincent Ah-Fat
Dominique Babuty
Arnaud Bisson
Publikationsdatum
16.08.2022
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 4/2023
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-022-01342-6

Weitere Artikel der Ausgabe 4/2023

Journal of Interventional Cardiac Electrophysiology 4/2023 Zur Ausgabe

Nach Herzinfarkt mit Typ-1-Diabetes schlechtere Karten als mit Typ 2?

29.05.2024 Herzinfarkt Nachrichten

Bei Menschen mit Typ-2-Diabetes sind die Chancen, einen Myokardinfarkt zu überleben, in den letzten 15 Jahren deutlich gestiegen – nicht jedoch bei Betroffenen mit Typ 1.

Erhöhtes Risiko fürs Herz unter Checkpointhemmer-Therapie

28.05.2024 Nebenwirkungen der Krebstherapie Nachrichten

Kardiotoxische Nebenwirkungen einer Therapie mit Immuncheckpointhemmern mögen selten sein – wenn sie aber auftreten, wird es für Patienten oft lebensgefährlich. Voruntersuchung und Monitoring sind daher obligat.

GLP-1-Agonisten können Fortschreiten diabetischer Retinopathie begünstigen

24.05.2024 Diabetische Retinopathie Nachrichten

Möglicherweise hängt es von der Art der Diabetesmedikamente ab, wie hoch das Risiko der Betroffenen ist, dass sich sehkraftgefährdende Komplikationen verschlimmern.

TAVI versus Klappenchirurgie: Neue Vergleichsstudie sorgt für Erstaunen

21.05.2024 TAVI Nachrichten

Bei schwerer Aortenstenose und obstruktiver KHK empfehlen die Leitlinien derzeit eine chirurgische Kombi-Behandlung aus Klappenersatz plus Bypass-OP. Diese Empfehlung wird allerdings jetzt durch eine aktuelle Studie infrage gestellt – mit überraschender Deutlichkeit.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.