Free-breathing simultaneous native myocardial T1, T2 and T1ρ mapping with Cartesian acquisition and dictionary matching

verfasst von: Zhenfeng Lyu, Sha Hua, Jian Xu, Yiwen Shen, Rui Guo, Peng Hu, Haikun Qi

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2023

Abstract

Background

T1, T2 and T1ρ are well-recognized parameters for quantitative cardiac MRI. Simultaneous estimation of these parameters allows for comprehensive myocardial tissue characterization, such as myocardial fibrosis and edema. However, conventional techniques either quantify the parameters individually with separate breath-hold acquisitions, which may result in unregistered parameter maps, or estimate multiple parameters in a prolonged breath-hold acquisition, which may be intolerable to patients. We propose a free-breathing multi-parametric mapping (FB-MultiMap) technique that provides co-registered myocardial T1, T2 and T1ρ maps in a single efficient acquisition.

Methods

The proposed FB-MultiMap performs electrocardiogram-triggered single-shot Cartesian acquisition over 16 consecutive cardiac cycles, where inversion, T2 and T1ρ preparations are introduced for varying contrasts. A diaphragmatic navigator was used for prospective through-plane motion correction and the in-plane motion was corrected retrospectively with a group-wise image registration method. Quantitative mapping was conducted through dictionary matching of the motion corrected images, where the subject-specific dictionary was created using Bloch simulations for a range of T1, T2 and T1ρ values, as well as B1 factors to account for B1 inhomogeneities. The FB-MultiMap was optimized and validated in numerical simulations, phantom experiments, and in vivo imaging of 15 healthy subjects and six patients with suspected cardiac diseases.

Results

The phantom T1, T2 and T1ρ values estimated with FB-MultiMap agreed well with reference measurements with no dependency on heart rate. In healthy subjects, FB-MultiMap T1 was higher than MOLLI T1 mapping (1218 ± 50 ms vs. 1166 ± 38 ms, p < 0.001). The myocardial T2 and T1ρ estimated with FB-MultiMap were lower compared to the mapping with T2- or T1ρ-prepared 2D balanced steady-state free precession (T2: 41.2 ± 2.8 ms vs. 42.5 ± 3.1 ms, p = 0.06; T1ρ: 45.3 ± 4.4 ms vs. 50.2 ± 4.0, p < 0.001). The pathological changes in myocardial parameters measured with FB-MultiMap were consistent with conventional techniques in all patients.

Conclusion

The proposed free-breathing multi-parametric mapping technique provides co-registered myocardial T1, T2 and T1ρ maps in 16 heartbeats, achieving similar mapping quality to conventional breath-hold mapping methods.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1186/​s12968-023-00973-6.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
AHA
American Heart Association
BH-MultiMap
Breath-hold multi-parametric mapping
bSSFP
Balanced steady-state free precession
dNAV
Diaphragmatic navigator
FA
Flip angle
FB-MultiMap
Free-breathing multi-parametric mapping
FOV
Field-of-view
IR
Inversion recovery
LGE
Late gadolinium enhancement
MOLLI
MOdified Look Locker Inversion recovery
MR
Magnetic resonance
MRF
Magnetic resonance fingerprinting
RRMSE
Relative root mean squared error
RSD
Relative standard deviation
SD
Standard deviation
sFB-MultiMap
FB-MultiMap with simplified post-processing
SL
Spin lock
SSIR
Slice-selective IR
T1ρ-prep
T1ρ preparation
T2-prep
T2 preparation
TSL
Spin lock time
VFA
Variable flip angle

Background

Quantitative cardiac magnetic resonance (MR) parametric mapping has emerged as a promising non-invasive tool for characterizing myocardial tissue, diagnosing various cardiovascular diseases, and monitoring treatment effects [13]. The relaxation times or parameters, such as commonly used T1 and T2, represent intrinsic tissue characteristics in a magnetic field and can reflect pathological changes of tissue. Inflammation, fibrosis and amyloid deposition lead to increased native T1 values [46], while the opposite can be observed in conditions such as iron deposition or substantial fat accumulation [79]. Elevated T2 values reflect increased free water content in the tissue, typically caused by myocardial edema or inflammation [10]. T1ρ, on the other hand, represents the longitudinal relaxation in a rotating frame, under a spin lock (SL) radiofrequency pulse which has a low frequency of several hundred Hertz. T1ρ has been shown to provide endogenous contrast for changes in macromolecular content [11]. Previous studies found elevated T1ρ in acute and chronic myocardial infarction [1217], as well as in nonischemic myocardial diseases, such as dilated cardiomyopathy [18] and hypertrophic cardiomyopathy [19]. Compared with native T1, T1ρ may be a more promising endogenous contrast for myocardial fibrosis, as increased T1 is not specific to fibrosis and the ability of native T1 to detect myocardial fibrosis is still controversial.
T1, T2 and T1ρ parameters provide complementary information about the myocardium, and a combination of multiple parameters may improve diagnostic sensitivity and boost diagnostic confidence for suspected cardiomyopathy [3, 20]. However, myocardial parameters are commonly measured individually in separate breath-hold acquisitions [1, 2125], which may result in nonregistered maps, prolonged scan time, and cause patient discomfort due to repeated breath-holds throughout the sequential acquisitions. In addition, the separate mapping methods such as MOdified Look Locker Inversion recovery (MOLLI) for T1 mapping [21] or T2-prepared balanced steady-state free precession (bSSFP) for T2 mapping [24] require dummy cardiac cycles to allow for longitudinal magnetization recovery needed for better exponential curve fitting, which results in inefficient acquisition and heart rate dependency.
Simultaneous multi-parametric mapping techniques have been proposed to overcome the limitations of separate mapping methods. Previous cardiac multi-parametric mapping approaches are mostly for T1 and T2 mapping, which can be generally classified into two categories. The first category involves sophisticated sequence design by interleaving saturation, inversion and T2 preparation pulses so that the acquired multi-contrast images can be fitted to an analytical formula for T1 and T2 mapping [2631]. The second category of techniques, such as cardiac MR fingerprinting (MRF) [32], involves designing a sequence that generates different signal evolutions for different T1 and T2 combinations [3338]. Subsequently, the acquired signals are matched to a dictionary generated according to the imaging sequence in order to derive the quantitative parameters. The dictionary matching techniques are not limited to acquiring images that conform to certain analytical function and thus can be easily extended to mapping more parameters. Velascol et al. have extended the cardiac MRF technique for simultaneous T1, T2 and T1ρ mapping [39]. However, besides the long breath-hold of 16 heartbeats, the MRF technique adopts highly undersampled spiral trajectory to sample the signal evolution and requires complex iterative reconstruction before dictionary matching to remove aliasing artifacts. Recently, the feasibility of single-shot Cartesian dictionary-based mapping has been demonstrated for myocardial T1 and T2 mapping, where only 10 time-points along the magnetization evolution were sampled. Cartesian dictionary-based mapping involves simplified post-processing and is more ready for clinical translation. However, the current Cartesian dictionary-based T1 and T2 mapping technique still requires breath-hold to mitigate respiratory motion and assumes homogeneous B1 across the left ventricle. Adding more parameters to be quantified would lead to increased scan time that may exceed the limit of breath-hold duration. Furthermore, there may be B1 inhomogeneities, especially at high field scanners. If not accounted for, the inhomogeneous B1 will influence the accuracy of dictionary-based parameter estimations.
The objective of this study is to propose a free-breathing multi-parametric mapping (FB-MultiMap) technique that enables B1 inhomogeneity-corrected simultaneous myocardial T1, T2 and T1ρ mapping with single-shot Cartesian acquisition and dictionary matching. Inspired by previous studies that leverage the diaphragmatic navigator (dNAV) for free-breathing parameter mapping [31, 40, 41], the dNAV was adopted in FB-MultiMap for through-plane motion correction with special design to reduce the influence of preparation pulses on the dNAV signal. The residual in-plane motion was corrected using a multi-contrast group-wise registration method [42]. The effects of spatial inhomogeneities of the radiofrequency transmit field, slice profile and respiratory motion-induced deviations between the measured signal and the simulated signal evolutions were modelled as an effective B1 factor in the dictionary of FB-MultiMap to achieve B1-corrected multi-parametric mapping. The proposed technique was optimized and validated in numerical simulations, phantoms, healthy subjects and patients with suspected cardiomyopathy.

Methods

An overview of the proposed free-breathing multi-parametric mapping technique is provided in Fig. 1, which includes single-shot acquisition with the optimized FB-MultiMap sequence, multi-contrast image registration, dictionary generation with Bloch simulations and T1, T2, T1ρ and B1 mapping with dictionary matching. Details are provided in the following sections.

Sequence optimization

The proposed free-breathing multi-parametric mapping sequence performs electrocardiogram-triggered single-shot Cartesian acquisition with 2D bSSFP readout including five ramp-up pulses to reach steady-state and linear k-space ordering, where inversion recovery (IR), T2 preparation (T2-prep) and T1ρ preparation (T1ρ-prep) pulses are included for T1, T2 and T1ρ sensitization, respectively. The T2-prep includes hard 90° tip-down and tip-up pulses and two adiabatic refocusing pulses [24]. The T1ρ-prep module adopts the 90°x-SLy-SL-y-180°-x-SL-y-SLy-90°x pulses cluster, where the SL pulse is equally divided into four segments. As shown in the previous T1ρ-prep optimization studies [43, 44], this T1ρ-prep is robust to B1 and B0 field inhomogeneities and performs better than the totally balanced spin-lock module with two 180° refocusing pulses [14, 45].
The accuracy and precision of quantitative measurements can be affected by the number and timings of preparation pulses and an exhaustive search for the optimal design may be intractable. Several candidate sequences with different preparation pulse configurations were empirically designed for investigation. Details of sequence optimization are provided in the Additional file 1.
FB-MultiMap adopts the bSSFP readout with large flip angles, and the inhomogeneous transmit B1 field, if not accounted for, would influence the dictionary-based mapping accuracy [36, 46]. However, with the Look-Locker type acquisition, the effect of excitation pulses with constant flip angle and T1 relaxation on the inversion recovered signal cannot be reliably separated [47]. Therefore, we proposed to change the flip angle after IR [48] and estimate the B1 factor along with T1, T2 and T1ρ to correct the nominal flip angle. To determine appropriate flip angle combinations, numerical simulations, phantom and in vivo experiments were conducted. In simulations, the flip angle was varied from 35° to 70° with a step of 5°. The target T1, T2, T1ρ and B1 were set to 1200 ms, 40 ms, 50 ms and 0.8, respectively. The relative root mean squared error (RRMSE) for T1, T2 and T1ρ, and the mean RRMSE of the three parameters were computed to evaluate the performance of FB-MultiMap with different flip angle combinations in simulation and phantom studies. The calculation equation of RRMSE is provided in the Additional file 1. For in vivo imaging, the relative spatial variability (RSD) was calculated by dividing the standard deviation (SD) with the mean value in the myocardium for T1, T2 and T1ρ.

Motion correction

For respiratory motion correction, the cross-pair diaphragm navigator was utilized to adjust the imaging plane in real-time to correct the through-plane motion with an empirical tracking factor of 0.6 [41, 49]. dNAV based respiratory motion correction has previously been adopted for free-breathing myocardial mapping [31, 40]. The preparation pulses in FB-MultiMap are all non-selective. To reduce the influence of IR on the dNAV signal, a slice-selective IR was performed covering the imaging location of dNAV along with the non-selective IR to restore the magnetization in the right diaphragm region. Also, the dNAV was applied before T2-prep and T1ρ-prep to avoid their influence. After image acquisition, a group-wise registration method proposed to correct motion for MOLLI T1 mapping [50] was adopted to correct in-plane motion among the multi-contrast images.

Parametric mapping

Dictionary matching was performed for parametric mapping. The subject-specific dictionary was generated with recorded R-wave intervals and trigger delays using the Bloch equation for ranges of T1, T2, T1ρ and B1. To evaluate the effect of correcting B1, FB-MultiMap with simplified post-processing (sFB-MultiMap) without modelling B1 in the dictionary simulation was also performed. Different from MRF which simulates the magnetization for each readout, for the Cartesian sampling, only one image was acquired per cardiac cycle with the contrast dominated by the k-space center and therefore only the transverse magnetization simulated corresponding to the center of k-space was used for the dictionary. Considering that online reconstructed images are usually magnitude images without phase information, the magnitude of the simulated magnetization was recorded for the dictionary to avoid determining the polarity of the magnitude images. In our initial experiments, we found dictionary matching with and without polarity achieved similar accuracy and precision for the proposed multi-parametric mapping technique.

Phantom experiments

The proposed FB-MultiMap sequence was implemented and evaluated on a 3T United Imaging scanner (uMR 890, United Imaging Healthcare, Shanghai, China) with 12-channel body and 48-channel spine coils. The phantoms with T1, T2 and T1ρ corresponding to physiological ranges were made of different concentrations of agarose and gadolinium-based contrast agent. Reference T1 values were obtained using IR spin echo with 11 inversion recovery times from 35 to 3000 ms and TR of 7000 ms. Reference T2 values were measured using multi-echo spin echo with 10 TEs ranging from 14 to 140 ms and TR of 7000 ms. Reference T1ρ values were estimated using a T1ρ-prep gradient echo technique [51] with spin lock time (TSL) = [2, 16, 30, 50, 80] ms and spin-lock frequency = 350 Hz. The reference images were acquired with 2 × 2 mm2 spatial resolution and 8 mm slice thickness. The multi-parametric mapping scans were performed with simulated heart rates from 20 to 120 bpm with 20 bpm increments, to test heart rate dependency. Other imaging parameters were: field-of-view (FOV) = 320 × 280 mm2; pixel size = 2.08 × 1.67 mm2; slice thickness = 8 mm; TR/TE = 2.82 ms/1.33 ms; readout bandwidth = 1200 Hz/pixel; inversion recovery time = 255 ms; generalized autocalibrating partially parallel acquisition (GRAPPA) [52] parallel imaging acceleration = 2, resulting in 77 lines acquired per heartbeat.
To reduce computation cost, the dictionary in the phantom studies was generated considering the actual parameters ranges. Specifically, the dictionary contained approximately 6,037,000 T1/T2/T1ρ/B1 combinations: [10:10:1800] ms for T1; [10:2:160] for T2 and T1ρ; [0.4:0.05:0.55, 0.55:0.02:0.8, 0.8:0.05:1.2] for B1. Combinations of T2 > T1 and T1ρ > T1 were excluded from the dictionary simulation. The Bloch simulations were implemented using MATLAB (The MathWorks, Natick, Massachusetts) and the dictionary generation took about 6 min on a workstation (Intel Xeon Gold 6226R 2.9 GHz processor, 384 Gb RAM) with parallel computing.

In vivo experiments

The in vivo imaging was approved by the local institutional review board. Fifteen healthy subjects (6 females, age: 26 ± 3.3 years, heart rate: 71 ± 7.7 bpm) and 6 patients with suspected cardiomyopathy (1 females, age: 58 ± 19.9 years, heart rate: 72.5 ± 15.3 bpm) were imaged after obtaining written informed consent. The demographic information and referral reasons of the patients are summarized in Table 1.
Table 1
Demographic information of healthy subjects and patients
 
n
Age (year)
Height (cm)
Weight (kg)
BMI
Volunteers
15 (F = 6)
26 ± 3.3
171 ± 6.7
69 ± 11
24 ± 2.4
Patients
6 (F = 1)
58 ± 19.9
170 ± 6.0
66 ± 13
23 ± 4.5
Patient referral reasons
Fabry disease
3
Hypertensive cardiomyopathy
1
Old myocardial infarction
1
Cardiac amyloidosis
1
The optimized FB-MultiMap was evaluated against the breath-hold separate mapping methods: MOLLI 5(3)3 for T1 mapping [21], T2-prep bSSFP for T2 mapping [24] and T1ρ-prep bSSFP for T1ρ mapping [51]. T2-prep bSSFP acquired three T2-weighted images with T2-prep durations = [0, 35, 55] ms, while T1ρ-prep bSSFP acquired four T1ρ-weighted images with TSL = [2, 16, 30, 50] ms and spin-lock frequency of 350 Hz. Both T2-prep bSSFP and T1ρ-prep bSSFP employed three idle cardiac cycles between each readout for signal recovery. Furthermore, to assess any respiratory motion induced parameter estimation variability of FB-MultiMap, the optimized multi-parameter mapping sequence was also performed under breath-holding in the healthy subjects, who are able to hold their breath for a relatively long time. Three short-axis slices at basal, middle and apical left ventricle were acquired for all mapping techniques. Other imaging parameters, including the matrix size, slice thickness and readout bandwidth, were the same to those of FB-MultiMap in phantom studies. The FOV in the phase-encoding direction was adjusted according to the subject’s size in the range of 240–280 mm. In addition to the mapping scans, the late gadolinium enhancement (LGE) imaging was performed in the patients after the mapping acquisitions to detect any focal enhancement. The breath-hold T1ρ-prep bSSFP was not included in the patients imaging protocol to reduce the number of breath-holds.
The dictionary generated for in vivo studies contained about 2,330,000 T1/T2/T1ρ/B1 combinations: [500:100:800, 810:10:1700, 1710:100:2000] ms for T1, [10:10:30, 31:2:70, 71:5:91, 91:10:121] for T2 and T1ρ, and [0.4:0.05:0.5, 0.5:0.03:1.1, 1.1:0.05:1.2] for B1. It took about 1.5 min for the subject-specific dictionary generation and 10 s for extracting parametric maps per slice using dictionary matching.

Image analysis

For phantoms, the mean and SD were calculated for each tube. The estimations of FB-MultiMap and sFB-MultiMap were then respectively compared with the reference methods using Pearson correlation. Bland–Altman analysis was also conducted to evaluate the agreement of FB-MultiMap with the reference methods. For in vivo images, region-of-interests were manually defined in the T1, T2 and T1ρ maps according to the 16-segment of the American Heart Association (AHA) model [53]. Subsequently, the mean and SD were calculated for each short-axis slice to evaluate the estimation accuracy and precision in healthy subjects. The slice-wise mean and SD were compared between the four measurement methods including separate breath-hold mapping, breath-hold simultaneous multi-parametric mapping (BH-MultiMap), FB-MultiMap and sFB-MultiMap using one-way ANOVA (One-way analysis of variance) with Bonferroni post-hoc correction. p < 0.05 is considered statistically significant.

Results

Sequence optimization

The optimized FB-MultiMap is demonstrated in Fig. 1, where the IR pulses are played at the first and the ninth cardiac cycles; T2-preps are played in the fifth to eighth cardiac cycles with durations = [35, 45, 55, 65] ms; T1ρ-prep with TSLs = [16, 30, 40, 50] ms are played in the 13th to 16th cardiac cycles. Detailed sequence optimization results are provided in the Additional file 1.
For the variable flip angle strategy, the dual flip angle scheme was initially tested with the excitation flip angle changed after the second IR (FA1/FA2 in Fig. 2A). However, as shown in the RRMSE maps in Fig. 2B, no FA1/FA2 combinations resulted in the lowest RRMSE for all parameters, with T1 favoring 45°/70°, while T2 and T1ρ favoring low FA1 of 35° and moderate FA2 of 50°. Therefore, we subsequently proposed to also alter the flip angles after the first T2-prep and the first T1ρ-prep pulse, resulting in a four-flip-angle strategy, FA1-1, FA1-2, FA2-1 and FA2-2 in Fig. 2A. With the observation that T1 and B1 RRMSE were mostly influenced by FA1-1 and FA2-1, while T2 and T1ρ RRMSE was respectively more influenced by FA1-2 and FA2-2, the appropriate four flip angle combination can be inferred from the dual flip angle results with FA1-1 (45°) and FA2-1 (70°) being the optimal flip angles for T1 and B1 and FA1-2 (35°) and FA2-2 (50°) being the optimal flip angles for T2 and T1ρ. The inferred four-flip-angle strategy (45°/35°/70°/50°) avoids the brute-force search for the optimal combination of four flip angles that would require prohibitively long computation time. Figure 3 shows example T1, T2, T1ρ and B1 maps of a healthy subject estimated with FB-MultiMap with different flip angle combinations. The variable flip angle schemes resulted in smoother B1 maps than the constant flip angle scheme, and the optimized four flip angle scheme outperformed the dual flip angle scheme by achieving lower spatial variability (RSD) for all three parameters. It is noted that B1 in ventricular blood pool cannot be reliably estimated due to blood flowing. Additional results of simulations and phantoms regarding variable flip angle optimization are included in the Additional file 1.

Phantom experiments

Example phantom T1, T2 and T1ρ maps estimated with FB-MultiMap at heart rate of 80 bpm are shown in Fig. 4A. FB-MultiMap estimations were overall consistent for heart rates ranging from 40 to 120 bpm (Fig. 4B) and showed strong correlation with the reference values (correlation coefficients R2 > 0.99 for all parameters, Fig. 4C). Compared with the reference, the proposed method tends to underestimate longer T1 (> 1400 ms) and slightly overestimates longer T1ρ (> 110 ms). Without B1 correction, sFB-MultiMap caused significant deviation of measured T2 from the reference method and overestimated longer T1ρ. T1 estimations with sFB-MultiMap were also slightly degraded compared with FB-MultiMap. Bland–Altman analyses (Fig. 4D) reveal that the T1 values estimated using the proposed method were a bit lower than the reference (T1 bias = − 71.9 ms, upper 95% limits of agreement = 73.6 ms, lower 95% limits of agreement = − 217.4 ms), whereas T2 and T1ρ values showed better agreement (T2: bias = 1.8 ms, upper 95% limits of agreement = 6.7 ms, lower 95% limits of agreement = − 3.1 ms; T1ρ: bias = 3.9 ms, upper 95% limits of agreement = 13.2 ms, lower 95% limits of agreement = − 5.4 ms).

In vivo imaging

Example T1, T2, and T1ρ maps at three short-axis slices of a representative healthy subject are provided in Fig. 5, where good mapping quality can be observed for FB-MultiMap without discernable motion artifacts compared with conventional mapping methods and BH-MultiMap. Like the phantom results, sFB-MultiMap overestimated T2. The segment-wise analysis results are visualized in Fig. 6 using bullseye plots, where the mean and SD for each segment were calculated by averaging across the 15 healthy subjects. The mapping quality was similar between BH- and FB-MultiMap with overall homogeneous parametric maps while sFB-MultiMap resulted in increased T2 measurement variability compared to FB-MultiMap. The mean and SD of myocardial T1, T2 and T1ρ values of each short-axis slice in the healthy subjects are compared between the four mapping methods in Figs. 7 and 8. There is no significant difference of parameter estimations between BH- and FB-MultiMap for all short-axis slices, while the mean and SD of FB-MultiMap T1 were significantly higher than MOLLI (mean T1 at mid-cavity: 1218 ± 50 ms vs. 1166 ± 38 ms, p < 0.001). The mean T2 and T1ρ estimated with FB-MultiMap was respectively lower than T2-prep bSSFP and T1ρ-prep bSSFP (mean T1ρ at mid-cavity: 45.3 ± 4.4 ms vs. 50.2 ± 4.0, p < 0.001), whereas the difference between FB-MultiMap and T2-prep bSSFP was not significant at the middle short-axis slice (41.2 ± 2.8 ms vs. 42.5 ± 3.1 ms, p = 0.06). The T2 SD of sFB-MultiMap was significantly higher than the other three measurement methods (p < 0.001).
Figure 9 shows the parametric maps and slice-matched LGE images of two patients: a 79-year-old female with hypertensive cardiomyopathy and intramural LGE in the antero-septal segment; a 71-year-old male patient with subendocardial myocardial infarction in the basal and mid-cavity inferoseptal segments. For Patient #1, elevated T1, T2 and T1ρ can be observed in the enhancement area (red contour) compared with the remote myocardium (green contour). In this patient, the T1 and T2 values in the remote myocardium were also higher than the healthy subjects, which is not the case for T1ρ, indicating there may be non-fibrotic myocardial abnormality in the remote myocardium. For Patient #2, the T1, T2 and T1ρ values were also increased in the scar region compared with the remote myocardium. This patient was unable to hold his breath well, leading to motion artifacts and motion blurring in the breath-hold LGE images and parameter maps, while the proposed free-breathing multi-parametric mapping technique can still generate good quality parameter maps. The alterations of parameters in the diseased myocardium were consistent between FB-MultiMap and conventional mapping methods.
The images of two male patients diagnosed with Fabry disease, aged 37 and 31 years respectively, are demonstrated in Fig. 10. No significant LGE was observed in the two subjects. In Patient #3, except for the basal inferoseptal and middle inferior, the myocardial T1 values were ~ 100 ms lower compared with the average of healthy subjects for both FB-MultiMap and MOLLI. Myocardial T1 values were also significantly shorter as measured with MOLLI and FB-MultiMap in the second Fabry patient (Fig. 10B). The mean T2 and T1ρ of the two Fabry patients were similar to values of healthy subjects in this study. The FB-MultiMap images of the two additional patients with one diagnosed as Fabry cardiomyopathy and another diagnosed as amyloid cardiomyopathy are provided in the Additional file 1.

Discussion

In this study, a free-breathing multi-parametric mapping technique was proposed for simultaneous myocardial T1, T2, and T1ρ mapping with B1 correction using a single-shot Cartesian acquisition and dictionary matching. The sequence was optimized for the number of preparation pulses, their timing, and variable excitation flip angles in order to obtain accurate T1, T2 and T1ρ estimations in the presence of B1 in homogeneities. In phantoms, the optimized sequence showed good agreement with reference measurements with no discernable dependency on heart rate. In healthy subjects, the free-breathing multi-parametric mapping quality was comparable to standard breath-hold mapping techniques. In preliminary validations in patients, the FB-MultiMap showed promising results for detecting various cardiomyopathies without contrast agent.
The single-shot Cartesian dictionary-based mapping technique has been recently been proposed for simultaneous cardiac T1 and T2 mapping under a breath-hold of ten heartbeats [36]. However, extending this technique to include more relaxation parameters is not straightforward. Firstly, more cardiac cycles are needed for additional parameter sensitization and the scan time may be too long for breath holding. In the proposed technique, a free-breathing acquisition was achieved with dNAV for prospective motion correction and multi-contrast image registration for retrospective motion correction, breaking the limit of breath-hold duration. Secondly, in addition to IR and T2-prep, the T1ρ-prep should be added for T1ρ sensitization, which requires further modification of the configuration of preparation pulses to ensure estimation accuracy and precision. Several candidate sequences with different IR, T2- and T1ρ-prep settings were investigated. The sequence with acquisition of 16 heartbeats (2 IRs, 4 recovery heartbeats between IR and T2/T1ρ-prep, and 4 T2/T1ρ-prep acquisitions) achieved overall lower RRMSE for all parameters in simulations and phantoms and the lowest spatial variability for in vivo parameter maps which indicates better accuracy and precision. Furthermore, the previously proposed Cartesian dictionary-based mapping technique [36] assumed homogeneous B1 across the left ventricle at 1.5T which is not true for 3T. According to our results, the effective B1 was much lower in the septum than in the lateral region. The non-uniformB1 factors, if not considered in dictionary simulations, leaded to inaccurate parameter estimations, especially for T2. The variable flip angle strategy was proposed with flip angles optimized in simulations, phantoms and in vivo imaging, achieving B1 corrected T1, T2 and T1ρ measurements. Similar to MRF, FB-MultiMap generates parameter maps based on dictionary matching. However, the proposed technique involves no complex image reconstruction to remove aliasing artifacts as required in MRF and thus may be more ready adopted in clinical settings for T1, T2 and T1ρ mapping. In addition, most cardiac MRF techniques require breath-holding, including the recent cardiac T1, T2 and T1ρ MRF [39], while the proposed technique allows for free-breathing acquisition. Compared with conventional mapping techniques, such as MOLLI and T2/T1ρ-prep bSSFP, which assumes an analytical model for parameter fitting, the dictionary based multi-parametric mapping method, such as FB-MultiMap, is advantageous in that it requires no empty cardiac cycles for signal recovery, being less dependent on heart rate, and could consider inter-parameter confounders in the signal model, leading to improved quantification accuracy. However, the subject-specific dictionary generation can be time-consuming. To reduce the computation time, simplified Bloch simulations were adopted in the dictionary generation with the B1 factor correcting the excitation flip angle only, without considering imperfect slice profile and preparation pulses. These confounders may be considered for improved accuracy at the expense of significantly increased dictionary simulation times [54, 55].
Similar to MOLLI and T2-prep bSSFP, 2D bSSFP readout was adopted in FB-MultiMap for the benefit signal-to-noise ratio. However, it is noted that bSSFP is sensitive to B0 inhomogeneity which may influence parameter estimation. The influence of off-resonance on the parameter estimations of FB-MultiMap was investigated in the Additional file 1 (Section 4). The off-resonance within ± 100 Hz has little influence on T1, T2 and T1ρ estimation, while B1 is more vulnerable to the B0 inhomogeneity. When the off-resonance exceeds the range of ± 50 Hz, B1 tends to be overestimated. Seeing from the typical in vivo B0 field maps in the Additional file 1: Fig. S7, the B0 offset in the heart region is well below 100 Hz, while in the liver and chest, it could exceed the range of ± 150 Hz. The large off-resonance leads to erroneously high B1 estimations in the non-Cardiac region. Considering the cardiac T1, T2 and T1ρ estimations with FB-MultiMap are robust to off-resonance within ± 100 Hz which can be achieved for most commercial 3T scanners with proper B0 shimming in the heart, the off-resonance was not modelled in the dictionary simulation. The spoiled gradient echo readout is more robust to non-uniform B0, albeit with reduced signal-to-noise ratio. Its feasibility for the multi-parametric mapping technique is worth exploring in future studies.
In phantoms, the proposed FB-MultiMap achieved good accuracy for short to moderate relaxation times. However, it tended to underestimate T1. The underestimation was caused by the uncorrected inversion efficiency in FB-MultiMap. In T1 fitting of IR spin echo, the three-parameter model was used to consider the imperfect inversion. However, in the dictionary simulation of FB-MultiMap, the inversion efficiency was assumed to be 100%, which leaded to the T1 underestimation of FB-MultiMap. FB-MultiMap overestimated longer T1ρ (> 110 ms). Increasing the TSL in T1ρ-prep may help to improve FB-MultiMap estimation accuracy for longer T1ρ.
In healthy subjects, FB-MultiMap achieved similar mapping quality to BH-MultiMap, indicating that after motion correction, respiratory motion has little influence on parameter quantification of FB-MultiMap. The FB-MultiMap T1 was higher than MOLLI, which may suggest improved T1 accuracy of FB-MultiMap as MOLLI is known to underestimate T1 [22, 56]. However, compared with MOLLI, the SD of and BH- and FB-MultiMap T1 was higher. As can be seen in the measurement homogeneity analysis results in the Additional file 1 (Section 5), the septal and inferior T1 was about 30 ms longer compared to the lateral and anterior T1 for both MOLLI and the multi-parametric mapping methods. This difference is likely caused by the non-uniform inversion efficiency. For T1 mapping with MOLLI and FB-MultiMap, the inversion efficiency was assumed to be evenly 100% across the left ventricle myocardium. However, the inversion efficiency can be different due to the non-uniform transmit B1, leading to the slightly different T1 estimations in different myocardial segments. Optimizing the inversion pulse design using for example, the tangent/hyperbolic tangent adiabatic pulse [57] can mitigate the problem of non-uniform inversion efficiency in the presence of inhomogeneous transmitting B1. More details about the influence of inversion efficiency on FB-MultiMap are provided in the Additional file 1 (Section 6). The myocardial T2 and T1ρ values measured by the proposed technique were lower compared to conventional T2 and T1ρ mapping methods that adopt two-parameter fitting, while it is known that the latter methods tend to overestimate T2 and T1ρ due to unaccounted T1 relaxation during readout [58, 59].
Cardiac diseases are usually complex, involving more than one type of pathological change in myocardium. There is where multi-parametric mapping techniques could come into play to aid in diagnosis with comprehensive myocardial tissue characterization. The feasibility of FB-MultiMap in detecting cardiomyopathies has been demonstrated in several patients. The pathological changes detected with FB-MultiMap agreed with conventional breath-hold mapping methods. Specifically, increased T1 and T1ρ values were observed for all the LGE-positive patients in the enhancement area, indicating their potential for detecting myocardial fibrosis without contrast agent. The deposition of lysosomal sphingolipids in Fabry disease leads to low T1 values, which were detected by both FB-MultiMap and MOLLI. In one Fabry patient (Fig. 10A), the basal inferoseptal and middle inferior segments had exceptionally longer T1 (1253 ms) when compared to other segments. There was no LGE in the long T1 regions, while much higher T2 (48.8 ms) and slightly increased T1ρ (47.2 ms) were observed, which may indicate inflammation and early fibrosis. The fibrosis and inflammation may alter the T1 measurement in Fabry disease as reported in previous studies [59, 60]. T1ρ mapping in Fabry disease has not been reported before. FB-MultiMap detected no marke alterations in T1ρ in the three Fabry patients. However, this finding is inconclusive due to the small sample size, and more Fabry patients are needed to validate this point.

Limitations

There are several limitations in this study. Although efforts were made to optimize the variable flip angle three-parameter mapping technique, the optimization was limited to several empirically designed candidate sequences. Exploring more complex preparation pulse configurations and variable flip angle schemes may help to further improve accuracy and precision. The diaphragm navigator based through-plane motion correction was performed with fixed slice-tracking factor. Adopting a patient-specific motion model trained in a calibration scan to characterize the heart motion relative to dNAV [61] may help to further improve motion correction, especially for apical slices. The proposed technique involves subject-specific dictionary generation which is computationally demanding. Deep learning neural networks may be explored to reduce post-processing time. Finally, this study focused on technical development and only a small number of patients were included for proof-of-concept. Further validation in more patients with cardiovascular diseases is warranted to assess the clinical value of the proposed technique.

Conclusions

The proposed approach enables simultaneous native myocardial T1, T2 and T1ρ mapping with B1 correction using single-shot Cartesian acquisition and dictionary matching during free breathing. FB-MultiMap demonstrated good agreement with reference techniques in phantoms and achieved similar mapping quality to conventional breath-hold mapping techniques. The multi-parametric mapping technique also demonstrated promising results in detecting pathological changes in T1, T2 and T1ρ. With efficient free-breathing acquisition and simple post-processing, FB-MultiMap has great potential for clinical applications.

Acknowledgements

Not applicable.

Declarations

The study was approved by the institutional review board at ShanghaiTech University, Shanghai, China. All participants provided written informed consent.
Not applicable.

Competing interests

Jian Xu is an employee of United Imaging Healthcare. All the other authors declare that they have no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Messroghli DR, Moon JC, Ferreira VM, Grosse-Wortmann L, He T, Kellman P, et al. Clinical recommendations for cardiovascular magnetic resonance mapping of T1, T2, T2* and extracellular volume: a consensus statement by the Society for Cardiovascular Magnetic Resonance (SCMR) endorsed by the European Association for Cardiovascular Imaging (EACVI). J Cardiovasc Magn Reson. 2017;19(1):75.PubMedPubMedCentralCrossRef Messroghli DR, Moon JC, Ferreira VM, Grosse-Wortmann L, He T, Kellman P, et al. Clinical recommendations for cardiovascular magnetic resonance mapping of T1, T2, T2* and extracellular volume: a consensus statement by the Society for Cardiovascular Magnetic Resonance (SCMR) endorsed by the European Association for Cardiovascular Imaging (EACVI). J Cardiovasc Magn Reson. 2017;19(1):75.PubMedPubMedCentralCrossRef
2.
Zurück zum Zitat Fotaki A, Velasco C, Prieto C, Botnar RM. Quantitative MRI in cardiometabolic disease: from conventional cardiac and liver tissue mapping techniques to multi-parametric approaches. Front Cardiovasc Med. 2022;9: 991383.PubMedCrossRef Fotaki A, Velasco C, Prieto C, Botnar RM. Quantitative MRI in cardiometabolic disease: from conventional cardiac and liver tissue mapping techniques to multi-parametric approaches. Front Cardiovasc Med. 2022;9: 991383.PubMedCrossRef
3.
Zurück zum Zitat Warnica W, Al-Arnawoot A, Stanimirovic A, Thavendiranathan P, Wald RM, Pakkal M, et al. Clinical impact of cardiac MRI T1 and T2 parametric mapping in patients with suspected cardiomyopathy. Radiology. 2022;305(2):319–26.PubMedCrossRef Warnica W, Al-Arnawoot A, Stanimirovic A, Thavendiranathan P, Wald RM, Pakkal M, et al. Clinical impact of cardiac MRI T1 and T2 parametric mapping in patients with suspected cardiomyopathy. Radiology. 2022;305(2):319–26.PubMedCrossRef
4.
Zurück zum Zitat Ferreira VM, Schulz-Menger J, Holmvang G, Kramer CM, Carbone I, Sechtem U, et al. Cardiovascular magnetic resonance in nonischemic myocardial inflammation: expert recommendations. J Am Coll Cardiol. 2018;72(24):3158–76.PubMedCrossRef Ferreira VM, Schulz-Menger J, Holmvang G, Kramer CM, Carbone I, Sechtem U, et al. Cardiovascular magnetic resonance in nonischemic myocardial inflammation: expert recommendations. J Am Coll Cardiol. 2018;72(24):3158–76.PubMedCrossRef
5.
Zurück zum Zitat Diao KY, Yang ZG, Xu HY, Liu X, Zhang Q, Shi K, et al. Histologic validation of myocardial fibrosis measured by T1 mapping: a systematic review and meta-analysis. J Cardiovasc Magn Reson. 2016;18(1):92.PubMedPubMedCentralCrossRef Diao KY, Yang ZG, Xu HY, Liu X, Zhang Q, Shi K, et al. Histologic validation of myocardial fibrosis measured by T1 mapping: a systematic review and meta-analysis. J Cardiovasc Magn Reson. 2016;18(1):92.PubMedPubMedCentralCrossRef
6.
Zurück zum Zitat Puntmann VO, Carr-White G, Jabbour A, Yu CY, Gebker R, Kelle S, et al. T1-Mapping and outcome in nonischemic cardiomyopathy: all-cause mortality and heart failure. JACC Cardiovasc Imaging. 2016;9(1):40–50.PubMedCrossRef Puntmann VO, Carr-White G, Jabbour A, Yu CY, Gebker R, Kelle S, et al. T1-Mapping and outcome in nonischemic cardiomyopathy: all-cause mortality and heart failure. JACC Cardiovasc Imaging. 2016;9(1):40–50.PubMedCrossRef
7.
Zurück zum Zitat Hanneman K, Nguyen ET, Thavendiranathan P, Ward R, Greiser A, Jolly MP, et al. Quantification of myocardial extracellular volume fraction with cardiac MR imaging in thalassemia major. Radiology. 2016;279(3):720–30.PubMedCrossRef Hanneman K, Nguyen ET, Thavendiranathan P, Ward R, Greiser A, Jolly MP, et al. Quantification of myocardial extracellular volume fraction with cardiac MR imaging in thalassemia major. Radiology. 2016;279(3):720–30.PubMedCrossRef
8.
Zurück zum Zitat Karur GR, Robison S, Iwanochko RM, Morel CF, Crean AM, Thavendiranathan P, et al. Use of myocardial T1 mapping at 3.0 T to differentiate Anderson-Fabry disease from hypertrophic cardiomyopathy. Radiology. 2018;288(2):398–406.PubMedCrossRef Karur GR, Robison S, Iwanochko RM, Morel CF, Crean AM, Thavendiranathan P, et al. Use of myocardial T1 mapping at 3.0 T to differentiate Anderson-Fabry disease from hypertrophic cardiomyopathy. Radiology. 2018;288(2):398–406.PubMedCrossRef
9.
Zurück zum Zitat Anan I, Sakuma T, Fukuro E, Morimoto S, Nojiri A, Kawai M, et al. The role of native T1 values on the evaluation of cardiac manifestation in Japanese Fabry disease patients. Mol Genet Metab Rep. 2022;31: 100858.PubMedPubMedCentral Anan I, Sakuma T, Fukuro E, Morimoto S, Nojiri A, Kawai M, et al. The role of native T1 values on the evaluation of cardiac manifestation in Japanese Fabry disease patients. Mol Genet Metab Rep. 2022;31: 100858.PubMedPubMedCentral
10.
Zurück zum Zitat O’Brien AT, Gil KE, Varghese J, Simonetti OP, Zareba KM. T2 mapping in myocardial disease: a comprehensive review. J Cardiovasc Magn Reson. 2022;24(1):33.PubMedPubMedCentralCrossRef O’Brien AT, Gil KE, Varghese J, Simonetti OP, Zareba KM. T2 mapping in myocardial disease: a comprehensive review. J Cardiovasc Magn Reson. 2022;24(1):33.PubMedPubMedCentralCrossRef
11.
Zurück zum Zitat Muthupillai R, Flamm SD, Wilson JM, Pettigrew RI, Dixon WT. Acute myocardial infarction: tissue characterization with T1rho-weighted MR imaging–initial experience. Radiology. 2004;232(2):606–10.PubMedCrossRef Muthupillai R, Flamm SD, Wilson JM, Pettigrew RI, Dixon WT. Acute myocardial infarction: tissue characterization with T1rho-weighted MR imaging–initial experience. Radiology. 2004;232(2):606–10.PubMedCrossRef
12.
Zurück zum Zitat Witschey WR, Zsido GA, Koomalsingh K, Kondo N, Minakawa M, Shuto T, et al. In vivo chronic myocardial infarction characterization by spin locked cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2012;14(1):37.PubMedPubMedCentralCrossRef Witschey WR, Zsido GA, Koomalsingh K, Kondo N, Minakawa M, Shuto T, et al. In vivo chronic myocardial infarction characterization by spin locked cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2012;14(1):37.PubMedPubMedCentralCrossRef
13.
14.
Zurück zum Zitat Qi H, Bustin A, Kuestner T, Hajhosseiny R, Cruz G, Kunze K, et al. Respiratory motion-compensated high-resolution 3D whole-heart T1rho mapping. J Cardiovasc Magn Reson. 2020;22(1):12.PubMedPubMedCentralCrossRef Qi H, Bustin A, Kuestner T, Hajhosseiny R, Cruz G, Kunze K, et al. Respiratory motion-compensated high-resolution 3D whole-heart T1rho mapping. J Cardiovasc Magn Reson. 2020;22(1):12.PubMedPubMedCentralCrossRef
15.
Zurück zum Zitat Bustin A, Toupin S, Sridi S, Yerly J, Bernus O, Labrousse L, et al. Endogenous assessment of myocardial injury with single-shot model-based non-rigid motion-corrected T1rho mapping. J Cardiovasc Magn Reson. 2021;23(1):119.PubMedPubMedCentralCrossRef Bustin A, Toupin S, Sridi S, Yerly J, Bernus O, Labrousse L, et al. Endogenous assessment of myocardial injury with single-shot model-based non-rigid motion-corrected T1rho mapping. J Cardiovasc Magn Reson. 2021;23(1):119.PubMedPubMedCentralCrossRef
16.
Zurück zum Zitat Stoffers RH, Madden M, Shahid M, Contijoch F, Solomon J, Pilla JJ, et al. Assessment of myocardial injury after reperfused infarction by T1rho cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2017;19(1):17.PubMedPubMedCentralCrossRef Stoffers RH, Madden M, Shahid M, Contijoch F, Solomon J, Pilla JJ, et al. Assessment of myocardial injury after reperfused infarction by T1rho cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2017;19(1):17.PubMedPubMedCentralCrossRef
17.
Zurück zum Zitat Yin Q, Abendschein D, Muccigrosso D, O’Connor R, Goldstein T, Chen R, et al. A non-contrast CMR index for assessing myocardial fibrosis. Magn Reson Imaging. 2017;42:69–73.PubMedPubMedCentralCrossRef Yin Q, Abendschein D, Muccigrosso D, O’Connor R, Goldstein T, Chen R, et al. A non-contrast CMR index for assessing myocardial fibrosis. Magn Reson Imaging. 2017;42:69–73.PubMedPubMedCentralCrossRef
18.
Zurück zum Zitat van Oorschot JW, Guclu F, de Jong S, Chamuleau SA, Luijten PR, Leiner T, et al. Endogenous assessment of diffuse myocardial fibrosis in patients with T1rho -mapping. J Magn Reson Imaging. 2017;45(1):132–8.PubMedCrossRef van Oorschot JW, Guclu F, de Jong S, Chamuleau SA, Luijten PR, Leiner T, et al. Endogenous assessment of diffuse myocardial fibrosis in patients with T1rho -mapping. J Magn Reson Imaging. 2017;45(1):132–8.PubMedCrossRef
19.
Zurück zum Zitat Thompson EW, Kamesh Iyer S, Solomon MP, Li Z, Zhang Q, Piechnik S, et al. Endogenous T1rho cardiovascular magnetic resonance in hypertrophic cardiomyopathy. J Cardiovasc Magn Reson. 2021;23(1):120.PubMedPubMedCentralCrossRef Thompson EW, Kamesh Iyer S, Solomon MP, Li Z, Zhang Q, Piechnik S, et al. Endogenous T1rho cardiovascular magnetic resonance in hypertrophic cardiomyopathy. J Cardiovasc Magn Reson. 2021;23(1):120.PubMedPubMedCentralCrossRef
20.
Zurück zum Zitat Sanchez Tijmes F, Thavendiranathan P, Udell JA, Seidman MA, Hanneman K. Cardiac MRI assessment of nonischemic myocardial inflammation: State of the art review and update on myocarditis associated with COVID-19 vaccination. Radiol Cardiothorac Imaging. 2021;3(6): e210252.PubMedPubMedCentralCrossRef Sanchez Tijmes F, Thavendiranathan P, Udell JA, Seidman MA, Hanneman K. Cardiac MRI assessment of nonischemic myocardial inflammation: State of the art review and update on myocarditis associated with COVID-19 vaccination. Radiol Cardiothorac Imaging. 2021;3(6): e210252.PubMedPubMedCentralCrossRef
21.
Zurück zum Zitat Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP. Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magn Reson Med. 2004;52(1):141–6.PubMedCrossRef Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP. Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magn Reson Med. 2004;52(1):141–6.PubMedCrossRef
23.
Zurück zum Zitat Sprinkart AM, Luetkens JA, Traber F, Doerner J, Gieseke J, Schnackenburg B, et al. Gradient Spin echo (GraSE) imaging for fast myocardial T2 mapping. J Cardiovasc Magn Reson. 2015;17(1):12.PubMedPubMedCentralCrossRef Sprinkart AM, Luetkens JA, Traber F, Doerner J, Gieseke J, Schnackenburg B, et al. Gradient Spin echo (GraSE) imaging for fast myocardial T2 mapping. J Cardiovasc Magn Reson. 2015;17(1):12.PubMedPubMedCentralCrossRef
24.
Zurück zum Zitat Huang TY, Liu YJ, Stemmer A, Poncelet BP. T2 measurement of the human myocardium using a T2-prepared transient-state TrueFISP sequence. Magn Reson Med. 2007;57(5):960–6.PubMedCrossRef Huang TY, Liu YJ, Stemmer A, Poncelet BP. T2 measurement of the human myocardium using a T2-prepared transient-state TrueFISP sequence. Magn Reson Med. 2007;57(5):960–6.PubMedCrossRef
25.
Zurück zum Zitat Baessler B, Schaarschmidt F, Stehning C, Schnackenburg B, Maintz D, Bunck AC. Cardiac T2-mapping using a fast gradient echo spin echo sequence—first in vitro and in vivo experience. J Cardiovasc Magn Reson. 2015;17(1):67.PubMedPubMedCentralCrossRef Baessler B, Schaarschmidt F, Stehning C, Schnackenburg B, Maintz D, Bunck AC. Cardiac T2-mapping using a fast gradient echo spin echo sequence—first in vitro and in vivo experience. J Cardiovasc Magn Reson. 2015;17(1):67.PubMedPubMedCentralCrossRef
26.
Zurück zum Zitat Blume U, Lockie T, Stehning C, Sinclair S, Uribe S, Razavi R, et al. Interleaved T1 and T2 relaxation time mapping for cardiac applications. J Magn Reson Imaging. 2009;29(2):480–7.PubMedCrossRef Blume U, Lockie T, Stehning C, Sinclair S, Uribe S, Razavi R, et al. Interleaved T1 and T2 relaxation time mapping for cardiac applications. J Magn Reson Imaging. 2009;29(2):480–7.PubMedCrossRef
27.
Zurück zum Zitat Kvernby S, Warntjes MJ, Haraldsson H, Carlhall CJ, Engvall J, Ebbers T. Simultaneous three-dimensional myocardial T1 and T2 mapping in one breath hold with 3D-QALAS. J Cardiovasc Magn Reson. 2014;16(1):102.PubMedPubMedCentralCrossRef Kvernby S, Warntjes MJ, Haraldsson H, Carlhall CJ, Engvall J, Ebbers T. Simultaneous three-dimensional myocardial T1 and T2 mapping in one breath hold with 3D-QALAS. J Cardiovasc Magn Reson. 2014;16(1):102.PubMedPubMedCentralCrossRef
28.
Zurück zum Zitat Akcakaya M, Weingartner S, Basha TA, Roujol S, Bellm S, Nezafat R. Joint myocardial T1 and T2 mapping using a combination of saturation recovery and T2 -preparation. Magn Reson Med. 2016;76(3):888–96.PubMedCrossRef Akcakaya M, Weingartner S, Basha TA, Roujol S, Bellm S, Nezafat R. Joint myocardial T1 and T2 mapping using a combination of saturation recovery and T2 -preparation. Magn Reson Med. 2016;76(3):888–96.PubMedCrossRef
29.
Zurück zum Zitat Guo R, Chen Z, Herzka DA, Luo J, Ding H. A three-dimensional free-breathing sequence for simultaneous myocardial T1 and T2 mapping. Magn Reson Med. 2019;81(2):1031–43.PubMedCrossRef Guo R, Chen Z, Herzka DA, Luo J, Ding H. A three-dimensional free-breathing sequence for simultaneous myocardial T1 and T2 mapping. Magn Reson Med. 2019;81(2):1031–43.PubMedCrossRef
30.
Zurück zum Zitat Santini F, Kawel-Boehm N, Greiser A, Bremerich J, Bieri O. Simultaneous T1 and T2 quantification of the myocardium using cardiac balanced-SSFP inversion recovery with interleaved sampling acquisition (CABIRIA). Magn Reson Med. 2015;74(2):365–71.PubMedCrossRef Santini F, Kawel-Boehm N, Greiser A, Bremerich J, Bieri O. Simultaneous T1 and T2 quantification of the myocardium using cardiac balanced-SSFP inversion recovery with interleaved sampling acquisition (CABIRIA). Magn Reson Med. 2015;74(2):365–71.PubMedCrossRef
31.
Zurück zum Zitat Guo R, Cai X, Kucukseymen S, Rodriguez J, Paskavitz A, Pierce P, et al. Free-breathing simultaneous myocardial T1 and T2 mapping with whole left ventricle coverage. Magn Reson Med. 2021;85(3):1308–21.PubMedCrossRef Guo R, Cai X, Kucukseymen S, Rodriguez J, Paskavitz A, Pierce P, et al. Free-breathing simultaneous myocardial T1 and T2 mapping with whole left ventricle coverage. Magn Reson Med. 2021;85(3):1308–21.PubMedCrossRef
32.
Zurück zum Zitat Liu Y, Hamilton J, Rajagopalan S, Seiberlich N. Cardiac magnetic resonance fingerprinting: technical overview and initial results. JACC Cardiovasc Imaging. 2018;11(12):1837–53.PubMedPubMedCentralCrossRef Liu Y, Hamilton J, Rajagopalan S, Seiberlich N. Cardiac magnetic resonance fingerprinting: technical overview and initial results. JACC Cardiovasc Imaging. 2018;11(12):1837–53.PubMedPubMedCentralCrossRef
33.
Zurück zum Zitat Cruz G, Jaubert O, Qi H, Bustin A, Milotta G, Schneider T, et al. 3D free-breathing cardiac magnetic resonance fingerprinting. NMR Biomed. 2020;33(10): e4370.PubMedCrossRef Cruz G, Jaubert O, Qi H, Bustin A, Milotta G, Schneider T, et al. 3D free-breathing cardiac magnetic resonance fingerprinting. NMR Biomed. 2020;33(10): e4370.PubMedCrossRef
34.
Zurück zum Zitat Qi H, Bustin A, Cruz G, Jaubert O, Chen H, Botnar RM, et al. Free-running simultaneous myocardial T1/T2 mapping and cine imaging with 3D whole-heart coverage and isotropic spatial resolution. Magn Reson Imaging. 2019;63:159–69.PubMedCrossRef Qi H, Bustin A, Cruz G, Jaubert O, Chen H, Botnar RM, et al. Free-running simultaneous myocardial T1/T2 mapping and cine imaging with 3D whole-heart coverage and isotropic spatial resolution. Magn Reson Imaging. 2019;63:159–69.PubMedCrossRef
35.
Zurück zum Zitat Phair A, Cruz G, Qi H, Botnar RM, Prieto C. Free-running 3D whole-heart T1 and T2 mapping and cine MRI using low-rank reconstruction with non-rigid cardiac motion correction. Magn Reson Med. 2023;89(1):217–32.PubMedCrossRef Phair A, Cruz G, Qi H, Botnar RM, Prieto C. Free-running 3D whole-heart T1 and T2 mapping and cine MRI using low-rank reconstruction with non-rigid cardiac motion correction. Magn Reson Med. 2023;89(1):217–32.PubMedCrossRef
36.
Zurück zum Zitat Henningsson M. Cartesian dictionary-based native T1 and T2 mapping of the myocardium. Magn Reson Med. 2022;87(5):2347–62.PubMedCrossRef Henningsson M. Cartesian dictionary-based native T1 and T2 mapping of the myocardium. Magn Reson Med. 2022;87(5):2347–62.PubMedCrossRef
37.
Zurück zum Zitat Hamilton JI, Jiang Y, Chen Y, Ma D, Lo WC, Griswold M, et al. MR fingerprinting for rapid quantification of myocardial T1, T2, and proton spin density. Magn Reson Med. 2017;77(4):1446–58.PubMedCrossRef Hamilton JI, Jiang Y, Chen Y, Ma D, Lo WC, Griswold M, et al. MR fingerprinting for rapid quantification of myocardial T1, T2, and proton spin density. Magn Reson Med. 2017;77(4):1446–58.PubMedCrossRef
38.
Zurück zum Zitat Milotta G, Bustin A, Jaubert O, Neji R, Prieto C, Botnar RM. 3D whole-heart isotropic-resolution motion-compensated joint T1/T2 mapping and water/fat imaging. Magn Reson Med. 2020;84(6):3009–26.PubMedCrossRef Milotta G, Bustin A, Jaubert O, Neji R, Prieto C, Botnar RM. 3D whole-heart isotropic-resolution motion-compensated joint T1/T2 mapping and water/fat imaging. Magn Reson Med. 2020;84(6):3009–26.PubMedCrossRef
39.
Zurück zum Zitat Velasco C, Cruz G, Lavin B, Hua A, Fotaki A, Botnar RM, et al. Simultaneous T1, T2, and T1rho cardiac magnetic resonance fingerprinting for contrast agent-free myocardial tissue characterization. Magn Reson Med. 2022;87(4):1992–2002.PubMedCrossRef Velasco C, Cruz G, Lavin B, Hua A, Fotaki A, Botnar RM, et al. Simultaneous T1, T2, and T1rho cardiac magnetic resonance fingerprinting for contrast agent-free myocardial tissue characterization. Magn Reson Med. 2022;87(4):1992–2002.PubMedCrossRef
40.
Zurück zum Zitat Giri S, Shah S, Xue H, Chung YC, Pennell ML, Guehring J, et al. Myocardial T2 mapping with respiratory navigator and automatic nonrigid motion correction. Magn Reson Med. 2012;68(5):1570–8.PubMedPubMedCentralCrossRef Giri S, Shah S, Xue H, Chung YC, Pennell ML, Guehring J, et al. Myocardial T2 mapping with respiratory navigator and automatic nonrigid motion correction. Magn Reson Med. 2012;68(5):1570–8.PubMedPubMedCentralCrossRef
41.
Zurück zum Zitat Basha TA, Roujol S, Kissinger KV, Goddu B, Berg S, Manning WJ, et al. Free-breathing cardiac MR stress perfusion with real-time slice tracking. Magn Reson Med. 2014;72(3):689–98.PubMedCrossRef Basha TA, Roujol S, Kissinger KV, Goddu B, Berg S, Manning WJ, et al. Free-breathing cardiac MR stress perfusion with real-time slice tracking. Magn Reson Med. 2014;72(3):689–98.PubMedCrossRef
42.
Zurück zum Zitat Huizinga W, Poot DH, Guyader JM, Klaassen R, Coolen BF, van Kranenburg M, et al. PCA-based groupwise image registration for quantitative MRI. Med Image Anal. 2016;29:65–78.PubMedCrossRef Huizinga W, Poot DH, Guyader JM, Klaassen R, Coolen BF, van Kranenburg M, et al. PCA-based groupwise image registration for quantitative MRI. Med Image Anal. 2016;29:65–78.PubMedCrossRef
43.
Zurück zum Zitat Qi H, Lv Z, Xu J, Hu P. Optimization of spin-lock preparation pulses for B1 and B0 insensitive cardiac T1ρ mapping. Toronto: ISMRM; 2023. p. 4003. Qi H, Lv Z, Xu J, Hu P. Optimization of spin-lock preparation pulses for B1 and B0 insensitive cardiac T1ρ mapping. Toronto: ISMRM; 2023. p. 4003.
44.
Zurück zum Zitat Kim J, Peng Q, Wu C, Li X. MR T1ρ preparations: B1 and B0 inhomogeneity and T2ρ evaluation with Bloch equation-based simulation. London: ISMRM; 2022. p. 4999. Kim J, Peng Q, Wu C, Li X. MR T1ρ preparations: B1 and B0 inhomogeneity and T2ρ evaluation with Bloch equation-based simulation. London: ISMRM; 2022. p. 4999.
45.
Zurück zum Zitat Gram M, Seethaler M, Gensler D, Oberberger J, Jakob PM, Nordbeck P. Balanced spin-lock preparation for B1-insensitive and B0-insensitive quantification of the rotating frame relaxation time T1rho. Magn Reson Med. 2021;85(5):2771–80.PubMedCrossRef Gram M, Seethaler M, Gensler D, Oberberger J, Jakob PM, Nordbeck P. Balanced spin-lock preparation for B1-insensitive and B0-insensitive quantification of the rotating frame relaxation time T1rho. Magn Reson Med. 2021;85(5):2771–80.PubMedCrossRef
46.
Zurück zum Zitat Bipin Mehta B, Coppo S, Frances McGivney D, Ian Hamilton J, Chen Y, Jiang Y, et al. Magnetic resonance fingerprinting: a technical review. Magn Reson Med. 2019;81(1):25–46.PubMedCrossRef Bipin Mehta B, Coppo S, Frances McGivney D, Ian Hamilton J, Chen Y, Jiang Y, et al. Magnetic resonance fingerprinting: a technical review. Magn Reson Med. 2019;81(1):25–46.PubMedCrossRef
47.
Zurück zum Zitat Crawley AP, Henkelman RM. A comparison of one-shot and recovery methods in T1 imaging. Magn Reson Med. 1988;7(1):23–34.PubMedCrossRef Crawley AP, Henkelman RM. A comparison of one-shot and recovery methods in T1 imaging. Magn Reson Med. 1988;7(1):23–34.PubMedCrossRef
48.
Zurück zum Zitat Zhou R, Weller DS, Yang Y, Wang J, Jeelani H, Mugler JP 3rd, et al. Dual-excitation flip-angle simultaneous cine and T1 mapping using spiral acquisition with respiratory and cardiac self-gating. Magn Reson Med. 2021;86(1):82–96.PubMedPubMedCentralCrossRef Zhou R, Weller DS, Yang Y, Wang J, Jeelani H, Mugler JP 3rd, et al. Dual-excitation flip-angle simultaneous cine and T1 mapping using spiral acquisition with respiratory and cardiac self-gating. Magn Reson Med. 2021;86(1):82–96.PubMedPubMedCentralCrossRef
49.
Zurück zum Zitat Kozerke S, Hasenkam JM, Nygaard H, Paulsen PK, Pedersen EM, Boesiger P. Heart motion-adapted MR velocity mapping of blood velocity distribution downstream of aortic valve prostheses: initial experience. Radiology. 2001;218(2):548–55.PubMedCrossRef Kozerke S, Hasenkam JM, Nygaard H, Paulsen PK, Pedersen EM, Boesiger P. Heart motion-adapted MR velocity mapping of blood velocity distribution downstream of aortic valve prostheses: initial experience. Radiology. 2001;218(2):548–55.PubMedCrossRef
50.
Zurück zum Zitat Tao Q, van der Tol P, Berendsen FF, Paiman EHM, Lamb HJ, van der Geest RJ. Robust motion correction for myocardial T1 and extracellular volume mapping by principle component analysis-based groupwise image registration. J Magn Reson Imaging. 2018;47(5):1397–405.PubMedCrossRef Tao Q, van der Tol P, Berendsen FF, Paiman EHM, Lamb HJ, van der Geest RJ. Robust motion correction for myocardial T1 and extracellular volume mapping by principle component analysis-based groupwise image registration. J Magn Reson Imaging. 2018;47(5):1397–405.PubMedCrossRef
51.
Zurück zum Zitat Qi H, Lv Z, Hu J, Xu J, Botnar R, Prieto C, et al. Accelerated 3D free-breathing high-resolution myocardial T1rho mapping at 3 Tesla. Magn Reson Med. 2022;88(6):2520–31.PubMedCrossRef Qi H, Lv Z, Hu J, Xu J, Botnar R, Prieto C, et al. Accelerated 3D free-breathing high-resolution myocardial T1rho mapping at 3 Tesla. Magn Reson Med. 2022;88(6):2520–31.PubMedCrossRef
52.
Zurück zum Zitat Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47(6):1202–10.PubMedCrossRef Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47(6):1202–10.PubMedCrossRef
53.
Zurück zum Zitat Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. Circulation. 2002;105(4):539–42.PubMedCrossRef Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. Circulation. 2002;105(4):539–42.PubMedCrossRef
54.
Zurück zum Zitat Ma D, Coppo S, Chen Y, McGivney DF, Jiang Y, Pahwa S, et al. Slice profile and B1 corrections in 2D magnetic resonance fingerprinting. Magn Reson Med. 2017;78(5):1781–9.PubMedPubMedCentralCrossRef Ma D, Coppo S, Chen Y, McGivney DF, Jiang Y, Pahwa S, et al. Slice profile and B1 corrections in 2D magnetic resonance fingerprinting. Magn Reson Med. 2017;78(5):1781–9.PubMedPubMedCentralCrossRef
55.
Zurück zum Zitat Liu Y, Hamilton J, Eck B, Griswold M, Seiberlich N. Myocardial T1 and T2 quantification and water-fat separation using cardiac MR fingerprinting with rosette trajectories at 3T and 1.5T. Magn Reson Med. 2021;85(1):103–19.PubMedCrossRef Liu Y, Hamilton J, Eck B, Griswold M, Seiberlich N. Myocardial T1 and T2 quantification and water-fat separation using cardiac MR fingerprinting with rosette trajectories at 3T and 1.5T. Magn Reson Med. 2021;85(1):103–19.PubMedCrossRef
56.
Zurück zum Zitat Roujol S, Weingartner S, Foppa M, Chow K, Kawaji K, Ngo LH, et al. Accuracy, precision, and reproducibility of four T1 mapping sequences: a head-to-head comparison of MOLLI, ShMOLLI, SASHA, and SAPPHIRE. Radiology. 2014;272(3):683–9.PubMedCrossRef Roujol S, Weingartner S, Foppa M, Chow K, Kawaji K, Ngo LH, et al. Accuracy, precision, and reproducibility of four T1 mapping sequences: a head-to-head comparison of MOLLI, ShMOLLI, SASHA, and SAPPHIRE. Radiology. 2014;272(3):683–9.PubMedCrossRef
57.
Zurück zum Zitat Kellman P, Herzka DA, Hansen MS. Adiabatic inversion pulses for myocardial T1 mapping. Magn Reson Med. 2014;71(4):1428–34.PubMedCrossRef Kellman P, Herzka DA, Hansen MS. Adiabatic inversion pulses for myocardial T1 mapping. Magn Reson Med. 2014;71(4):1428–34.PubMedCrossRef
58.
Zurück zum Zitat Akcakaya M, Basha TA, Weingartner S, Roujol S, Berg S, Nezafat R. Improved quantitative myocardial T2 mapping: Impact of the fitting model. Magn Reson Med. 2015;74(1):93–105.PubMedCrossRef Akcakaya M, Basha TA, Weingartner S, Roujol S, Berg S, Nezafat R. Improved quantitative myocardial T2 mapping: Impact of the fitting model. Magn Reson Med. 2015;74(1):93–105.PubMedCrossRef
59.
Zurück zum Zitat Kampmann C, Linhart A, Baehner F, Palecek T, Wiethoff CM, Miebach E, et al. Onset and progression of the Anderson-Fabry disease related cardiomyopathy. Int J Cardiol. 2008;130(3):367–73.PubMedCrossRef Kampmann C, Linhart A, Baehner F, Palecek T, Wiethoff CM, Miebach E, et al. Onset and progression of the Anderson-Fabry disease related cardiomyopathy. Int J Cardiol. 2008;130(3):367–73.PubMedCrossRef
60.
Zurück zum Zitat Perry R, Shah R, Saiedi M, Patil S, Ganesan A, Linhart A, et al. The role of cardiac imaging in the diagnosis and management of Anderson-Fabry disease. JACC Cardiovasc Imaging. 2019;12(7 Pt 1):1230–42.PubMedCrossRef Perry R, Shah R, Saiedi M, Patil S, Ganesan A, Linhart A, et al. The role of cardiac imaging in the diagnosis and management of Anderson-Fabry disease. JACC Cardiovasc Imaging. 2019;12(7 Pt 1):1230–42.PubMedCrossRef
61.
Zurück zum Zitat Bush MA, Ahmad R, Jin N, Liu Y, Simonetti OP. Patient specific prospective respiratory motion correction for efficient, free-breathing cardiovascular MRI. Magn Reson Med. 2019;81(6):3662–74.PubMedPubMedCentralCrossRef Bush MA, Ahmad R, Jin N, Liu Y, Simonetti OP. Patient specific prospective respiratory motion correction for efficient, free-breathing cardiovascular MRI. Magn Reson Med. 2019;81(6):3662–74.PubMedPubMedCentralCrossRef
Metadaten
Titel
Free-breathing simultaneous native myocardial T1, T2 and T1ρ mapping with Cartesian acquisition and dictionary matching
verfasst von
Zhenfeng Lyu
Sha Hua
Jian Xu
Yiwen Shen
Rui Guo
Peng Hu
Haikun Qi
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2023
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-023-00973-6

Weitere Artikel der Ausgabe 1/2023

Journal of Cardiovascular Magnetic Resonance 1/2023 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.