Open Access 01.12.2023 | Research

Fully automated film mounting in dental radiography: a deep learning model

verfasst von: Yu-Chun Lin, Meng-Chi Chen, Cheng-Hsueh Chen, Mu-Hsiung Chen, Kang-Yi Liu, Cheng-Chun Chang

Erschienen in: BMC Medical Imaging | Ausgabe 1/2023

Abstract

Background

Dental film mounting is an essential but time-consuming task in dental radiography, with manual methods often prone to errors. This study aims to develop a deep learning (DL) model for accurate automated classification and mounting of both intraoral and extraoral dental radiography.

Method

The present study employed a total of 22,334 intraoral images and 1,035 extraoral images to train the model. The performance of the model was tested on an independent internal dataset and two external datasets from different institutes. Images were categorized into 32 tooth areas. The VGG-16, ResNet-18, and ResNet-101 architectures were used for pretraining, with the ResNet-101 ultimately being chosen as the final trained model. The model’s performance was evaluated using metrics of accuracy, precision, recall, and F1 score. Additionally, we evaluated the influence of misalignment on the model’s accuracy and time efficiency.

Results

The ResNet-101 model outperformed VGG-16 and ResNet-18 models, achieving the highest accuracy of 0.976, precision of 0.969, recall of 0.984, and F1-score of 0.977 (p < 0.05). For intraoral images, the overall accuracy remained consistent across both internal and external datasets, ranging from 0.963 to 0.972, without significant differences (p = 0.348). For extraoral images, the accuracy consistently achieved the highest value of 1 across all institutes. The model’s accuracy decreased as the tilt angle of the X-ray film increased. The model achieved the highest accuracy of 0.981 with correctly aligned films, while the lowest accuracy of 0.937 was observed for films exhibiting severe misalignment of ± 15° (p < 0.001). The average time required for the tasks of image rotation and classification for each image was 0.17 s, which was significantly faster than that of the manual process, which required 1.2 s (p < 0.001).

Conclusion

This study demonstrated the potential of DL-based models in automating dental film mounting with high accuracy and efficiency. The proper alignment of X-ray films is crucial for accurate classification by the model.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Dental radiography, a crucial diagnostic tool in dentistry, provides detailed images of the teeth, jaw, and surrounding structures [1]. Both intraoral and extraoral radiographic images are instrumental in dental practice. While intraoral images primarily focus on individual teeth or small groups of teeth, extraoral images capture a comprehensive view of the larger anatomical structures in the maxillofacial region, including the jaws, temporomandibular joints, sinuses, and other adjacent structures [2].
The accurate interpretation of dental X-rays requires the proper alignment and positioning of the film, which is generally conducted manually by dental radiographers or dentists [3]. This process involves rotating and identifying the correct position of the film and placing it in the appropriate tooth area. Because this process relies on the subjective judgement of the radiographer, it is time-consuming and prone to error [4].
In recent years, deep learning (DL) has emerged as a powerful tool for automating various image-related processes, such as identification and classification [57]. Studies have established the ability of DL models to analyze complex patterns and relationships in data, thereby advancing the accuracy and efficiency of various processes [8] and providing a promising solution for enhancing the interpretation process in dental radiography [9]. For instance, Lee et al. [10] and Bayrakdar et al. [11] leveraged the potential of convolutional neural network (CNN) algorithms for the detection and diagnosis of dental caries. Murata et al. [12] employed a CNN for the evaluation of maxillary sinusitis on panoramic radiography. Despite these advancements, the field has yet to explore one critical area — automated film mounting in dental radiography. This process, encompassing both intraoral and extraoral images, holds the potential to significantly enhance X-ray interpretation. By automating the time-consuming and error-prone task of manual film mounting, clinicians can focus on X-ray interpretation, resulting in increased accuracy and efficiency. To our knowledge, this study is the first to propose an automatic film mounting method for dental X-rays.
In this study, we developed and assessed a DL model for the accurate automated identification, rotation, and mounting of both dental intraoral and extraoral films. The goal is to enhance the efficiency and accuracy of the dental radiography interpretation process.

Methods

Patients and datasets

The Institutional Review Board of Chang Gung Medical Foundation approved this study (IRB number: 201900816B0C501), and also granted a waiver for the requirement of written informed consent. This study retrospectively enrolled a total of 1,500 patients at the Taipei branch of CGMH from July 2019 to June 2021 to train the model. The training dataset comprised a total of 23,379 images, including 22,344 intraoral images and 1,035 extraoral images. To enhance the diversity of the dataset, the training data were augmented four times by rotating the films to 0°, 90°, 180°, and 270°. The data were divided into training and validation sets using 5-fold cross-validation to prevent overfitting. An additional 2,333 independent images were employed to test the model’s performance; these included 2,221 intraoral and 112 extraoral images (Table 1).
Table 1
Number of images in training and testing datasets
Institution
Internal dataset Taipei CGMH
External dataset
Linkou
Taoyuan
Training Data
23,379
-
-
 Intraoral
22,344 (95.6%)
-
-
 Extraoral
1035 (4.4%)
-
-
Testing Data
2333
1916
1645
 Intraoral
2221 (95.2%)
1828 (95.3%)
1565 (95.1%)
 Extraoral
112 (4.8%)
88 (4.6%)
80 (4.9%)
To test the model’s generalization capabilities, external testing was performed using independent datasets obtained from two additional hospitals. The first hospital, the Linkou branch of CGMH, provided 1,828 intraoral images and 88 extraoral images. The second hospital, the Taoyuan branch of CGMH, supplied 1,565 intraoral images and 80 extraoral images. These images were not included in the training phase of the DL model, thereby offering a more rigorous test of the model’s ability to generalize to unseen data, a characteristic critical for real-world applications.

Data labelling

The matrix of intraoral images included Dental CR#0 (380 × 400 pixels), Dental DR#2 (800 × 800 pixels), Dental CR#2 (550 × 700 pixels), and Dental CR#4 (1100 × 800 pixels). The matrix of extraoral images included Panorex (1200 × 800 pixels), temporomandibular joint (TMJ; 1200 × 2400 pixels), and cephalometric (1500 × 3000 pixels) images.
We collected intraoral films with corresponding labels for the correct tooth position for each film. We categorized the data into 32 dental regions in accordance with the standard positioning guidelines for dental radiography, comprising 28 intraoral and 4 extraoral categories. The intraoral images included 14 categories of periapical images, 2 categories of bitewing (BW) images, 4 categories of vertical BW (VBW) images, 2 categories of occlusal images, 2 categories of pediatric upper (52–62) and lower (72–82) arch images, and 4 categories of pediatric upper (53 − 16, 63 − 26) and lower (73 − 36, 83 − 46) images. The extraoral images included one category of Panorex images, one category of TMJ images, and two categories of cephalometric (posterior–anterior and lateral) images. The criteria used to categorise the images are detailed in Table 2.
Table 2
Categorisation criteria of intraoral images
Teeth area
Intraoral film direction
Image selection criteria
Upper and lower teeth 2 ~ 2
Straight, Horizontal
Image contains upper and lower center incisors and lateral incisors, a few contain children’s upper and lower incisors.
Upper and lower teeth 3 ~ 5
Straight
Image contains upper and lower canines, first premolars and second premolars.
Upper and lower teeth 4 ~ 6
Straight, Horizontal
Image contains upper and lower first premolars, second premolars and first molars.
Upper and lower teeth 6 ~ 8
Straight, Horizontal
Image contains the upper and lower first molars, second molars and the third molars (wisdom teeth).
BW: Left and right sides
Horizontal
The adult image includes the upper and lower first premolars, second premolars, first molars and the second molars. The child image includes the upper and lower canines(3), first molars(4) and the second molars(5) .
VBW
Left and right front
Straight
The adult image includes the upper and lower first premolars, second premolars and a few of the first molars. The child image includes the upper and lower first molars(4) and second molars(5).
VBW
Left and right behind
Straight
Image contains the upper and lower first molars and second molars; a few contain second premolars.
Occlusal
Upper and lower
Straight, Horizontal
Image contains adult and children Upper and lower front teeth.
Pediatric upper (52 ~ 62) and lower (72 ~ 82) arch images
Straight, Horizontal
Image contains upper and lower center incisors and lateral incisors. A few contain adult upper and lower center incisors and lateral incisors.
Pediatric upper (53 ~ 16, 63 ~ 26) and lower (73 ~ 36, 83 ~ 46) images
Straight, Horizontal
The image includes upper and lower canines(3), first molars(4) and the second molars(5). A few include adult upper and lower canines, first premolars, second premolars and the first molars(6).

Network training

We initially trained the DL models using three different networks: VGG-16, ResNet-18, and ResNet-101 [13]. After comparing their preliminary accuracies, we adopted ResNet-101 as the final model due to its superior performance. During the training process, we implemented the Adam optimization algorithm and the categorical cross-entropy loss function. Other hyperparameters included the following: number of epochs = 100; learning rate = 0.1; batch size = 32; and weight decay = 0.001.
The network was trained on an Intel Xeon E5-2650 with 16GB DRAM, using a GTX-1080 GPU. The software, which was written in Python 3.5.4, used Keras 2.1.4 and TensorFlow 1.5.0.
After categorizing the images, a visualization tool was developed that automatically oriented and positioned the films on a standard template. The template, a standardized grid or reference image, helped align the films to their correct positions, ensuring a consistent and uniform presentation of the radiographs.

Workflow of the DL model inference

Figure 1 presents the workflow of our proposed DL model tool. This process begins with a patient undergoing dental radiography, yielding intraoral or extraoral images. The DL model takes these raw images as input, identifying and classifying each one into a specific tooth area. Subsequent automatic rotation ensures all images align correctly. The tool then executes digital film mounting, arranging the images in their appropriate positions to create a holistic view of the dental structures, echoing the conventional physical film mounting process. The final product is a set of well-organized, mounted images ready for clinical review and interpretation.

Evaluation of misalignment

To assess the effect of film misalignment on the performance of the DL model, we conducted experiments using a real human dental skull model by tilting the films at various angles relative to the X-ray tube. The tilting angles ranged from − 15° to + 15°, with increments of − 10°, − 5°, 0°, 5°, and 10°, with 0° indicating perfect alignment between the film and the X-ray tube. Three intraoral films were obtained for each angle and tooth position. This process aimed to evaluate the model’s ability to detect subtle changes in image orientation caused by film tilt and its effect on the accuracy of tooth position recognition.

Performance evaluation

We utilized the trained model to classify the test images into 32 tooth position classes and aligned the radiographs based on the predicted tooth positions. By comparing the results with the actual labels, we were able to determine the performance of the trained DL model. The performances were assessed for each fold in the five-fold cross-validation, and the average across these folds was deemed the final performance of the DL model for each network. The performance evaluation metrics employed for the models included (1) Accuracy, (2) Precision, (3) Recall, (4) F1 score.
$$Accuracy= \frac{TP+TN}{TP+TN+FP+FN}$$
(1)
$$Precision= \frac{TP}{TP+FP}$$
(2)
$$Recall= \frac{TP}{TP+FN}$$
(3)
$$F1 score=2\times \frac{Recall\times Precision}{Recall+Precision}$$
(4)
Where TP, FP, FN and TN represent true positive, false positive, false negative, and true negative, respectively.

Time of tasks

To assess the efficiency of the DL model, we calculated the time required for the model to complete the tasks of image identification, rotation, and mounting. To provide a comparative context, we randomly selected a sample of 50 patients, and measured the time duration consumed in performing the same tasks manually. This facilitated a direct comparison of the time efficiencies between the manual process and the DL model’s operation.

Statistics

Statistical analysis was performed using GraphPad Prism (version 8.0). We employed descriptive statistics to summarise the data and determine the mean accuracy, standard deviation, and range of scores. The variations in accuracy across different hospitals and among the distinct tilt angle groups were examined using an analysis of variance (ANOVA) test. To assess efficiency, we employed Student’s t-test to compare the time required by manual processing with that of the trained model. Statistical significance was indicated at p < 0.05.

Results

Model performance

This study employed a total of 5,894 images to test the performance of the trained model, which included 5,614 intraoral images and 280 extraoral images from 3 institutes (Table 1). Table 3 presents the classification performance of the pre-trained DL models on the internal test dataset. The ResNet-101 model demonstrated superior performance with the highest accuracy of 0.976 (95% CI: 0.968–0.983), precision of 0.969 (95% CI: 0.951–0.981), recall of 0.984 (95% CI: 0.969–0.991), and F1-score of 0.977 (95% CI: 0.969–0.984). These results significantly outperformed the performances of the VGG-16 and ResNet-18 models (p < 0.05 for all).
Table 3
Performances of the pre-trained DL models on the internal test dataset
Network
Accuracy
Precision
Recall
F1-Score
VGG-16
0.934
(0.912, 0.957)
0.916
(0.895, 0.932)
0.956
(0.933, 0.973)
0.937
(0.917, 0.959)
ResNet-18
0.963
(0.948, 0.978)
0.952
(0.941, 0.963)
0.977
(0.962, 0.989)
0.965
(0.951, 0.979)
ResNet-101
0.976
(0.968, 0.983)
0.969
(0.951, 0.981)
0.984
(0.969, 0.991)
0.977
(0.969, 0.984)
Data are presented as means with 95% confidence intervals
Table 4 displays the accuracies of ResNet-101 model’s image classification for each tooth position on both internal and external test datasets. For intraoral images, the overall accuracy was 0.972 (95% CI: 0.965–0.98) for Taipei CGMH, 0.963 (95% CI: 0.955–0.972) for Linkou CGMH, and 0.967 (95% CI: 0.961–0.974) for Taoyuan CGMH. The differences were not statistically significant (p = 0.348). For extraoral images, the accuracy consistently achieved the highest value of 1 across all institutes.
Table 4
Accuracy of image classification of ResNet-101 model for each tooth position
Tooth position
Accuracy
Internal testing Taipei CGMH
External testing
Linkou
Taoyuan
16 ~ 18
0.954
0.946
0.961
16 ~ 14
0.942
0.932
0.941
13 ~ 15
0.950
0.934
0.945
12 ~ 22
0.972
0.952
0.962
23 ~ 25
0.962
0.958
0.956
24 ~ 26
0.984
0.975
0.971
26 ~ 28
1.000
0.986
0.975
46 ~ 48
0.992
1.000
0.986
46 ~ 44
0.946
0.938
0.952
43 ~ 45
0.979
0.962
0.963
42 ~ 32
0.983
0.952
0.964
33 ~ 35
0.963
0.952
0.973
34 ~ 36
0.972
0.983
0.971
36 ~ 38
0.976
0.962
0.973
Occlusal Upper
1.000
1.000
1.000
Occlusal Lower
1.000
1.000
1.000
53 ~ 16
0.946
0.933
0.958
52 ~ 62
0.992
0.982
0.985
63 ~ 26
0.968
0.963
0.965
83 ~ 46
0.984
0.956
0.978
72 ~ 82
1.000
0.976
0.985
73 ~ 36
0.978
0.964
0.963
BW-Right
0.992
0.994
0.986
BW-Left
0.975
0.986
0.971
VBW-Right anterior
0.967
0.952
0.946
VBW-Right posterior
0.976
0.963
0.966
VBW-Left anterior
0.942
0.952
0.943
VBW-Left posterior
0.934
0.921
0.938
Panorex
1.000
1.000
1.000
TMJ
1.000
1.000
1.000
Cephalometric posterior-anterior
1.000
1.000
1.000
Cephalometric Lateral
1.000
1.000
1.000
The accuracy among the three hospitals did not differ, with accuracies of 0.976 (95% CI: 0.969–0.983) for Taipei CGMH, 0.968 (95% CI: 0.959–0.977) for Linkou CGMH, and 0.971 (95% CI: 0.964–0.978) for Taoyuan CGMH (p = 0.348).

Influence of alignment tilt angles on model accuracy

Figure 2 illustrates the effect of film tilt angles on the accuracy of the model. The tilt angle of X-ray films affected the model’s accuracy. The control group, with X-ray films correctly aligned at 0°, achieved an accuracy of 0.981 (95% CI: 0.968–0.991). X-ray films with a slight misalignment of ± 5° achieved an accuracy of 0.964 (95% CI: 0.953–0.975, p = 0.02 compared with the control group), whereas those with a moderate misalignment of ± 10° achieved an accuracy of 0.95 (95% CI: 0.936–0.971, p < 0.05 compared with the control group). The group with a severe misalignment of ± 15° had the worst performance, achieving an accuracy of 0.937 (95% CI: 0.918–0.956, p < 0.001 compared with the control group).
Figures 3 and 4 illustrate the performance of the DL model in automating the dental film mounting process. The DL model is adept at correcting orientations and rotating images to achieve proper alignment. As depicted in Fig. 3, the developed DL model accurately classified and positioned intraoral films, even when misaligned or inverted. Conversely, Fig. 4 presents a case in which the DL model misclassified an intraoral film. This error occurred due to a substantial 15° tilt of the X-ray tube, causing the projected image to resemble a different position and leading the DL model to incorrectly classify the tooth area.

Time of tasks

The results underscored a substantial enhancement in time efficiency by adopting our DL model in contrast to the traditional manual method. In the testing phase, the DL model adeptly executed image rotation and classification tasks, demonstrating significant time savings (0.17 ± 0.02 s per image for the DL model vs. 1.2 ± 0.28 s manually; p < 0.001). Of particular note, the per-patient processing time further exemplified the efficiency of the model. While manual processing required 118.6 ± 28.5 s per patient, the DL model drastically cut this down to only 3.3 ± 0.41 s (p < 0.001). The DL model’s processing time remained consistent, regardless of the film type, whether BW or periapical (p = 0.125), attesting to its robust performance across diverse imaging modalities.

Discussion

The results of this study demonstrate the effectiveness of the developed DL model in automating dental film mounting. The model achieved a high accuracy of 97.2% for intraoral images and 100% for extraoral images, demonstrating consistent performance across internal and external institutions without significant differences. These findings suggest that the DL model can serve as a valuable tool in dental practice, streamlining the film mounting process and potentially reducing the risk of misdiagnosis or treatment errors stemming from incorrect film interpretation.
The high accuracy for intraoral images indicates the model effectively recognizes and classifies tooth positions within the oral cavity, which is crucial for accurate dental film mounting [10]. The consistent performance across institutions suggests robustness and generalizability, making it a reliable tool for dental practitioners [11]. The high accuracy for extraoral images demonstrates the model’s ability to differentiate between intraoral and extraoral images, preventing interchange errors during mounting [14]. The notable success of our study lies in our approach to model training and the inherent qualities of the images used. Our DL model was extensively trained on a diverse dataset, allowing it to effectively classify a wide range of unique features. In particular, the high accuracy in classifying extraoral images is attributed to their distinct anatomical landmarks, such as sinuses and nasal bones, which serve as reliable classification indicators. Further, the reduced variability in these images compared to intraoral ones simplifies the task, aiding in the DL model’s superior performance.
The study highlights the impact of tilt angle on the model’s accuracy, emphasizing the importance of proper X-ray film alignment for accurate classification by the DL model. Practitioners should ensure correct alignment during image acquisition to optimize performance. Robustly adapting models to varying film angles is crucial, and including diverse images with different angles in the training dataset could address this issue [15]. However, collecting a comprehensive dataset may be challenging. The DL model accurately classified and mounted most intraoral films in the dataset, suggesting that DL could significantly improve the film mounting process in dental radiography.
Although the model performed well in most cases, it struggled with classification in certain scenarios, such as when the X-ray tube was significantly tilted. This highlights the need for careful consideration when designing and implementing DL systems in clinical practice and emphasizes the importance of selecting and curating the dataset used for training and testing. Despite some limitations, the DL model detected subtle changes in angle deviation and generated results generally acceptable to clinical dentists.
In terms of efficiency, the DL model’s ability to process images with significantly reduced time relative to manual methods underscores its potential in streamlining workflow in dental radiography. While the per-image time savings might seem small, the DL model drastically reduced the per-patient processing time. In a manual setting, operators are required to handle each image individually, taking into account their correct alignment and position. Furthermore, there can often be pauses, hesitations, or fatigue-related slowdowns that occur when operators manually process a series of images. This can lengthen the overall processing time significantly, especially when scaled to a larger number of patients. The minutes saved through the use of our DL model can be reallocated to more critical aspects of patient care such as diagnosis and treatment planning, consequently enhancing overall dental healthcare efficiency.
Our study has limitations. Firstly, our focus was primarily centered on the accuracy of the film mounting process, and we did not explore the diagnostic accuracy of the images processed using the DL model in depth. To confirm its clinical impact, future research must examine how the DL model affects diagnostic accuracy. Secondly, the model’s performance may vary across different dental practices due to the training dataset’s limited diversity. Ensuring its broad clinical applicability requires a more comprehensive dataset, covering various patient demographics, tooth morphologies, clinical conditions, and imaging techniques. Finally, while the model demonstrated high accuracy, there may be instances of minor misalignments due to varied clinical practices. As such, a future avenue of improvement could include the introduction of data augmentation with more subtle rotational degrees, improving the model’s ability to manage minor misalignments and potentially enhancing its robustness.
In conclusion, the results of this study demonstrated the potential of DL in automating the dental film mounting process. The DL model exhibited a high level of accuracy and efficiency in classifying and mounting dental films, which could greatly enhance the workflow in dental radiography. The results also highlighted the importance of proper X-ray film alignment for accurate classification by the DL model.

Acknowledgements

Not applicable

Declarations

The Institutional Review Board of Chang Gung Medical Foundation (CGMH; IRB number: 201900816B0C501) approved this study, and all procedures involving human participants were conducted in accordance with the institutional and the Declaration of Helsinki’s ethical standards. The requirement for written informed consent was waived by the Institutional Review Board of Chang Gung Medical Foundation.
Not applicable.

Competing interests

The authors declare no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Praskalo J, Beganović A, Milanović J, Stanković K, INTRAORAL DENTAL X-RAY RADIOGRAPHY IN BOSNIA AND HERZEGOVINA: STUDY FOR REVISING DIAGNOSTIC REFERENCE LEVEL VALUE. Radiat Prot Dosimetry. 2020;190(1):90–9.CrossRefPubMed Praskalo J, Beganović A, Milanović J, Stanković K, INTRAORAL DENTAL X-RAY RADIOGRAPHY IN BOSNIA AND HERZEGOVINA: STUDY FOR REVISING DIAGNOSTIC REFERENCE LEVEL VALUE. Radiat Prot Dosimetry. 2020;190(1):90–9.CrossRefPubMed
2.
Zurück zum Zitat Favia G, Lacaita MG, Limongelli L, Tempesta A, Laforgia N, Cazzolla AP, Maiorano E. Hyperphosphatemic familial tumoral calcinosis: odontostomatologic management and pathological features. Am J Case Rep. 2014;15:569–75.CrossRefPubMedPubMedCentral Favia G, Lacaita MG, Limongelli L, Tempesta A, Laforgia N, Cazzolla AP, Maiorano E. Hyperphosphatemic familial tumoral calcinosis: odontostomatologic management and pathological features. Am J Case Rep. 2014;15:569–75.CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Zhang W, Huynh CP, Abramovitch K, Leon IL, Arvizu L. Comparison of technique errors of intraoral radiographs taken on film v photostimulable phosphor (PSP) plates. Tex Dent J. 2012;129(6):589–96.PubMed Zhang W, Huynh CP, Abramovitch K, Leon IL, Arvizu L. Comparison of technique errors of intraoral radiographs taken on film v photostimulable phosphor (PSP) plates. Tex Dent J. 2012;129(6):589–96.PubMed
6.
7.
Zurück zum Zitat Lin YC, Lin CH, Lu HY, Chiang HJ, Wang HK, Huang YT, Ng SH, Hong JH, Yen TC, Lai CH, et al. Deep learning for fully automated tumor segmentation and extraction of magnetic resonance radiomics features in cervical cancer. Eur Radiol. 2020;30(3):1297–305.CrossRefPubMed Lin YC, Lin CH, Lu HY, Chiang HJ, Wang HK, Huang YT, Ng SH, Hong JH, Yen TC, Lai CH, et al. Deep learning for fully automated tumor segmentation and extraction of magnetic resonance radiomics features in cervical cancer. Eur Radiol. 2020;30(3):1297–305.CrossRefPubMed
8.
Zurück zum Zitat Zhang B, Jia C, Wu R, Lv B, Li B, Li F, Du G, Sun Z, Li X. Improving rib fracture detection accuracy and reading efficiency with deep learning-based detection software: a clinical evaluation. Br J Radiol. 2021;94(1118):20200870.CrossRefPubMed Zhang B, Jia C, Wu R, Lv B, Li B, Li F, Du G, Sun Z, Li X. Improving rib fracture detection accuracy and reading efficiency with deep learning-based detection software: a clinical evaluation. Br J Radiol. 2021;94(1118):20200870.CrossRefPubMed
9.
Zurück zum Zitat Celik B, Celik ME. Automated detection of dental restorations using deep learning on panoramic radiographs. Dentomaxillofac Radiol. 2022;51(8):20220244.CrossRefPubMed Celik B, Celik ME. Automated detection of dental restorations using deep learning on panoramic radiographs. Dentomaxillofac Radiol. 2022;51(8):20220244.CrossRefPubMed
10.
Zurück zum Zitat Lee JH, Kim DH, Jeong SN, Choi SH. Detection and diagnosis of dental caries using a deep learning-based convolutional neural network algorithm. J Dent. 2018;77:106–11.CrossRefPubMed Lee JH, Kim DH, Jeong SN, Choi SH. Detection and diagnosis of dental caries using a deep learning-based convolutional neural network algorithm. J Dent. 2018;77:106–11.CrossRefPubMed
11.
Zurück zum Zitat Bayrakdar IS, Orhan K, Akarsu S, Çelik Ö, Atasoy S, Pekince A, Yasa Y, Bilgir E, Sağlam H, Aslan AF, et al. Deep-learning approach for caries detection and segmentation on dental bitewing radiographs. Oral Radiol. 2022;38(4):468–79.CrossRefPubMed Bayrakdar IS, Orhan K, Akarsu S, Çelik Ö, Atasoy S, Pekince A, Yasa Y, Bilgir E, Sağlam H, Aslan AF, et al. Deep-learning approach for caries detection and segmentation on dental bitewing radiographs. Oral Radiol. 2022;38(4):468–79.CrossRefPubMed
12.
Zurück zum Zitat Murata M, Ariji Y, Ohashi Y, Kawai T, Fukuda M, Funakoshi T, Kise Y, Nozawa M, Katsumata A, Fujita H, et al. Deep-learning classification using convolutional neural network for evaluation of maxillary sinusitis on panoramic radiography. Oral Radiol. 2019;35(3):301–7.CrossRefPubMed Murata M, Ariji Y, Ohashi Y, Kawai T, Fukuda M, Funakoshi T, Kise Y, Nozawa M, Katsumata A, Fujita H, et al. Deep-learning classification using convolutional neural network for evaluation of maxillary sinusitis on panoramic radiography. Oral Radiol. 2019;35(3):301–7.CrossRefPubMed
13.
Zurück zum Zitat He K, Zhang X, Ren S, Sun J. Deep Residual Learning for Image Recognition. In: 2016 IEEE Conference on Computer Vision and Pattern Recognition (CVPR) 2016: 770–778. He K, Zhang X, Ren S, Sun J. Deep Residual Learning for Image Recognition. In: 2016 IEEE Conference on Computer Vision and Pattern Recognition (CVPR) 2016: 770–778.
14.
Zurück zum Zitat Różyło-Kalinowska I. Panoramic radiography in dentistry. Clin Dentistry Reviewed. 2021;5(1):26.CrossRef Różyło-Kalinowska I. Panoramic radiography in dentistry. Clin Dentistry Reviewed. 2021;5(1):26.CrossRef
15.
Zurück zum Zitat Lee Y, Jung Y, Choi Y, Kim Y, Kim S, Hong SJ, Kim H, Pae A. Accuracy of impression methods through the comparison of 3D deviation between implant fixtures. Int J Comput Dent. 2023;0(0):0.PubMed Lee Y, Jung Y, Choi Y, Kim Y, Kim S, Hong SJ, Kim H, Pae A. Accuracy of impression methods through the comparison of 3D deviation between implant fixtures. Int J Comput Dent. 2023;0(0):0.PubMed
Metadaten
Titel
Fully automated film mounting in dental radiography: a deep learning model
verfasst von
Yu-Chun Lin
Meng-Chi Chen
Cheng-Hsueh Chen
Mu-Hsiung Chen
Kang-Yi Liu
Cheng-Chun Chang
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
BMC Medical Imaging / Ausgabe 1/2023
Elektronische ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-023-01064-9

Weitere Artikel der Ausgabe 1/2023

BMC Medical Imaging 1/2023 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.