Introduction
The introduction of amyloid-β (Aβ) tracers for positron emission tomography (PET) has enabled in vivo assessment of one of the earliest pathological markers of Alzheimer’s disease (AD). This has been a great step forward for the field of AD research, as it has allowed for extracting quantitative measures of disease state, measuring treatment effects, and differentiation between patients with AD or non-AD neurodegenerative disorders [
1‐
6]. Three of the Fluorine-18-labelled amyloid PET tracers, [
18F]flutemetamol, [
18F]florbetapir and [
18F]florbetaben, have been approved by the European Medicines Agency (EMA) and the Food and Drug Administration (FDA) and are now commonly used for visualising and quantifying Aβ pathology in clinical, research, and clinical trial settings.
Nonetheless, there are various challenges to the routine use of PET scans in these settings. First, PET examinations are expensive compared with commonly used imaging modalities, such as magnetic resonance imaging (MRI) and computed tomography (CT). A key contributing factor to these high costs is the complexity of the ligand synthesis and ligand availability [
7]. Second, amyloid PET examinations have an associated radiation burden, typically between 6-7 mSv for Aβ tracers [
8‐
10], and they are time-intensive, with acquisition protocols of 20 min in clinical routine and up to 110 min for research protocols [
1,
11,
12].
Radiation burden in particular has been an obstacle when imaging healthy individuals and can restrict longitudinal and multi-tracer studies because of radiation safety regulations. This last aspect is of special importance given the current availability of, e.g., tau and synaptic density PET tracers that allow for characterizing different pathological aspects of AD [
13,
14].
These issues emphasise the importance of investigating whether the currently recommended injected dose for PET examinations can be reduced without compromising diagnostic performance. The benefits of a reduced injected dose would be two-fold: first, it would reduce scanning costs and second, it would reduce radiation exposure for patients, study participants and staff. Investigating the effect of reduced injected doses also allows for indirectly assessing reduced acquisition time as they will scale with each other, although differences will occur due to small changes in tracer kinetics and detection events when using a reduced acquisition time as opposed to assessing reduced injected dose. In the context of AD, the possibility of having a shorter scan would be highly relevant given that patients with AD dementia are not always able to lay still in the scanner for the duration of the scan. Furthermore, shorter scans could also allow for increasing throughput.
PET scanners have constantly evolved in terms of hardware and software, leading to improved spatial resolution, sensitivity, and image quality [
15‐
17]. Consequently, acquisition protocols should be reviewed regularly to determine whether any adjustments are warranted. For the FDA-approved [
18F]flutemetamol tracer, recommendations regarding the injected dose have been set using images collected on older generation cameras such as the Siemens HR + scanner [
18], and the same principle holds true for the [
18F]florbetapir and [
18F]florbetaben tracers. Previous studies have examined reduced injected doses for [
18F]FDG, [
18F]florbetapir and [
18F]Genentech Tau Probe 1 (GTP1) in the context of AD. These studies demonstrated that it is feasible to reduce the injected dose, in some cases to as low as 10% of the original injected dose, without compromising the quantitative information extracted from the scan [
19‐
21]. However, the implications of reduced doses on the clinical assessment of Aβ PET scans, remain largely unexplored. Furthermore, it has been suggested that visual assessment of reduced dose scans might be particularly challenging in grey-zone scans (i.e., scans displaying intermediate levels of Aβ pathology) [
19].
The most straightforward method for evaluating the feasibility of a reduced dose is to scan a participant twice, once at the original dose and once at a reduced dose [
21], however, this increases the participants' radiation burden. Additionally, biological variation between repeated acquisitions will increase uncertainty in the effect of reductions in dose. An alternative approach involves acquiring list-mode data and reconstructing or simulating a scan of reduced dose using a subset of these data [
20,
22‐
26]. Currently, this method is not regularly employed because of increased processing complexity and large storage requirements for list-mode data. Finally, a third method is the use of bootstrapping to generate a larger set of low-dose images from a smaller dataset [
19,
27]
In this study, we investigated the impact of reducing the injected dose in a research and clinical setting for [18F]flutemetamol and [18F]florbetaben by reconstructing previously acquired list-mode data that has been edited to simulate reductions in the injected dose. This method was chosen as it allows for a true simulation of a reduced dose, assessing the effect of a range of dose reductions and does not require additional scanning. More specifically, we investigated how reductions in the injected dose impact the standardised uptake value ratio (SUVR), Centiloid (CL) and the ability to visually assess PET scans with varying degrees of amyloid-β pathology. The overall goal of this project was thus to determine whether the vendor’s recommended injected dose could be reduced for [18F]flutemetamol and [18F]florbetaben, and to provide insight as to what dose reduction could be considered feasible.
Methods
Subjects and study protocol
In this study, 250 participants were retrospectively included. Of these, 175 participants were scanned with [
18F]flutemetamol across two centres: Barcelonaβeta Brain Research Center (BBRC), Barcelona, Spain (
N = 68), and Geneva University Hospital, Geneva, Switzerland (
N = 107). Participants were part of the Amyloid Imaging to Prevent Alzheimer’s Disease (AMYPAD) Diagnostic and Patient Management Study (DPMS) [
28] or Prognostic and Natural History Study (PNHS) [
29] or were recruited from the Geneva Memory Clinic cohort (GMC). Another 75 participants were also part of the AMYPAD PNHS study, but scanned with [
18F]florbetaben at the Centre Hospitalier Universitaire de Toulouse (CHUT), France. All PET data were available in list-mode format, as required per the design of the present study. Further cohort details can be found in Table
1. All participants had undergone standard neurological screening and neuropsychological assessment of each recruiting site. Prior to enrolling in the study, participants provided written informed consent in accordance with the Declaration of Helsinki and the International Conference on Harmonization Good Clinical Practice. Study protocols were approved by the local Medical Ethics Review Committees and can be accessed as part of their clinical trial registrations (EudraCT Number AMYPAD DPMS: 2017–002527-21 (2018–04-24) and AMYPAD PNHS is registered at
www.clinicaltrialsregister.eu with EudraCT number AMYPAD PNHS: 2018–002277-22 (2018–06-25)).
Table 1
Demographics per cohort
N | | 68 | 107 | 75 |
Cohort Aβ + status (%) | | AMYPAD PNHS 28.3 | AMYPAD DPMS, PNHS and GMC 57.9* | EPAD LCS 12.0 |
Age (y) | | 63.3 ± 5.9 | 73.8 ± 8.9* | 67.7 ± 7.5 |
Sex (F) (%) | | 51.2 | 49.5 | 66.6 |
Diagnosis (%) | CN# MCI Dementia Other | 98.7 1.3 0.0 0.0 | 28.0 51.4 16.8 3.7 | 100.0 0.0 0.0 0.0 |
Scanner | | Siemens Biograph 64 mCT | Siemens Biograph 128 Edge mCT Flow and Siemens Biograph 128 Vision 600 Edge | Siemens Biograph 4 Emission Duo LSO and Siemens Truepoint HiRez |
Scan duration | | 20 min (90–110 min p.i.) | 20 min (90–110 min p.i.) | 20 min (90–110 min p.i.) |
Injected Dose | | 190.2 ± 14.3 MBq | 179.4 ± 16.1 MBq | 302.9 ± 23.3 |
Acquisition Type | | List-mode | List-mode | List-mode |
Reconstruction | | OSEM, PSF, ToF | OSEM, PSF, ToF | OSEM, PSF |
Reconstruction settings | | 3 iterations, 24 subsets, 5 mm gaussian filter | 3 iterations, 24 subsets, 5 mm gaussian filter | 3 iterations, 24 subsets, 5 mm gaussian filter |
Image acquisition
For [
18F]flutemetamol, participants were injected with an average dose of 190.2 ± 14.3 MBq (BBRC cohort) or 179.4 ± 16.1 MBq (GMC cohort), and PET scans were acquired on a Siemens Biograph mCT or a Siemens Biograph Vision scanner. The remaining 75 participants from CHUT were injected with 302.9 ± 23.3 MBq [
18F]florbetaben, and PET scans were acquired on a Siemens Biograph 4 Emission Duo LSO or Siemens Truepoint HiRez scanner (acquisition parameters can be found in Table
1). Prior to the PET scan, a non-diagnostic low-dose computed tomography (ldCT) scan was acquired on the same scanner for attenuation correction purposes. All scans were acquired in list-mode for 20 min, 90–110 min post radiotracer intravenous injection.
List-mode processing
Reduced injected doses were simulated by manipulating the original PET list-mode data. List-mode data were delisted using vendor-provided reconstruction tools named “JSRecon” and “e7tools” (Siemens Molecular Imaging, Knoxville, USA). Across the entire 20-min acquisition window, from each one-second interval, true and random detection events were removed as a proportion of each injected dose level to represent a reduction in detection events across the entire acquisition time and therefore simulate a reduced injected dose, rather than simply a reduction in scan time. For example, at 50% of the originally injected dose, each one-second interval would have 50% of the true and random detection events removed. This method was used to generate the reduced-dose images corresponding to 75, 50, 25, 12.5, and 5% of the original injected dose.
Post-processing
First, original, and reduced-dose list-mode data were reconstructed using an iterative 3D-ordered subset expectation maximisation (OSEM) algorithm with the default parameters in JSRecon. Point spread function (PSF) modelling, time-of-flight (ToF) corrections where available and attenuation correction using the corresponding ldCT scan were applied. Next, PET images were normalised to Montreal Neurological Institute (MNI) space using rPOP and the Statistical Parametric Mapping “Old Normalise Estimate & Write” function using a previously validated PET template consisting of [
18F]flutemetamol, [
18F]florbetaben and [
18F]florbetapir data (SPM12, Wellcome Centre for Human Neuroimaging, London, UK) [
30]. SUVRs were derived on a voxel-by-voxel basis using the cerebellar cortex as a reference region as defined by the Global Alzheimer’s Association Interactive Network (GAAIN) Centiloid project [
4,
31‐
33]. A composite global cortical volume-of-interest (VOI), also from the GAAIN Centiloid project, and three VOIs known to show early Aβ accumulation, i.e., the precuneus, posterior cingulate cortex (PCC) and the orbitofrontal cortex (OFC) derived from the Desikan-Killiany atlas [
34], were used as target regions for the quantitative analysis [
35,
36]. To facilitate interpretability of the impact of dose reductions in absolute terms, we converted the global SUVR (from each tracer and cohort) to CL and will report this as additional outcome measure. CL values were pre-established using a validated pipeline at IXICO (London, UK) and regression lines were established for each tracer and cohort between the original, 100% SUVR and the CL values. These equations were subsequently applied to the reduced dose SUVRs, to convert these to CL values. The equations were the following: 107.65x-112.11 (BBRC), 121.21x-126.5 (GMC), and 124.65x-129.32 (CHUT).
Visual assessment
Aβ status was determined by a trained nuclear medicine physician who visually assessed the original PET scans according to the respective tracer manufacturer’s reading guidelines [
37]. To assess the effect of reduced doses on visual assessments as performed in a clinical setting, 32 scans from each tracer dataset were selected ([
18F]flutemetamol: 16 Aβ-,16 Aβ + , range: -24 to 143CL, and [
18F]florbetaben: 23 Aβ-, 9 Aβ + , range: -10 to 105CL) by a researcher who did not assess the scans, based upon the original full-dose visual assessment. From the 32 scans, 16 were considered grey-zone for [
18F]flutemetamol and 15 for [
18F]florbetaben, with grey-zone defined as having a CL between 12 and 50. [
32,
38,
39]. Note, for [
18F]flutemetamol there was an equal number of visually Aβ- and Aβ + grey-zone scans, while for [
18F]florbetaben, the grey-zone group included 14 visually Aβ- scans and one Aβ + scan, as a result of the limited number of grey-zone and Aβ + scans in this dataset. For each of these scans, the original PET image (in counts) and the 75, 50, 25, 12.5 and 5% injected-dose images were assessed independently by two other trained readers for [
18F]flutemetamol and one trained reader for [
18F]florbetaben, without anatomical reference (i.e., MRI or CT scan) to replicate the most challenging clinical scenario. Scans were presented in random order and classified as either positive or negative and readers were blinded to the dose level and clinical information. In addition, a confidence score was provided based on a five-point scale (1 = very low confidence, 5 = very high confidence).
Statistics
Between-cohort differences in age, proportion of males/females and Aβ + /Aβ- scans were investigated using Mann–Whitney U-tests and chi-square tests, respectively. Cortical SUVRs stratified by Aβ-status were visualised across dose levels and Mann–Whitney U-tests were used to verify whether SUVR differed significantly between Aβ-groups. Corresponding effect sizes were calculated using Hedge’s G. For SUVR, absolute and percentage differences were calculated between dose levels for all VOIs, while for CL only the absolute difference was calculated. In addition, receiver operating characteristic (ROC) analyses were performed to assess diagnostic accuracy and to determine the optimal sensitivity and specificity for distinguishing Aβ- from Aβ + scans. The area under the curve (AUC), Youden's index, sensitivity, and specificity were derived from the ROC analyses. The coefficient of variation (CoV) was calculated as a measure of variance (CoV = standard deviation/mean) and compared between dose levels for all VOIs. Bland–Altman analyses were used to assess potential bias between cortical SUVR of the original dose (i.e., 100%) and of each reduced dose level, and the presence of proportional bias was determined by fitting a regression line through the Bland–Altman plot.
To determine whether the visual assessment of the scan was affected by the reduction in dose, intra-reader agreement was assessed by comparing the reader’s assessment of the reduced dose images against their own read of the 100% images. Both the percentage agreement and Cohen’s kappa (κ) are reported. Agreement was assessed per tracer, per reader, for all scans and separately for the group of grey-zone (12 ≥ CL ≤ 50) and non-grey-zone scans (CL < 12 & CL > 50). In addition, the number of false positives (FP) and false negatives (FN) was calculated. Finally, Wilcoxon matched-pair signed rank tests were used to compare the average reader confidence scores between grey-zone and non-grey-zone scans and between dose levels.
Discussion
Quantitative analyses showed that for both tracers, injecting only 5% of the original dose resulted in a maximum change of 3.72% in SUVR and 2.04 pp in its variability. The ability to quantitatively separate Aβ- from Aβ + scans was high (Hedge’s G > 2.0) across dose levels. Visual assessment of the scans showed that for both tracers and all dose levels, the intra-reader agreement was > 80% suggesting that scans with significant dose reductions could still be used for qualitative assessment of amyloid scans.
Quantitative analyses showed similar signal-to-noise ratios for SUVRs across dose levels, as shown by the effect sizes, apart from a small drop in effect size for [
18F]flutemetamol at 5%. Furthermore, high AUC values and stable Youden’s Indices were reported across doses levels, demonstrating the feasibility of discriminating between Aβ-groups at lower doses for both tracers. These results demonstrate that injecting 5% of the original dose does not have a meaningful impact on the ability to quantitatively differentiate Aβ- from Aβ + scans. The maximum effect of reduced doses on cortical SUVR, CoV and CL was 2.27%, 0.70 pp and 4.62CL for [
18F]flutemetamol and 1.71%, 1.28 pp and 3.35CL for [
18F]florbetaben, at the 5% injected dose level. This magnitude of change is comparable to annual rates of change (i.e., 3.5–5.2CL) reported in Aβ + individuals [
40], which implies that it could hinder separation of non-accumulators from true accumulators. At 12.5% injected dose, the absolute change in CL was only 1.31 and 0.87CL for [
18F]flutemetamol and [
18F]florbetaben, respectively. Furthermore, only at 5%, coregistration to the MNI template failed for five scans. These results agree to a certain extent with those reported by Herholz and colleagues for [
18F]florbetapir, whom reported that the effect of reduced doses (i.e., 50, 20 and 10% of the original dose) on the mean cortical signal was minimal and no systematic bias was observed compared with the original dose [
19]. However, their findings differed in that they reported a dose-dependent increase in CoV for lower injected doses, and in that their CoVs were considerably lower than ours [
19]. The differences with the present study could be explained by a number of factors, first, their results correspond to a different amyloid tracer (i.e., [
18F]florbetapir), second, their target ROI was slightly different from ours (comprising all cortical GM rather than the GAAIN cortical ROI), and they used a different scanner and reconstruction algorithm, which are factors known to affect the signal-to-noise ratio (SNR). Finally, Herholz and colleagues included list-mode datasets of only four participants, and used a bootstrapping procedure to generate a larger dataset, while our study included list-mode datasets from
N = 175 and
N = 75 participants for [
18F]flutemetamol and [
18F]florbetaben, respectively, contributing to the greater CoVs, and to a more realistic definition of Aβ-groups.
As smaller regions tend to have higher intrinsic noise levels than larger ones, we conducted the same quantitative analyses for three regions known to show early Aβ accumulation [
35,
36]. For these smaller regions, the impact of injecting reduced doses resulted in a maximum change in SUVR of 3.72% at 5% and 0.76% at 12.5% injected dose. Vandenberghe and colleagues reported test–retest variability of [
18F]flutemetamol cortical SUVR to be 1.5 ± 0.7%, using a cerebellar cortex reference region. At 5% injected dose, the maximum change in SUVR in the present study falls outside the reported test–retest variability, while at 12.5%, it falls within this range, suggesting that injecting 12.5% dose is still acceptable [
18]. Moreover, based on previous research, it would be reasonable to assume that test–retest variability of smaller cortical regions would be similar or higher than 1.5 ± 0.7%, suggesting that the maximum change in SUVR reported here for 12.5% can be considered insignificant [
41‐
43]. For [
18F]florbetaben, maximum change in SUVR for these smaller regions was 3.12% at 5% injected dose and 1.34% at 12.5% injected dose. A previous study reported average cortical SUVR test–retest variability of [
18F]florbetaben (using cerebellar cortex as reference region) CN participants to be 2.9% (range 0.1–9.0) [
44]. This suggests that the change induced in SUVR at 12.5% injected dose (observed in the present study) would still be acceptable. Nonetheless, it should be noted that these test–retest studies are from 2010 and 2009 respectively, and test–retest variability may have improved for the currently used scanners. Bland–Altman analyses showed no systematic bias between original and reduced doses, although variability of the bias increased for lower injected doses. A bias proportional to underlying levels of Aβ pathology was observed for the lowest dose levels, which indicated that both the negative and the positive bias increased in magnitude.
Regarding the visual assessment of the [
18F]flutemetamol scans, intra-reader agreement for all scans ranged from 81–97% with κ ranging from 0.63–0.94. Prior studies on intra-reader agreement for [
18F]flutemetamol have reported intra-reader kappa values ranging from κ = 0.71–0.96, with great differences across study populations [
45,
46]. Overall, our kappa values appear comparable to these previously reported values, except for the 25% and 5% dose level [
18F]flutemetamol reader II results, that fall just outside this window (κ = 0.63). This suggests that level of experience from the reader may also play a role in the ability to assess reduced dose scans [
47]. For [
18F]florbetaben, intra-reader agreement for all scans ranged from 88–100% with κ ranging from 0.81–1.00. Two other studies that assessed intra-reader agreement using [
18F]florbetaben reported an average κ = 0.78 in newly trained readers in a diverse study population and an agreement of 91–98% in CN participants and AD dementia patients [
48,
49]. Our results are comparable to these previous studies, except for those in the 25% grey-zone group (agreement 73% and κ = 0.61) that fall outside this window. Overall, intra-reader agreement was considered adequate for both tracers and all dose levels even in the most challenging clinical cases and without anatomical reference [
50]. It has been suggested that visual assessment of reduced dose scans might be more challenging for grey-zone cases [
19], however, the present study did not find significant differences between grey-zone and non-grey-zone scans.
The present results suggest that reducing the injected dose to 12.5% for both tracers has limited effects on the quantitative and visual assessment of the scans, potentially enabling acquisition of additional PET scans within the same individual. A reduction below 12.5% is not recommended due to the aforementioned co-registration issues and meaningful quantitative effect on SUVR and CL. Note that our study assessed reconstructions from simulated reduced tracer doses rather than actual different tracer dose injections. However, differences between these methods are expected to be minor and are therefore unlikely to affect the overall conclusions of this study. This proof-of-concept study should thus be used as a guide in addition to a specific centre’s clinical or research circumstances when considering using reduced injected doses or to infer the use of reduced acquisition times from these results. Effects such as motion correction improvements and slight changes to tracer uptake for reduced acquisition times as well as changes in reconstruction algorithms for different scanners should be taken into account.
Declarations
Competing interests
FB is supported by the NIHR biomedical research centre at UCLH. He is steering committee or Data Safety Monitoring Board member for Biogen, Merck, ATRI/ACTC and Prothena. Consultant for Roche, Celltrion, Rewind Therapeutics, Merck, IXICO, Jansen, Combinostics and has research agreements with Merck, Biogen, GE Healthcare, Roche. He is also co-founder and shareholder of Queen Square Analytics LTD. JDG holds a ‘Ramón y Cajal’ fellowship (RYC-2013–13054) from the Spanish Ministry of Science, Innovation and Universities, has received research support from GE Healthcare, Roche Diagnostics and Hoffmann–La Roche, and speaker or consultant fees from Philips Nederlands and Roche Diagnostics. VG reports grants from the Swiss National Science Foundation (projects:320030_169876, 320030_185028 and IZSEZ0_188355), the Velux foundation (project:1123) and she received financial support for research and/or speaker fees through her institution from Siemens Healthineers, GE Healthcare, Life Molecular Imaging, Cerveau Technologies, Roche, Merck. MS has served on advisory boards for Servier, Roche and Novo Nordisk (outside scope of submitted work) and receives funding from the Knut and Alice Wallenberg Foundation (Wallenberg Centre for Molecular and Translational Medicine; KAW2014.0363, KAW2023.0371), the Swedish Research Council (2017–02869, 2021–02678 and 2021–06545), the Swedish State under the Agreement Between the Swedish Government and the County Councils, the ALF-agreement (ALFGBG-813971 and ALFGBG-965326), the Swedish Brain Foundation (FO2021-0311) and the Swedish Alzheimer Foundation (AF-740191). GBF is funded by the following private donors under the supervision of the Private Foundation of Geneva University Hospitals: A.P.R.A.—Association Suisse pour la Recherche sur la Maladie d’Alzheimer, Genève; Fondation Segré, Genève; Race Against Dementia Foundation, London, UK; Fondation Child Care, Genève; Fondation Edmond J. Safra, Genève; FondationMinkoff, Genève; Fondazione Agusta, Lugano; McCall Macbain Foundation, Canada; Nicole et René Keller, Genève; Fondation AETAS, Genève. GBF has received funding for research projects from: H2020, Innovative Medicines Initiative (IMI), IMI2, Swiss National Science Foundation, VELUX Foundation; has received unrestricted grants and support for event organisation from ROCHE Pharmaceuticals, OM Pharma, EISAI Pharmaceuticals, Biogen Pharmaceuticals. The Clinical Research Center, at Geneva University Hospital and Faculty of Medicine provides valuable support for regulatory submissions and data management, and the Biobank at Geneva University Hospital for biofluid processing and storage. LC has received research support from GE Healthcare and Springer Healthcare (funded by Eli Lilly). Both paid to the institution.