Impact of super-resolution and image acquisition on the detection of calcifications in digital breast tomosynthesis

verfasst von: Bruno Barufaldi, Raymond J. Acciavatti, Emily F. Conant, Andrew D. A. Maidment

Erschienen in: European Radiology | Ausgabe 1/2024

Einloggen, um Zugang zu erhalten

Abstract

Objectives

A virtual clinical trial (VCT) method is proposed to determine the limit of calcification detection in tomosynthesis.

Methods

Breast anatomy, focal findings, image acquisition, and interpretation (n = 14 readers) were simulated using screening data (n = 660 patients). Calcifications (0.2–0.4 mm3) were inserted into virtual breast phantoms. Digital breast tomosynthesis (DBT) acquisitions were simulated assuming various acquisition geometries: source motion (continuous and step-and-shoot), detector element size (140 and 70 µm), and reconstructed voxel size (35–140 µm). VCT results were estimated using multiple-reader multiple-case analyses and d’ statistics. Signal-to-noise (SNR) analyses were also performed using BR3D phantoms.

Results

Source motion and reconstructed voxel size demonstrated significant changes in the performance of imaging systems. Acquisition geometries that use 70 µm reconstruction voxel size and step-and-shoot motion significantly improved calcification detection. Comparing 70 with 100 µm reconstructed voxel size for step-and-shoot, the ΔAUC was 0.0558 (0.0647) and d’ ratio was 1.27 (1.29) for 140 µm (70 µm) detector element size. Comparing step-and-shoot with a continuous motion for a 70 µm reconstructed voxel size, the ΔAUC was 0.0863 (0.0434) and the d’ ratio was 1.40 (1.19) for 140 µm (70 µm) detector element. Small detector element sizes (e.g., 70 µm) did not significantly improve detection. The SNR results with the BR3D phantom show that calcification detection is dependent upon reconstructed voxel size and detector element size, supporting VCT results with comparable agreement (ratios: d’ = 1.16 ± 0.11, SNR = 1.34 ± 0.13).

Conclusion

DBT acquisition geometries that use super-resolution (smaller reconstructed voxels than the detector element size) combined with step-and-shoot motion have the potential to improve the detection of calcifications.

Clinical relevance

Calcifications may not always be discernable in tomosynthesis because of differences in acquisition and reconstruction methods. VCTs can identify strategies to optimize acquisition and reconstruction parameters for calcification detection in tomosynthesis, most notably through super-resolution in the reconstruction.

Key Points

• Super-resolution improves calcification detection and SNR in tomosynthesis; specifically, with the use of smaller reconstruction voxels.
• Calcification detection using step-and-shoot motion is superior to that using continuous tube motion.
• A detector element size of 70 µm does not provide better detection than 140 µm for small calcifications at the threshold of detectability.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
5.
6.
Zurück zum Zitat Shaheen E, Marshall N, Bosmans H (2011) Investigation of the effect of tube motion in breast tomosynthesis: continuous or step and shoot? In: Proceedings Volume 7961, Medical Imaging 2011: Physics of Medical Imaging 79611E Shaheen E, Marshall N, Bosmans H (2011) Investigation of the effect of tube motion in breast tomosynthesis: continuous or step and shoot? In: Proceedings Volume 7961, Medical Imaging 2011: Physics of Medical Imaging 79611E
7.
Zurück zum Zitat Yaffe MJ, Mainprize JG (2004) Detectors for digital mammography. Technol Cancer Res Treat 3:309–324CrossRefPubMed Yaffe MJ, Mainprize JG (2004) Detectors for digital mammography. Technol Cancer Res Treat 3:309–324CrossRefPubMed
9.
Zurück zum Zitat Barufaldi B, Bakic PR, Maidment ADA (2019) Multiple-reader, multiple-case ROC analysis for determining the limit of calcification detection in tomosynthesis. In: Proc. SPIE 10948, Medical Imaging 2019: Physics of Medical Imaging, 109480N (1 March 2019). https://doi.org/10.1117/12.2512884 Barufaldi B, Bakic PR, Maidment ADA (2019) Multiple-reader, multiple-case ROC analysis for determining the limit of calcification detection in tomosynthesis. In: Proc. SPIE 10948, Medical Imaging 2019: Physics of Medical Imaging, 109480N (1 March 2019). https://​doi.​org/​10.​1117/​12.​2512884
13.
Zurück zum Zitat Zeng R, Samuelson FW, Sharma D et al (2020) Computational reader design and statistical performance evaluation of an in-silico imaging clinical trial comparing digital breast tomosynthesis with full-field digital mammography. J Med Imaging (Bellingham) (4):042802. Epub 2020 Feb 26. https://doi.org/10.1117/1.jmi.7.4.042802 Zeng R, Samuelson FW, Sharma D et al (2020) Computational reader design and statistical performance evaluation of an in-silico imaging clinical trial comparing digital breast tomosynthesis with full-field digital mammography. J Med Imaging (Bellingham) (4):042802. Epub 2020 Feb 26. https://​doi.​org/​10.​1117/​1.​jmi.​7.​4.​042802
14.
Zurück zum Zitat Chase JG, Preiser JC, Dickson JL et al (2018) Next-generation, personalised, model-based critical care medicine: a state-of-the art review of in silico virtual patient models, methods, and cohorts, and how to validation them. Biomed Eng Online 17(1):24. https://doi.org/10.1186/s12938-018-0455-y Chase JG, Preiser JC, Dickson JL et al (2018) Next-generation, personalised, model-based critical care medicine: a state-of-the art review of in silico virtual patient models, methods, and cohorts, and how to validation them. Biomed Eng Online 17(1):24. https://​doi.​org/​10.​1186/​s12938-018-0455-y
25.
Zurück zum Zitat Maidment ADA, Albert M, Conant EF, Feig SA (1996) Three-dimensional calcification reconstruction from a limited number of views. Proc. SPIE 2708, Medical Imaging 1996: Physics of Medical Imaging, (11 April 1996); 378–389. https://doi.org/10.1117/12.237800 Maidment ADA, Albert M, Conant EF, Feig SA (1996) Three-dimensional calcification reconstruction from a limited number of views. Proc. SPIE 2708, Medical Imaging 1996: Physics of Medical Imaging, (11 April 1996); 378–389. https://​doi.​org/​10.​1117/​12.​237800
27.
Zurück zum Zitat International Commission on Radiation Units and Measurements (ICRU) Report 44. Tissue substitutes in radiation dosimetry. Washington: International Commission on Radiation Units and Measurements; 1989 International Commission on Radiation Units and Measurements (ICRU) Report 44. Tissue substitutes in radiation dosimetry. Washington: International Commission on Radiation Units and Measurements; 1989
37.
Zurück zum Zitat Acciavatti RJ, Maidment ADA (2012) Optimization of continuous tube motion and step-and-shoot motion in digital breast tomosynthesis systems with patient motion. Proc. SPIE 8313, Medical Imaging 2012. Phys Med Imag 831306 (2 March2012). https://doi.org/10.1117/12.911016 Acciavatti RJ, Maidment ADA (2012) Optimization of continuous tube motion and step-and-shoot motion in digital breast tomosynthesis systems with patient motion. Proc. SPIE 8313, Medical Imaging 2012. Phys Med Imag 831306 (2 March2012). https://​doi.​org/​10.​1117/​12.​911016
Metadaten
Titel
Impact of super-resolution and image acquisition on the detection of calcifications in digital breast tomosynthesis
verfasst von
Bruno Barufaldi
Raymond J. Acciavatti
Emily F. Conant
Andrew D. A. Maidment
Publikationsdatum
12.08.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 1/2024
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-023-10103-6

Weitere Artikel der Ausgabe 1/2024

European Radiology 1/2024 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.