Open Access 01.12.2021 | Case report

Insufficient procedural anticoagulation during leadless pacing led to catheter-related thrombosis in a hemodialysis patient

verfasst von: Qiang Chen, Yong Jiang Ma, Chun Hong Zhang, Li Wei Zhang

Erschienen in: BMC Cardiovascular Disorders | Ausgabe 1/2021

Abstract

Background

Leadless pacemaker was a promising innovation than traditional transvenous pacemaker, the procedural complications were prone to be bleeding-related. However, very few reports also concerned about the thrombus formation during the procedure.

Case presentation

A hemodialysis patient with diabetic gangrene of right foot suffered from catheter-related thrombosis during leadless pacing, resulting in failure of recapture the pacemaker. A low activated clotting time (ACT) level of 104 s confirmed the insufficiency of anticoagulation. Finally, the whole delivery catheter had to be removed from the delivery sheath, another new pacemaker system was applied and successfully implanted after adjusting the ACT level to 248 s.

Conclusion

Catheter-related thrombosis could be a large obstacle for leadless pacemaker implantation. In addition to routine anticoagulation, ACT monitoring might be necessary during the procedure.
Begleitmaterial
Additional file 1. Failure of withdrawing the pacemaker into the catheter, the delivery cup could not be advanced due to great resistance.
Additional file 2. The whole delivery catheter was removed from the delivery sheath.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1186/​s12872-021-02318-6.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
ACT
Activated clotting time
BMI
Body mass index
ESRD
End-stage renal disease
UFH
Unfractionated heparin

Background

Leadless pacemaker was a promising innovation than traditional transvenous pacemaker with its miniaturized profile and superior safety. The procedural complications such as venous site hemorrhage, arteriovenous fistulas and pericardial effusion had been studied [1], which were prone to be bleeding-related. However, very few reports also concerned about the thrombus formation during the procedure. We presented a case for the first time that a hemodialysis patient with diabetic gangrene of right foot suffered from catheter-related thrombosis, resulted in difficulty of recapture and withdrawal the pacemaker. A low activated clotting time (ACT) level of 104 s confirmed the insufficiency of anticoagulation.

Case presentation

A 68-year-old man, who was 175 cm height and 75 kg weight with a body mass index (BMI) of 24.5, suffered from syncope for several times, the ECG showed paroxysmal atrial fibrillation with long RR intervals for up to 9.3 s (Fig. 1). The echocardiography showed mild-to-moderate tricuspid regurgitation, normal ventricular wall motion and normal left ventricular ejection fraction. The medical history included end-stage renal disease (ESRD) requiring hemodialysis for 6 years, diabetic gangrene of right foot for 5 months. Indication for pacemaker implantation was established according to the ECG results, in order to prevent transvenous pocket inflammation or systemic inflammation from the gangrenous foot, the Micra leadless pacemaker (Medtronic, Inc. Minneapolis, USA) was recommended and the informed consent was approved by the patient.
The right femoral vein was successfully punctured and inserted with a 6 French sheath, and a bolus of 3000 IU unfractionated heparin (UFH) was intravenous administrated at the beginning of the procedure. The access site was gradually dilated by 8- and 12-French dilators, and a 23-French delivery sheath was introduced into the inferior vena cava, then the pacemaker delivery catheter was connected to the sheath. All delivery devices were flushed and connected with heparinized saline properly. The Micra pacemaker was advanced and conducted into the mid-septum of right ventricle and the fluoroscopic location was acceptable. However, electrical measurement showed a low sensing amplitude of only 2.8 mV, suggesting that the pacemaker needed to be repositioned again.
For recapture, the tether connected to the handle was pulled with constant tension and the pacemaker was closed to the recapture cone, different fluoroscopic angles were viewed to confirm the devices were coaxial. Since then the delivery cup could not be advanced due to great resistance, the pacemaker was not able to be withdrew into the catheter (Fig. 2, Additional file 1). Finally, the whole delivery catheter had to be removed from the delivery sheath (Fig. 2, Additional file 2).
Luckily no valvular or myocardial tissues were found at the tip of the catheter, instead the lumen was fully filled with heavy thrombi (Fig. 3). An instant ACT level of 104 s revealed that procedural anticoagulant therapy was insufficient, thus, additional 4000 IU (a total amount of 7000 IU) UFH was administrated immediately and the ACT level turned to 248 s 10 min later. After being removed all the thrombi the structure and mobility of the delivery catheter recovered (Fig. 4), which in turn proved that the failure of recapture was caused by the thrombi within the lumen. For preventing thromboembolism, a new pacemaker system was applied and successfully implanted. The final electrical measurement was satisfied: pacing threshold was 0.5 V at 0.24 ms, sensing amplitude was 7.9 mV, and pacing impedance was 560 Ω. The final ACT level was 204 s.

Discussion

An increasing number of patients will no doubt benefit from the new pacing technology. The procedure seems to be easier and safer than traditional pacing with over 95% successful rate. The key point of the implantation is to settle the pacemaker with suitable electrical parameters, which sometimes requires retrieving and redeploying the device. In most of the time the retrieval is without difficulty unless the nearby myocardial tissues are hooked by the pacemaker. In this case, we proved that thrombosis within the catheter could also hinder the movement of delivery cup or somehow affect the coaxality and led to the failure of recapture.
According to the guideline on percutaneous coronary intervention UFH is the preferred anticoagulant [2], the recommended dose is 70–100 U/kg intravenous and need not to be adjusted in CKD patients. But it is not clear whether routine anticoagulation is mandatory during leadless pacemaker implantation. Thrombosis s very rare [3] as compared with the bleeding complications, some experienced centers recommended a dose of 3000–5000 IU UFH intravenous without monitoring the ACT [4] and continuously flushing the delivery devices with heparinized saline. But in hemodialysis patients the data was lacking.
In our case the patient was at both high thromboembolic and bleeding risks with 4 points of CHA2DS2VASc score and 3 points of HASBLED score, which made the strategy for anticoagulation challenging. Considering the hemodialysis background UFH seemed to be the optimal anticoagulant for the patient. But apparently 3000 IU UFH was insufficient according to the first ACT level, additional 4000 IU UFH turned the level into 248s which almost reached the ideal ACT level during coronary intervention (between 250s and 300s) and successfully prevented another thrombosis complication. This could be an important finding: in complex situations, ACT levels should be measured in order to evaluate the anticoagulant effects in time. But the proper ACT ranges for leadless pacemaker implantation were still unknown and needed to be discussed further.
The safety during periprocedural period and the 1-year follow-up was promising for hemodialysis patient [5]. The patient also recovered well and was discharged 3 days later. Post-procedural echocardiography showed no evidence of pericardial effusion.
In conclusion, Catheter-related thrombosis could be a large obstacle for leadless pacemaker implantation. In addition to routine anticoagulation, ACT monitoring might be necessary during the procedure, and the appropriate ACT levels deserve further study.

Acknowledgements

Not applicable.

Declarations

The study protocol was approved by the ethics committee of the sixth medical center of Chinese PLA general hospital.
The written informed consent to publish this information was obtained from the individual.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Anhänge

Supplementary Information

Additional file 1. Failure of withdrawing the pacemaker into the catheter, the delivery cup could not be advanced due to great resistance.
Additional file 2. The whole delivery catheter was removed from the delivery sheath.
Literatur
1.
Zurück zum Zitat Lee JZ, Mulpuru SK, Shen WK. Leadless pacemaker: Performance and complications. Trends Cardiovasc Med. 2018;28(2):130–41.CrossRef Lee JZ, Mulpuru SK, Shen WK. Leadless pacemaker: Performance and complications. Trends Cardiovasc Med. 2018;28(2):130–41.CrossRef
2.
Zurück zum Zitat Neumann FJ, Sousa-Uva M, Ahlsson A, Alfonso F, Banning AP, Benedetto U, et al. 2018 ESC/EACTS Guidelines on myocardial revascularization. Eur Heart J. 2019;40(2):87–165.CrossRef Neumann FJ, Sousa-Uva M, Ahlsson A, Alfonso F, Banning AP, Benedetto U, et al. 2018 ESC/EACTS Guidelines on myocardial revascularization. Eur Heart J. 2019;40(2):87–165.CrossRef
3.
Zurück zum Zitat Cipolletta L, Volpato G, Biffi M, Capucci A. An indissoluble knot: an unexpected troubleshooting during Micra implantation. Pacing Clin Electrophysiol PACE. 2019;42(6):747–8.CrossRef Cipolletta L, Volpato G, Biffi M, Capucci A. An indissoluble knot: an unexpected troubleshooting during Micra implantation. Pacing Clin Electrophysiol PACE. 2019;42(6):747–8.CrossRef
4.
Zurück zum Zitat Martínez-Sande JL, García-Seara J, Rodríguez-Mañero M, Fernández-López XA, González-Melchor L, Redondo-Diéguez A, et al. The micra leadless transcatheter pacemaker. Implantation and mid-term follow-up results in a single center. Revista espanola de cardiologia (English ed). 2017;70(4):275–81. Martínez-Sande JL, García-Seara J, Rodríguez-Mañero M, Fernández-López XA, González-Melchor L, Redondo-Diéguez A, et al. The micra leadless transcatheter pacemaker. Implantation and mid-term follow-up results in a single center. Revista espanola de cardiologia (English ed). 2017;70(4):275–81.
5.
Zurück zum Zitat El-Chami MF, Clementy N, Garweg C, Omar R, Duray GZ, Gornick CC, et al. Leadless pacemaker implantation in hemodialysis patients: experience with the micra transcatheter pacemaker. JACC Clin Electrophysiol. 2019;5(2):162–70.CrossRef El-Chami MF, Clementy N, Garweg C, Omar R, Duray GZ, Gornick CC, et al. Leadless pacemaker implantation in hemodialysis patients: experience with the micra transcatheter pacemaker. JACC Clin Electrophysiol. 2019;5(2):162–70.CrossRef
Metadaten
Titel
Insufficient procedural anticoagulation during leadless pacing led to catheter-related thrombosis in a hemodialysis patient
verfasst von
Qiang Chen
Yong Jiang Ma
Chun Hong Zhang
Li Wei Zhang
Publikationsdatum
01.12.2021
Verlag
BioMed Central
Erschienen in
BMC Cardiovascular Disorders / Ausgabe 1/2021
Elektronische ISSN: 1471-2261
DOI
https://doi.org/10.1186/s12872-021-02318-6

Weitere Artikel der Ausgabe 1/2021

BMC Cardiovascular Disorders 1/2021 Zur Ausgabe

Nach Herzinfarkt mit Typ-1-Diabetes schlechtere Karten als mit Typ 2?

29.05.2024 Herzinfarkt Nachrichten

Bei Menschen mit Typ-2-Diabetes sind die Chancen, einen Myokardinfarkt zu überleben, in den letzten 15 Jahren deutlich gestiegen – nicht jedoch bei Betroffenen mit Typ 1.

Erhöhtes Risiko fürs Herz unter Checkpointhemmer-Therapie

28.05.2024 Nebenwirkungen der Krebstherapie Nachrichten

Kardiotoxische Nebenwirkungen einer Therapie mit Immuncheckpointhemmern mögen selten sein – wenn sie aber auftreten, wird es für Patienten oft lebensgefährlich. Voruntersuchung und Monitoring sind daher obligat.

GLP-1-Agonisten können Fortschreiten diabetischer Retinopathie begünstigen

24.05.2024 Diabetische Retinopathie Nachrichten

Möglicherweise hängt es von der Art der Diabetesmedikamente ab, wie hoch das Risiko der Betroffenen ist, dass sich sehkraftgefährdende Komplikationen verschlimmern.

TAVI versus Klappenchirurgie: Neue Vergleichsstudie sorgt für Erstaunen

21.05.2024 TAVI Nachrichten

Bei schwerer Aortenstenose und obstruktiver KHK empfehlen die Leitlinien derzeit eine chirurgische Kombi-Behandlung aus Klappenersatz plus Bypass-OP. Diese Empfehlung wird allerdings jetzt durch eine aktuelle Studie infrage gestellt – mit überraschender Deutlichkeit.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.