Personalising treatment plan quality review with knowledge-based planning in the TROG 15.03 trial for stereotactic ablative body radiotherapy in primary kidney cancer

verfasst von: Nicholas Hardcastle, Olivia Cook, Xenia Ray, Alisha Moore, Kevin L. Moore, David Pryor, Alana Rossi, Farshad Foroudi, Tomas Kron, Shankar Siva

Erschienen in: Radiation Oncology | Ausgabe 1/2021

Abstract

Introduction

Quality assurance (QA) of treatment plans in clinical trials improves protocol compliance and patient outcomes. Retrospective use of knowledge-based-planning (KBP) in clinical trials has demonstrated improved treatment plan quality and consistency. We report the results of prospective use of KBP for real-time QA of treatment plan quality in the TROG 15.03 FASTRACK II trial, which evaluates efficacy of stereotactic ablative body radiotherapy (SABR) for kidney cancer.

Methods

A KBP model was generated based on single institution data. For each patient in the KBP phase (open to the last 31 patients in the trial), the treating centre submitted treatment plans 7 days prior to treatment. A treatment plan was created by using the KBP model, which was compared with the submitted plan for each organ-at-risk (OAR) dose constraint. A report comparing each plan for each OAR constraint was provided to the submitting centre within 24 h of receiving the plan. The centre could then modify the plan based on the KBP report, or continue with the existing plan.

Results

Real-time feedback using KBP was provided in 24/31 cases. Consistent plan quality was in general achieved between KBP and the submitted plan. KBP review resulted in replan and improvement of OAR dosimetry in two patients. All centres indicated that the feedback was a useful QA check of their treatment plan.

Conclusion

KBP for real-time treatment plan review was feasible for 24/31 cases, and demonstrated ability to improve treatment plan quality in two cases. Challenges include integration of KBP feedback into clinical timelines, interpretation of KBP results with respect to clinical trade-offs, and determination of appropriate plan quality improvement criteria.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1186/​s13014-021-01820-7.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
API
Application programming interface
DVH
Dose volume histogram
IMRT
Intensity modulated arc therapy
KBP
Knowledge based planning
NTCP
Normal tissue complication probability
OAR
Organ at risk
PTV
Planning target volume
QA
Quality assurance
RCC
Renal cell cancer
SABR
Stereotactic ablative body radiotherapy
VMAT
Volumetric modulated arc therapy

Introduction

Patient outcomes are dependent on the technical quality of radiation therapy in clinical trials [13]. Further, clinical trials in radiation oncology are often used to progress techniques or treatments with demonstrated efficacy in single institution Phase I trials to multi-institution implementations [4, 5]. Appropriate quality assurance (QA) is critical to ensure early positive results can be replicated in a multi-institution setting. QA such as real-time case review prior to treatment can not only reduce the risk profile of novel treatment techniques, but can provide invaluable improvements to treatment quality of each individual patient, in particular for institutions with limited experience with a given technique or anatomical site. Real time QA in clinical trials, however, is resource intensive, requiring substantial effort to de-identify and submit radiation therapy treatment planning data to a central repository and significant investment of time from volunteer reviewers. This is compounded by time constraints, where plan review must be performed efficiently, often in the context of timing with delivery of systemic agents. Typically, dose distributions are often reviewed only to ensure compliance with trial protocol, with limited scope for reviewers to provide input and feedback on whether a treatment plan is optimal for that particular patient.
TROG 15.03 FASTRACK II (ACTRN12615001181594) is a Phase II multi-centre single arm trial with the primary objective of estimating the efficacy of stereotactic ablative body radiation therapy (SABR) for primary renal cell cancer (RCC) (freedom from local progression) [6]. SABR for RCC is technically challenging due to the proximity of multiple organs at risk (OARs) to the target volume, tumour and OAR motion due to respiration, and the often poor image quality with online image guidance in this region [79].
Knowledge based planning (KBP) uses previous treatment plans to inform achievable plan quality for new patients [1012]. KBP estimates achievable OAR dose-volume-histograms (DVHs) for a given patient using a machine learning model. The model has learned achievable dose volume histograms from a library of plans, in which OAR dose depends on the geometric relationship between a target and OAR, beam geometry and prescription dose. KBP has been used to improve the consistency and efficiency of treatment planning in routine practice, to audit treatment plan quality within and external to an institution, and to retrospectively review treatment plan quality in clinical trials [10, 1321]. In retrospective review of clinical trial data, the potential of KBP to improve individual patient OAR dosimetry has been demonstrated [13, 16, 17, 2224]. In the current work, we report the use of KBP in real-time QA for the FASTRACK II trial. We used semi-automated KBP review to perform treatment plan quality assessment, in parallel with manual review for a subset of trial participants. This is the first report describing experience of KBP in real-time review of clinical trial plan quality.

Materials and methods

The workflow shown in Fig. 1, adapted from Tol et al. [23], shows the pathway of real-time review used in the current study. For a subset of patients enrolled in the FASTRACK II trial, the submitted treatment planning data (image, structures and treatment beam geometry) was planned using a KBP model. The achievable OAR DVHs with the KBP model was compared with those from the submitted plan, and feedback was provided to the submitting institution, along with the feedback from manual reviewers. The institution could then update the plan, or continue with the existing plan.

Model generation

In the FASTRACK II trial, participants receive a single fraction of 26 Gy for tumours smaller than 4 cm, or 42 Gy in three fractions for larger tumours. A sequential cohort of 53 previously treated primary kidney cancer patients were used as the training set for the KBP model. These had been treated at a single institution either as part of a prospective Phase I trial or off-trial following the FASTRACK II protocol (23 patients treated with single fraction, 30 treated with three fractions). These patients were selected as they represented the expected cohort that would be recruited in the FASTRACK II trial. This technique was previously demonstrated to be safe with a signal for efficacy [9, 25]. All patients were planned and treated with 3DCRT, using 7–9 beams typically including 1–2 non-coplanar beams. The plans were added into an initial KBP model (RapidPlan, v13.6, Varian Medical Systems), and a series of objectives were created using this KBP model (Additional file 1: Supplementary Table 1). It was expected the majority of plans in the present trial would be treated with a Volumetric Modulated Arc Therapy (VMAT) technique, therefore all patients in this cohort were optimised with a class solution VMAT plan created with two ipsilateral arcs spanning 210° from the posterior-anterior direction around to 30° past midline. The plans were optimised, and refinements were made to objectives to meet the FASTRACK II protocol requirements. In particular, the priority of maximum dose objectives to organs at risk were adjusted to ensure the plans met constraints, and further reduction of mid-range OAR doses was achieved through use of the maximum generalised Equivalent Uniform Dose (gEUD) objective. A new KBP model was created with the final plans. Due insufficient numbers to create two separate models, we created one model for both fractionations. For most OAR constraints, the two fractionations have equivalent OAR constraints as a percentage of the prescription dose. Where this was not the case, we set the objectives to meet the lower constraint of the two fractionations. Validation is described in the Additional file 1: supplementary material.

Manual review

Real-time review was performed for all patients treated on the trial, and involved review of the CT imaging used for treatment planning, including relevant 4DCT phases, diagnostic imaging used for contouring, structure sets, treatment plans and dose distributions. The plan was submitted ideally seven days prior to treatment start date. The contouring and dose distributions were reviewed by an independent radiation oncologist, and the treatment plan, dose distributions were reviewed by a medical physicist. The review ensured adherence to protocol for contouring, treatment planning parameters, motion management and target and OAR dosimetry. Feedback was provided to the site via a PDF case report with itemised violation criteria, and the site had the opportunity to correct any violations prior to treatment delivery. Subsequent revisions underwent the same review as an initial treatment plan.

KBP review

KBP review was performed centrally by TROG Cancer Research RTQA Services. In the Eclipse treatment planning system (v13.6, Varian Medical Systems), a generic linac model was created using representative beam data and machine characteristics for each linac type expected to be used in the trial. Only C-arm linear accelerators were used in this trial, and included Varian TrueBeam, Varian TrueBeam STx, Elekta VersaHD and Elekta Synergy linacs. For each patient, a plan was created in Eclipse using the same beam angles and a generic linac model matching that submitted by the centre. The KBP model was applied and the plan was run through one full optimisation without user interaction. An application programing interface (API) script then generated a PDF report which was sent to the site within 24 h of receiving the plan.
For each of the OAR constraints, the KBP generated plan was compared with the submitted plan; the submitted plan value and the KBP plan values were presented, along with the minor and major constraints. Possible improvements were flagged in the report if the submitted plan violated the trial constraint and the KBP plan met the constraint, or if the KBP plan OAR dose was less than 90% of the submitted plan dose. The individual centre then had the opportunity to revise the plan, or continue with the submitted plan, based on their clinical judgement. No plan alterations were mandated. The centre was then invited to provide feedback on the process, using the form presented as Additional file 2: Supplementary Table 2.
For all patients treated in this phase of the FASTRACK II trial, the output of the API script was collected which included the target and OAR metrics for the submitted and KBP plans. Comparison was performed per patient and per metric based on classifications in the report as described above.

Results

The last 31 patients in the trial were part of the real-time KBP phase, out of 70 patients recruited for the trial. Plans were from 8 centres and included IMRT (1), VMAT (28), dynamic conformal arc (1) and 3D conformal (1) from both Varian and Elekta linacs, and Pinnacle, Eclipse and Monaco treatment planning systems. Only 10/31 cases were provided with at least 7 days prior to radiotherapy start date; real-time reports were however provided in 24/31 cases. Six weren’t provided due to insufficient time between plan submission and treatment for QA staff to provide a report, and one was not provided due to technical issues with the KBP report generation (Additional file 3: Supplementary Table 3). Resubmission of the plan was performed in 10/24 plans. The reason for resubmission was contouring violations in five cases, OAR dosimetry in four cases and insufficient motion management in one case.

Target dosimetry

The PTV near-minimum and near-maximum doses were similar between the submitted and KBP plans. In the 26 Gy cohort, the PTV_2600 D99% was at or above prescription dose for 13/15 submitted plans, and 11/15 KBP plans, but the D95% was at or above prescription dose for all submitted and KBP plans. In the 42 Gy cohort, the PTV_4200 D99% was at or above prescription dose for 10/16 submitted plans, and 3/16 KBP plans, but the D95% was at or above prescription dose for 15/16 submitted and KBP plans. In one patient the large bowel was adjacent to the target and the PTV coverage was compromised to ensure large bowel constraints were met.

OAR dosimetry

Figure 2 classifies the KBP result for each OAR constraint. Figure 3 shows the near-maximum doses for the small bowel, large bowel and spinal cord for both fractionation regimes, and Fig. 4 shows the mid-range DVH metrics for the stomach and small bowel. The near-maximum doses and the DVH constraints were however well below constraints for all plans. There was a relatively even split between KBP being within 10%, or either KBP or the submitted being superior by more than 10%. Improvement was suggested most often for the large bowel D1.5 cc, and Kidney_I V50%. In the three fraction schedule, the near-maximum and mid-range DVH metrics were close to constraint for many patients (Figs. 3, 4). As a result, there were five instances where the submitted plan met the constraint but the KBP plan did not; three were near-maximum doses (small bowel, large bowel and spinal cord) and two were small bowel D30cc. For the small bowel D30cc, there were three submitted plans that did not meet constraints, but the KBP plan did (Fig. 4).
The small bowel D30cc violation was detected by the reviewer for Patient #51 (see Additional files 4, 5: supplementary reports), with the KBP suggesting this constraint could be met. Figure 5a shows the initial plan, with intermediate isodose lines extending across the bowel structures in the submitted plan. The replan (Fig. 5b) shows significant improvement in the intermediate dose spillage, and all constraints met. Patient #54 showed improvement was possible such that the small bowel D30cc constraint was met, but the centre elected to not replan. In this case, objective functions for the small bowel were only used for the near-maximum doses, not the intermediate doses.
There were three plans that were resubmitted based on dosimetric review. For #46, the reviewers recommended reducing target coverage to keep the prescription isodose line off the large bowel. The KBP showed minor improvement, but due to its proximity to the PTV, minimal gains could be made. For patient #60 the reviewers indicated there was a high risk of bowel constraints being exceeded if bowel loops were in a different position at time of treatment. The KBP plan had a similar dose distribution to the submitted plan. In #71, small and large bowel doses were within constraints, but the KBP was more than 10% less than the submitted plan. The site elected to replan this case, in the process reducing small and large bowel dose (Fig. 5c, d), with the aim of improving robustness to bowel position variation at time of treatment.
Feedback was provided by the treating centre for five of 30 participants, and is provided in Additional file 2: Supplementary Table 2. One centre (two patients) indicated the plans could be improved based on the KBP report, however due to time constraints they did not replan. Another centre (two patients) indicated they preferred to retain their chosen clinical trade-offs prioritising one OAR over another. Despite no changes being made in the subset of plans for which feedback was provided, in each case the centre indicated that KBP based QA was a valuable tool for plan quality improvement in clinical trials. Specifically, the participating centres who provided feedback thought analysis of plan quality was useful for their own practice, and for the overall quality of the plans in the trial.

Discussion

Maintaining quality in clinical trials is important yet resource-intensive. Use of KBP in real-time plan review aims to rapidly identify dosimetric outliers that could be resolved prior to treatment, and provide guidance on optimisation to improve both compliant and non-compliant plans. For clinical trials of new indications or technology, KBP has two key roles. Firstly, this trial was aimed at measuring safety and efficacy of SABR for primary renal cell carcinoma. This was based on a single institution Phase 1 trial which demonstrated this was safe and effective, therefore ideally the treatment planning quality from this Phase 1 trial is carried on to the multi-centre Phase II trial. In this aspect, ensuring the KBP model was at least as good as that used in the Phase I trial maintains this safety, through rapid identification of violations that could be resolved. Secondly, for centres who have little or no experience with kidney SABR, KBP provides a useful reassurance that they are achieving sufficient and consistent plan quality in each of their patients. Plan quality was very consistent between KBP and submitted plans, despite variations in experience, and available software and hardware.
KBP has been used for retrospective review of treatment plan quality in head and neck [16, 22, 23], lung [13] and prostate cancer [24], and vertebral SABR [17]. In all cases, plans performed with KBP were shown to improve the consistency of treatment plan quality, and result in reduced doses to OARs. KBP real-time review processes have been developed for NRG-HN001, facilitated by the Imaging and Radiation Oncology Core [16], and in NRG-GY006 [10, 26]. To our knowledge, this is the first report of prospective use of KBP for real-time review in clinical trial QA. We have demonstrated feasibility of the process, and in two cases in this trial, replanning was performed directly as a result of KBP feedback. In particular, in this trial we had both manual review and KBP based review of the plans; manual reviewers were able to determine whether plans exceeded constraints, but the KBP provided added value by demonstrating that these constraints could in fact be met. The demonstrated improvement in treatment plan quality has the potential to result in improved clinical outcomes for these patients.
We don’t expect that KBP results in superior or equivalent plans in all cases, however we do expect to identify plans with inferior plan quality to the plans upon which the KBP model is based. There were instances where KBP resulted in inferior dosimetry to the submitted plan, in particular where the near-maximum dose in both submitted and KBP plans was close to constraints. High near-maximum dose was due to OAR-target proximity, and manual fine-tuning is likely required, which wasn’t part of the KBP plan generation process. This trade-off mainly affects the near-maximum of the adjacent OAR(s), and the PTV coverage percentage, rather than the intermediate OAR doses. Therefore there may be limited utility in providing feedback based on near-maximum OAR doses and target coverage from KBP-derived plans, with the benefit of KBP mainly arising from ensuring intermediate OAR doses are appropriately optimised. For compliant plans, improvements identified by KBP were in intermediate doses of bowel and ipsilateral kidney, which likely reflects a number of scenarios. Firstly in SABR planning there may not be as much emphasis placed on reducing low and intermediate dose wash in OARs. Secondly, the KBP plan reflects trade-offs between OARs that exist in the model plans, which might be different to trade-offs made by the submitting institution. Lastly, more uniform dose fall-off outside the target may have been preferred over sharp fall off at specific bowel loops. Providing images of dose distributions may assist in interpretation of the KBP plan and help in these scenarios, as recommended by one centre who provided feedback. A further limitation of our approach was to use a single KBP model for both fractionations. Although the relative OAR constraints as a percentage of the prescription dose was very similar in most OARs, the constraints were typically more restrictive for the three fraction regime. Model objectives were set to meet the three fraction constraints, but further improvement may be achievable by having fraction-specific models.
There are a number of challenges with interpretation of KBP generated estimates for deciding whether to re-optimise a plan. Firstly, we provided per-structure recommendations. Certain OARs may have been prioritised over others at the discretion of the treating team, and this is not taken into account by the KBP-based recommendations. Secondly, we elected to suggest OAR sparing when the KBP achieved metric was at least 10% lower than that in the submitted plan, as opposed to any gain. This was performed in attempt to ensure suggestion was made when clinically meaningful. Giaddui et al. suggest 5% is suitable for head and neck cancer, based on analysis of potential gains and link to calibration dosimetry. Ideally improvement is suggested based on normal tissue complication probability (NTCP) [24]. That is, if the level of OAR dose is where the NTCP curve is flat, a large change in OAR dose would be required to determine a clinically relevant gain, and where the NTCP curve is steep, a small change in dose would lead to a large gain. In abdominal SABR however, there is a lack of dose response data for relevant OARs, and constraints are mainly derived from serial structure toxicities. Lastly, it may also be argued that plan modulation metrics should also be considered as part of real-time review, both to ensure the submitted plan is not over-modulated, and to ensure that any KBP suggested improvements are not at the expense of over-modulation which may impact deliverability.
Short turnaround time impacts feasibility of KBP QA, as indicated by the lack of real-time KBP review provided for six cases and the feedback from participating centres. The 7 day criteria was an arbitrary timeline selected to allow manual handling of datasets and central transfer with a sufficient time period to enable the start date of radiotherapy within a clinical trial timeframe. An alternative approach is provision of a KBP model to participating centres at trial inception, facilitating estimated DVHs or objectives at time of rather than after treatment planning, such that that the KBP report is incorporated within the final plan check process. This would not necessitate such long lead-times as was designed in this trial. Further, improvements in plan quality must be balanced with additional resources required with plan modification. Further, it is proposed that a KBP model or OAR DVHs are summarised from the treatment plans used for treatment, to facilitate replication of trial results in the real-world setting [27].

Conclusion

In this prospective trial evaluating safety and efficacy of SABR for kidney cancer, we have demonstrated KBP for real-time review of treatment plan quality can be feasible. Challenges were faced with timelines between plan submission, provision of the KBP report, and treatment start date. KBP demonstrated utility in highlighting plans that exceeded tolerance and may be improved through further optimisation. There are however some potential improvements to this process, such as provision of KBP models to participating sites at the start of the trial, and improved quality improvement thresholds. Such improvements may result in substantial potential for automation of real-time review adding further emphasis on plan quality.

Acknowledgements

Not applicable.

Declarations

All participants consented to participate in the FASTRACK II clinical trial. Ethics approval was provided by Peter MacCallum Cancer Centre.
Not applicable.

Competing interests

KLM reports grants, personal fees and non-financial support from Varian Medical Systems, outside the submitted work. In addition, KLM has a patent Developing Predictive Dose-Volume Relationships for a Radiotherapy Treatment licensed to Varian Medical Systems, and a patent Knowledge-based prediction of three-dimensional dose distributions licensed to Varian Medical Systems. SS has received speaker fee for educational event or scientific meeting from Varian Medical Systems.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
2.
Zurück zum Zitat Peters LJ, O’Sullivan B, Giralt J, Fitzgerald TJ, Trotti A, Bernier J, et al. Critical impact of radiotherapy protocol compliance and quality in the treatment of advanced head and neck cancer: Results from TROG 02.02. J Clin Oncol. [Internet]. 2010 [cited 2020 Aug 10];28(18):2996–3001. https://pubmed.ncbi.nlm.nih.gov/20479390/ Peters LJ, O’Sullivan B, Giralt J, Fitzgerald TJ, Trotti A, Bernier J, et al. Critical impact of radiotherapy protocol compliance and quality in the treatment of advanced head and neck cancer: Results from TROG 02.02. J Clin Oncol. [Internet]. 2010 [cited 2020 Aug 10];28(18):2996–3001. https://​pubmed.​ncbi.​nlm.​nih.​gov/​20479390/​
3.
Zurück zum Zitat Fairchild A, Straube W, Laurie F, Followill D. Does quality of radiation therapy predict outcomes of multicenter cooperative group trials? A literature review [Internet]. Vol. 87, International Journal of Radiation Oncology Biology Physics. Int J Radiat Oncol Biol Phys.; 2013 [cited 2020 Aug 10]. p. 246–60. https://pubmed.ncbi.nlm.nih.gov/23683829/ Fairchild A, Straube W, Laurie F, Followill D. Does quality of radiation therapy predict outcomes of multicenter cooperative group trials? A literature review [Internet]. Vol. 87, International Journal of Radiation Oncology Biology Physics. Int J Radiat Oncol Biol Phys.; 2013 [cited 2020 Aug 10]. p. 246–60. https://​pubmed.​ncbi.​nlm.​nih.​gov/​23683829/​
4.
Zurück zum Zitat Kron T, Chesson B, Hardcastle N, Crain M, Clements N, Burns M, et al. Credentialing of radiotherapy centres in Australasia for TROG 09.02 (Chisel), a Phase III clinical trial on stereotactic ablative body radiotherapy of early stage lung cancer. Br J Radiol. [Internet]. 2018 [cited 2019 Sep 11];91(1085):20170737. http://www.ncbi.nlm.nih.gov/pubmed/29446317 Kron T, Chesson B, Hardcastle N, Crain M, Clements N, Burns M, et al. Credentialing of radiotherapy centres in Australasia for TROG 09.02 (Chisel), a Phase III clinical trial on stereotactic ablative body radiotherapy of early stage lung cancer. Br J Radiol. [Internet]. 2018 [cited 2019 Sep 11];91(1085):20170737. http://​www.​ncbi.​nlm.​nih.​gov/​pubmed/​29446317
6.
Zurück zum Zitat Siva S, Chesson B, Bressel M, Pryor D, Higgs B, Reynolds HM, et al. TROG 15.03 phase II clinical trial of Focal Ablative STereotactic Radiosurgery for Cancers of the Kidney—FASTRACK II 11 Medical and Health Sciences 1103 Clinical Sciences 11 Medical and Health Sciences 1112 Oncology and Carcinogenesis. BMC Cancer [Internet]. 2018 [cited 2019 Jun 12];18(1):1030. http://www.ncbi.nlm.nih.gov/pubmed/30352550 Siva S, Chesson B, Bressel M, Pryor D, Higgs B, Reynolds HM, et al. TROG 15.03 phase II clinical trial of Focal Ablative STereotactic Radiosurgery for Cancers of the Kidney—FASTRACK II 11 Medical and Health Sciences 1103 Clinical Sciences 11 Medical and Health Sciences 1112 Oncology and Carcinogenesis. BMC Cancer [Internet]. 2018 [cited 2019 Jun 12];18(1):1030. http://​www.​ncbi.​nlm.​nih.​gov/​pubmed/​30352550
7.
Zurück zum Zitat Siva S, Correa RJM, Warner A, Staehler M, Ellis RJ, Ponsky L, et al. Stereotactic ablative radiotherapy for ≥T1b primary renal cell carcinoma: a report from the International Radiosurgery Oncology Consortium for Kidney (IROCK). Int J Radiat Oncol Biol Phys. 2020;108:941–9.CrossRef Siva S, Correa RJM, Warner A, Staehler M, Ellis RJ, Ponsky L, et al. Stereotactic ablative radiotherapy for ≥T1b primary renal cell carcinoma: a report from the International Radiosurgery Oncology Consortium for Kidney (IROCK). Int J Radiat Oncol Biol Phys. 2020;108:941–9.CrossRef
8.
Zurück zum Zitat Siva S, Jackson P, Kron T, Bressel M, Lau E, Hofman M, et al. Impact of stereotactic radiotherapy on kidney function in primary renal cell carcinoma: establishing a dose-response relationship. Radiother Oncol. 2016;118(3):540–6.CrossRef Siva S, Jackson P, Kron T, Bressel M, Lau E, Hofman M, et al. Impact of stereotactic radiotherapy on kidney function in primary renal cell carcinoma: establishing a dose-response relationship. Radiother Oncol. 2016;118(3):540–6.CrossRef
9.
Zurück zum Zitat Pham D, Kron T, Bressel M, Foroudi F, Hardcastle N, Schneider M, et al. Image guidance and stabilization for stereotactic ablative body radiation therapy (SABR) treatment of primary kidney cancer. Pract Radiat Oncol. 2015;5(6):e597-605.CrossRef Pham D, Kron T, Bressel M, Foroudi F, Hardcastle N, Schneider M, et al. Image guidance and stabilization for stereotactic ablative body radiation therapy (SABR) treatment of primary kidney cancer. Pract Radiat Oncol. 2015;5(6):e597-605.CrossRef
10.
Zurück zum Zitat Li N, Carmona R, Sirak I, Kasaova L, Followill D, Michalski J, et al. Highly efficient training, refinement, and validation of a knowledge-based planning quality-control system for radiation therapy clinical trials. Int J Radiat Oncol Biol Phys. 2017;97(1):164–72.CrossRef Li N, Carmona R, Sirak I, Kasaova L, Followill D, Michalski J, et al. Highly efficient training, refinement, and validation of a knowledge-based planning quality-control system for radiation therapy clinical trials. Int J Radiat Oncol Biol Phys. 2017;97(1):164–72.CrossRef
13.
Zurück zum Zitat Kavanaugh JA, Holler S, DeWees TA, Robinson CG, Bradley JD, Iyengar P, et al. Multi-institutional validation of a knowledge-based planning model for patients enrolled in RTOG 0617: implications for plan quality controls in cooperative group trials. Pract Radiat Oncol. 2019;9(2):e218–27.CrossRef Kavanaugh JA, Holler S, DeWees TA, Robinson CG, Bradley JD, Iyengar P, et al. Multi-institutional validation of a knowledge-based planning model for patients enrolled in RTOG 0617: implications for plan quality controls in cooperative group trials. Pract Radiat Oncol. 2019;9(2):e218–27.CrossRef
16.
Zurück zum Zitat Giaddui T, Geng H, Chen Q, Linnemann N, Radden M, Lee NY, et al. Offline quality assurance for intensity modulated radiation therapy treatment plans for NRG-HN001 head and neck clinical trial using knowledge-based planning. Adv Radiat Oncol [Internet]. 2020. https://doi.org/10.1016/j.adro.2020.05.005 Giaddui T, Geng H, Chen Q, Linnemann N, Radden M, Lee NY, et al. Offline quality assurance for intensity modulated radiation therapy treatment plans for NRG-HN001 head and neck clinical trial using knowledge-based planning. Adv Radiat Oncol [Internet]. 2020. https://​doi.​org/​10.​1016/​j.​adro.​2020.​05.​005
19.
Zurück zum Zitat Berry SL, Ma R, Boczkowski A, Jackson A, Zhang P, Hunt M. Evaluating inter-campus plan consistency using a knowledge based planning model. Radiother Oncol. 2016;120(2):349–55.CrossRef Berry SL, Ma R, Boczkowski A, Jackson A, Zhang P, Hunt M. Evaluating inter-campus plan consistency using a knowledge based planning model. Radiother Oncol. 2016;120(2):349–55.CrossRef
22.
Zurück zum Zitat Tol JP, Dahele M, Gregoire V, Overgaard J, Slotman BJ, Verbakel WFAR. Analysis of EORTC-1219-DAHANCA-29 trial plans demonstrates the potential of knowledge-based planning to provide patient-specific treatment plan quality assurance. Radiother Oncol. 2019;1(130):75–81.CrossRef Tol JP, Dahele M, Gregoire V, Overgaard J, Slotman BJ, Verbakel WFAR. Analysis of EORTC-1219-DAHANCA-29 trial plans demonstrates the potential of knowledge-based planning to provide patient-specific treatment plan quality assurance. Radiother Oncol. 2019;1(130):75–81.CrossRef
24.
Zurück zum Zitat Moore KL, Schmidt R, Moiseenko V, Olsen LA, Tan J, Xiao Y, et al. Quantifying unnecessary normal tissue complication risks due to suboptimal planning: a secondary study of RTOG 0126. Int J Radiat Oncol Biol Phys [Internet]. 2015 [cited 2020 Aug 10];92(2):228–35. https://pubmed.ncbi.nlm.nih.gov/25847605/ Moore KL, Schmidt R, Moiseenko V, Olsen LA, Tan J, Xiao Y, et al. Quantifying unnecessary normal tissue complication risks due to suboptimal planning: a secondary study of RTOG 0126. Int J Radiat Oncol Biol Phys [Internet]. 2015 [cited 2020 Aug 10];92(2):228–35. https://​pubmed.​ncbi.​nlm.​nih.​gov/​25847605/​
26.
Zurück zum Zitat Yusufaly TI, Meyers SM, Mell LK, Moore KL. Knowledge-Based Planning for Intact Cervical Cancer. Vol. 30, Seminars in Radiation Oncology. W.B. Saunders; 2020. p. 328–39. Yusufaly TI, Meyers SM, Mell LK, Moore KL. Knowledge-Based Planning for Intact Cervical Cancer. Vol. 30, Seminars in Radiation Oncology. W.B. Saunders; 2020. p. 328–39.
27.
Zurück zum Zitat Knutson NC, Samson P, Hugo G, Goddu SM, Reynoso F, Kavanaugh J, et al. Radiotherapy Workflow and Dosimetric Analysis from a Phase I/II Trial of Noninvasive Cardiac Radioablation for Ventricular Tachycardia. Int J Radiat Oncol [Internet]. 2019 [cited 2019 May 12]. http://www.ncbi.nlm.nih.gov/pubmed/31002942 Knutson NC, Samson P, Hugo G, Goddu SM, Reynoso F, Kavanaugh J, et al. Radiotherapy Workflow and Dosimetric Analysis from a Phase I/II Trial of Noninvasive Cardiac Radioablation for Ventricular Tachycardia. Int J Radiat Oncol [Internet]. 2019 [cited 2019 May 12]. http://​www.​ncbi.​nlm.​nih.​gov/​pubmed/​31002942
Metadaten
Titel
Personalising treatment plan quality review with knowledge-based planning in the TROG 15.03 trial for stereotactic ablative body radiotherapy in primary kidney cancer
verfasst von
Nicholas Hardcastle
Olivia Cook
Xenia Ray
Alisha Moore
Kevin L. Moore
David Pryor
Alana Rossi
Farshad Foroudi
Tomas Kron
Shankar Siva
Publikationsdatum
01.12.2021
Verlag
BioMed Central
Erschienen in
Radiation Oncology / Ausgabe 1/2021
Elektronische ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-021-01820-7

Weitere Artikel der Ausgabe 1/2021

Radiation Oncology 1/2021 Zur Ausgabe

Erhebliches Risiko für Kehlkopfkrebs bei mäßiger Dysplasie

29.05.2024 Larynxkarzinom Nachrichten

Fast ein Viertel der Personen mit mäßig dysplastischen Stimmlippenläsionen entwickelt einen Kehlkopftumor. Solche Personen benötigen daher eine besonders enge ärztliche Überwachung.

15% bedauern gewählte Blasenkrebs-Therapie

29.05.2024 Urothelkarzinom Nachrichten

Ob Patienten und Patientinnen mit neu diagnostiziertem Blasenkrebs ein Jahr später Bedauern über die Therapieentscheidung empfinden, wird einer Studie aus England zufolge von der Radikalität und dem Erfolg des Eingriffs beeinflusst.

Erhöhtes Risiko fürs Herz unter Checkpointhemmer-Therapie

28.05.2024 Nebenwirkungen der Krebstherapie Nachrichten

Kardiotoxische Nebenwirkungen einer Therapie mit Immuncheckpointhemmern mögen selten sein – wenn sie aber auftreten, wird es für Patienten oft lebensgefährlich. Voruntersuchung und Monitoring sind daher obligat.

Costims – das nächste heiße Ding in der Krebstherapie?

28.05.2024 Onkologische Immuntherapie Nachrichten

„Kalte“ Tumoren werden heiß – CD28-kostimulatorische Antikörper sollen dies ermöglichen. Am besten könnten diese in Kombination mit BiTEs und Checkpointhemmern wirken. Erste klinische Studien laufen bereits.

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.