Erschienen in: Journal of Cardiovascular Magnetic Resonance 1/2023

Open Access 01.12.2023 | Research

Repeatability and reproducibility of various 4D Flow MRI postprocessing software programs in a multi-software and multi-vendor cross-over comparison study

verfasst von: Thekla H. Oechtering, André Nowak, Malte M. Sieren, Andreas M. Stroth, Nicolas Kirschke, Franz Wegner, Maren Balks, Inke R. König, Ning Jin, Joachim Graessner, Hendrik Kooijman-Kurfuerst, Anja Hennemuth, Jörg Barkhausen, Alex Frydrychowicz

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2023

Abstract

Background

Different software programs are available for the evaluation of 4D Flow cardiovascular magnetic resonance (CMR). A good agreement of the results between programs is a prerequisite for the acceptance of the method. Therefore, the goal was to compare quantitative results from a cross-over comparison in individuals examined on two scanners of different vendors analyzed with four postprocessing software packages.

Methods

Eight healthy subjects (27 ± 3 years, 3 women) were each examined on two 3T CMR systems (Ingenia, Philips Healthcare; MAGNETOM Skyra, Siemens Healthineers) with a standardized 4D Flow CMR sequence. Six manually placed aortic contours were evaluated with Caas (Pie Medical Imaging, SW-A), cvi42 (Circle Cardiovascular Imaging, SW-B), GTFlow (GyroTools, SW-C), and MevisFlow (Fraunhofer Institute MEVIS, SW-D) to analyze seven clinically used parameters including stroke volume, peak flow, peak velocity, and area as well as typically scientifically used wall shear stress values. Statistical analysis of inter- and intrareader variability, inter-software and inter-scanner comparison included calculation of absolute and relative error (ER), intraclass correlation coefficient (ICC), Bland–Altman analysis, and equivalence testing based on the assumption that inter-software differences needed to be within 80% of the range of intrareader differences.

Results

SW-A and SW-C were the only software programs showing agreement for stroke volume (ICC = 0.96; ER = 3 ± 8%), peak flow (ICC: 0.97; ER = −1 ± 7%), and area (ICC = 0.81; ER = 2 ± 22%). Results from SW-A/D and SW-C/D were equivalent only for area and peak flow. Other software pairs did not yield equivalent results for routinely used clinical parameters. Especially peak maximum velocity yielded poor agreement (ICC ≤ 0.4) between all software packages except SW-A/D that showed good agreement (ICC = 0.80). Inter- and intrareader consistency for clinically used parameters was best for SW-A and SW-D (ICC = 0.56–97) and worst for SW-B (ICC = -0.01–0.71). Of note, inter-scanner differences per individual tended to be smaller than inter-software differences.

Conclusions

Of all tested software programs, only SW-A and SW-C can be used equivalently for determination of stroke volume, peak flow, and vessel area. Irrespective of the applied software and scanner, high intra- and interreader variability for all parameters have to be taken into account before introducing 4D Flow CMR in clinical routine. Especially in multicenter clinical trials a single image evaluation software should be applied.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
4D Flow CMR
Time-resolved 3-dimensional cardiovascular magnetic resonance
bwSV
Backward stroke value = Reflux
CMR
Cardiovascular magnetic resonance
MRI1
Ingenia, Philips Healthcare, Best, the Netherlands
MRI2
MAGNETOM Skyra, Siemens Healthineers, Erlangen, Germany
fwSV
Forward stroke volume
LLOA
Lower limit of agreement
LOA
95% Limit of agreement
netSV
Net stroke volume
PC
Phase contrast
SW-A
Caas (v5.01, Pie Medical, Maastricht, Netherlands)
SW-B
Cvi42 (v5.9.4, Circle Cardiovascular Imaging, Alberta, Canada, Canada)
SW-C
GTFlow (v3.1.13, GyroTools, Zurich, Switzerland)
SW-D
MEVISFlow (v10.3, MEVIS Fraunhofer, Germany)
ULOA
Upper limit of agreement
Vavg
Peak velocity averaged over a contour
Vmax
Peak maximum velocity of a contour, i.e. the highest velocity that was measured at one voxel in a contour during the cardiac cycle
WSS
Wall shear stress
WSScon
Peak wall shear stress value of wall shear stress averaged over a contour (“peak WSS per contour”)
WSSseg
Peak value of wall shear stress per segment, i.e., the highest wall shear stress that was measured in one of eight segments per contour (“peak WSS per segment”)

Background

4D Flow cardiovascular magnetic resonance (CMR) offers the unique opportunity to gather time-resolved, 3-dimensional and 3-directional flow information of the heart and blood vessels non-invasively and without contrast agent. Using 4D Flow CMR, routinely used basic flow parameters such as stroke volume, velocities, and flow volumes can be evaluated simultaneously with matching geometric information and advanced derived parameters such as wall shear stress, pressure gradients, and turbulent kinetic energy [1]. Furthermore, 4D Flow CMR allows retrospective analysis of a 3D volume with complex vascular anatomies and flow patterns without the necessity of repeated scans and planning. This makes it an ideal technique for the evaluation of patients with complex anatomy such as congenital heart disease [2].
Reproducibility of results between both different scanners and post-processing software remains to be addressed [1, 3]. Unconfirmed reproducibility limits the possibility to perform meta-analyses and imposes bias to any multicenter trials in order to determine the clinical relevance of 4D Flow CMR and its vast number of available parameters. And while most studies nowadays comply to the scan recommendations of the consensus paper by Dyverfeldt et al. [1], there is a plethora of measured and deducted parameters extracted by a considerable number of different software applications, some commercially available, some home-built lacking thorough comparison or standardization. However, software programs must provide repeatable and reproducible results to be used interchangeably. While the variability of measurements from different sequences and vendors has been recognized [48], there is a relevant lack of comparative studies analyzing interchangeability of 4D Flow CMR postprocessing software.
Hence, the goal of this study was to perform a cross-over comparative study of 4D Flow CMR software analyzing data of healthy individuals’ thoracic aorta scanned on two 3 T MRI scanners of different vendors. As primary endpoint, repeatability and reproducibility of quantitative 4D Flow CMR results from four different postprocessing software programs as well as reproducibility of results between different software programs were to be evaluated. As secondary endpoint, reproducibility between MRI systems by different vendors was analyzed.

Methods

CMR scan

Thoracic aortic 4D Flow CMR was conducted in eight healthy subjects (27 ± 3 years, full demographics in Table 1) scanned on two 3 T CMR scanners (MRI1 = Ingenia, Philips Healthcare, Best, The Netherlands; MRI2 = MAGNETOM Skyra, Siemens Healthcare, Erlangen, Germany) using scan parameters per guideline recommendations [1]. On MRI2, a prototype 4D Flow CMR sequence was used. Acquisition settings for both sequences were carefully adapted to match each other closely (Table 2). The adaption of these parameters resulted in a nominal echo time (TE)/repetition time (TR) of 1.7/3.0 ms for MRI1 and 2.2/4.8 ms for MRI2. To exclude circadian effects on hemodynamics, both scans were acquired at the same time of day. To exclude effects of digestion, all participants fasted 2 h prior to CMR scans.
Table 1
Demographics
Variables
Healthy subjects (n = 8)
Age (years)
27 ± 3
Height (cm)
176 ± 6
Weight (kg)
80 ± 15
Body mass index (m2)
25.5 ± 3.1
Gender ratio (male: female)
5:3
Systolic blood pressure (mmHg)
131 ± 16
Diastolic blood pressure (mmHg)
80 ± 8
Heart rate (bpm)
66 ± 8
Table 2
Typical scan parameters
Parameter
Unit
MRI1
MRI2
Field-of-view
mm3
290 × 290 × 56
312 × 384 × 50
Acquired spatial resolution
mm3
2.5 × 2.5 × 2.5
2.5 × 2.0 × 2.5
Reconstructed spatial resolution
mm3
2.0 × 2.0 × 2.0
2.0 × 2.0 ×  2.0
Number of reconstructed heart phases
 
24
24
ECG synchronization
 
Retrospective
Retrospective
Respiratory motion compensation
 
Gating window 8 mm
Gating window 8 mm
Flip Angle
°
8
7
Parallel Imaging
 
SENSE acceleration factor: 2.2
GRAPPA acceleration factor: 3.0
Velocity-encoding factor
cm/s
200
200
ECG, electrocardiogram

Processing of CMR data

The resulting 16 datasets were evaluated with four different software packages that were available at our hospital: Three programs were commercially available [SW-A = Caas (v5.01, Pie Medical, Maastricht, the Netherlands), SW-B = cvi42 (v5.9.4, Circle Cardiovascular Imaging, Calgary, Alberta, Canada), and SW-C = GTFlow (v3.1.13, GyroTools, Zurich, Switzerland), alphabetical order]. One additional program was only available through a research collaboration [SW-D = MEVISFlow (v10.3, MEVIS Fraunhofer, Germany)]. Programs were installed on the same computer (Intel® Xeon® CPU E5-1620 v3 @ 3.50 GHz processor, 16 GB RAM, NVIDIA Quadro K4200 graphic card). Sample screenshots of the graphical user interface of the individual software programs can be seen in Fig. 1.
All datasets were independently evaluated with each software by two readers who were blinded to the other software programs’ and the other reader’s results. To account for real-life scenario and potential errors, each reader underwent the entire data evaluation process for each software including data filtering, analysis plane positioning and segmentation. Both readers did not have prior experience with 4D flow CMR analysis to exclude bias by different levels of familiarity with the programs. They were trained by experienced readers and software representatives for the purpose of this study. For intrareader comparison, all datasets were re-evaluated by one reader after three to six months to avoid recognition effects. Every postprocessing workflow started with loading the DICOM data into the respective software. In each program, phase background offsets caused by eddy currents were corrected. SW-B, SW-C, and SW-D allowed manual thresholding for the selection of static tissue. Differences in background phase offset correction between the programs included different grades of polynomial fitting (SW-A and SW-B: linear; SW-C: quadratic; SW-D: cubic).
In SW-A, SW-B, and SW-D, a 3D vessel contour of the thoracic aorta was delineated semi-manually, and voxels outside this region of interest were ignored during the analysis. In SW-B, manual changes of the contour delineation were inevitably corrected automatically. There was no way to turn off the auto-correction. This made it impossible for users to delineate the vessel exactly as they intended (Fig. 1). In SW-C, a noise masking was applied. Six contours were carefully placed manually at six predefined sites of the thoracic aorta using B-splines in multiplanar reformatted planes (Fig. 2). Contours were manually adjusted for each time frame. Typically used clinical parameters were evaluated: forward, backward, and net stroke volumes (fwSV, bwSV, netSV, respectively [ml]), peak flow [ml/s], peak maximum velocity of a contour, i.e. the highest velocity that was measured at one voxel in a contour (Vmax [cm/s]), peak velocity averaged over a contour (Vavg [cm/s]), and area at time of peak flow (Area [mm2]). Moreover, peak values of wall shear stress (WSS [mPa]) were evaluated averaged over a contour (WSScon) as well as for the maximum segmental WSS of the contour that was subdivided into 8 segments (WSSseg). SW-A calculated WSS as described by Perinajova et al. [9]. SW-B and SW-C calculated WSS according to Stalder et al. [10]. Zimmermann et al. [6] described the approach to calculating WSS as implemented by SW-D. In SW-A, SW-C and SW-D, the viscosity of blood assumed for the WSS calculations could be manually adapted and was set to 3.2 mPa*s to match the fixed value of SW-B. SW-B offered no segmental maximum WSS values, but only WSS values averaged over the whole contour.
Different software issued different names for the evaluated parameters (Table 3). Duration of evaluation was noted for the second evaluation with every software, excluding the time needed for the import of digital imaging and communications in medicine (DICOM) data.
Table 3
Different software issue different names for parameters
Nomenclature of this publication
SW-A
SW-B
SW-C
SW-D
Stroke volume (ml)
Pumped blood volume (ml)
Total Volume (ml)
Stroke Volume (ml)
Flow (l)
Forward stroke volume (ml)
Forward flow (ml)
Total Forward
Volume (ml)
Net Forward Flow Volume (ml)
Forward (l)
Backward stroke volume (ml)
Backward flow (ml)
Total Backward Volume (ml)
Net Backward Flow Volume (ml)
Backward (l)
Peak flow (ml/s)
Flow Graph (ml/s)
Maximum Flow (ml/s)
peak value of time resolved “Net Flow” (ml/s)
peak value of time resolved “Flow” (l/s)
Peak maximum velocity (cm/s)
Max Velocity Graph (cm/s)
peak value of time resolved “Max Mag” (cm/s)
peak value of time resolved “Velocity Max” (cm/s)
peak value of time resolved “Velocity Max” (m/s)
Peak average velocity (cm/s)
Mean Velocity Graph (cm/s)
Maximum Mean Velocity (cm/s)
peak value of time resolved „Velocity Avg “ (cm/s)
peak value of time resolved “Velocity Mean” (m/s)
Area (mm2)
Contour Area Graph (mm2)
Area (mm2)
Area (mm2)
Area (mm2)
Wall shear stress (WSS)
WSS (mPa)
WSS1,2 (mPa)
WSS1,2 (Pa)
WSS Magnitude1 (N/m2)
WSS No projection1,2,3 (Pa)
Temporal resolution
Yes
Yes
Yes
Yes
Spatial resolution
1 contour with n (90) Vectors
1 contour, no segments
1 contour with n (8) segments
1 contour with n (8) segments
Peak WSS per segment
Calculated:
11 vectors got grouped together. Peak of those mean values
Peak value of time resolved, “Segment 1 WSS Mag.” to “Segment 8 WSS Mag.”
Peak value of segmentally resolved “Max WSS(mean) No Projection”
Peak WSS per contour
Calculated:
Mean value of all WSS vectors
“Max Wall Shear Stress”
Peak value of time resolved “Avg WSS Mag”
Calculated:
Peak value of the calculated mean WSS No Projection value of segments 1–8 calculated separately for each time point
1 Axial WSS also available; 2circumferential WSS also available; 3 oscillatory shear index also available

Statistics

Repeatability and interreader reproducibility of 4D Flow CMR results were evaluated by intra- and interreader comparison, respectively. Reproducibility of results by different software programs was evaluated by inter-software comparison. To better place these results into context, reproducibility between scanners was evaluated by inter-scanner comparison determined with the software that showed best repeatability and reproducibility. As statistical measures for repeatability and reproducibility, the agreement between measurement pairs (e.g., between reader, software, scanner) was tested using Bland–Altman analysis and intraclass correlation coefficient (ICC) [1113].
For all parameters, average absolute and relative differences, and standard deviation (SD) between two comparative pairs (methods), i.e., readers, software, or scanners, were calculated. The relative difference ΔR(X–Y) for each parameter measured by methods X and Y was calculated as follows:
$${\Delta }_{R(X-Y)}=\frac{X-Y}{(X+Y)/2}$$
(1)
where the difference between the results of methods X and Y is divided by the mean of methods X and Y. Values are presented as mean ± SD.
Bland–Altman analyses were performed to determine the average bias and the 95% limits of agreement between each software combination for each parameter. Limits of agreement (LOA) were calculated as bias ± 1.96 * SD. In the same fashion, inter-scanner and intra- and interreader variability were statistically evaluated.
ICC estimates and their 95% confidence intervals (CI) for intrareader comparison were calculated based on a single rating, absolute-agreement, 2-way mixed-effects model. For interreader, inter-software, and inter-scanner comparisons, ICC estimates and their 95% CI were calculated based on a single rating, absolute-agreement, 2-way random-effects model [14]. ICC values less than 0.5, between 0.5 and 0.75, between 0.75 and 0.9, and greater than 0.9 are indicative of poor, moderate, good, and excellent agreement, respectively [14].
Using the described ICC estimates and results from the Bland–Altman analysis, we aimed at determining equivalence between software programs. For this, we adapted the equivalence test proposed by Zange et al. [15] to include the degree of intrareader variability in the assessment of reproducibility between software packages. Specifically, we based our assessment of differences between software on the intrareader variability: inter-software variability cannot be lower than intrareader variability of one single software. We considered software programs to be equivalent if X% of comparisons between two software programs were within the limits of agreement obtained from the intrareader variability. We determined the threshold X based on results from the interreader comparison: We determined the percentage of measurement comparisons between two readers that were within the 95% limits of agreement from the intrareader comparison, focusing only on the clinically used parameters area, stroke volume, flow, and velocity. Therefore, the narrowest limits of agreement obtained from the intrareader variability derived from Bland–Altman analysis were chosen and defined as equivalence limits. Equivalence was concluded for a parameter if X% of the absolute bias results between two software packages were completely within the limits of equivalence and ICC analysis displayed at least moderate agreement. SPSS (version 26.0 Statistical Package for the Social Sciences, International Business Machines, Inc., Armonk, New York, USA) was used for statistical analysis.

Results

All eight subjects were successfully scanned at both scanners. Acquisition time was 12.4 ± 2.5 min for MRI1 and 15.7 ± 5.5 min for MRI2. Respiratory navigator acceptance rate was 55 ± 17%. SW-B and SW-D failed to open one of eight scans from MRI1, although not the same dataset concerned. There were no problems evaluating these datasets with the other software programs. Standard customer support of the respective software programs was contacted but could not resolve the issue. All other datasets could be analyzed. There was no aliasing and therefore no need for aliasing correction. Figure 3 depicts data of the ascending aortic contour for all volunteers, scanners, and software.

Intrareader variability as measure of repeatability

Intrareader variability between software programs varied considerably. As detailed in Table 4, the intrareader analysis of SW-A revealed best repeatability with good to excellent agreement for all parameters and lowest standard deviation of relative errors between 7 and 15%. Bland–Altman analysis revealed smallest bias and narrowest limits of agreement for intrareader variability of most parameters using SW-A (e.g., LOA for stroke volume: − 12 to 12 ml, peak maximum velocity: − 23 to 21 cm/s, area: − 175 to 205 mm2). SW-C and SW-D showed moderate to excellent agreement except for SW-C concerning peak maximum velocity and WSS values, where ICC revealed poor agreement.
Table 4
Intrareader variability for all contours combined
 
SW-A
SW-B
SW-C
SW-D
Stroke volume
Absolute error [ml]
0 ± 6
 − 1 ± 18
2 ± 6
0 ± 12
Relative error [%]
1 ± 7
 − 2 ± 26
3 ± 8
0 ± 16
ICC
0.967***
0.648*
0.964***
0.790**
Peak flow
Absolute error [ml/s]
4 ± 27
 − 3 ± 89
26 ± 92
5 ± 41
Relative error [%]
1 ± 7
 − 1 ± 27
8 ± 27
1 ± 11
ICC
0.969***
0.644*
0.663*
0.917***
Peak maximum velocity
Absolute error [cm/s]
 − 1 ± 11
 − 1 ± 28
5 ± 26
 − 2 ±  − 20
Relative error [%]
 − 1 ± 8
 − 1 ± 21
4 ± 19
 − 1 ± 13
ICC
0.883**
0.219
0.348
0.746*
Peak average velocity
Absolute error [cm/s]
 − 1 ± 10
 − 2 ± 17
9 ± 10
3 ± 7
Relative error [%]
 − 1 ± 14
 − 2 ± 21
12 ± 13
4 ± 8
ICC
0.809**
0.394
0.541*
0.839**
Area
Absolute error [mm2]
15 ± 97
13 ± 190
 − 80 ± 143
 − 21 ± 87
Relative error [%]
3 ± 15
2 ± 29
 − 12 ± 19
 − 4 ± 15
ICC
0.919***
0.705*
0.788**
0.915***
Peak WSS per segment
Absolute error [mPa]
8 ± 133
737 ± 847
 − 137 ± 614
Relative error [%]
1 ± 9
44 ± 50
 − 5 ± 29
ICC
0.921***
0.068
0.677*
Peak WSS per contour
Absolute error [mPa]
14 ± 121
 − 32 ± 54
567 ± 563
 − 12 ± 156
Relative error [%]
2 ± 11
 − 7 ± 12
60 ± 56
 − 1 ± 22
ICC
0.921***
0.944***
0.068
0.706*
Absolute and relative error are given as mean ± standard deviation. Values of bias and standard deviation yielding to smallest limits of agreement within one row are marked as bold indicating that they were used to calculate equivalence limits. Asterisks indicate ***excellent, **good, and *moderate intraclass correlation (ICC). No asterisk indicates poor agreement. WSS, wall shear stress
SW-B yielded the worst repeatability with poor to moderate agreement for most parameters, with standard deviations for stroke volume, peak flow, and peak velocities varying between 21 and 27%. This is underlined by broadest limits of agreement that were found with SW-B for stroke volume (− 36 to 34 ml), peak velocities (e.g., peak maximum velocity: − 56 to 54 cm/s), and area (− 359 to 385 mm2). Relative error of bwSV was high in all cases due to small absolute values, while absolute error of bwSV was low with every software.

Interreader variability as measure of reproducibility

As expected, interreader agreement was lower than intrareader agreement with few exceptions. Lowest interreader variability was found with SW-A with moderate to excellent agreement for all parameters yielding to standard deviations of differences between readers between 10 and 22% (Table 5). Smallest limits of agreement between two readers were detected with SW-A for stroke volumes (SV: − 19 to 17 ml), peak flow (F: − 81 to 67 ml/s), and peak velocities (Vmax: − 32 to 30 cm/s), closely followed by SW-D, which displayed narrowest limits of agreement for area (A: − 65 to 229 mm2).
Table 5
Interreader variability for all contours combined
 
SW-A
SW-B
SW-C
SW-D
Stroke volume
Absolute error [ml]
 − 1 ± 9
0 ± 20
7 ± 12
 − 1 ± 12
Relative error [%]
 − 1 ± 11
0 ± 28
9 ± 15
 − 2 ± 16
ICC
0.920***
0.569*
0.823**
0.809**
Peak flow
Absolute error [ml/s]
 − 7 ± 38
 − 5 ± 100
37 ± 45
1 ± 45
Relative error [%]
 − 2 ± 10
2 ± 32
11 ± 12
1 ± 16
ICC
0.937***
0.565*
0.862**
0.907***
Peak maximum velocity
Absolute error [cm/s]
 − 1 ± 16
14 ± 27
3 ± 28
 − 2 ± 19
Relative error [%]
 − 1 ± 13
12 ± 23
2 ± 19
 − 1 ± 13
ICC
0.705*
0.274
0.359
0.730*
Peak average velocity
Absolute error [cm/s]
2 ± 14
1 ± 17
 − 6 ± 10
2 ± 10
Relative error [%]
2 ± 22
1 ± 22
 − 7 ± 13
3 ± 12
ICC
0.556*
0.304
0.699*
0.754**
Area
Absolute error [mm2]
 − 18 ± 156
15 ± 192
92 ± 146
82 ± 75
Relative error [%]
 − 6 ± 22
0 ± 30
17 ± 22
 − 6 ± 19
ICC
0.757**
0.605*
0.671*
0.861**
Peak WSS per segment
Absolute error [mPa]
4 ± 190
599 ± 718
591 ± 755
Relative error [%]
0 ± 14
32 ± 42
18 ± 37
ICC
0.827**
0.227
0.480
Peak WSS per contour
Absolute error [mPa]
31 ± 160
50 ± 78
357 ± 513
118 ± 135
Relative error [%]
3 ± 15
10 ± 18
29 ± 52
 − 3 ± 26
ICC
0.857**
0.832**
0.221
0.573*
Absolute and relative error are given as mean ± standard deviation. Asterisks indicate ***excellent, **good, and *moderate intraclass correlation (ICC). No asterisk indicates poor agreement
Highest interreader variability was found with SW-B presenting poor to moderate agreement concerning clinical parameters. Broadest limits of agreement were found with SW-B for stroke volumes (SV: − 39 to 39 ml), peak flow (F: − 201 to 191 ml/s), peak velocities (Vmax: − 39 to 67 cm/s) and area (A: − 361 to 391 mm2). Highest bias was found with SW-C for stroke volumes, peak flow and area indicating systematic error between readers (Table 5).
In general, stroke volume and peak flow presented better reproducibility (moderate to good agreement) compared to peak velocities, which mostly displayed poor to moderate agreement between readers. WSS values displayed good agreement between both readers when they were using SW-A or SW-B, while SW-C and SW-D yielded mostly poor agreement.

Inter-software variability as measure of reproducibility

Stroke volumes

Only SW-A/C showed an excellent agreement for netSV and fwSV with smallest bias and narrowest limits of agreement (SV: LOA = − 14 to 10 ml, ER = 3 ± 8%) in the Bland–Altman analysis (Table 6, Fig. 4). In contrast, poor agreement was found with SW-B/D associated with broadest limits of agreement (SV: LOA = − 41 to 41 ml; ER = − 1 ± 28%), whereas other software pairs reached moderate agreement. SW-A/D, SW-B/C and SW-C/D presented with relatively broad LOA (lower limits of agreement LLOA: − 45 to − 14 ml; upper limits of agreement ULOA: 10–33 ml). Absolute errors for bwSV were small with high relative errors due to small absolute values. Limits of agreement for bwSV varied between − 4 and 4 ml.
Table 6
Software comparison: absolute and relative error and correlation coefficient
 
SW-A/B
SW-A/C
SW-A/D
SW-B/C
SW-B/D
SW-C/D
Net stroke volume
Absolute error [ml]
7 ± 19
2 ± 6
5 ± 17
6 ± 20
0 ± 21
3 ± 17
Relative error [%]
9 ± 26
3 ± 8
6 ± 21
 − 7 ± 27
 − 1 ± 28
3 ± 21
ICC
0.591*
0.959***
0.634*
0.589*
0.466
0.664*
Peak flow
Absolute error [ml/s]
24 ± 92
5 ± 27
17 ± 25
 − 29 ± 92
4 ± 78
22 ± 25
Relative error [%]
7 ± 26
 − 1 ± 7
5 ± 7
 − 8 ± 26
1 ± 23
6 ± 7
ICC
0.631*
0.971***
0.972***
0.645*
0.718*
0.971***
Peak maximum velocity
Absolute error [cm/s]
 − 5 ± 27
 − 13 ± 28
 − 5 ± 14
 − 10 ± 26
2 ± 28
8 ± 29
Relative error [%]
 − 4 ± 21
 − 10 ± 20
 − 4 ± 10
 − 4 ± 19
1 ± 20
6 ± 20
ICC
0.258
0.248
0.797**
0.401
0.352
0.328
Peak average velocity
Absolute error [cm/s]
 − 8 ± 18
 − 6 ± 15
 − 8 ± 10
1 ± 17
 − 1 ± 17
 − 3 ± 11
Relative error [%]
 − 12 ± 25
 − 9 ± 21
 − 13 ± 15
1 ± 21
 − 2 ± 22
 − 4 ± 14
ICC
0.276
0.505*
0.697*
0.314
0.314
0.688*
Area
Absolute error [mm2]
31 ± 174
21 ± 147
66 ± 84
1 ± 207
43 ± 160
48 ± 131
Relative error [%]
5 ± 26
2 ± 22
12 ± 14
 − 1 ± 32
8 ± 27
10 ± 21
ICC
0.745*
0.805**
0.898**
0.614*
0.740*
0.807**
Peak WSS per segment
Absolute error [mPa]
395 ± 760
212 ± 437
505 ± 789
Relative error [%]
 − 17 ± 45
20 ± 31
32 ± 49
ICC
 
0.168
0.371
  
0.104
Peak WSS per contour
Absolute error [mPa]
716 ± 195
64 ± 553
595 ± 244
696 ± 496
 − 105 ± 126
502 ± 463
Relative error [%]
88 ± 16
14 ± 52
68 ± 22
 − 77 ± 46
 − 23 ± 26
50 ± 47
ICC
0.134
0.192
0.103
0.085
0.542*
0.092
Absolute and relative error are given as mean ± standard deviation. Asterisks indicate ***excellent, **good, and *moderate intraclass correlation (ICC). No asterisk indicates poor agreement

Peak flow

There was an excellent agreement for peak flow between SW-A/C, SW-C/D, and SW-A/D with relative errors in the range of − 1 to 6 ± 7% (Table 6). Bland–Altman analysis revealed narrowest limits of agreement for SW-A/D (− 32 to 66 ml/s), followed by SW-CD and SW-A/C (− 48 to 58 ml/s; Fig. 5). Compared to SW-B, all other software agreed moderately with relative differences of up to − 8 ± 26% (SW-A/B).

Peak velocities

There was poor agreement between all software packages for peak maximum velocity, except for SW-A/D, which agreed well and presented with narrowest limits of agreement (LOA: − 32 to 22 cm/s, ER: − 4 ± 10%; Table 6, Fig. 5). All other software pairs displayed high standard deviation in the range of 19–21%.
Moderate agreement was found for peak average velocity among SW-A, SW-C, and SW-D. In contrast, there was poor agreement between SW-B and the other software programs. Limits of agreement for peak average velocity were narrower than for peak maximum velocity (e.g., for SW-A/C; Vavg: − 23 to 35 cm/s, Vmax: − 42 to 68 cm/s).

Area

There was good agreement for maximum area among SW-A, SW-C, and SW-D (Table 6). SW-B showed moderate agreement with the other software programs. The smallest absolute bias but broadest limits of agreement for the contour area was between SW-B and SW-C (EA = − 1 ± 207 mm2; LOA: − 407 to 405 mm2; Fig. 6). Narrowest limits of agreement were found for SW-A/D (− 99 to 230 mm2), followed by SW-C/D. Both readers noted independently that SW-B inevitably automatically adapted the manually corrected contour. Due to this automated correction, it was not possible to delineate the vessel contour in exactly the way they wanted to.

Wall shear stress (WSS)

There was a moderate positive correlation for the peak WSS per contour between SW-B/D with high relative error (ER = − 23 ± 26%; Table 6). All other software pairs showed poor agreement for both WSS parameters (ICC ≤ 0.37) with relative bias of up to 88 ± 16% (WSScon with SW-A/B). Bland–Altman analysis showed smaller bias and smaller limits of agreement for peak WSS per contour compared to peak WSS per segment (Fig. 6).

Equivalence test

81–85% of measurements between two readers for the clinically used parameters (area, SV, flow, velocity) were within 95% limits of agreement from intrareader comparison when they used the same software. We have therefore chosen a clear-cut threshold of X = 80%. Equivalence was concluded for a parameter if values between the 10th and 90th percentile of absolute bias between two software packages, i.e., 80% of values, were completely within the limits of equivalence.
SW-A, SW-C and SW-D all together reached equivalence regarding area and peak flow (Fig. 7). Less than 80% of comparisons including SW-B were within the equivalence limits, thus not reaching equivalence for area and peak flow. Only SW-A/C reached equivalence for stroke volume but failed to prove equivalent for peak velocities. No software pair could prove equivalent for both, peak average, and maximum velocity. 80% of differences between SW-C and SW-D were within equivalence levels for peak average velocity, while for peak maximum velocity this was the case with SW-A/D. There was no equivalence between software for WSS.

Duration of evaluation

Evaluation was fastest with SW-A (22 ± 7 min) and SW-D (23 ± 6 min), followed by SW-C (36 ± 11 min). It must be noted that SW-C crashed 1.4 times per evaluation and had to be restarted, increasing overall evaluation time by an average of 12 ± 4 min. This error was fixed in the following software version which should therefore allow for an evaluation of about 24 min. Based on these results, there would be no significant differences between SW-A, SW-C, and SW-D in terms of evaluation duration (for all, p > 0.05). Processing time using SW-B was significantly prolonged due to an inevitable segmentation algorithm that repeatedly changed manually drawn contours (51 ± 10 min, for all, p ≤ 0.01).

Inter-scanner variability as measure of reproducibility

Heart rate and blood pressure for both measurements showed only minor differences that revealed no statistical significance (MRI1: 66 ± 8 bpm, 131 ± 16/80 ± 8 mmHg; MRI2: 66 ± 13 bpm, 133 ± 16/83 ± 8 mmHg; for all, p > 0.05). Scan duration was 12 ± 3 min at MRI1 and 14 ± 1 min at MRI2 (p = 0.23). Time between both scans was 104 ± 59 days. For inter-scanner comparison, SW-A was chosen due to its comparably small intra- and interreader variability.
There was good agreement for netSV and fwSV between MRI1 and MRI2 (Table 7) with a relative error of 0 ± 15% and limits of agreement between − 22 and 26 ml for netSV (Fig. 8). Limits of agreement of bwSV were − 3 to 1 ml. Between scans, there was a good agreement for peak flow (ER = 3 ± 16%, LOA − 96 to 140 ml/s). Regarding peak velocities, there was a low agreement between both scans with relatively high relative error of 8 ± 17%. There was better agreement, lower bias and narrower limits of agreement for peak average velocity compared to peak maximum velocity (LOA; Vmax: − 32 to 50 cm/s, Vavg: − 22 to 36 cm/s). There was good agreement for the vessel area between scans with a systematic underestimation of − 8 ± 20% on MRI1 data compared to MRI2. Between both scans, there was good and moderate agreement for peak WSS per contour and per segment, respectively, with a relative error of 10 ± 16% and 11 ± 17%, respectively.
Table 7
Scanner comparison: absolute and relative error and correlation coefficient, evaluated with SW-A
 
Stroke volume
Forward stroke volume
Backward stroke volume
Absolute error
2 ± 12 ml
2 ± 11 ml
 − 1 ± 1 ml
Relative error [%]
0 ± 15
1 ± 13
38 ± 120
ICC
0.862**
0.874**
0.337
 
Peak flow
Peak maximum velocity
Peak average velocity
Absolute error
22 ± 60 ml/s
9 ± 21 cm/s
7 ± 15 cm/s
Relative error [%]
3 ± 16
8 ± 17
11 ± 22
ICC
0.832**
0.474
0.491
 
Area
Peak WSS per segment
Peak WSS per contour
Absolute error
 − 49 ± 137 mm2
144 ± 239 mPa
110 ± 179 mPa
Relative error [%]
 − 8 ± 20
11 ± 17
10 ± 16
ICC
0.828**
0.659*
0.764**
Absolute error = MRI1–MRI2. Relative error = absolute error / 0.5* (MRI1 + MRI2). Absolute and relative errors are given as mean ± standard deviation. Asterisks indicate ***excellent, **good, and *moderate intraclass correlation (ICC). No asterisk indicates poor agreement

Discussion

Intrareader repeatability and interreader reproducibility

Focus on software

There were relevant differences of intra- and interreader variability between software packages with limits of agreement varying by factors between 1.9 and 4.5. SW-A presented with highest repeatability and reproducibility for most parameters, followed by SW-D and SW-C. Repeatability and reproducibility were worst with SW-B except for WSS. We hypothesize that the inevitable automated correction of vessel contours plays a major role in the explanation for the poor performance of SW-B. Ideally, semi-automated contour detection should decrease evaluation time as well as intra- and interreader variability. In this study, the opposite was the case for SW-B. Interestingly, SW-B showed good agreement for WSS in the intra- and interreader comparisons although repeatability of contour delineation was low. This was unexpected as WSS is very sensitive to contour delineation [10].

Focus on parameters

SV, peak flow, and area showed good to excellent repeatability and reproducibility with most software packages. In comparison, peak velocities and WSS displayed less repeatability and reproducibility. A higher variability of maximum velocity and WSS results was expected. Both parameters are highly dependent on a quantity of parameters such as noise, temporal and spatial resolution, plane positioning, as well as contour delineation [10, 16]. Moreover, maximum velocity values depend on one single voxel that presents with the highest velocity value. There is no averaging in time nor space to mitigate these interfering factors. This may explain why peak maximum velocity and WSS showed worse repeatability and reproducibility than averaged parameters.

Comparison to the literature

Other 4D Flow CMR studies found dramatically better interreader agreement. Typically, bias between two readers for SV varied between 0 and 3 ml with a standard deviation between 2 and 4 ml [10, 17, 18], compared to the here presented bias of − 1 to 7 ml (SD 9–20 ml). Peak flow interreader variability was published to be − 8 ± 12 ml/s [18] compared to this study’s best result with SW-A of − 7 ± 38 ml/s. Another group found an interreader bias for vessel area of 45 ± 14 mm2 [10] compared to this study’s best result of 82 ± 75 mm2 with SW-D.
However, in these studies [10, 17, 18] both readers used the same analysis plane and same eddy current corrections for the vessel segmentation. In contrast, we did not use the same plane for comparisons. Every reader applied background phase offset correction and noise reduction individually and used multiplanar reformatting with every single software to find the adequate planes and adjusted the contour for every single timestep. This much closer reflects “real life” and is better suited for comparison of results from different groups than inter- and intrareader comparisons undertaken on exactly the same image.
We found only one paper that explicitly described “background correction and measurements” undertaken by two readers independently to determine the interreader reproducibility of thoracic 4D Flow CMR. In contrast to our study, Chelu et al. used gadolinium-based contrast agents, General Electric Healthcare CMR scanners (General Electric Healthcare, Chicago, Illinois, USA), and another post-processing software (Arterys Inc, San Francisco, California, USA). However, they found a comparable correlation for stroke volume in patients’ ascending aorta with an unspecified intraclass coefficient of 0.975 [19], compared to an ICC of 0.920 for stroke volume of all contours in the thoracic aorta with SW-A. Another study of intracranial 4D Flow CMR using Arterys software (Arterys, Inc.) found excellent interreader agreement (ICC > 0.9) for blood flow and peak velocity [20]. Using SW-C and SW-D, we could reproduce closely matching results in the thoracic aorta for blood flow but not for peak velocity. However, the methods cannot be compared directly since Wen and colleagues [20] measured carotid and intracranial flow, which are inherently less prone to motion artifacts than thoracic or abdominal exams, and they averaged three consecutive slices.
A different study found good to excellent intra- and interreader reproducibility for regional aortic WSSseg (ICC ≥ 0.78) and WSScon (ICC ≥ 0.86) using SW-A [21]. We registered similar reproducibility for peak WSSseg only with SW-A (ICC = 0.83) and for peak WSS of a contour only with SW-A and SW-B (ICC ≥ 0.706).
Given the high variability of measurements, plane-wise analysis seems not optimal for evaluation of complex 4D Flow CMR data since it can miss peak values. This affects particularly parameters that have a high variation within one vessel segment and are sensitive to noise, such as the peak velocity and WSS [10, 16]. One option to mitigate the effect of noise might be to use the median value of measurement on three consecutive slices [20], as it was also recommended by the Quantitative Imaging Biomarker Alliance (QIBA), [12]. On top of that, 3D analysis with voxel-by-voxel analysis and maximum intensity projection should perform better for detection and localization of peak values [22]. However, this was not available with the tested software programs.

Reproducibility of different postprocessing software

Focus on software

While SW-A, SW-C, and SW-D produced equivalent results for area and peak flow, only SW-A/C proved equivalent for stroke volume. The reason for the high deviation of SW-B from results with other software packages lies probably in the automated correction of contours that did not allow an undisturbed manual delineation of the contour. In general, we suspect differences between software packages at least partly to originate from different background phase correction methods, temporal and spatial smoothing, and different interpolation algorithms of each software.
It must be noted that the equivalence test allowed a broad spectrum of errors due to the relatively high intrareader variability of all software packages. However, given the relatively high intrareader variability, it would not have made sense to choose narrower limits of equivalence.

Focus on parameters

Peak flow, stroke volume, and area were the parameter with the best reproducibility among the tested software packages, matching their good repeatability and interreader reproducibility. For the same reasons, i.e. noise, resolution, plane positioning and contour delineation, peak velocities and WSS were the parameters with the worst reproducibility among software programs.

Comparison to the literature

Compared to a recently published 2D phase contrast (PC) CMR on differences between various software programs, the bias and limits of agreement of the herein presented results were remarkably high. Typical 2D PC CMR values [bias ± SD] were reported: maximum velocity: 0–5 ± 3 cm/s for stroke volume: 0–3 ± 3 ml between three software programs [15]. However, hemodynamic measurements are dependent on the positioning of the plane [23]. This influencing factor was not considered in the 2D PC CMR study since the same plane was analyzed with different software packages. In our study, plane positioning remained an influencing factor adding to the potential sources of error. Additionally, other sources of error including elaborate postprocessing in 4D Flow CMR do not apply for 2D PC CMR. Further, noise is higher in 4D Flow CMR data because of the typically two- to three-fold smaller voxel size compared to 2D PC CMR. Therefore, differences between software programs or readers using the same DICOM data set were expected to be smaller for 2D PC CMR compared to 4D Flow CMR.

Reproducibility of different scanners

Of note and counterintuitively, data from different scanners evaluated with a single software typically resulted in better comparability than data from a single scanner evaluated with different software programs. This was especially true for WSS: There was moderate to good agreement between both scans while—apart from SW-B/D—WSS values did poorly agree between different software packages. In general, the scanner comparison yielded similar or lower bias, but higher standard deviation compared to software comparison. Higher standard deviations were expected since both scans were done months apart and it was anticipated that physiological changes would alter the results between both scans.
However, the presented smaller interscan bias is surprising, given that previous studies showed significant differences between different scanners of the same vendor and between vendors. For 2D PC measurements, the velocity offset between measurements with the same scanner and sequence at different sites is well acknowledged [24]. Similarly, Bock et al. have shown a decreased bias for kinetic energy evaluation if scans were repeated with an interval of 14 days at the same scanner as opposed to a repeat scan using a scanner from a different vendor on the same day [4].
Scan-rescan abdominal 4D Flow CMR studies revealed neglectable bias for stroke volume (0 ± 3 ml) [25] and flow (2 ± 5 ml/min) [26], evaluating scans performed on the same day using the same scanner, while there were larger limits of agreement of scans performed on different days (flow 0 ± 11 ml/min), [26] - presumably due to physiological fluctuations. The impact of physiological variability to the inter-scanner variability in this study should therefore not be underestimated.
Van der Palen et al. [21] found moderate to good agreement for mean and maximum WSS between two consecutive scans performed on the same scanner. Similar agreement was achieved in our study for WSS values derived from two different scanners underlining good comparability of the sequences at both scanners.

Clinical relevance

Ultimately, our results underline that only SW-A and SW-C can be used equivalently for determination of stroke volume, peak flow, and area. One can neither compare results from the other tested 4D Flow CMR analysis software programs nor can one compare the other tested parameters between software packages. Unexpectedly, results from different scanners often showed better agreement than results from different software packages. This is promising, since careful adaption of sequences at scanners of different manufactures may allow for multicenter studies with scans from different scanners in the future using the same postprocessing software.
Inter- and intrareader consistency was best with SW-A and SW-D that also permitted fastest evaluation times. WSS results should be interpreted cautiously due to low repeatability and reproducibility between software programs. Relatively high intra- and interreader variability do not allow the calculation of a conversion factor between programs at the moment. Overall high variability of peak velocities with all software packages are alarming since this parameter is frequently used in clinical routine for stenosis classification.
The authors of this paper identified the following issues that should be addressed by the software vendors to improve repeatability and reproducibility of 4D Flow CMR results:
1)
Making analysis less susceptible to minor changes of contour placement and vessel wall delineation. This could be achieved by intelligent averaging algorithms over several contours and improved vessel wall detection. Further, the analysis of peak values by volume maximum intensity projection analyses might reduce the risk of underestimated peak values due to misplacement of the analysis plane. An automatic averaging of results from three consecutive planes as recommended by the Quantitative Imaging Biomarker Alliance (QIBA) [12] might reduce bias related to noise.
 
2)
Homogenizing masking and correction methods for background phase offsets and interpolation methods for oblique cut planes. Disclosure of underlying concepts and equations for basic and advanced postprocessing steps would facilitate comparability between software.
 
Additional studies are needed to identify the sources of the differences in results between software solutions. For example, flow analysis in a standardized imaging plane could help to assess the impact of plane selection on results. Further optimization and standardization of software results and workflows is necessary to achieve effective comparison and expand the applicability of 4D Flow CMR findings.

Limitations

We acknowledge that results from a small group of young and healthy subjects cannot be directly transferred to patients. Future studies should comprise patients and the analysis of variations such as diurnal or postprandial changes. Moreover, we regarded the error in different areas of the thoracic aorta as equivalent—in a larger cohort, it would be interesting to see whether reproducibility varies between different regions in the thoracic aorta.
We point out that the results of our analysis refer to measurements performed by beginners. Experienced readers might have obtained a lower variability of results. However, we have tried to exclude bias due to familiarity with one or another software to create fair comparison conditions for all programs.
We modified the equivalence test proposed by Zange et al. [15] to assess equivalence of different software solutions. Neither of the tests are validated, and different thresholds could significantly change the results.
The scans on both 3 T CMR scanners were performed on different days. Hence, the impact of physiological changes cannot be differentiated from the measurement error between both scanners. However, a dedicated analysis of the measurement error between scanners was not the main goal of this study. This comparison was only undertaken to allow interpretation of software comparison results in a larger context. A dedicated study focusing on scanner comparison should focus on consecutive scans in random order.
A final limitation of our study is the lack of a reference standard. After all, the aim of this study was to analyze comparability of software, not to evaluate accuracy of software using a reference standard.

Conclusion

This study confirms that only SW-A and SW-C can be used equivalently for determination of SV, peak flow, and area. SW-D yielded also equivalent values for peak flow and area in comparison to SW-A and SW-C. Other software packages and parameters did not yield comparable results. Moreover, results from different scanners often showed better agreement than results from different software packages. High intra- and interreader variability for all parameters, especially for peak velocities, needs to be addressed. Before introducing 4D Flow CMR into clinical routine, not only scanning protocols but also postprocessing software need to be synchronized to allow for cross-vendor comparison at least of clinically relevant results.

Acknowledgements

None.

Declarations

This HIPAA-compliant study was approved by the ethics committee of the University of Luebeck, reference number 17-252. All participants gave written informed consent.
Not applicable.

Competing interests

There were research collaborations between the Department of Radiology and Nuclear Medicine, Universität zu Lübeck, Lübeck, with all scanner and software companies mentioned in this paper: Circle Cardiovascular Imaging, CAN; GyroTools, CH; MEVIS Fraunhofer, Germany; Philips, NL; Pie Medical, NL; Siemens, Germany (alphabetic order).
N. Jin and J. Graessner work for Siemens. H. Kooijman-Kurfuerst works for Philips. A. Hennemuth works for MEVIS Fraunhofer.
The other authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Dyverfeldt P, Bissell M, Barker AJ, Bolger AF, Carlhall CJ, Ebbers T, et al. 4D flow cardiovascular magnetic resonance consensus statement. J Cardiovasc Magn Reson. 2015;17(1):72.CrossRefPubMedPubMedCentral Dyverfeldt P, Bissell M, Barker AJ, Bolger AF, Carlhall CJ, Ebbers T, et al. 4D flow cardiovascular magnetic resonance consensus statement. J Cardiovasc Magn Reson. 2015;17(1):72.CrossRefPubMedPubMedCentral
2.
Zurück zum Zitat Rizk J. 4D flow MRI applications in congenital heart disease. Eur Radiol. 2021;31(2):1160–74.CrossRefPubMed Rizk J. 4D flow MRI applications in congenital heart disease. Eur Radiol. 2021;31(2):1160–74.CrossRefPubMed
3.
Zurück zum Zitat Markl M, Frydrychowicz A, Kozerke S, Hope M, Wieben O. 4D flow MRI. J Magn Reson Imaging. 2012;36(5):1015–36.CrossRefPubMed Markl M, Frydrychowicz A, Kozerke S, Hope M, Wieben O. 4D flow MRI. J Magn Reson Imaging. 2012;36(5):1015–36.CrossRefPubMed
4.
Zurück zum Zitat Bock J, Toger J, Bidhult S, Markenroth Bloch K, Arvidsson P, Kanski M, et al. Validation and reproducibility of cardiovascular 4D-flow MRI from two vendors using 2 x 2 parallel imaging acceleration in pulsatile flow phantom and in vivo with and without respiratory gating. Acta Radiol. 2019;60(3):327–37.CrossRefPubMed Bock J, Toger J, Bidhult S, Markenroth Bloch K, Arvidsson P, Kanski M, et al. Validation and reproducibility of cardiovascular 4D-flow MRI from two vendors using 2 x 2 parallel imaging acceleration in pulsatile flow phantom and in vivo with and without respiratory gating. Acta Radiol. 2019;60(3):327–37.CrossRefPubMed
5.
Zurück zum Zitat Montalba C, Urbina J, Sotelo J, Andia ME, Tejos C, Irarrazaval P, et al. Variability of 4D flow parameters when subjected to changes in MRI acquisition parameters using a realistic thoracic aortic phantom. Magn Reson Med. 2018;79(4):1882–92.CrossRefPubMed Montalba C, Urbina J, Sotelo J, Andia ME, Tejos C, Irarrazaval P, et al. Variability of 4D flow parameters when subjected to changes in MRI acquisition parameters using a realistic thoracic aortic phantom. Magn Reson Med. 2018;79(4):1882–92.CrossRefPubMed
6.
Zurück zum Zitat Zimmermann J, Demedts D, Mirzaee H, Ewert P, Stern H, Meierhofer C, et al. Wall shear stress estimation in the aorta: Impact of wall motion, spatiotemporal resolution, and phase noise. J Magn Reson Imaging. 2018;48(3):718–28.CrossRef Zimmermann J, Demedts D, Mirzaee H, Ewert P, Stern H, Meierhofer C, et al. Wall shear stress estimation in the aorta: Impact of wall motion, spatiotemporal resolution, and phase noise. J Magn Reson Imaging. 2018;48(3):718–28.CrossRef
7.
Zurück zum Zitat David A, Le Touze D, Warin-Fresse K, Paul-Gilloteaux P, Bonnefoy F, Idier J, et al. In-vitro validation of 4D flow MRI measurements with an experimental pulsatile flow model. Diagn Interv Imaging. 2019;100(1):17–23.CrossRefPubMed David A, Le Touze D, Warin-Fresse K, Paul-Gilloteaux P, Bonnefoy F, Idier J, et al. In-vitro validation of 4D flow MRI measurements with an experimental pulsatile flow model. Diagn Interv Imaging. 2019;100(1):17–23.CrossRefPubMed
8.
Zurück zum Zitat Watanabe T, Isoda H, Fukuyama A, Takahashi M, Amano T, Takehara Y, et al. Accuracy of the flow velocity and three-directional velocity profile measured with three-dimensional cine phase-contrast mr imaging: verification on scanners from different manufacturers. Magn Reson Med Sci. 2019;18(4):265–71.CrossRefPubMedPubMedCentral Watanabe T, Isoda H, Fukuyama A, Takahashi M, Amano T, Takehara Y, et al. Accuracy of the flow velocity and three-directional velocity profile measured with three-dimensional cine phase-contrast mr imaging: verification on scanners from different manufacturers. Magn Reson Med Sci. 2019;18(4):265–71.CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Perinajova R, Juffermans JF, Mercado JL, Aben JP, Ledoux L, Westenberg JJM, et al. Assessment of turbulent blood flow and wall shear stress in aortic coarctation using image-based simulations. Biomed Eng Online. 2021;20(1):84.CrossRefPubMedPubMedCentral Perinajova R, Juffermans JF, Mercado JL, Aben JP, Ledoux L, Westenberg JJM, et al. Assessment of turbulent blood flow and wall shear stress in aortic coarctation using image-based simulations. Biomed Eng Online. 2021;20(1):84.CrossRefPubMedPubMedCentral
10.
Zurück zum Zitat Stalder AF, Russe MF, Frydrychowicz A, Bock J, Hennig J, Markl M. Quantitative 2D and 3D phase contrast MRI: optimized analysis of blood flow and vessel wall parameters. Magn Reson Med. 2008;60(5):1218–31.CrossRefPubMed Stalder AF, Russe MF, Frydrychowicz A, Bock J, Hennig J, Markl M. Quantitative 2D and 3D phase contrast MRI: optimized analysis of blood flow and vessel wall parameters. Magn Reson Med. 2008;60(5):1218–31.CrossRefPubMed
11.
12.
Zurück zum Zitat Sullivan DC, Obuchowski NA, Kessler LG, Raunig DL, Gatsonis C, Huang EP, et al. Metrology standards for quantitative imaging biomarkers. Radiology. 2015;277(3):813–25.CrossRefPubMed Sullivan DC, Obuchowski NA, Kessler LG, Raunig DL, Gatsonis C, Huang EP, et al. Metrology standards for quantitative imaging biomarkers. Radiology. 2015;277(3):813–25.CrossRefPubMed
13.
Zurück zum Zitat Raunig DL, McShane LM, Pennello G, Gatsonis C, Carson PL, Voyvodic JT, et al. Quantitative imaging biomarkers: a review of statistical methods for technical performance assessment. Stat Methods Med Res. 2015;24(1):27–67.CrossRefPubMed Raunig DL, McShane LM, Pennello G, Gatsonis C, Carson PL, Voyvodic JT, et al. Quantitative imaging biomarkers: a review of statistical methods for technical performance assessment. Stat Methods Med Res. 2015;24(1):27–67.CrossRefPubMed
14.
Zurück zum Zitat Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;15(2):155–63.CrossRefPubMedPubMedCentral Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;15(2):155–63.CrossRefPubMedPubMedCentral
15.
Zurück zum Zitat Zange L, Muehlberg F, Blaszczyk E, Schwenke S, Traber J, Funk S, et al. Quantification in cardiovascular magnetic resonance: agreement of software from three different vendors on assessment of left ventricular function, 2D flow and parametric mapping. J Cardiovasc Magn Reson. 2019;21(1):12.CrossRefPubMedPubMedCentral Zange L, Muehlberg F, Blaszczyk E, Schwenke S, Traber J, Funk S, et al. Quantification in cardiovascular magnetic resonance: agreement of software from three different vendors on assessment of left ventricular function, 2D flow and parametric mapping. J Cardiovasc Magn Reson. 2019;21(1):12.CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Petersson S, Dyverfeldt P, Ebbers T. Assessment of the accuracy of MRI wall shear stress estimation using numerical simulations. J Magn Reson Imaging. 2012;36(1):128–38.CrossRefPubMed Petersson S, Dyverfeldt P, Ebbers T. Assessment of the accuracy of MRI wall shear stress estimation using numerical simulations. J Magn Reson Imaging. 2012;36(1):128–38.CrossRefPubMed
17.
Zurück zum Zitat Moersdorf R, Treutlein M, Kroeger JR, Ruijsink B, Wong J, Maintz D, et al. Precision, reproducibility and applicability of an undersampled multi-venc 4D flow MRI sequence for the assessment of cardiac hemodynamics. Magn Reson Imaging. 2019;61:73–82.CrossRefPubMed Moersdorf R, Treutlein M, Kroeger JR, Ruijsink B, Wong J, Maintz D, et al. Precision, reproducibility and applicability of an undersampled multi-venc 4D flow MRI sequence for the assessment of cardiac hemodynamics. Magn Reson Imaging. 2019;61:73–82.CrossRefPubMed
18.
Zurück zum Zitat Markl M, Wallis W, Harloff A. Reproducibility of flow and wall shear stress analysis using flow-sensitive four-dimensional MRI. J Magn Reson Imaging. 2011;33(4):988–94.CrossRefPubMed Markl M, Wallis W, Harloff A. Reproducibility of flow and wall shear stress analysis using flow-sensitive four-dimensional MRI. J Magn Reson Imaging. 2011;33(4):988–94.CrossRefPubMed
19.
Zurück zum Zitat Chelu RG, Horowitz M, Sucha D, Kardys I, Ingremeau D, Vasanawala S, et al. Evaluation of atrial septal defects with 4D flow MRI-multilevel and inter-reader reproducibility for quantification of shunt severity. MAGMA. 2019;32(2):269–79.CrossRefPubMed Chelu RG, Horowitz M, Sucha D, Kardys I, Ingremeau D, Vasanawala S, et al. Evaluation of atrial septal defects with 4D flow MRI-multilevel and inter-reader reproducibility for quantification of shunt severity. MAGMA. 2019;32(2):269–79.CrossRefPubMed
20.
Zurück zum Zitat Wen B, Tian S, Cheng J, Li Y, Zhang H, Xue K, et al. Test-retest multisite reproducibility of neurovascular 4D flow MRI. J Magn Reson Imaging. 2019;49(6):1543–52.CrossRefPubMed Wen B, Tian S, Cheng J, Li Y, Zhang H, Xue K, et al. Test-retest multisite reproducibility of neurovascular 4D flow MRI. J Magn Reson Imaging. 2019;49(6):1543–52.CrossRefPubMed
21.
Zurück zum Zitat van der Palen RLF, Roest AAW, van den Boogaard PJ, de Roos A, Blom NA, Westenberg JJM. Scan-rescan reproducibility of segmental aortic wall shear stress as assessed by phase-specific segmentation with 4D flow MRI in healthy volunteers. MAGMA. 2018;31(5):653–63.CrossRefPubMedPubMedCentral van der Palen RLF, Roest AAW, van den Boogaard PJ, de Roos A, Blom NA, Westenberg JJM. Scan-rescan reproducibility of segmental aortic wall shear stress as assessed by phase-specific segmentation with 4D flow MRI in healthy volunteers. MAGMA. 2018;31(5):653–63.CrossRefPubMedPubMedCentral
22.
Zurück zum Zitat van Ooij P, Powell AL, Potters WV, Carr JC, Markl M, Barker AJ. Reproducibility and interobserver variability of systolic blood flow velocity and 3D wall shear stress derived from 4D flow MRI in the healthy aorta. J Magn Reson Imaging. 2016;43(1):236–48.CrossRefPubMed van Ooij P, Powell AL, Potters WV, Carr JC, Markl M, Barker AJ. Reproducibility and interobserver variability of systolic blood flow velocity and 3D wall shear stress derived from 4D flow MRI in the healthy aorta. J Magn Reson Imaging. 2016;43(1):236–48.CrossRefPubMed
23.
Zurück zum Zitat Lotz J, Meier C, Leppert A, Galanski M. Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation. Radiographics. 2002;22(3):651–71.CrossRefPubMed Lotz J, Meier C, Leppert A, Galanski M. Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation. Radiographics. 2002;22(3):651–71.CrossRefPubMed
24.
Zurück zum Zitat Gatehouse PD, Rolf MP, Graves MJ, Hofman MB, Totman J, Werner B, et al. Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements. J Cardiovasc Magn Reson. 2010;12:5.CrossRefPubMedPubMedCentral Gatehouse PD, Rolf MP, Graves MJ, Hofman MB, Totman J, Werner B, et al. Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements. J Cardiovasc Magn Reson. 2010;12:5.CrossRefPubMedPubMedCentral
25.
Zurück zum Zitat Wentland AL, Grist TM, Wieben O. Repeatability and internal consistency of abdominal 2D and 4D phase contrast MR flow measurements. Acad Radiol. 2013;20(6):699–704.CrossRefPubMedPubMedCentral Wentland AL, Grist TM, Wieben O. Repeatability and internal consistency of abdominal 2D and 4D phase contrast MR flow measurements. Acad Radiol. 2013;20(6):699–704.CrossRefPubMedPubMedCentral
26.
Zurück zum Zitat Macdonald JA, Corrado PA, Nguyen SM, Johnson KM, Francois CJ, Magness RR, et al. Uteroplacental and Fetal 4D Flow MRI in the Pregnant Rhesus Macaque. J Magn Reson Imaging. 2019;49(2):534–45.CrossRefPubMed Macdonald JA, Corrado PA, Nguyen SM, Johnson KM, Francois CJ, Magness RR, et al. Uteroplacental and Fetal 4D Flow MRI in the Pregnant Rhesus Macaque. J Magn Reson Imaging. 2019;49(2):534–45.CrossRefPubMed
Metadaten
Titel
Repeatability and reproducibility of various 4D Flow MRI postprocessing software programs in a multi-software and multi-vendor cross-over comparison study
verfasst von
Thekla H. Oechtering
André Nowak
Malte M. Sieren
Andreas M. Stroth
Nicolas Kirschke
Franz Wegner
Maren Balks
Inke R. König
Ning Jin
Joachim Graessner
Hendrik Kooijman-Kurfuerst
Anja Hennemuth
Jörg Barkhausen
Alex Frydrychowicz
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2023
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-023-00921-4

Weitere Artikel der Ausgabe 1/2023

Journal of Cardiovascular Magnetic Resonance 1/2023 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.