03.11.2023 | Cardiovascular Critical Care (A Higgins and R Lee, Section Editors)

Stroke and Mechanical Circulatory Support in Adults

verfasst von: David J. Bae, Joshua Z. Willey, Chinwe Ibeh, Melana Yuzefpolskaya, Paolo C. Colombo

Erschienen in: Current Cardiology Reports | Ausgabe 12/2023

Einloggen, um Zugang zu erhalten

Abstract

Purpose of the Review

Short-term and durable mechanical circulatory support (MCS) devices represent life-saving interventions for patients with cardiogenic shock and end-stage heart failure. This review will cover the epidemiology, risk factors, and treatment of stroke in this patient population.

Recent Findings

Short-term devices such as intra-aortic balloon pump, Impella, TandemHeart, and Venoatrial Extracorporal Membrane Oxygenation, as well as durable continuous-flow left ventricular assist devices (LVADs), improve cardiac output and blood flow to the vital organs. However, MCS use is associated with high rates of complications, including ischemic and hemorrhagic strokes which carry a high risk for death and disability. Improvements in MCS technology have reduced but not eliminated the risk of stroke. Mitigation strategies focus on careful management of anti-thrombotic therapies. While data on therapeutic options for stroke are limited, several case series reported favorable outcomes with thrombectomy for ischemic stroke patients with large vessel occlusions, as well as with reversal of anticoagulation for those with hemorrhagic stroke.

Summary

Stroke in patients treated with MCS is associated with high morbidity and mortality. Preventive strategies are targeted based on the specific form of MCS. Improvements in the design of the newest generation device have reduced the risk of ischemic stroke, though hemorrhagic stroke remains a serious complication.
Literatur
1.
Zurück zum Zitat Mozaffarian D, Benjamin EJ, Go AS, et al. Heart Disease and Stroke Statistics-2016 update: a report from the American Heart Association. Circulation. 2016;133(4):e38-60.PubMed Mozaffarian D, Benjamin EJ, Go AS, et al. Heart Disease and Stroke Statistics-2016 update: a report from the American Heart Association. Circulation. 2016;133(4):e38-60.PubMed
2.
Zurück zum Zitat Kohno H, Matsumiya G, Sawa Y, et al. The Jarvik 2000 left ventricular assist device as a bridge to transplantation: Japanese Registry for Mechanically Assisted Circulatory Support. J Heart Lung Transplant. 2018;37(1):71–8.PubMed Kohno H, Matsumiya G, Sawa Y, et al. The Jarvik 2000 left ventricular assist device as a bridge to transplantation: Japanese Registry for Mechanically Assisted Circulatory Support. J Heart Lung Transplant. 2018;37(1):71–8.PubMed
3.
Zurück zum Zitat Mehra MR, Naka Y, Uriel N, et al. A fully magnetically levitated circulatory pump for advanced heart failure. N Engl J Med. 2017;376(5):440–50.PubMed Mehra MR, Naka Y, Uriel N, et al. A fully magnetically levitated circulatory pump for advanced heart failure. N Engl J Med. 2017;376(5):440–50.PubMed
4.
Zurück zum Zitat Mehra MR, Goldstein DJ, Uriel N, et al. Two-year outcomes with a magnetically levitated cardiac pump in heart failure. N Engl J Med. 2018;378(15):1386–95.PubMed Mehra MR, Goldstein DJ, Uriel N, et al. Two-year outcomes with a magnetically levitated cardiac pump in heart failure. N Engl J Med. 2018;378(15):1386–95.PubMed
5.
Zurück zum Zitat Rose EA, Gelijns AC, Moskowitz AJ, et al. Long-term use of a left ventricular assist device for end-stage heart failure. N Engl J Med. 2001;345(20):1435–43.PubMed Rose EA, Gelijns AC, Moskowitz AJ, et al. Long-term use of a left ventricular assist device for end-stage heart failure. N Engl J Med. 2001;345(20):1435–43.PubMed
6.
Zurück zum Zitat •• Mehra MR, Goldstein DJ, Cleveland JC, et al. Five-year outcomes in patients with fully magnetically levitated vs axial-flow left ventricular assist devices in the MOMENTUM 3 Randomized Trial. JAMA. 2022;328(12):1233–42. This manuscript highlights the 5 -year outcomes comparing the Heartmate 3 to Heartmate II, with improved survival and lower risk of stroke in the Heartmate3 device.PubMedPubMedCentral •• Mehra MR, Goldstein DJ, Cleveland JC, et al. Five-year outcomes in patients with fully magnetically levitated vs axial-flow left ventricular assist devices in the MOMENTUM 3 Randomized Trial. JAMA. 2022;328(12):1233–42. This manuscript highlights the 5 -year outcomes comparing the Heartmate 3 to Heartmate II, with improved survival and lower risk of stroke in the Heartmate3 device.PubMedPubMedCentral
7.
Zurück zum Zitat Tsukui H, Abla A, Teuteberg JJ, et al. Cerebrovascular accidents in patients with a ventricular assist device. J Thorac Cardiovasc Surg. 2007;134(1):114–23.PubMed Tsukui H, Abla A, Teuteberg JJ, et al. Cerebrovascular accidents in patients with a ventricular assist device. J Thorac Cardiovasc Surg. 2007;134(1):114–23.PubMed
8.
Zurück zum Zitat Yuan N, Arnaoutakis GJ, George TJ, et al. The spectrum of complications following left ventricular assist device placement. J Card Surg. 2012;27(5):630–8.PubMed Yuan N, Arnaoutakis GJ, George TJ, et al. The spectrum of complications following left ventricular assist device placement. J Card Surg. 2012;27(5):630–8.PubMed
9.
Zurück zum Zitat Kato TS, Schulze PC, Yang J, et al. Pre-operative and post-operative risk factors associated with neurologic complications in patients with advanced heart failure supported by a left ventricular assist device. J Heart Lung Transplant. 2012;31(1):1–8.PubMed Kato TS, Schulze PC, Yang J, et al. Pre-operative and post-operative risk factors associated with neurologic complications in patients with advanced heart failure supported by a left ventricular assist device. J Heart Lung Transplant. 2012;31(1):1–8.PubMed
10.
Zurück zum Zitat Kirklin JK, Naftel DC, Pagani FD, et al. Sixth INTERMACS annual report: a 10,000-patient database. J Heart Lung Transplant. 2014;33(6):555–64.PubMed Kirklin JK, Naftel DC, Pagani FD, et al. Sixth INTERMACS annual report: a 10,000-patient database. J Heart Lung Transplant. 2014;33(6):555–64.PubMed
11.
Zurück zum Zitat Kirklin JK, Naftel DC, Kormos RL, et al. Fifth INTERMACS annual report: risk factor analysis from more than 6,000 mechanical circulatory support patients. J Heart Lung Transplant. 2013;32(2):141–56.PubMed Kirklin JK, Naftel DC, Kormos RL, et al. Fifth INTERMACS annual report: risk factor analysis from more than 6,000 mechanical circulatory support patients. J Heart Lung Transplant. 2013;32(2):141–56.PubMed
12.
Zurück zum Zitat Starling RC, Estep JD, Horstmanshof DA, et al. Risk assessment and comparative effectiveness of left ventricular assist device and medical management in ambulatory heart failure patients: the ROADMAP Study 2-year results. JACC Heart Fail. 2017;5(7):518–27.PubMed Starling RC, Estep JD, Horstmanshof DA, et al. Risk assessment and comparative effectiveness of left ventricular assist device and medical management in ambulatory heart failure patients: the ROADMAP Study 2-year results. JACC Heart Fail. 2017;5(7):518–27.PubMed
13.
Zurück zum Zitat Stone GW, Ohman EM, Miller MF, et al. Contemporary utilization and outcomes of intra-aortic balloon counterpulsation in acute myocardial infarction: the benchmark registry. J Am Coll Cardiol. 2003;41(11):1940–5.PubMed Stone GW, Ohman EM, Miller MF, et al. Contemporary utilization and outcomes of intra-aortic balloon counterpulsation in acute myocardial infarction: the benchmark registry. J Am Coll Cardiol. 2003;41(11):1940–5.PubMed
14.
Zurück zum Zitat Huckaby LV, Seese LM, Mathier MA, Hickey GW, Kilic A. Intra-aortic balloon pump bridging to heart transplantation: impact of the 2018 allocation change. Circ Heart Fail. 2020;13(8):e006971.PubMedPubMedCentral Huckaby LV, Seese LM, Mathier MA, Hickey GW, Kilic A. Intra-aortic balloon pump bridging to heart transplantation: impact of the 2018 allocation change. Circ Heart Fail. 2020;13(8):e006971.PubMedPubMedCentral
15.
Zurück zum Zitat Estep JD, Cordero-Reyes AM, Bhimaraj A, et al. Percutaneous placement of an intra-aortic balloon pump in the left axillary/subclavian position provides safe, ambulatory long-term support as bridge to heart transplantation. JACC Heart Fail. 2013;1(5):382–8.PubMed Estep JD, Cordero-Reyes AM, Bhimaraj A, et al. Percutaneous placement of an intra-aortic balloon pump in the left axillary/subclavian position provides safe, ambulatory long-term support as bridge to heart transplantation. JACC Heart Fail. 2013;1(5):382–8.PubMed
16.
Zurück zum Zitat Russo MJ, Jeevanandam V, Stepney J, et al. Intra-aortic balloon pump inserted through the subclavian artery: a minimally invasive approach to mechanical support in the ambulatory end-stage heart failure patient. J Thorac Cardiovasc Surg. 2012;144(4):951–5.PubMed Russo MJ, Jeevanandam V, Stepney J, et al. Intra-aortic balloon pump inserted through the subclavian artery: a minimally invasive approach to mechanical support in the ambulatory end-stage heart failure patient. J Thorac Cardiovasc Surg. 2012;144(4):951–5.PubMed
17.
Zurück zum Zitat Lauten A, Engstrom AE, Jung C, et al. Response to letter regarding article. Percutaneous left-ventricular support with the Impella-2.5-assist device in acute cardiogenic shock results of the Impella-EUROSHOCK registry. Circ Heart Fail. 2013;6(4):e56. Lauten A, Engstrom AE, Jung C, et al. Response to letter regarding article. Percutaneous left-ventricular support with the Impella-2.5-assist device in acute cardiogenic shock results of the Impella-EUROSHOCK registry. Circ Heart Fail. 2013;6(4):e56.
18.
Zurück zum Zitat Ouweneel DM, Eriksen E, Sjauw KD, et al. Percutaneous mechanical circulatory support versus intra-aortic balloon pump in cardiogenic shock after acute myocardial infarction. J Am Coll Cardiol. 2017;69(3):278–87.PubMed Ouweneel DM, Eriksen E, Sjauw KD, et al. Percutaneous mechanical circulatory support versus intra-aortic balloon pump in cardiogenic shock after acute myocardial infarction. J Am Coll Cardiol. 2017;69(3):278–87.PubMed
19.
Zurück zum Zitat Doersch KM, Tong CW, Gongora E, Konda S, Sareyyupoglu B. Temporary left ventricular assist device through an axillary access is a promising approach to improve outcomes in refractory cardiogenic shock patients. ASAIO J. 2015;61(3):253–8.PubMedPubMedCentral Doersch KM, Tong CW, Gongora E, Konda S, Sareyyupoglu B. Temporary left ventricular assist device through an axillary access is a promising approach to improve outcomes in refractory cardiogenic shock patients. ASAIO J. 2015;61(3):253–8.PubMedPubMedCentral
20.
Zurück zum Zitat Ramzy D, Anderson M, Batsides G, et al. Early outcomes of the first 200 US patients treated with Impella 5.5: a novel temporary left ventricular assist device. Innovations (Phila). 2021;16(4):365–372. Ramzy D, Anderson M, Batsides G, et al. Early outcomes of the first 200 US patients treated with Impella 5.5: a novel temporary left ventricular assist device. Innovations (Phila). 2021;16(4):365–372.
21.
Zurück zum Zitat Geller BJ, Sinha SS, Kapur NK, et al. Escalating and de-escalating temporary mechanical circulatory support in cardiogenic shock: a scientific statement from the American Heart Association. Circulation. 2022;146(6):e50–68.PubMed Geller BJ, Sinha SS, Kapur NK, et al. Escalating and de-escalating temporary mechanical circulatory support in cardiogenic shock: a scientific statement from the American Heart Association. Circulation. 2022;146(6):e50–68.PubMed
22.
Zurück zum Zitat Cheng R, Hachamovitch R, Kittleson M, et al. Complications of extracorporeal membrane oxygenation for treatment of cardiogenic shock and cardiac arrest: a meta-analysis of 1,866 adult patients. Ann Thorac Surg. 2014;97(2):610–6.PubMed Cheng R, Hachamovitch R, Kittleson M, et al. Complications of extracorporeal membrane oxygenation for treatment of cardiogenic shock and cardiac arrest: a meta-analysis of 1,866 adult patients. Ann Thorac Surg. 2014;97(2):610–6.PubMed
23.
Zurück zum Zitat Le Guennec L, Cholet C, Huang F, et al. Ischemic and hemorrhagic brain injury during venoarterial-extracorporeal membrane oxygenation. Ann Intensive Care. 2018;8(1):129.PubMedPubMedCentral Le Guennec L, Cholet C, Huang F, et al. Ischemic and hemorrhagic brain injury during venoarterial-extracorporeal membrane oxygenation. Ann Intensive Care. 2018;8(1):129.PubMedPubMedCentral
24.
Zurück zum Zitat • Cho SM, Canner J, Caturegli G, et al. Risk factors of ischemic and hemorrhagic strokes during venovenous extracorporeal membrane oxygenation: analysis of data from the Extracorporeal Life Support Organization Registry. Crit Care Med. 2021;49(1):91–101. This study describes the main risk factors for ischemic and hemorrhagic stroke in patients treated with ECMO. The risk factors for ischemic stroke were low pre-treatment pH, high pO2, circuit failure, and renal replacement therapy. PubMedPubMedCentral • Cho SM, Canner J, Caturegli G, et al. Risk factors of ischemic and hemorrhagic strokes during venovenous extracorporeal membrane oxygenation: analysis of data from the Extracorporeal Life Support Organization Registry. Crit Care Med. 2021;49(1):91–101. This study describes the main risk factors for ischemic and hemorrhagic stroke in patients treated with ECMO. The risk factors for ischemic stroke were low pre-treatment pH, high pO2, circuit failure, and renal replacement therapy. PubMedPubMedCentral
25.
Zurück zum Zitat Cevasco MR, Li B, Han J, et al. Adverse event profile associated with prolonged use of CentriMag ventricular assist device for refractory cardiogenic shock. ASAIO J. 2019;65(8):806–11.PubMed Cevasco MR, Li B, Han J, et al. Adverse event profile associated with prolonged use of CentriMag ventricular assist device for refractory cardiogenic shock. ASAIO J. 2019;65(8):806–11.PubMed
26.
Zurück zum Zitat Mohite PN, Zych B, Popov AF, et al. CentriMag short-term ventricular assist as a bridge to solution in patients with advanced heart failure: use beyond 30 days. Eur J Cardiothorac Surg. 2013;44(5):e310-315.PubMed Mohite PN, Zych B, Popov AF, et al. CentriMag short-term ventricular assist as a bridge to solution in patients with advanced heart failure: use beyond 30 days. Eur J Cardiothorac Surg. 2013;44(5):e310-315.PubMed
27.
Zurück zum Zitat Worku B, Pak SW, van Patten D, et al. The CentriMag ventricular assist device in acute heart failure refractory to medical management. J Heart Lung Transplant. 2012;31(6):611–7.PubMed Worku B, Pak SW, van Patten D, et al. The CentriMag ventricular assist device in acute heart failure refractory to medical management. J Heart Lung Transplant. 2012;31(6):611–7.PubMed
28.
Zurück zum Zitat Alli OO, Singh IM, Holmes DR Jr, Pulido JN, Park SJ, Rihal CS. Percutaneous left ventricular assist device with TandemHeart for high-risk percutaneous coronary intervention: the Mayo Clinic experience. Catheter Cardiovasc Interv. 2012;80(5):728–34.PubMed Alli OO, Singh IM, Holmes DR Jr, Pulido JN, Park SJ, Rihal CS. Percutaneous left ventricular assist device with TandemHeart for high-risk percutaneous coronary intervention: the Mayo Clinic experience. Catheter Cardiovasc Interv. 2012;80(5):728–34.PubMed
29.
Zurück zum Zitat Schwartz BG, Ludeman DJ, Mayeda GS, Kloner RA, Economides C, Burstein S. Treating refractory cardiogenic shock with the tandemheart and Impella devices: a single center experience. Cardiol Res. 2012;3(2):54–66.PubMedPubMedCentral Schwartz BG, Ludeman DJ, Mayeda GS, Kloner RA, Economides C, Burstein S. Treating refractory cardiogenic shock with the tandemheart and Impella devices: a single center experience. Cardiol Res. 2012;3(2):54–66.PubMedPubMedCentral
30.
Zurück zum Zitat Roy SK, Howard EW, Panza JA, Cooper HA. Clinical implications of thrombocytopenia among patients undergoing intra-aortic balloon pump counterpulsation in the coronary care unit. Clin Cardiol. 2010;33(1):30–5.PubMedPubMedCentral Roy SK, Howard EW, Panza JA, Cooper HA. Clinical implications of thrombocytopenia among patients undergoing intra-aortic balloon pump counterpulsation in the coronary care unit. Clin Cardiol. 2010;33(1):30–5.PubMedPubMedCentral
31.
Zurück zum Zitat Takano AM, Iwata H, Miyosawa K, et al. Reduced number of platelets during intra-aortic balloon pumping counterpulsation predicts higher cardiovascular mortality after device removal in association with systemic inflammation. Int Heart J. 2020;61(1):89–95.PubMed Takano AM, Iwata H, Miyosawa K, et al. Reduced number of platelets during intra-aortic balloon pumping counterpulsation predicts higher cardiovascular mortality after device removal in association with systemic inflammation. Int Heart J. 2020;61(1):89–95.PubMed
32.
Zurück zum Zitat Roka-Moiia Y, Li M, Ivich A, Muslmani S, Kern KB, Slepian MJ. Impella 5.5 versus Centrimag: a head-to-head comparison of device hemocompatibility. ASAIO J. 2020;66(10):1142–1151. Roka-Moiia Y, Li M, Ivich A, Muslmani S, Kern KB, Slepian MJ. Impella 5.5 versus Centrimag: a head-to-head comparison of device hemocompatibility. ASAIO J. 2020;66(10):1142–1151.
33.
Zurück zum Zitat Hassett CE, Cho SM, Hasan S, et al. Ischemic stroke and intracranial hemorrhages during Impella cardiac support. ASAIO J. 2020;66(8):e105–9.PubMed Hassett CE, Cho SM, Hasan S, et al. Ischemic stroke and intracranial hemorrhages during Impella cardiac support. ASAIO J. 2020;66(8):e105–9.PubMed
34.
Zurück zum Zitat Saeed O, Jakobleff WA, Forest SJ, et al. Hemolysis and nonhemorrhagic stroke during venoarterial extracorporeal membrane oxygenation. Ann Thorac Surg. 2019;108(3):756–63.PubMedPubMedCentral Saeed O, Jakobleff WA, Forest SJ, et al. Hemolysis and nonhemorrhagic stroke during venoarterial extracorporeal membrane oxygenation. Ann Thorac Surg. 2019;108(3):756–63.PubMedPubMedCentral
35.
Zurück zum Zitat Acharya P, Jakobleff WA, Forest SJ, et al. Fibrinogen albumin ratio and ischemic stroke during venoarterial extracorporeal membrane oxygenation. ASAIO J. 2020;66(3):277–82.PubMedPubMedCentral Acharya P, Jakobleff WA, Forest SJ, et al. Fibrinogen albumin ratio and ischemic stroke during venoarterial extracorporeal membrane oxygenation. ASAIO J. 2020;66(3):277–82.PubMedPubMedCentral
36.
Zurück zum Zitat Cavayas YA, Munshi L, Del Sorbo L, Fan E. The early change in PaCO2 after extracorporeal membrane oxygenation initiation is associated with neurological complications. Am J Respir Crit Care Med. 2020;201(12):1525–35.PubMed Cavayas YA, Munshi L, Del Sorbo L, Fan E. The early change in PaCO2 after extracorporeal membrane oxygenation initiation is associated with neurological complications. Am J Respir Crit Care Med. 2020;201(12):1525–35.PubMed
37.
Zurück zum Zitat Shou BL, Ong CS, Zhou AL, et al. Arterial carbon dioxide and acute brain injury in venoarterial extracorporeal membrane oxygenation. ASAIO J. 2022. Shou BL, Ong CS, Zhou AL, et al. Arterial carbon dioxide and acute brain injury in venoarterial extracorporeal membrane oxygenation. ASAIO J. 2022.
38.
Zurück zum Zitat Shou BL, Wilcox C, Florissi I, et al. Early low pulse pressure in VA-ECMO is associated with acute brain injury. Neurocrit Care. 2022. Shou BL, Wilcox C, Florissi I, et al. Early low pulse pressure in VA-ECMO is associated with acute brain injury. Neurocrit Care. 2022.
39.
Zurück zum Zitat Hogan M, Berger JS. Heparin-induced thrombocytopenia (HIT): review of incidence, diagnosis, and management. Vasc Med. 2020;25(2):160–73.PubMed Hogan M, Berger JS. Heparin-induced thrombocytopenia (HIT): review of incidence, diagnosis, and management. Vasc Med. 2020;25(2):160–73.PubMed
40.
Zurück zum Zitat Sokolovic M, Pratt AK, Vukicevic V, Sarumi M, Johnson LS, Shah NS. Platelet count trends and prevalence of heparin-induced thrombocytopenia in a cohort of extracorporeal membrane oxygenator patients. Crit Care Med. 2016;44(11):e1031–7.PubMed Sokolovic M, Pratt AK, Vukicevic V, Sarumi M, Johnson LS, Shah NS. Platelet count trends and prevalence of heparin-induced thrombocytopenia in a cohort of extracorporeal membrane oxygenator patients. Crit Care Med. 2016;44(11):e1031–7.PubMed
41.
Zurück zum Zitat Kelly J, Malloy R, Knowles D. Comparison of anticoagulated versus non-anticoagulated patients with intra-aortic balloon pumps. Thromb J. 2021;19(1):46.PubMedPubMedCentral Kelly J, Malloy R, Knowles D. Comparison of anticoagulated versus non-anticoagulated patients with intra-aortic balloon pumps. Thromb J. 2021;19(1):46.PubMedPubMedCentral
42.
Zurück zum Zitat Burzotta F, Trani C, Doshi SN, et al. Impella ventricular support in clinical practice: collaborative viewpoint from a European expert user group. Int J Cardiol. 2015;201:684–91.PubMed Burzotta F, Trani C, Doshi SN, et al. Impella ventricular support in clinical practice: collaborative viewpoint from a European expert user group. Int J Cardiol. 2015;201:684–91.PubMed
43.
Zurück zum Zitat Jennings DL, Nemerovski CW, Kalus JS. Effective anticoagulation for a percutaneous ventricular assist device using a heparin-based purge solution. Ann Pharmacother. 2013;47(10):1364–7.PubMed Jennings DL, Nemerovski CW, Kalus JS. Effective anticoagulation for a percutaneous ventricular assist device using a heparin-based purge solution. Ann Pharmacother. 2013;47(10):1364–7.PubMed
44.
Zurück zum Zitat Seyfarth M, Sibbing D, Bauer I, et al. A randomized clinical trial to evaluate the safety and efficacy of a percutaneous left ventricular assist device versus intra-aortic balloon pumping for treatment of cardiogenic shock caused by myocardial infarction. J Am Coll Cardiol. 2008;52(19):1584–8.PubMed Seyfarth M, Sibbing D, Bauer I, et al. A randomized clinical trial to evaluate the safety and efficacy of a percutaneous left ventricular assist device versus intra-aortic balloon pumping for treatment of cardiogenic shock caused by myocardial infarction. J Am Coll Cardiol. 2008;52(19):1584–8.PubMed
45.
Zurück zum Zitat Nishikawa M, Willey J, Takayama H, et al. Stroke patterns and cannulation strategy during veno-arterial extracorporeal membrane support. J Artif Organs. 2022;25(3):231–7.PubMed Nishikawa M, Willey J, Takayama H, et al. Stroke patterns and cannulation strategy during veno-arterial extracorporeal membrane support. J Artif Organs. 2022;25(3):231–7.PubMed
46.
Zurück zum Zitat Omar HR, Mirsaeidi M, Shumac J, Enten G, Mangar D, Camporesi EM. Incidence and predictors of ischemic cerebrovascular stroke among patients on extracorporeal membrane oxygenation support. J Crit Care. 2016;32:48–51.PubMed Omar HR, Mirsaeidi M, Shumac J, Enten G, Mangar D, Camporesi EM. Incidence and predictors of ischemic cerebrovascular stroke among patients on extracorporeal membrane oxygenation support. J Crit Care. 2016;32:48–51.PubMed
47.
Zurück zum Zitat Lee Y, Weeks PA. Effectiveness of protocol guided heparin anticoagulation in patients with the TandemHeart percutaneous ventricular assist device. ASAIO J. 2015;61(2):207–8.PubMed Lee Y, Weeks PA. Effectiveness of protocol guided heparin anticoagulation in patients with the TandemHeart percutaneous ventricular assist device. ASAIO J. 2015;61(2):207–8.PubMed
48.
Zurück zum Zitat Le Guennec L, Schmidt M, Clarencon F, et al. Mechanical thrombectomy in acute ischemic stroke patients under venoarterial extracorporeal membrane oxygenation. J Neurointerv Surg. 2020;12(5):486–8.PubMed Le Guennec L, Schmidt M, Clarencon F, et al. Mechanical thrombectomy in acute ischemic stroke patients under venoarterial extracorporeal membrane oxygenation. J Neurointerv Surg. 2020;12(5):486–8.PubMed
49.
Zurück zum Zitat Ziser A, Adir Y, Lavon H, Shupak A. Hyperbaric oxygen therapy for massive arterial air embolism during cardiac operations. J Thorac Cardiovasc Surg. 1999;117(4):818–21.PubMed Ziser A, Adir Y, Lavon H, Shupak A. Hyperbaric oxygen therapy for massive arterial air embolism during cardiac operations. J Thorac Cardiovasc Surg. 1999;117(4):818–21.PubMed
50.
Zurück zum Zitat John R, Panch S, Hrabe J, et al. Activation of endothelial and coagulation systems in left ventricular assist device recipients. Ann Thorac Surg. 2009;88(4):1171–9.PubMed John R, Panch S, Hrabe J, et al. Activation of endothelial and coagulation systems in left ventricular assist device recipients. Ann Thorac Surg. 2009;88(4):1171–9.PubMed
51.
Zurück zum Zitat Kirklin JK, Naftel DC, Myers SL, Pagani FD, Colombo PC. Quantifying the impact from stroke during support with continuous flow ventricular assist devices: an STS INTERMACS analysis. J Heart Lung Transplant. 2020;39(8):782–94.PubMed Kirklin JK, Naftel DC, Myers SL, Pagani FD, Colombo PC. Quantifying the impact from stroke during support with continuous flow ventricular assist devices: an STS INTERMACS analysis. J Heart Lung Transplant. 2020;39(8):782–94.PubMed
52.
Zurück zum Zitat Kirklin JK, Pagani FD, Kormos RL, et al. Eighth annual INTERMACS report: special focus on framing the impact of adverse events. J Heart Lung Transplant. 2017;36(10):1080–6.PubMed Kirklin JK, Pagani FD, Kormos RL, et al. Eighth annual INTERMACS report: special focus on framing the impact of adverse events. J Heart Lung Transplant. 2017;36(10):1080–6.PubMed
53.
Zurück zum Zitat Izzy S, Rubin DB, Ahmed FS, et al. Cerebrovascular accidents during mechanical circulatory support: new predictors of ischemic and hemorrhagic strokes and outcome. Stroke. 2018;49(5):1197–203.PubMed Izzy S, Rubin DB, Ahmed FS, et al. Cerebrovascular accidents during mechanical circulatory support: new predictors of ischemic and hemorrhagic strokes and outcome. Stroke. 2018;49(5):1197–203.PubMed
54.
Zurück zum Zitat Milano CA, Rogers JG, Tatooles AJ, et al. HVAD: the ENDURANCE Supplemental Trial. JACC Heart Fail. 2018;6(9):792–802.PubMed Milano CA, Rogers JG, Tatooles AJ, et al. HVAD: the ENDURANCE Supplemental Trial. JACC Heart Fail. 2018;6(9):792–802.PubMed
55.
Zurück zum Zitat Teuteberg JJ, Slaughter MS, Rogers JG, et al. The HVAD left ventricular assist device: risk factors for neurological events and risk mitigation strategies. JACC Heart Fail. 2015;3(10):818–28.PubMed Teuteberg JJ, Slaughter MS, Rogers JG, et al. The HVAD left ventricular assist device: risk factors for neurological events and risk mitigation strategies. JACC Heart Fail. 2015;3(10):818–28.PubMed
56.
Zurück zum Zitat Colombo PC, Mehra MR, Goldstein DJ, et al. comprehensive analysis of stroke in the long-term cohort of the MOMENTUM 3 Study. Circulation. 2019;139(2):155–68.PubMed Colombo PC, Mehra MR, Goldstein DJ, et al. comprehensive analysis of stroke in the long-term cohort of the MOMENTUM 3 Study. Circulation. 2019;139(2):155–68.PubMed
57.
Zurück zum Zitat Fried J, Garan AR, Shames S, et al. Aortic root thrombosis in patients supported with continuous-flow left ventricular assist devices. J Heart Lung Transplant. 2018. Fried J, Garan AR, Shames S, et al. Aortic root thrombosis in patients supported with continuous-flow left ventricular assist devices. J Heart Lung Transplant. 2018.
58.
Zurück zum Zitat Chivukula VK, Beckman JA, Prisco AR, et al. Left ventricular assist device inflow cannula angle and thrombosis risk. Circ Heart Fail. 2018;11(4):e004325.PubMed Chivukula VK, Beckman JA, Prisco AR, et al. Left ventricular assist device inflow cannula angle and thrombosis risk. Circ Heart Fail. 2018;11(4):e004325.PubMed
59.
Zurück zum Zitat Cho SM, Moazami N, Frontera JA. Stroke and intracranial hemorrhage in HeartMate II and HeartWare left ventricular assist devices: a systematic review. Neurocrit Care. 2017;27(1):17–25.PubMed Cho SM, Moazami N, Frontera JA. Stroke and intracranial hemorrhage in HeartMate II and HeartWare left ventricular assist devices: a systematic review. Neurocrit Care. 2017;27(1):17–25.PubMed
60.
Zurück zum Zitat Boehme AK, Pamboukian SV, George JF, et al. Anticoagulation control in patients with ventricular assist devices. ASAIO J. 2017;63(6):759–65.PubMedPubMedCentral Boehme AK, Pamboukian SV, George JF, et al. Anticoagulation control in patients with ventricular assist devices. ASAIO J. 2017;63(6):759–65.PubMedPubMedCentral
61.
Zurück zum Zitat Starling RC, Moazami N, Silvestry SC, et al. Unexpected abrupt increase in left ventricular assist device thrombosis. N Engl J Med. 2014;370(1):33–40.PubMed Starling RC, Moazami N, Silvestry SC, et al. Unexpected abrupt increase in left ventricular assist device thrombosis. N Engl J Med. 2014;370(1):33–40.PubMed
62.
Zurück zum Zitat Rogers JG. Changing the carburetor without changing the plugs: the intersection of stock car racing with mechanically assisted circulation. JACC Heart Fail. 2015;3(9):712–4.PubMed Rogers JG. Changing the carburetor without changing the plugs: the intersection of stock car racing with mechanically assisted circulation. JACC Heart Fail. 2015;3(9):712–4.PubMed
63.
Zurück zum Zitat Frontera JA, Starling R, Cho SM, et al. Risk factors, mortality, and timing of ischemic and hemorrhagic stroke with left ventricular assist devices. J Heart Lung Transplant. 2017;36(6):673–83.PubMed Frontera JA, Starling R, Cho SM, et al. Risk factors, mortality, and timing of ischemic and hemorrhagic stroke with left ventricular assist devices. J Heart Lung Transplant. 2017;36(6):673–83.PubMed
64.
Zurück zum Zitat John R, Naka Y, Park SJ, et al. Impact of concurrent surgical valve procedures in patients receiving continuous-flow devices. J Thorac Cardiovasc Surg. 2014;147(2):581–589; discussion 589. John R, Naka Y, Park SJ, et al. Impact of concurrent surgical valve procedures in patients receiving continuous-flow devices. J Thorac Cardiovasc Surg. 2014;147(2):581–589; discussion 589.
65.
Zurück zum Zitat Deshmukh A, Bhatia A, Sayer GT, et al. Left atrial appendage occlusion with left ventricular assist device decreases thromboembolic events. Ann Thorac Surg. 2019;107(4):1181–6.PubMed Deshmukh A, Bhatia A, Sayer GT, et al. Left atrial appendage occlusion with left ventricular assist device decreases thromboembolic events. Ann Thorac Surg. 2019;107(4):1181–6.PubMed
66.
Zurück zum Zitat Cho SM, Moazami N, Katz S, Bhimraj A, Shrestha NK, Frontera JA. Stroke risk following infection in patients with continuous-flow left ventricular assist device. Neurocrit Care. 2019;31(1):72–80.PubMed Cho SM, Moazami N, Katz S, Bhimraj A, Shrestha NK, Frontera JA. Stroke risk following infection in patients with continuous-flow left ventricular assist device. Neurocrit Care. 2019;31(1):72–80.PubMed
67.
Zurück zum Zitat Aggarwal A, Gupta A, Kumar S, et al. Are blood stream infections associated with an increased risk of hemorrhagic stroke in patients with a left ventricular assist device? ASAIO J. 2012;58(5):509–13.PubMed Aggarwal A, Gupta A, Kumar S, et al. Are blood stream infections associated with an increased risk of hemorrhagic stroke in patients with a left ventricular assist device? ASAIO J. 2012;58(5):509–13.PubMed
68.
Zurück zum Zitat Willey JZ, Boehme AK, Castagna F, et al. Hypertension and stroke in patients with left ventricular assist devices (LVADs). Curr Hypertens Rep. 2016;18(2):12.PubMedPubMedCentral Willey JZ, Boehme AK, Castagna F, et al. Hypertension and stroke in patients with left ventricular assist devices (LVADs). Curr Hypertens Rep. 2016;18(2):12.PubMedPubMedCentral
69.
Zurück zum Zitat Tabit CE, Chen P, Kim GH, et al. Elevated angiopoietin-2 level in patients with continuous-flow left ventricular assist devices leads to altered angiogenesis and is associated with higher nonsurgical bleeding. Circulation. 2016;134(2):141–52.PubMedPubMedCentral Tabit CE, Chen P, Kim GH, et al. Elevated angiopoietin-2 level in patients with continuous-flow left ventricular assist devices leads to altered angiogenesis and is associated with higher nonsurgical bleeding. Circulation. 2016;134(2):141–52.PubMedPubMedCentral
70.
Zurück zum Zitat Murase S, Okazaki S, Yoshioka D, et al. Abnormalities of brain imaging in patients after left ventricular assist device support following explantation. J Heart Lung Transplant. 2020;39(3):220–7.PubMed Murase S, Okazaki S, Yoshioka D, et al. Abnormalities of brain imaging in patients after left ventricular assist device support following explantation. J Heart Lung Transplant. 2020;39(3):220–7.PubMed
71.
Zurück zum Zitat Willey JZ, Demmer RT, Takayama H, Colombo PC, Lazar RM. Cerebrovascular disease in the era of left ventricular assist devices with continuous flow: risk factors, diagnosis, and treatment. J Heart Lung Transplant. 2014;33(9):878–87.PubMed Willey JZ, Demmer RT, Takayama H, Colombo PC, Lazar RM. Cerebrovascular disease in the era of left ventricular assist devices with continuous flow: risk factors, diagnosis, and treatment. J Heart Lung Transplant. 2014;33(9):878–87.PubMed
72.
Zurück zum Zitat Mehra MR, Crandall DL, Gustafsson F, et al. Aspirin and left ventricular assist devices: rationale and design for the international randomized, placebo-controlled, non-inferiority ARIES HM3 trial. Eur J Heart Fail. 2021;23(7):1226–37.PubMed Mehra MR, Crandall DL, Gustafsson F, et al. Aspirin and left ventricular assist devices: rationale and design for the international randomized, placebo-controlled, non-inferiority ARIES HM3 trial. Eur J Heart Fail. 2021;23(7):1226–37.PubMed
73.
Zurück zum Zitat Netuka I, Ivak P, Tucanova Z, et al. Evaluation of low-intensity anti-coagulation with a fully magnetically levitated centrifugal-flow circulatory pump-the MAGENTUM 1 study. J Heart Lung Transplant. 2018;37(5):579–86.PubMed Netuka I, Ivak P, Tucanova Z, et al. Evaluation of low-intensity anti-coagulation with a fully magnetically levitated centrifugal-flow circulatory pump-the MAGENTUM 1 study. J Heart Lung Transplant. 2018;37(5):579–86.PubMed
74.
Zurück zum Zitat • Eisen HJ, Flack JM, Atluri P, et al. Management of hypertension in patients with ventricular assist devices: a scientific statement from the American Heart Association. Circ Heart Fail. 2022;15(5):e000074. This study provides an overview of the data surrounding risks associated with hypertension in patients with an LVAD, and appropriate strategies to treat hypertension.PubMed • Eisen HJ, Flack JM, Atluri P, et al. Management of hypertension in patients with ventricular assist devices: a scientific statement from the American Heart Association. Circ Heart Fail. 2022;15(5):e000074. This study provides an overview of the data surrounding risks associated with hypertension in patients with an LVAD, and appropriate strategies to treat hypertension.PubMed
75.
Zurück zum Zitat Feldman D, Pamboukian SV, Teuteberg JJ, et al. The 2013 International Society for Heart and Lung Transplantation Guidelines for mechanical circulatory support: executive summary. J Heart Lung Transplant. 2013;32(2):157–87.PubMed Feldman D, Pamboukian SV, Teuteberg JJ, et al. The 2013 International Society for Heart and Lung Transplantation Guidelines for mechanical circulatory support: executive summary. J Heart Lung Transplant. 2013;32(2):157–87.PubMed
76.
Zurück zum Zitat Potapov EV, Antonides C, Crespo-Leiro MG, et al. 2019 EACTS Expert Consensus on long-term mechanical circulatory support. Eur J Cardiothorac Surg. 2019;56(2):230–70.PubMedPubMedCentral Potapov EV, Antonides C, Crespo-Leiro MG, et al. 2019 EACTS Expert Consensus on long-term mechanical circulatory support. Eur J Cardiothorac Surg. 2019;56(2):230–70.PubMedPubMedCentral
77.
Zurück zum Zitat Castagna F, Stohr EJ, Pinsino A, et al. The unique blood pressures and pulsatility of LVAD patients: current challenges and future opportunities. Curr Hypertens Rep. 2017;19(10):85.PubMedPubMedCentral Castagna F, Stohr EJ, Pinsino A, et al. The unique blood pressures and pulsatility of LVAD patients: current challenges and future opportunities. Curr Hypertens Rep. 2017;19(10):85.PubMedPubMedCentral
78.
Zurück zum Zitat Colombo PC, Lanier GM, Orlanes K, Yuzefpolskaya M, Demmer RT. Usefulness of a standard automated blood pressure monitor in patients with continuous-flow left ventricular assist devices. J Heart Lung Transplant. 2015. Colombo PC, Lanier GM, Orlanes K, Yuzefpolskaya M, Demmer RT. Usefulness of a standard automated blood pressure monitor in patients with continuous-flow left ventricular assist devices. J Heart Lung Transplant. 2015.
79.
Zurück zum Zitat Lanier GM, Orlanes K, Hayashi Y, et al. Validity and reliability of a novel slow cuff-deflation system for noninvasive blood pressure monitoring in patients with continuous-flow left ventricular assist device. Circ Heart Fail. 2013;6(5):1005–12.PubMed Lanier GM, Orlanes K, Hayashi Y, et al. Validity and reliability of a novel slow cuff-deflation system for noninvasive blood pressure monitoring in patients with continuous-flow left ventricular assist device. Circ Heart Fail. 2013;6(5):1005–12.PubMed
80.
Zurück zum Zitat McCullough M, Caraballo C, Ravindra NG, et al. Neurohormonal blockade and clinical outcomes in patients with heart failure supported by left ventricular assist devices. JAMA Cardiol. 2020;5(2):175–82.PubMed McCullough M, Caraballo C, Ravindra NG, et al. Neurohormonal blockade and clinical outcomes in patients with heart failure supported by left ventricular assist devices. JAMA Cardiol. 2020;5(2):175–82.PubMed
81.
Zurück zum Zitat Rettenmaier LA, Garg A, Limaye K, Leira EC, Adams HP, Shaban A. Management of ischemic stroke following left ventricular assist device. J Stroke Cerebrovasc Dis. 2020;29(12):105384.PubMed Rettenmaier LA, Garg A, Limaye K, Leira EC, Adams HP, Shaban A. Management of ischemic stroke following left ventricular assist device. J Stroke Cerebrovasc Dis. 2020;29(12):105384.PubMed
82.
Zurück zum Zitat Yoshioka D, Okazaki S, Toda K, et al. Prevalence of cerebral microbleeds in patients with continuous-flow left ventricular assist devices. J Am Heart Assoc. 2017;6(9). Yoshioka D, Okazaki S, Toda K, et al. Prevalence of cerebral microbleeds in patients with continuous-flow left ventricular assist devices. J Am Heart Assoc. 2017;6(9).
83.
Zurück zum Zitat Al-Mufti F, Bauerschmidt A, Claassen J, Meyers PM, Colombo PC, Willey JZ. Neuroendovascular interventions for acute ischemic strokes in patients supported with left ventricular assist devices: a single-center case series and review of the literature. World Neurosurg. 2016;88:199–204.PubMed Al-Mufti F, Bauerschmidt A, Claassen J, Meyers PM, Colombo PC, Willey JZ. Neuroendovascular interventions for acute ischemic strokes in patients supported with left ventricular assist devices: a single-center case series and review of the literature. World Neurosurg. 2016;88:199–204.PubMed
84.
Zurück zum Zitat Ibeh C, Tirschwell DL, Mahr C, Creutzfeldt CJ. Medical and surgical management of left ventricular assist device-associated intracranial hemorrhage. J Stroke Cerebrovasc Dis. 2021;30(10):106053.PubMedPubMedCentral Ibeh C, Tirschwell DL, Mahr C, Creutzfeldt CJ. Medical and surgical management of left ventricular assist device-associated intracranial hemorrhage. J Stroke Cerebrovasc Dis. 2021;30(10):106053.PubMedPubMedCentral
85.
Zurück zum Zitat Cho SM, Moazami N, Katz S, Starling R, Frontera JA. Reversal and resumption of antithrombotic therapy in LVAD-associated intracranial hemorrhage. Ann Thorac Surg. 2019;108(1):52–8.PubMed Cho SM, Moazami N, Katz S, Starling R, Frontera JA. Reversal and resumption of antithrombotic therapy in LVAD-associated intracranial hemorrhage. Ann Thorac Surg. 2019;108(1):52–8.PubMed
86.
Zurück zum Zitat Santos CD, Matos NL, Asleh R, et al. The dilemma of resuming antithrombotic therapy after intracranial hemorrhage in patients with left ventricular assist devices. Neurocrit Care. 2020;32(3):822–7.PubMed Santos CD, Matos NL, Asleh R, et al. The dilemma of resuming antithrombotic therapy after intracranial hemorrhage in patients with left ventricular assist devices. Neurocrit Care. 2020;32(3):822–7.PubMed
87.
Zurück zum Zitat Cho SM, Canner J, Chiarini G, et al. Modifiable risk factors and mortality from ischemic and hemorrhagic strokes in patients receiving venoarterial extracorporeal membrane oxygenation: results from the Extracorporeal Life Support Organization Registry. Crit Care Med. 2020;48(10):e897–905.PubMedPubMedCentral Cho SM, Canner J, Chiarini G, et al. Modifiable risk factors and mortality from ischemic and hemorrhagic strokes in patients receiving venoarterial extracorporeal membrane oxygenation: results from the Extracorporeal Life Support Organization Registry. Crit Care Med. 2020;48(10):e897–905.PubMedPubMedCentral
88.
Zurück zum Zitat Thiele H, Zeymer U, Neumann FJ, et al. Intraaortic balloon support for myocardial infarction with cardiogenic shock. N Engl J Med. 2012;367(14):1287–96.PubMed Thiele H, Zeymer U, Neumann FJ, et al. Intraaortic balloon support for myocardial infarction with cardiogenic shock. N Engl J Med. 2012;367(14):1287–96.PubMed
89.
Zurück zum Zitat Tanaka A, Tuladhar SM, Onsager D, et al. The subclavian intraaortic balloon pump: a compelling bridge device for advanced heart failure. Ann Thorac Surg. 2015;100(6):2151–2157; discussion 2157–2158. Tanaka A, Tuladhar SM, Onsager D, et al. The subclavian intraaortic balloon pump: a compelling bridge device for advanced heart failure. Ann Thorac Surg. 2015;100(6):2151–2157; discussion 2157–2158.
90.
Zurück zum Zitat O'Neill WW, Kleiman NS, Moses J, et al. A prospective, randomized clinical trial of hemodynamic support with Impella 2.5 versus intra-aortic balloon pump in patients undergoing high-risk percutaneous coronary intervention: the PROTECT II study. Circulation. 2012;126(14):1717–1727. O'Neill WW, Kleiman NS, Moses J, et al. A prospective, randomized clinical trial of hemodynamic support with Impella 2.5 versus intra-aortic balloon pump in patients undergoing high-risk percutaneous coronary intervention: the PROTECT II study. Circulation. 2012;126(14):1717–1727.
91.
Zurück zum Zitat Ouweneel DM, de Brabander J, Karami M, et al. Real-life use of left ventricular circulatory support with Impella in cardiogenic shock after acute myocardial infarction: 12 years AMC experience. Eur Heart J Acute Cardiovasc Care. 2019;8(4):338–49.PubMed Ouweneel DM, de Brabander J, Karami M, et al. Real-life use of left ventricular circulatory support with Impella in cardiogenic shock after acute myocardial infarction: 12 years AMC experience. Eur Heart J Acute Cardiovasc Care. 2019;8(4):338–49.PubMed
92.
Zurück zum Zitat Cho SM, Mehaffey JH, Meyers SL, et al. Cerebrovascular events in patients with centrifugal-flow left ventricular assist devices: propensity score-matched analysis from the Intermacs Registry. Circulation. 2021;144(10):763–72.PubMed Cho SM, Mehaffey JH, Meyers SL, et al. Cerebrovascular events in patients with centrifugal-flow left ventricular assist devices: propensity score-matched analysis from the Intermacs Registry. Circulation. 2021;144(10):763–72.PubMed
93.
Zurück zum Zitat Xie A, Phan K, Tsai YC, Yan TD, Forrest P. Venoarterial extracorporeal membrane oxygenation for cardiogenic shock and cardiac arrest: a meta-analysis. J Cardiothorac Vasc Anesth. 2015;29(3):637–45.PubMed Xie A, Phan K, Tsai YC, Yan TD, Forrest P. Venoarterial extracorporeal membrane oxygenation for cardiogenic shock and cardiac arrest: a meta-analysis. J Cardiothorac Vasc Anesth. 2015;29(3):637–45.PubMed
94.
Zurück zum Zitat Chang CH, Chen HC, Caffrey JL, et al. Survival analysis after extracorporeal membrane oxygenation in critically ill adults: a nationwide cohort study. Circulation. 2016;133(24):2423–33.PubMed Chang CH, Chen HC, Caffrey JL, et al. Survival analysis after extracorporeal membrane oxygenation in critically ill adults: a nationwide cohort study. Circulation. 2016;133(24):2423–33.PubMed
Metadaten
Titel
Stroke and Mechanical Circulatory Support in Adults
verfasst von
David J. Bae
Joshua Z. Willey
Chinwe Ibeh
Melana Yuzefpolskaya
Paolo C. Colombo
Publikationsdatum
03.11.2023
Verlag
Springer US
Erschienen in
Current Cardiology Reports / Ausgabe 12/2023
Print ISSN: 1523-3782
Elektronische ISSN: 1534-3170
DOI
https://doi.org/10.1007/s11886-023-01985-5

Weitere Artikel der Ausgabe 12/2023

Current Cardiology Reports 12/2023 Zur Ausgabe

Echocardiography (JM Gardin and AH Waller, Section Editors)

The Role of Artificial Intelligence in Echocardiography: A Clinical Update

Cardiovascular Critical Care (A Higgins and R Lee, Section Editors)

Shock Teams: A Contemporary Review

Cardiac Biomarkers (AA Quyyumi, Section Editor)

Biomarkers of Hepatic Dysfunction and Cardiovascular Risk

Nach Herzinfarkt mit Typ-1-Diabetes schlechtere Karten als mit Typ 2?

29.05.2024 Herzinfarkt Nachrichten

Bei Menschen mit Typ-2-Diabetes sind die Chancen, einen Myokardinfarkt zu überleben, in den letzten 15 Jahren deutlich gestiegen – nicht jedoch bei Betroffenen mit Typ 1.

Erhöhtes Risiko fürs Herz unter Checkpointhemmer-Therapie

28.05.2024 Nebenwirkungen der Krebstherapie Nachrichten

Kardiotoxische Nebenwirkungen einer Therapie mit Immuncheckpointhemmern mögen selten sein – wenn sie aber auftreten, wird es für Patienten oft lebensgefährlich. Voruntersuchung und Monitoring sind daher obligat.

GLP-1-Agonisten können Fortschreiten diabetischer Retinopathie begünstigen

24.05.2024 Diabetische Retinopathie Nachrichten

Möglicherweise hängt es von der Art der Diabetesmedikamente ab, wie hoch das Risiko der Betroffenen ist, dass sich sehkraftgefährdende Komplikationen verschlimmern.

TAVI versus Klappenchirurgie: Neue Vergleichsstudie sorgt für Erstaunen

21.05.2024 TAVI Nachrichten

Bei schwerer Aortenstenose und obstruktiver KHK empfehlen die Leitlinien derzeit eine chirurgische Kombi-Behandlung aus Klappenersatz plus Bypass-OP. Diese Empfehlung wird allerdings jetzt durch eine aktuelle Studie infrage gestellt – mit überraschender Deutlichkeit.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.