Erschienen in: Radiation Oncology 1/2015

Open Access 01.12.2015 | Research

Texture analysis on the edge-enhanced fluence of VMAT

verfasst von: So-Yeon Park, Jong Min Park, Wonmo Sung, Il Han Kim, Sung-Joon Ye

Erschienen in: Radiation Oncology | Ausgabe 1/2015

Abstract

Background

Textural features of edge-enhanced fluence were analysed to quantify modulation degree of volumetric modulated arc therapy (VMAT) plans.

Methods

Twenty prostate and twenty head and neck VMAT plans were retrospectively selected. Fluences of VMAT plans were generated by integration of monitor units shaped by multi-leaf collimators (MLCs) at each control point. When generating fluences, the values of pixels representing MLC tips were doubled to prevent smearing out of small or irregular fields (edge-enhancement). Six kinds of textural features, including angular second moment, inverse difference moment, contrast, variance, correlation and entropy, were calculated with particular displacement distances (d) of 1, 5 and 10. Plan delivery accuracy was evaluated by gamma-index method, mechanical parameter differences between plan and delivery and differences in dose-volumetric parameters between plan and delivery. Spearman’s correlation coefficients (r s ) were calculated between the values of textural features and VMAT delivery accuracy.

Results

The r s values of contrast (d = 1) with edge-enhancement to global gamma passing rates with 2%/2 mm, 1%/2 mm and 2%/1 mm were 0.546 (p < 0.001), 0.744 (p < 0.001) and 0.487 (p = 0.001), respectively. Those with local 2%/2 mm, 1%/2 mm and 2%/1 mm were 0.588, 0.640 and 0.644, respectively (all with p < 0.001). The r s values of contrast (d = 1) to MLC and gantry angle errors were -0.853 and 0.655, respectively (all with p < 0.001). The contrast (d = 1) showed statistically significant r s values in 11 dose-volumetric parameter differences from a total of 35 cases, and generally showed better correlations to plan delivery accuracy than did previously suggested textural features with non-edge-enhanced fluences, as well as conventional modulation indices.

Conclusions

Contrast (d = 1) with edge-enhanced fluences could be used as modulation index for VMAT.
Hinweise

Competing interests

This work was in part supported by the National Research Foundation of Korea (NRF) grant (no. 490-20140029 and no. 490-20130047) funded by the Korea government.

Authors’ contributions

SYP and JMP conceived of the study concept, compiled and analysed data, drafted the manuscript and participated in all aspects of the study. WMS helped to measure data. IHK analysed the data from a clinical perspective and discussed to improve the significance of this study. SJY oversaw and verified the completion of study. All authors read and approved the final manuscript.
Abkürzungen
VMAT
Volumetric modulated arc therapy
MLC
Multi-leaf collimator
IMRT
Intensity-modulated radiation therapy
TPS
Treatment planning system
QA
Quality assurance
MCSv
Modulation complexity score for volumetric modulated arc therapy
LTMCS
Leaf travel modulation complexity score
MCS
Modulation complexity score
MISPORT
Modulation index supporting station parameter optimized radiation therapy
MU
Monitor unit
CP
Control point
MIt
Modulation index of total consideration of modulating parameters of volumetric modulated arc therapy
H&N
Head and neck
PRO3
Progressive resolution optimizer 3
AAA
Anisotropic analytic algorithm
SIB
Simultaneously integrated boost
GLCM
Gray level co-occurrence matrix
d
Particular displacement distance
ASM
Angular second moment
IDM
Inverse difference moment
OAR
organ at risk
AAPM
American association of physicists in medicine
TG
Task group
CBCT
Cone beam computed tomography
Dn%
Dose received by n% volume of a structure
rs
Spearman’s rank correlation coefficient

Background

Volumetric modulated arc therapy (VMAT) enables rapid delivery of intensity-modulated photon beams by simultaneous modulations of mechanical parameters, i.e. multi-leaf collimator (MLC) positions, gantry rotation speed and dose-rate [1,2]. Since VMAT can deliver comparable or better dose distributions to a patient faster than intensity-modulated radiation therapy (IMRT), it has been widely adopted in the clinic [2-4]. However, just as with IMRT, excessive modulation of photon beam intensity of VMAT results in discrepancies in the dose distributions between treatment plans and actual delivery due to increased uncertainty in the mechanical operation of the linac [5,6]. Excessive modulation may also lead to increased small or irregular field usage, which can potentially cause inaccurate calculation of dose distributions in commercial treatment planning systems (TPS) [5,6]. Therefore, various verification methods have been suggested for IMRT and VMAT since they were introduced in the field of radiation therapy [7-13].
Pre-treatment quality assurance (QA) using the gamma-index method with a planar dose distribution measured using a detector array is a popular verification method for both IMRT and VMAT, and widely used clinically [7,13-15]. However, several recent studies have demonstrated the weak clinical relevance of gamma passing rates [8,16-18]. Nelms et al. demonstrated weak correlations between gamma passing rates and anatomy-based dose-volumetric parameters by introducing intentional errors in IMRT plans [16]. Another option for plan verification that has been suggested is the analysis of log files registered by the linac control system during IMRT or VMAT [8,9,12,18]. This method is limited in that it is not an independent verification system of delivery. Calculation based methods, such as modulation indices, for verifying the accuracy of IMRT or VMAT plan delivery have also been suggested [5,6,19-23]. For VMAT, Masi et al. suggested the modulation complexity score for VMAT (MCSv) and the leaf travel modulation complexity score (LTMCS) by modifying modulation complexity score (MCS) which was originally suggested by McNiven et al. for IMRT [20,22]. Li and Xing suggested another modulation index for VMAT (MISPORT) by quantifying MLC positional variations weighted by segmental monitor units (MU) at each control point (CP) [21]. Those studies tried to quantify modulation degree of VMAT with variations of MLC positions. Park et al. suggested modulation index (MIt) by quantifying MLC speeds, MLC accelerations, gantry rotation accelerations and dose-rate variations simultaneously [6]. Considerable correlations between the values of MIt and the gamma passing rates as well as the results of linac log file analysis were shown with statistical significances in that study. We analysed textural features calculated from fluences of VMAT plans to quantify the modulation degree of VMAT in a previous study [23]. In that study, textural features were calculated from a single fluence per VMAT plan. The fluence was generated by integration of all MUs shaped by MLC apertures (MU maps) at each CP. Although we demonstrated considerable correlations of textural features to the discrepancy between plan and delivery, some small or irregular fields at different CPs could be potentially smeared out when fluences were generated by the whole integration of various shaped MU maps. For example, if several small fields with the same MU, which could potentially cause discrepancy between plan and delivery, are contained in a VMAT plan at different CPs, and if those small fields make a single large field when they are integrated, those small fields cannot be identified with a fluence generated by the whole integration of every MU map. If we can distinguish every small field in a fluence, textural features calculated with that fluence might have better power to quantify the modulation degree of VMAT.
In this study, we tried to distinguish all the small fields in the fluence of a VMAT plan in order to calculate textural features considering every aperture at each CP. To this end, we enhanced the values of edges shaped by MLCs at each CP in a fluence. We tested the performance of the textural features calculated with edge-enhanced fluences using correlation analysis, and compared the results to the indices suggested in our previous study, as well as to the conventional set of modulation indices suggested for VMAT [6,21-23].

Methods

Sampling of VMAT plans

Twenty VMAT plans for head and neck (H&N) cancer and twenty VMAT plans for prostate cancer which were selected in our previous study were used again for this study to compare textural features calculated with edge-enhanced fluences to those calculated with non-enhanced fluences [23]. All VMAT plans were generated with 6 MV photon beams of Trilogy™ with Millennium™ 120 MLC (Varian Medical Systems, Palo Alto, CA) and used two full arcs. All the VMAT plans were optimized with the progressive resolution optimizer 3 (PRO3, ver.10, Varian Medical Systems, Palo Alto, CA) and dose distributions were calculated with the anisotropic analytic algorithm (AAA, ver.10, Varian Medical Systems, Palo Alto, CA) in the Eclipse™ system (Varian Medical Systems, Palo Alto, CA). The dose calculation grid of patient CT images was 2.5 mm. All VMAT plans were clinically acceptable, showing global gamma passing rates with gamma criterion of 2%/2 mm of higher than 90% as recommended by Heilemann et al [13]. In our institution, prostate cancer is treated with sequential delivery of a primary plan delivering 50.4 Gy to both the prostate and seminal vesicles in 28 fractions, and a boost plan delivering 30.6 Gy to the prostate in 17 fractions. Primary plans were analysed in this study. For H&N VMAT plans, prescription doses of 67.5 Gy, 54 Gy and 48 Gy were delivered to a total of 3 target volumes in 30 fractions with simultaneous integrated boost (SIB) technique.

Fluence generation with edge-enhancement

Each VMAT plan was exported in DICOM-RT format from the Eclipse™ system. Using an in-house program written in MATLAB (ver.8.1, Mathworks, Inc., Natick, MA), fluences for each plan were generated by integration of every MU shaped by the MLCs at each CP. Although the width of Millennium™ 120 MLC is 5 mm in the central region, and 10 mm for periphery region, the resolution of the fluences was set to be 1 mm for detailed analysis in the direction of MLC movement. When integrating MU maps to make a single fluence for each VMAT plan, the values (MUs) of pixels (size of 1 mm × 1 mm) representing MLC tips were doubled (edge-enhancement of fluence). In other words, the field apertures defined by MLC tips were highlighted by doubling the values of pixels representing MLC tips at each CP. The goal of this was to distinguish individual small fields at different CPs contained in a single VMAT plan. By doing this, we could reduce the probability of smearing out of some small fields at different CPs when generating a fluence which was a superposition of every MU map at each CP. Unlike IMRT, MLCs of VMAT moves in and out continuously during beam delivery [24], therefore, if we enhance the edges of MU maps parallel to the direction of MLC movement, excessively high values would be assigned at that region in a fluence. Since this could be a disturbance factor of texture analysis, and small (or irregular) fields could be identified without edge-enhancement of this region, the edges parallel to the MLC moving direction were not enhanced, but the edges perpendicular to the direction of MLC movement (i.e. MLC tip) were enhanced. Consequently, the edge-enhanced fluences showed a lot of short discrete lines perpendicular to the direction of MLC movement, in contrast to the relatively smoother fluences without edge-enhancement. The length and width of those lines were 5 mm (or 10 mm) by 1 mm due to the width of MLCs and the resolution of fluence in this study, respectively.

Calculation of textural features

The methods used to calculate textural features were the same as those in our previous study [23]. The difference in textural features between this study and our previous study was that textural features in this study were calculated with edge-enhanced fluences. First, gray level co-occurrence matrices (GLCMs) were calculated from each edge-enhanced fluence of VMAT plans in order to calculate textural features. The GLCM is a matrix or distribution indicating the co-occurring values at a given offset (i.e. particular displacement distance, d) [23,25,26]. When finding co-occurring values, the angles of searching directions were 0°, 45°, 90° and 135° for each value of d. The values of d in this study were 1, 5 and 10. Since the resolution of the fluences in this study was 1 mm, the fluence was investigated at the distance of 1 mm, 5 mm and 10 mm in the horizontal and vertical directions (\( \sqrt{2}, \) \( 5\sqrt{2} \) and \( 10\sqrt{2} \) for diagonal directions). Just as in our previous study, textural features such as angular second moment (ASM), inverse difference moment (IDM), contrast, variance, correlation and entropy were calculated with from the GLCM [23]. Since 6 kinds of textural features were calculated with 3 values of d (1, 5 and 10), a total of 18 textural features were calculated for each VMAT plan.

Quantification of plan delivery accuracy of VMAT

The data indicating VMAT plan delivery accuracy were the same as those in our previous study [23]. Three kinds of methods for each VMAT plan were adopted to verify VMAT plan delivery accuracy, which were the gamma-index method with a planar dose distribution, mechanical parameter differences between original treatment plan and linac log files registered during delivery, and differences in dose-volumetric parameters of each organ at risk (OAR) as well as target volumes between original treatment plan and the plan reconstructed with linac log files registered during delivery.
For the gamma-index method, the calculated planar dose distribution in the Eclipse™ system was compared to the dose distributions measured using a MapCHECK2™ detector array (Sun Nuclear Corporation, Melbourne, FL) inserted in the MapPHAN™ (Sun Nuclear Corporation, Melbourne, FL). For accurate measurements, the output of the linac was calibrated based on American Association of Physicists in Medicine (AAPM) Task Group (TG) 51 protocol and the readings of detectors in the MapCHECK2™ detector array were calibrated following the guidelines provided by the manufacturer, before measurements of planar dose distributions [27]. For exact setup of the device, a cone beam computed tomography (CBCT) image of the device was taken and setup was corrected by matching the CT images and CBCT images of the device before measurements. Both global and local gamma evaluations were performed with gamma criteria of 2%/2 mm, 1%/2 mm and 2%/1 mm. Following recommendations of previous studies on the gamma-index method for VMAT, gamma criteria of 3%/3 mm and 1%/1 mm were not used in this study [13,28]. Since we used gamma criterion of 2%/1 mm, the calculation grid of planar dose distribution in the Eclipse™ system was set to be 1 mm. The points of doses less than 10% of the maximum dose were not evaluated as often cited in the literature [13,15,29,30].
During measurements of planar dose distributions, both dynamic log files and DynaLog files were acquired for each VMAT plan. The information of gantry angles and delivered MUs at each CP during delivery was acquired from dynamic log files while the information of MLC positions was acquired from DynaLog files. With an in-house program written in MATLAB, dynamic log files and DynaLog files were combined into DICOM-RT format. After that, the differences in MLC positions, gantry angles and delivered MUs between original treatment plans and those recorded during delivery were calculated at each CP and averaged for each VMAT plan.
The DICOM-RT format files were imported into the Eclipse™ system and dose distributions were calculated with patient CT images under the same conditions as treatment planning for patient treatment. The differences in the values of dose-volumetric parameters between original treatment plans and plans recalculated using log files were calculated. As dose-volumetric parameters for target volumes, the dose received by 95% of target volume (D95%), D5%, the minimum, maximum and mean dose were calculated. For OARs of prostate VMAT plans, D20% and mean dose to rectal wall, D20% and mean dose to bladder and D50% and mean dose to femoral heads were calculated. For OARs of H&N VMAT plans, mean dose to each parotid gland and the maximum dose to the spinal cord, brain stem, each lens, optic chiasm and each optic nerve were calculated [23].

Correlation analysis

To investigate the correlation of the values of textural features to the VMAT plan delivery accuracy, correlation analysis between the textural features and results of 3 kinds of VMAT verification methods mentioned above was performed individually. Spearman’s rank correlation coefficients (r s ) and corresponding p values were calculated. The p values were calculated under the two-tailed unpaired parameter condition.

Results

Values of textural features

The fluences with and without edge-enhancement of prostate and H&N VMAT plans are shown in Figure 1. The GLCMs generated with edge-enhancement of those prostate and H&N VMAT plans are shown in Figure 2. The textural features of prostate and H&N VMAT plans calculated from the GLCMs, and p values showing the statistical significances of their differences are shown in Table 1. All textural features of prostate VMAT plans were different from those of H&N VMAT plans with statistical significances (all with p < 0.003). The values of ASM, contrast and variance of prostate VMAT plans were higher than those of H&N VMAT plans, while the values of IDM, correlation and entropy of H&N VMAT plans were higher than those of prostate VMAT plans. This tendency of the values of textural features calculated with edge-enhanced fluences was the same as that of the textural features with non-edge-enhanced fluences in our previous study, although the values were different from each other [23].
Table 1
The values of textural features calculated from edge-enhanced fluences
 
d= 1
d= 5
d= 10
 
Prostate
H&N
p
Prostate
H&N
p
Prostate
H&N
p
ASM (×10−3)
1.433 ± 0.170
0.842 ± 0.264
<0.001
2.183 ± 0.497
0.837 ± 0.270
<0.001
2.972 ± 0.650
0.989 ± 0.439
<0.001
IDM
0.236 ± 0.021
0.271 ± 0.027
<0.001
0.115 ± 0.013
0.167 ± 0.020
<0.001
0.082 ± 0.011
0.126 ± 0.015
<0.001
Contrast
264.73 ± 70.21
113.52 ± 51.79
<0.001
960.75 ± 236.74
341.90 ± 152.29
<0.001
1737.04 ± 484.32
537.56 ± 239.04
<0.001
Variance
45.48 ± 4.60
37.48 ± 7.74
0.001
46.41 ± 4.49
38.59 ± 8.08
0.001
46.32 ± 4.64
39.06 ± 8.10
0.002
Correlation
0.874 ± 0.021
0.921 ± 0.017
<0.001
0.583 ± 0.050
0.791 ± 0.035
<0.001
0.308 ± 0.132
0.686 ± 0.059
<0.001
Entropy
2.971 ± 0.056
3.243 ± 0.118
<0.001
2.875 ± 0.071
3.308 ± 0.124
<0.001
2.730 ± 0.092
3.287 ± 0.151
<0.001
Abbreviations: d = particular displacement distance, prostate = volumetric modulated arc therapy plans for prostate cancer, H&N = volumetric modulated arc therapy plans for head and neck cancer, ASM = angular second moment, IDM = inverse difference moment.

Correlations between textural features and gamma passing rates

The values of both global and local gamma passing rates of the 40 VMAT plans are shown in our previous study [23]. The values of r s and corresponding p values of each textural feature to global gamma passing rates are shown in Table 2. With the exceptions of r s values between global 2%/2 mm and entropy (d = 1), global 2%/1 mm and ASM (d = 1, 5 and 10), and correlation (d = 1, 5 and 10) and entropy (d = 1, 5 and 10), every value of r s was statistically significant, showing p values less than 0.05. The highest correlation was observed in contrast (d = 1) and global 1%/2 mm (r s  = 0.744 with p < 0.001). The values of r s of contrast (d = 1) calculated with edge-enhanced fluences to global gamma passing rates were generally higher than those of contrast (d = 1) and variance (d = 1) with non-edge-enhanced fluences, which showed the best performance in our previous study [23]. In addition, the values of r s of contrast (d = 1) with edge-enhanced fluences to global gamma passing rates were always higher than those of MCSv, LTMCS and MISPORT. Contrast (d = 1) with edge-enhancement showed higher values of r s than MIt to passing rates with 2%/2 mm (0.546 with p < 0.001 for contrast vs. -0.536 with p < 0.001 for MIt) and 2%/1 mm (0.487 with p < 0.001 for contrast vs. -0.361 with p = 0.022 for MIt). However, it showed lower value of r s to passing rates with 1%/2 mm than MIt (0.744 with p < 0.001 for contrast vs. -0.764 with p < 0.001 for MIt).
Table 2
The values of r s between textural features and global gamma passing rates
  
2%/2 mm
1%/2 mm
2%/1 mm
 
d
r s
p
r s
p
r s
p
ASM
1
0.373
0.018
0.623
<0.001
0.174
0.282
5
0.400
0.011
0.569
<0.001
0.133
0.415
10
0.381
0.015
0.571
<0.001
0.187
0.249
IDM
1
−0.610
<0.001
−0.690
<0.001
−0.503
0.001
5
−0.582
<0.001
−0.713
<0.001
−0.417
0.007
10
−0.540
<0.001
−0.668
<0.001
−0.362
0.022
Contrast
1
0.546
<0.001
0.744
<0.001
0.487
0.001
5
0.570
<0.001
0.725
<0.001
0.450
0.004
10
0.554
<0.001
0.740
<0.001
0.402
0.010
Variance
1
0.486
0.001
0.623
<0.001
0.485
0.002
5
0.507
0.001
0.630
<0.001
0.490
0.001
10
0.486
0.001
0.587
<0.001
0.530
<0.001
Correlation
1
−0.390
0.013
−0.648
<0.001
−0.279
0.082
5
−0.486
0.001
−0.678
<0.001
−0.247
0.125
10
−0.374
0.017
−0.633
<0.001
−0.093
0.569
Entropy
1
−0.303
0.057
−0.581
<0.001
−0.210
0.194
5
−0.363
0.021
−0.594
<0.001
−0.201
0.213
10
−0.312
0.050
−0.597
<0.001
−0.200
0.216
Abbreviations: d = particular displacement distance, rs = Spearman’s rank correlation coefficient, ASM = angular second moment, IDM = inverse difference moment.
The values of r s and corresponding p values of each textural feature to local gamma passing rates are shown in Table 3. All values of r s to local gamma passing rates were statistically significant, showing p values less than 0.05. The highest value of r s was observed between contrast (d = 1) and gamma passing rates with 2%/1 mm (r s  = 0.644 with p < 0.001). Contrast (d = 1) always showed higher values of r s with statistical significances than the other textural features to local gamma passing rates with every gamma criterion tested in this study. The values of r s of contrast (d = 1) with edge-enhanced fluences were always higher than those of contrast (d = 1) and variance (d = 1) with non-edge-enhanced fluences [23]. In addition, contrast (d = 1) with edge-enhanced fluences always showed higher correlations than did MCSv, LTMCS and MISPORT. However, compared to MIt, contrast (d = 1) with edge-enhanced fluences showed higher correlations to local gamma passing rates with 2%/1 mm, but lower values of r s to gamma passing rates with 2%/2 mm and 1%/2 mm than did MIt.
Table 3
The values of r s between textural features and local gamma passing rates
  
2%/2 mm
1%/2 mm
2%/1 mm
 
d
r s
p
r s
p
r s
p
ASM
1
0.522
0.001
0.603
<0.001
0.469
0.002
5
0.421
0.007
0.496
0.001
0.356
0.024
10
0.389
0.013
0.489
0.001
0.334
0.035
IDM
1
−0.504
0.001
−0.605
<0.001
−0.451
0.003
5
−0.528
<0.001
−0.571
<0.001
−0.494
0.001
10
−0.462
0.003
−0.528
<0.001
−0.402
0.010
Contrast
1
0.588
<0.001
0.640
<0.001
0.644
<0.001
5
0.593
<0.001
0.629
<0.001
0.613
<0.001
10
0.603
<0.001
0.642
<0.001
0.584
<0.001
Variance
1
0.519
0.001
0.525
0.001
0.594
<0.001
5
0.538
<0.001
0.538
<0.001
0.599
<0.001
10
0.464
0.003
0.477
0.002
0.549
<0.001
Correlation
1
−0.473
0.002
−0.585
<0.001
−0.458
0.003
5
−0.565
<0.001
−0.633
<0.001
−0.492
0.001
10
−0.561
<0.001
−0.605
<0.001
−0.430
0.006
Entropy
1
−0.467
0.002
−0.527
<0.001
−0.470
0.002
5
−0.447
0.004
−0.529
<0.001
−0.437
0.005
10
−0.443
0.004
−0.534
<0.001
−0.418
0.007
Abbreviations: d = particular displacement distance, rs = Spearman’s rank correlation coefficient, ASM = angular second moment, IDM = inverse difference moment.

Correlations between textural features and MLC positional errors

The r s values and corresponding p values of textural features to the differences in mechanical parameters are shown in Table 4. No statistically significant correlations were observed between textural features and the differences in MU.
Table 4
The values of r s between textural features and mechanical parameter differences
  
MLC error
Gantry angle error
MU error
 
d
r s
p
r s
p
r s
p
ASM
1
−0.747
<0.001
0.445
0.004
0.108
0.506
5
−0.749
<0.001
0.499
0.001
0.165
0.309
10
−0.785
<0.001
0.600
<0.001
0.106
0.513
IDM
1
0.604
<0.001
−0.444
0.004
−0.243
0.131
5
0.821
<0.001
−0.705
<0.001
−0.217
0.179
10
0.787
< 0.001
−0.716
< 0.001
−0.206
0.203
Contrast
1
−0.853
< 0.001
0.655
< 0.001
0.145
0.372
5
−0.843
< 0.001
0.643
< 0.001
0.160
0.323
10
−0.825
< 0.001
0.636
< 0.001
0.217
0.179
Variance
1
−0.699
< 0.001
0.526
< 0.001
0.110
0.501
5
−0.690
< 0.001
0.510
0.001
0.090
0.579
10
−0.648
< 0.001
0.507
0.001
0.040
0.808
Correlation
1
0.766
< 0.001
−0.510
0.001
−0.050
0.761
5
0.758
< 0.001
−0.540
< 0.001
−0.132
0.416
10
0.728
< 0.001
−0.533
< 0.001
−0.157
0.332
Entropy
1
0.820
< 0.001
−0.516
0.001
−0.095
0.559
5
0.823
< 0.001
−0.538
< 0.001
−0.169
0.297
10
0.841
< 0.001
−0.592
< 0.001
−0.117
0.474
Abbreviations: MLC = multi-leaf collimator, MU = monitor unit, d = particular displacement distance, rs = Spearman’s rank correlation coefficient, ASM = angular second moment, IDM = inverse difference moment.
For MLC positional errors, the highest r s value was observed between contrast (d = 1) and MLC errors (r s  = -0.853 with p < 0.001). In the case of contrast (d = 1) with edge-enhanced fluences, the r s value to MLC errors was smaller than that of contrast (d = 1) with non-edge-enhanced fluences (r s  = -0.863 with p < 0.001), LTMCS (r s  = -0.857 with p < 0.001) and MIt (r s  = 0.917 with p < 0.001) while it was larger than that of variance (d = 1) with non-edge-enhanced fluences (r s  = -0.828 with p < 0.001), MCSv (r s  = -0.635 with p < 0.001) and MISPORT (r s  = 0.795 with p < 0.001).
For gantry angle errors, the highest correlation was observed between IDM (d = 10) and gantry angles (r s  = -0.716 with p < 0.001). In the case of contrast (d = 1) with edge-enhanced fluences, r s value to gantry angle error (0.655 with p < 0.001) was smaller than those of LTMCS (r s  = -0.714 with p < 0.001) and MISPORT (r s  = 0.721 with p < 0.001) while it was larger than those of MCSv (r s  = -0.620 with p < 0.001), MIt (r s  = 0.630 with p < 0.001) and contrast (d = 1) and variance (d = 1) with non-edge-enhanced fluences (r s  = 0.639 with p < 0.001 and r s  = 0.628 with p < 0.001, respectively) [23].

Correlations between textural features and dose-volumetric parameters

The statistically significant r s values of textural features calculated with edge-enhanced fluences generated from prostate and H&N VMAT plans to differences in the clinically relevant dose-volumetric parameters between plan and delivery are shown in Tables 5 and 6, respectively. Statistically significant values of r s were found more frequently between variance (d = 1, 5 and 10) and the differences in dose-volumetric parameters (13 cases from a total of 35 cases), than between other textural features and the dose-volumetric differences. Contrast (d = 1) showed statistically significant r s values in 11 cases to the dose-volumetric parameter differences. The numbers of statistically significant r s values to the dose-volumetric parameter differences of MCSv, LTMCS, MISPORT, MIt and contrast (d = 1) and variance (d = 1) with non-edge-enhanced fluences were 3, 2, 4, 15, 4 and 10, respectively [23]. Therefore, the performance of contrast (d = 1) with edge-enhanced fluences was better than those of MCSv, LTMCS, MISPORT, contrast (d = 1) and variance (d = 1) with non-edge-enhanced fluences while it was inferior to that of MIt.
Table 5
The values of statistically significant r s of textural features to dose-volumetric parameter differences of prostate VMAT plans
 
Contrast
Correlation
 
d= 1
d= 10
d= 1
d= 10
 
d= 1
d= 10
d= 1
d= 10
Dose-volumetric parameter
r s
p
r s
p
r s
p
r s
p
D20% of rectal wall
-
-
-
-
0.485
0.030
-
-
Mean dose to rectal wall
−0.493
0.027
−0.473
0.035
0.446
0.048
-
-
Mean dose to bladder
−0.456
0.043
-
-
-
-
-
-
D50% of femoral heads
-
-
-
-
-
-
0.448
0.047
Abbreviations: d = particular displacement distance, rs = Spearman’s rank correlation coefficient, Dn% = dose received by n% volume of structure.
Table 6
The values of statistically significant r s of textural features to dose-volumetric parameter differences of head and neck VMAT plans
 
d= 1
d= 5
d= 10
 
r s
p
r s
p
r s
p
 
ASM
D5% of target 2
-
-
0.501
0.026
-
-
Mean dose to target 2
0.532
0.016
0.565
0.009
-
-
D95% of target 3
-
-
0.641
0.003
-
-
D5% of target 3
0.558
0.013
0.635
0.003
-
-
Mean dose to target 3
0.489
0.034
0.635
0.003
-
-
 
IDM
D5% of target 1
-
-
0.511
0.021
0.478
0.033
Mean dose to target 1
-
-
0.562
0.010
0.520
0.019
D95% of target 2
-
-
0.490
0.028
-
-
Minimum dose to target 2
0.447
0.048
-
-
-
-
Mean dose to target 2
0.501
0.024
0.593
0.006
0.548
0.012
D5% of target 3
-
-
0.512
0.025
-
-
Mean dose to right parotid gland
0.540
0.014
0.619
0.004
-
-
Mean dose to left parotid gland
-
-
0.506
0.023
-
-
Maximum dose to right optic nerve
0.498
0.025
0.517
0.020
-
-
 
Contrast
D5% of target 1
-
-
−0.454
0.044
−0.502
0.024
Mean dose to target 1
−0.486
0.030
−0.539
0.014
−0.488
0.029
D95% of target 2
−0.494
0.027
−0.651
0.002
-
-
D5% of target 2
−0.534
0.017
−0.618
0.004
−0.477
0.035
Minimum dose to target 2
−0.478
0.033
-
-
-
-
Mean dose to target 2
−0.626
0.003
−0.704
0.001
−0.521
0.018
D95% of target 3
-
-
−0.508
0.026
-
-
D5% of target 3
−0.576
0.010
−0.635
0.003
-
-
Mean dose to target 3
−0.484
0.036
−0.493
0.032
-
-
Mean dose to right parotid gland
−0.595
0.006
−0.516
0.020
-
-
Maximum dose to right optic nerve
−0.631
0.003
−0.477
0.033
-
-
 
Variance
D95% of target 1
−0.553
0.012
−0.558
0.011
−0.589
0.006
D5% of target 1
−0.693
0.001
−0.689
0.001
−0.688
0.001
Mean dose to target 1
−0.705
0.001
−0.708
< 0.001
−0.727
< 0.001
D95% of target 2
−0.630
0.003
−0.617
0.004
−0.626
0.003
D5% of target 2
−0.668
0.002
−0.672
0.002
−0.645
0.003
Mean dose to target 2
−0.734
< 0.001
−0.733
< 0.001
−0.734
< 0.001
D95% of target 3
−0.553
0.014
−0.548
0.015
−0.519
0.023
D5% of target 3
−0.665
0.002
−0.670
0.002
−0.691
0.001
Maximum dose to target 3
−0.511
0.025
−0.490
0.033
−0.497
0.030
Mean dose to target 3
−0.487
0.035
−0.502
0.029
−0.519
0.023
Mean dose to right parotid gland
−0.705
0.001
−0.675
0.001
−0.709
< 0.001
Mean dose to left parotid gland
−0.479
0.033
−0.444
0.050
−0.458
0.042
Maximum dose to right optic nerve
−0.568
0.009
−0.560
0.010
−0.560
0.010
 
Correlation
D5% of target 1
−0.589
0.006
−0.546
0.013
-
-
Maximum dose to target 1
-
-
−0.490
0.028
-
-
 
Entropy
D5% of target 1
-
-
−0.487
0.029
-
-
Mean dose to target 1
-
-
−0.466
0.038
-
-
D5% of target 2
-
-
−0.519
0.021
−0.451
0.047
Mean dose to target 2
−0.465
0.039
−0.537
0.015
-
-
D95% of target 3
−0.522
0.022
−0.584
0.009
−0.497
0.030
D5% of target 3
−0.598
0.007
−0.656
0.002
−0.543
0.016
Mean dose to target 3
−0.487
0.035
−0.552
0.014
-
-
Abbreviations: d = particular displacement distance, rs = Spearman’s rank correlation coefficient, Dn% = dose received by n% volume of structure, ASM = angular second moment, IDM = inverse difference moment.

Discussion

In a previous study, we demonstrated the potential of textural features calculated from fluences generated from VMAT plans as a modulation index, showing considerable correlations to VMAT delivery accuracy as quantified with gamma-index method, quantification of mechanical parameter differences between plan and delivery using linac log file and analysis on the differences in dose-volumetric parameters between plan and delivery with linac log files [23]. In that study, contrast (d = 1) and variance (d = 1) showed stronger correlations to VMAT delivery accuracy as compared to MCSv, LTMCS and MISPORT. However, as mentioned above, the effect of some small or irregular fields on the values of textural features might be smeared out because every MU map was simply integrated to generate the fluences in our previous study. Therefore, we doubled the values in the pixels representing MLC tips in a fluence to identify small or irregular fields in a single fluence, and performed correlation analysis between the textural features calculated from that fluence and VMAT delivery accuracy in this study. We found generally stronger correlations of contrast (d = 1) to VMAT delivery accuracy than those of textural features in our previous study, as well as conventional modulation indices. By enhancement of values in the region of MLC tips at each CP in a fluence, we improved the performance of contrast (d = 1) as a modulation index for VMAT.
Just as in our previous study, the values of contrast (d = 1) of lowly-modulated VMAT plans (prostate VMAT plans) were higher than those of highly-modulated VMAT plans (H&N VMAT plans) [23]. Due to the enhancement of values in the region of MLC tips in this study, values of contrast (d = 1) of both prostate and H&N VMAT plans increased compared to those calculated with non-enhanced fluences. The most noticeable improvement of contrast (d = 1) by edge-enhancement of fluences were observed in the number of statistically significant r s values to the differences in dose-volumetric parameters (4 cases with non-edge-enhancement vs. 11 cases with edge-enhancement) [23]. Besides that, performance improvements of contrast (d = 1) by edge-enhancement were observed in both global and local gamma passing rates with every gamma criterion tested in this study and gantry angle errors. Although a lower value of r s was observed between contrast (d = 1) and MLC errors by edge-enhancement of fluences (-0.853 with edge-enhanced fluences vs. -0.863 with non-edge-enhanced fluences), contrast (d = 1) still showed strong correlation to MLC errors, with a value higher than 0.8 (p < 0.001). Comparing contrast (d = 1) with edge-enhancement and variance (d = 1) with non-edge-enhancement, with the exception of correlation to global gamma passing rates with 2%/2 mm, contrast (d = 1) with enhancement always showed stronger correlations than did variance (d = 1) with non-enhancement to every method of VMAT delivery accuracy verification tested in this study [23]. Contrast (d = 1) with enhancement always showed stronger correlations to plan delivery accuracy than did MCSv [23]. In the case of LTMCS, with the exception of correlations to mechanical parameter differences, contrast (d = 1) with enhancement of fluences showed stronger correlations to every method of plan delivery accuracy verification [23]. For MISPORT, contrast (d = 1) showed stronger correlations to every method of plan delivery accuracy verification except correlation to gantry angle errors [23]. To compare contrast (d = 1) with edge-enhancement to MIt, better performance was shown by contrast (d = 1) in global gamma passing rates with 2%/2 mm and 2%/1 mm, local gamma passing rates with 2%/1 mm and gantry angle errors than MIt, while it showed inferior performance in global gamma passing rates with 1%/2 mm, local gamma passing rates with 2%/2 mm and 1%/2 mm, MLC errors and number of statistically significant dose-volumetric parameters between plan and delivery than MIt [6]. Since we quantified plan delivery accuracy with various verification methods and those results were not always consistent in this study, similar to the findings of Nelms et al. (data are not shown), neither contrast (d = 1) nor MIt always showed stronger correlations to the results of every method of plan delivery accuracy verification [16]. To determine which indicator is superior, further analysis by increasing sample size and collecting various types of samples (VMAT plans generated with various types of TPS, linacs or treatment sites and gamma evaluation with various types of detectors) should be done. This will be performed as a future work.
We acquired gamma passing rates using a single detector array (MapCHECK2™ detector array) which has a spatial resolution of 7.07 mm. The insufficient spatial resolution might cause weak correlations between gamma passing rates and the textural features in this study. However, we believe that those weak correlations came from the intrinsic limitation of the 2D gamma-index method rather than the poor resolution of MapCHECK2™ detector array considering the study by Kim et al [15]. As mentioned above, further study on this adopting various types of detectors will be done as a future work.
We could not suggest a tolerance level for contrast (d = 1) with edge-enhanced fluences in this study since the sample size was only 40. Moreover, all the VMAT plans in this study were clinically acceptable. Since no excessively-modulated VMAT plans which were clinically unacceptable were included in this study, we couldn’t acquire tolerance level for contrast (d = 1) to identify clinically unacceptable VMAT plans. As mentioned above, by increasing samples with various types of VMAT plans and by the inclusion of excessively-modulated VMAT plans, a tolerance level for contrast (d = 1) with edge-enhanced fluences will be suggested as a future work.
As shown in our previous study, the variations of gamma passing rates, mechanical parameter differences between plan and delivery and dose-volumetric parameter differences were small, as every VMAT plan in this study was clinically acceptable and used for patient treatment [23]. The global gamma passing rates with 2%/2 mm criterion recommended by Heilemann et al. for VMAT pre-treatment QA were 98.6% for prostate VMAT plans and 97.0% for H&N VMAT plans on average [13]. The mean errors in MLC positions, gantry angles and MUs were 0.24 mm, 0.39° and 0.16 MU, respectively, for prostate VMAT plans and 0.80 mm, 0.37° and 0.14 MU, respectively, for H&N VMAT plans, showing minimal differences. Within this small variation, contrast (d = 1) with edge-enhanced fluences showed considerable correlations with statistical significances to every type of verification method for VMAT delivery accuracy. Therefore, contrast (d = 1) with edge-enhanced fluences could be used as a modulation index for VMAT and it could possibly reject highly-modulated VMAT plans at the planning stage.
We could not guarantee accuracy of patient treatment with only the value of contrast (d = 1) since there are various factors affecting patient treatment accuracy, such as patient respiratory motion, setup uncertainty and anatomy changes during treatment. Although we could not predict treatment accuracy by evaluating the value of contrast (d = 1), at least, we could predict the delivery accuracy of VMAT using that value. We believe this has some value in the clinic.

Conclusions

Contrast (d = 1) calculated from fluences with enhancement of values at the tips of MLCs to prevent potential smearing out of small or irregular fields showed considerable correlations with statistical significances to gamma passing rates, mechanical errors during delivery and differences in dose-volumetric parameters between plan and delivery of VMAT. It showed stronger correlations to plan delivery accuracy than previously suggested textural features, including contrast (d = 1) and variance (d = 1) with non-edge-enhancement as well as MCSv, LTMCS and MISPORT [23]. Contrast (d = 1) with edge-enhancement could be used as a modulation index for VMAT to predict plan delivery accuracy.

Acknowledgements

The authors would like to thank Mr. Joel N.K Carlson for editing this manuscript and advising for this work.
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​4.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Competing interests

This work was in part supported by the National Research Foundation of Korea (NRF) grant (no. 490-20140029 and no. 490-20130047) funded by the Korea government.

Authors’ contributions

SYP and JMP conceived of the study concept, compiled and analysed data, drafted the manuscript and participated in all aspects of the study. WMS helped to measure data. IHK analysed the data from a clinical perspective and discussed to improve the significance of this study. SJY oversaw and verified the completion of study. All authors read and approved the final manuscript.
Literatur
1.
Zurück zum Zitat Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35:310–7.CrossRefPubMed Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35:310–7.CrossRefPubMed
2.
Zurück zum Zitat Park JM, Kim K, Chie EK, Choi CH, Ye SJ, Ha SW. RapidArc vs intensity-modulated radiation therapy for hepatocellular carcinoma: a comparative planning study. Br J Radiol. 2012;85:e323–9.CrossRefPubMedCentralPubMed Park JM, Kim K, Chie EK, Choi CH, Ye SJ, Ha SW. RapidArc vs intensity-modulated radiation therapy for hepatocellular carcinoma: a comparative planning study. Br J Radiol. 2012;85:e323–9.CrossRefPubMedCentralPubMed
3.
Zurück zum Zitat Kim H, Kim JW, Hong SJ, Rha KH, Lee CG, Yang SC, et al. Treatment outcome of localized prostate cancer by 70 Gy hypofractionated intensity-modulated radiotherapy with a customized rectal balloon. Radiat Oncol J. 2014;32:187–97.CrossRefPubMedCentralPubMed Kim H, Kim JW, Hong SJ, Rha KH, Lee CG, Yang SC, et al. Treatment outcome of localized prostate cancer by 70 Gy hypofractionated intensity-modulated radiotherapy with a customized rectal balloon. Radiat Oncol J. 2014;32:187–97.CrossRefPubMedCentralPubMed
4.
Zurück zum Zitat Yea JW. Comparison of the dose distributions with beam arrangements in the stereotactic body radiotherapy (SBRT) for primary lung cancer. Prog Med Phys. 2014;25:110–5.CrossRef Yea JW. Comparison of the dose distributions with beam arrangements in the stereotactic body radiotherapy (SBRT) for primary lung cancer. Prog Med Phys. 2014;25:110–5.CrossRef
5.
Zurück zum Zitat Du W, Cho SH, Zhang X, Hoffman KE, Kudchadker RJ. Quantification of beam complexity in intensity-modulated radiation therapy treatment plans. Med Phys. 2014;41:021716.CrossRefPubMed Du W, Cho SH, Zhang X, Hoffman KE, Kudchadker RJ. Quantification of beam complexity in intensity-modulated radiation therapy treatment plans. Med Phys. 2014;41:021716.CrossRefPubMed
6.
Zurück zum Zitat Park JM, Park SY, Kim H, Kim JH, Carlson J, Ye SJ. Modulation indices for volumetric modulated arc therapy. Phys Med Biol. 2014;59:7315–40.CrossRefPubMed Park JM, Park SY, Kim H, Kim JH, Carlson J, Ye SJ. Modulation indices for volumetric modulated arc therapy. Phys Med Biol. 2014;59:7315–40.CrossRefPubMed
7.
Zurück zum Zitat Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25:656–61.CrossRefPubMed Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25:656–61.CrossRefPubMed
8.
Zurück zum Zitat Schreibmann E, Dhabaan A, Elder E, Fox T. Patient-specific quality assurance method for VMAT treatment delivery. Med Phys. 2009;36:4530–5.CrossRefPubMed Schreibmann E, Dhabaan A, Elder E, Fox T. Patient-specific quality assurance method for VMAT treatment delivery. Med Phys. 2009;36:4530–5.CrossRefPubMed
9.
Zurück zum Zitat Teke T, Bergman AM, Kwa W, Gill B, Duzenli C, Popescu IA. Monte Carlo based, patient-specific RapidArc QA using Linac log files. Med Phys. 2010;37:116–23.CrossRefPubMed Teke T, Bergman AM, Kwa W, Gill B, Duzenli C, Popescu IA. Monte Carlo based, patient-specific RapidArc QA using Linac log files. Med Phys. 2010;37:116–23.CrossRefPubMed
10.
Zurück zum Zitat Chandraraj V, Stathakis S, Manickam R, Esquivel C, Supe SS, Papanikolaou N. Consistency and reproducibility of the VMAT plan delivery using three independent validation methods. J Appl Clin Med Phys. 2011;12:3373. Chandraraj V, Stathakis S, Manickam R, Esquivel C, Supe SS, Papanikolaou N. Consistency and reproducibility of the VMAT plan delivery using three independent validation methods. J Appl Clin Med Phys. 2011;12:3373.
11.
Zurück zum Zitat Agnew CE, King RB, Hounsell AR, McGarry CK. Implementation of phantom-less IMRT delivery verification using Varian DynaLog files and R/V output. Phys Med Biol. 2012;57:6761–77.CrossRefPubMed Agnew CE, King RB, Hounsell AR, McGarry CK. Implementation of phantom-less IMRT delivery verification using Varian DynaLog files and R/V output. Phys Med Biol. 2012;57:6761–77.CrossRefPubMed
12.
Zurück zum Zitat Manikandan A, Sarkar B, Holla R, Vivek TR, Sujatha N. Quality assurance of dynamic parameters in volumetric modulated arc therapy. Br J Radiol. 2012;85:1002–10.CrossRefPubMedCentralPubMed Manikandan A, Sarkar B, Holla R, Vivek TR, Sujatha N. Quality assurance of dynamic parameters in volumetric modulated arc therapy. Br J Radiol. 2012;85:1002–10.CrossRefPubMedCentralPubMed
13.
Zurück zum Zitat Heilemann G, Poppe B, Laub W. On the sensitivity of common gamma-index evaluation methods to MLC misalignments in Rapidarc quality assurance. Med Phys. 2013;40:031702.CrossRefPubMed Heilemann G, Poppe B, Laub W. On the sensitivity of common gamma-index evaluation methods to MLC misalignments in Rapidarc quality assurance. Med Phys. 2013;40:031702.CrossRefPubMed
14.
Zurück zum Zitat Fredh A, Scherman JB, Fog LS, Munck AF, Rosenschold P. Patient QA systems for rotational radiation therapy: a comparative experimental study with intentional errors. Med Phys. 2013;40:031716.CrossRefPubMed Fredh A, Scherman JB, Fog LS, Munck AF, Rosenschold P. Patient QA systems for rotational radiation therapy: a comparative experimental study with intentional errors. Med Phys. 2013;40:031716.CrossRefPubMed
15.
Zurück zum Zitat Kim JI, Park SY, Kim HJ, Kim JH, Ye SJ, Park JM. The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT. Radiat Oncol. 2014;9:167.CrossRefPubMedCentralPubMed Kim JI, Park SY, Kim HJ, Kim JH, Ye SJ, Park JM. The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT. Radiat Oncol. 2014;9:167.CrossRefPubMedCentralPubMed
16.
Zurück zum Zitat Nelms BE, Zhen H, Tome WA. Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors. Med Phys. 2011;38:1037–44.CrossRefPubMedCentralPubMed Nelms BE, Zhen H, Tome WA. Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors. Med Phys. 2011;38:1037–44.CrossRefPubMedCentralPubMed
17.
Zurück zum Zitat Hussein M, Rowshanfarzad P, Ebert MA, Nisbet A, Clark CH. A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems. Radiother Oncol. 2013;109:370–6.CrossRefPubMed Hussein M, Rowshanfarzad P, Ebert MA, Nisbet A, Clark CH. A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems. Radiother Oncol. 2013;109:370–6.CrossRefPubMed
18.
Zurück zum Zitat Peng J, Zhang Z, Zhou L, Zhao J, Wang J, Kong L, et al. A study on investigating the delivery parameter error effect on the variation of patient quality assurance during RapidArc treatment. Med Phys. 2013;40:031703.CrossRefPubMed Peng J, Zhang Z, Zhou L, Zhao J, Wang J, Kong L, et al. A study on investigating the delivery parameter error effect on the variation of patient quality assurance during RapidArc treatment. Med Phys. 2013;40:031703.CrossRefPubMed
19.
Zurück zum Zitat Webb S. Use of a quantitative index of beam modulation to characterize dose conformality: illustration by a comparison of full beamlet IMRT, few-segment IMRT (fsIMRT) and conformal unmodulated radiotherapy. Phys Med Biol. 2003;48:2051–62.CrossRefPubMed Webb S. Use of a quantitative index of beam modulation to characterize dose conformality: illustration by a comparison of full beamlet IMRT, few-segment IMRT (fsIMRT) and conformal unmodulated radiotherapy. Phys Med Biol. 2003;48:2051–62.CrossRefPubMed
20.
Zurück zum Zitat McNiven AL, Sharpe MB, Purdie TG. A new metric for assessing IMRT modulation complexity and plan deliverability. Med Phys. 2010;37:505–15.CrossRefPubMed McNiven AL, Sharpe MB, Purdie TG. A new metric for assessing IMRT modulation complexity and plan deliverability. Med Phys. 2010;37:505–15.CrossRefPubMed
21.
Zurück zum Zitat Li R, Xing L. An adaptive planning strategy for station parameter optimized radiation therapy (SPORT): Segmentally boosted VMAT. Med Phys. 2013;40:050701.CrossRefPubMedCentralPubMed Li R, Xing L. An adaptive planning strategy for station parameter optimized radiation therapy (SPORT): Segmentally boosted VMAT. Med Phys. 2013;40:050701.CrossRefPubMedCentralPubMed
22.
Zurück zum Zitat Masi L, Doro R, Favuzza V, Cipressi S, Livi L. Impact of plan parameters on the dosimetric accuracy of volumetric modulated arc therapy. Med Phys. 2013;40:071718.CrossRefPubMed Masi L, Doro R, Favuzza V, Cipressi S, Livi L. Impact of plan parameters on the dosimetric accuracy of volumetric modulated arc therapy. Med Phys. 2013;40:071718.CrossRefPubMed
23.
Zurück zum Zitat Park SY, Kim IH, Ye SJ, Carlson J, Park JM. Texture analysis on the fluence map to evaluate the degree of modulation for volumetric modulated arc therapy. Med Phys. 2014;41:111718.CrossRefPubMed Park SY, Kim IH, Ye SJ, Carlson J, Park JM. Texture analysis on the fluence map to evaluate the degree of modulation for volumetric modulated arc therapy. Med Phys. 2014;41:111718.CrossRefPubMed
24.
Zurück zum Zitat Vanetti E, Nicolini G, Nord J, Peltola J, Clivio A, Fogliata A, et al. On the role of the optimization algorithm of RapidArc((R)) volumetric modulated arc therapy on plan quality and efficiency. Med Phys. 2011;38:5844–56.CrossRefPubMed Vanetti E, Nicolini G, Nord J, Peltola J, Clivio A, Fogliata A, et al. On the role of the optimization algorithm of RapidArc((R)) volumetric modulated arc therapy on plan quality and efficiency. Med Phys. 2011;38:5844–56.CrossRefPubMed
25.
Zurück zum Zitat Gong P, Marceau DJ, Howarth PJ. A comparison of spatial feature-extraction algorithms for land-Use classification with spot Hrv data. Remote Sens Environ. 1992;40:137–51.CrossRef Gong P, Marceau DJ, Howarth PJ. A comparison of spatial feature-extraction algorithms for land-Use classification with spot Hrv data. Remote Sens Environ. 1992;40:137–51.CrossRef
26.
Zurück zum Zitat Scalco E, Fiorino C, Cattaneo GM, Sanguineti G, Rizzo G. Texture analysis for the assessment of structural changes in parotid glands induced by radiotherapy. Radiother Oncol. 2013;109:384–7.CrossRefPubMed Scalco E, Fiorino C, Cattaneo GM, Sanguineti G, Rizzo G. Texture analysis for the assessment of structural changes in parotid glands induced by radiotherapy. Radiother Oncol. 2013;109:384–7.CrossRefPubMed
27.
Zurück zum Zitat Almond PR, Biggs PJ, Coursey BM, Hanson WF, Huq MS, Nath R, et al. AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys. 1999;26:1847–70.CrossRefPubMed Almond PR, Biggs PJ, Coursey BM, Hanson WF, Huq MS, Nath R, et al. AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys. 1999;26:1847–70.CrossRefPubMed
28.
Zurück zum Zitat Poppe B, Blechschmidt A, Djouguela A, Kollhoff R, Rubach A, Willborn KC, et al. Two-dimensional ionization chamber arrays for IMRT plan verification. Med Phys. 2006;33:1005–15.CrossRefPubMed Poppe B, Blechschmidt A, Djouguela A, Kollhoff R, Rubach A, Willborn KC, et al. Two-dimensional ionization chamber arrays for IMRT plan verification. Med Phys. 2006;33:1005–15.CrossRefPubMed
29.
Zurück zum Zitat Gloi AM, Buchana RE, Zuge CL, Goettler AM. RapidArc quality assurance through MapCHECK. J Appl Clin Med Phys. 2011;12:3251.PubMed Gloi AM, Buchana RE, Zuge CL, Goettler AM. RapidArc quality assurance through MapCHECK. J Appl Clin Med Phys. 2011;12:3251.PubMed
30.
Zurück zum Zitat Iftimia I, Cirino ET, Xiong L, Mower HW. Quality assurance methodology for Varian RapidArc treatment plans. J Appl Clin Med Phys. 2010;11:3164.PubMed Iftimia I, Cirino ET, Xiong L, Mower HW. Quality assurance methodology for Varian RapidArc treatment plans. J Appl Clin Med Phys. 2010;11:3164.PubMed
Metadaten
Titel
Texture analysis on the edge-enhanced fluence of VMAT
verfasst von
So-Yeon Park
Jong Min Park
Wonmo Sung
Il Han Kim
Sung-Joon Ye
Publikationsdatum
01.12.2015
Verlag
BioMed Central
Erschienen in
Radiation Oncology / Ausgabe 1/2015
Elektronische ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-015-0382-z

Weitere Artikel der Ausgabe 1/2015

Radiation Oncology 1/2015 Zur Ausgabe

Erhebliches Risiko für Kehlkopfkrebs bei mäßiger Dysplasie

29.05.2024 Larynxkarzinom Nachrichten

Fast ein Viertel der Personen mit mäßig dysplastischen Stimmlippenläsionen entwickelt einen Kehlkopftumor. Solche Personen benötigen daher eine besonders enge ärztliche Überwachung.

15% bedauern gewählte Blasenkrebs-Therapie

29.05.2024 Urothelkarzinom Nachrichten

Ob Patienten und Patientinnen mit neu diagnostiziertem Blasenkrebs ein Jahr später Bedauern über die Therapieentscheidung empfinden, wird einer Studie aus England zufolge von der Radikalität und dem Erfolg des Eingriffs beeinflusst.

Erhöhtes Risiko fürs Herz unter Checkpointhemmer-Therapie

28.05.2024 Nebenwirkungen der Krebstherapie Nachrichten

Kardiotoxische Nebenwirkungen einer Therapie mit Immuncheckpointhemmern mögen selten sein – wenn sie aber auftreten, wird es für Patienten oft lebensgefährlich. Voruntersuchung und Monitoring sind daher obligat.

Costims – das nächste heiße Ding in der Krebstherapie?

28.05.2024 Onkologische Immuntherapie Nachrichten

„Kalte“ Tumoren werden heiß – CD28-kostimulatorische Antikörper sollen dies ermöglichen. Am besten könnten diese in Kombination mit BiTEs und Checkpointhemmern wirken. Erste klinische Studien laufen bereits.

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.