Skip to main content
Erschienen in: Clinical Neuroradiology 4/2023

Open Access 04.07.2023 | Original Article

Ultra-High-Resolution Time-of-Flight MR-Angiography for the Noninvasive Assessment of Intracranial Aneurysms, Alternative to Preinterventional DSA?

verfasst von: Tilman Schubert, Hakim Shakir Husain, Patrick Thurner, Jawid Madjidyar, Isabelle Barnaure, Marco Piccirelli, Markus Klarhöfer, Michaela Schmidt, Peter Speier, Christoph Forman, Zsolt Kulcsar

Erschienen in: Clinical Neuroradiology | Ausgabe 4/2023

Abstract

Purpose

The 3D time-of-flight (TOF) magnetic resonance angiography (MRA) at 3T shows high sensitivity for intracranial aneurysms but is inferior to three-dimensional digital subtraction angiography (3D-DSA) regarding aneurysm characteristics. We applied an ultra-high-resolution (UHR) TOF-MRA using compressed sensing reconstruction to investigate the diagnostic performance in preinterventional evaluation of intracranial aneurysms compared to conventional TOF-MRA and 3D-DSA.

Methods

In this study 17 patients with unruptured intracranial aneurysms were included. Aneurysm dimensions, configuration, image quality and sizing of endovascular devices were compared between conventional TOF-MRA at 3T and UHR-TOF with 3D-DSA as gold standard. Quantitatively, contrast-to-noise ratios (CNR) were compared between TOF-MRAs.

Results

On 3D-DSA, 25 aneurysms in 17 patients were detected. On conventional TOF, 23 aneurysms were detected (sensitivity: 92.6%). On UHR-TOF, 25 aneurysms were detected (sensitivity: 100%). Image quality was not significantly different between TOF and UHR-TOF (p = 0.17). Aneurysm dimension measurements were significantly different between conventional TOF (3.89 mm) and 3D-DSA (4.2 mm, p = 0.08) but not between UHR-TOF (4.12 mm) and 3D-DSA (p = 0.19). Irregularities and small vessels at the aneurysm neck were more frequently correctly depicted on UHR-TOF compared to conventional TOF. Comparison of the planned framing coil diameter and flow-diverter (FD) diameter revealed neither a statistically significant difference between TOF and 3D-DSA (coil p = 0.19, FD p = 0.45) nor between UHR-TOF and 3D-DSA (coil: p = 0.53, FD 0.33). The CNR was significantly higher in conventional TOF (p = 0.009).

Conclusion

In this pilot study, ultra-high-resolution TOF-MRA visualized all aneurysms and accurately depicted aneurysm irregularities and vessels at the base of the aneurysm comparably to DSA, outperforming conventional TOF. UHR-TOF with compressed sensing reconstruction seems to represent a non-invasive alternative to pre-interventional DSA for intracranial aneurysms.
Abkürzungen
CS
Compressed sensing
DSA
Digital subtraction angiography
FD
Flow-diverter
MRA
Magnetic resonance angiography
UHR
Ultra-high resolution

Introduction

Non-contrast time-of-flight magnetic resonance angiography (TOF-MRA) has become the standard of care in evaluation of unruptured intracranial aneurysms due to its high specificity and sensitivity as well as the lack of ionizing radiation. The limited spatial resolution and flow-related artifacts may, however, prevent detailed analysis of aneurysm morphology and of neighboring arteries [1]. Therefore, 3D digital subtraction angiography (DSA) is still considered the gold standard of treatment planning of intracranial aneurysms despite its invasive nature potentially associated with complications [2, 3]. A noninvasive, high spatial resolution imaging modality for this purpose would therefore be welcome in the diagnostic arsenal for cerebral aneurysms.
Compressed sensing (CS) is a method to accelerate MR image acquisition by k‑space undersampling and recovering the missing data through exploiting sparsity in an appropriate domain [4, 5]. CS has been applied to MRI preferentially to reduce scan time [69]. It can also be applied to increase spatial resolution while keeping scan time constant. The potential benefit of very high spatial resolution CS TOF MRA over conventional TOF MRA for treatment planning of intracranial aneurysms has however not yet been exploited.
The purpose of the present study was to compare the diagnostic performance of ultra-high-resolution (UHR) TOF MRA using compressed sensing reconstruction with conventional TOF MRA and the gold standard 3D-DSA in the preinterventional assessment of intracranial aneurysms to evaluate if UHR TOF may be an alternative to preinterventional DSA.

Material and Methods

Patients

A total of 17 patients (12 female, mean age 52 years) were included in this study. All patients signed a general informed consent form for retrospective research at the University Hospital Zurich. The study was approved by the regional ethics committee. All patients were diagnosed with an incidental intracranial aneurysm and had received diagnostic work-up including a 3T MRI examination. All patients received a dedicated MRI examination including an ultra-high-resolution CS TOF the day prior to treatment. The interval from initial MRI to UHR-TOF ranged from 3 to 207 days (average: 90 days). The UHR-TOF was performed 1 day prior to DSA. The interval from conventional TOF to DSA ranged from 4 to 208 days (average: 90.1 days).

MR Imaging Parameters

All UHR-TOF examinations were performed on a MAGNETOM Skyra 3T MR imaging scanner (Siemens Healthineers, Erlangen, Germany) using a 20-channel head-neck coil and a prototype sequence (ASP 1005E Sparse MRA [10] Siemens Healthineers). Scanning parameters were TR 21.6 ms, TE 3.3 ms, flip angle (FA) 18°, acquisition matrix 544/494, field of view 218/198 mm, slice thickness 0.27 mm, resulting in an acquired spatial resolution of 0.4 × 0.4 × 0.27 mm3 that was reconstructed to an isotropic 0.3 × 0.3 × 0.3 mm3 voxel size. The compressed sensing acceleration factor of UHR-TOF was set at 10.4. The total acquisition time was 5 min and 43 s.
Of the conventional TOF acquisitions, 10 were performed at our institution either on a MAGNETOM Skyra 3T (Siemens Healthineers) in 7 cases or on a Philips Ingenia 3T (Philips Medical, Best, Netherlands) in 3 cases. The remaining seven patients were scanned at referring institutions on a Philips Ingenia 3T (three cases), Philips Achieva 3T (two cases), Siemens Healthineers Skyra 3T and a GE Signa 3T (GE medical, Waukeesha, WI, USA) in one case each.
Mean in-plane resolution of the conventional TOF acquisition was 0.53 × 0.67 mm (range 0.35–0.6 × 0.57–0.8 mm) with an average slice thickness of 0.6 mm (range 0.55–0.7 mm).
All imaging parameters for conventional TOF examinations are listed in Table 1.
Table 1
Summary of the FOV, matrix, slice thickness, TR, TE and flip angle (FA) of all conventional TOF acquisitions as well as the in-plane unreconstructed resolution. Patients scanned at the study institution are marked “Inhouse” and were scanned either on a Siemens Magnetom Skyra 3T or Philips Ingenia 3T
 
Conventional TOF
FOVx
FOVy
Matrix x
Matrix y
In-plane resolution x
In-plane resolution y
Slice thickness
TR/TE (ms)
FA
Patient 1
Philips Achieva 3T
200
200
368
244
0.54
0.82
0.6
23/3.4
18
Patient 2
Siemens Skyra 3T
200
180
384
313
0.52
0.58
0.6
21/3.4
18
Patient 3
Siemens Skyra Inhouse
190
165
320
238
0.59
0.69
0.6
21/3.4
20
Patient 4
Philips Ingenia 3T
200
200
400
308
0.5
0.65
0.55
25/3.4
20
Patient 5
Philips Ingenia 3T
200
200
572
329
0.35
0.61
0.55
25/3.4
20
Patient 6
Philips Ingenia Inhouse
200
200
400
308
0.5
0.65
0.55
25/3.4
20
Patient 7
Philips Ingenia Inhouse
200
200
400
308
0.5
0.65
0.55
25/3.4
20
Patient 8
Philips Ingenia 3T
200
200
500
333
0.4
0.6
0.6
25/3.4
20
Patient 9
Philips Ingenia 3T
200
200
400
308
0.5
0.65
0.55
23/3.5
18
Patient 10
Siemens Skyra Inhouse
190
165
320
238
0.59
0.69
0.6
21/3.4
20
Patient 11
Philips Achieva 3T
200
200
368
244
0.54
0.82
0.6
25/6.2
20
Patient 12
Siemens Skyra Inhouse
190
165
320
238
0.59
0.69
0.6
21/3.4
20
Patient 13
GE Signa 3T
220
150
416
256
0.53
0.59
0.7
30/2.7
15
Patient 14
Siemens Skyra Inhouse
190
165
320
238
0.59
0.69
0.6
21/3.4
20
Patient 15
Siemens Skyra Inhouse
190
165
320
238
0.59
0.69
0.6
21/3.4
20
Patient 16
Siemens Skyra Inhouse
190
165
320
238
0.59
0.69
0.6
21/3.4
20
Patient 17
Siemens Skyra Inhouse
190
165
320
238
0.59
0.69
0.6
21/3.4
20
FOV Field-of-View, TR Relaxation Time, TE Echo Time, ms Milliseconds, x x-axis direction, y y-axis direction

Image Evaluation

Aneurysm dimensions were quantified on multiplanar reconstructions by a senior interventional neuroradiologist. Height, maximum and minimum diameters and maximum neck diameter in both MRA acquisitions and 3‑D DSA were measured side by side in all three modalities to adapt measurement orientations (Fig. 1). Measurements were performed on a commercially available PACS system (DeepUnity, Dedalus Healthcare, Bonn, Germany). For qualitative assessment and device sizing, images were assessed by two senior interventional neuroradiologists in consensus reading where also the number of aneurysms per patient was registered. To minimize bias, conventional TOF images were read first, followed by UHR-TOF and 3D-DSA with a minimum interval of 4 weeks between modalities. Image quality was assessed on a 3-point scale with 3 indicating completely blurred arteries and severe artifacts (poor, nondiagnostic), 2 indicating slight to moderate artifacts (sufficient) and 1 indicating excellent arterial delineation and no artifacts (excellent image quality). Aneurysm configuration (regular vs. irregular, daughter aneurysms, blebs) shape and small vessels originating at the aneurysm neck were registered as anatomical features. Regarding treatment type (coiling/flow diversion), the sizing of the framing coil or flow-diverter diameter were defined on each MRA acquisition and on DSA. In case both coiling and flow-diverter treatment were deemed feasible, both the framing coil diameter and flow-diverter diameter were recorded.
Contrast-to-noise ratios (CNR) were assessed by drawing a 3 mm2 ROI in the vertical petrous segment of the internal carotid artery and a 10 mm2 ROI in the temporal muscle on the same slice. The CNR was calculated using the following formula:
$$\text{CNR}=\frac{\text{SI}_\text{ica}-\text{SI}_\text{tm}}{\sqrt{{\text{StD}_\text{ica}}^{2}-{\text{StD}_\text{tm}}^{2}}}$$
Where SIica refers to the signal intensity in the internal carotid artery, SItm to the signal intensity in the temporal muscle and StDica refers to the standard deviation within the ROI in the internal carotid artery.

Statistical Analysis

Aneurysm dimensions between the three modalities were analyzed using one-way ANOVA. The diameter of the framing coil or flow-diverter diameter were compared using either paired 2‑sided Studentʼs t‑test or Wilcoxon-rank test depending on normal distribution. Furthermore, the sensitivity regarding aneurysm detection compared to DSA was evaluated. In order to minimize bias derived from different MR systems, we compared aneurysm dimension measurements in the subgroup that received conventional TOF and UHR TOF at different MRI scanners and in patients that received both conventional TOF and UHR TOF at the same scanner.
Significance levels were set at 0.1. Statistical analyses were performed with SPSS Version 27 (IBM, Armonk, NY, USA).

Results

On 3D digital subtraction angiography, 25 aneurysms were detected in 17 patients. On conventional TOF imaging, 23 aneurysms were detected (sensitivity: 92.6%), 2 aneurysms were not detected, 1 was a small aneurysm of the inferior wall of the internal carotid artery adjacent to a larger aneurysm (Fig. 2), the second was a recurrent, previously ruptured carotid T aneurysm treated by clipping and coiling (Fig. 3). Conventional TOF parameters of the missed aneurysm cases are added to the figure legends. On UHR-TOF imaging, 25 aneurysms were detected (sensitivity: 100%).

Image Evaluation

Image Quality

There were no poor (nondiagnostic) examinations present in either group. Average image quality was rated 1.17 for conventional TOF and 1.06 for UHR-TOF (3-point scale with 3 indicating completely blurred arteries and severe artifacts (poor, nondiagnostic), 2 indicating slight to moderate artifacts (sufficient) and 1 indicating excellent arterial delineation and no artifacts (excellent image quality)).

Anatomical Features

Irregular shapes were registered in 10 out of 23 aneurysms on conventional TOF, in 13 out of 25 aneurysms on UHR-TOF and in 12 out of 25 aneurysms on 3D-DSA. Small vessels originating close to the aneurysm neck were detected on conventional TOF in 8 out of 23 aneurysms, on UHR-TOF in 13 out of 25 aneurysms and on 3D-DSA in 12 out of 25 aneurysms. One small vessel that was detected on UHR-TOF but not on 3D-DSA was confirmed on 2D DSA (Fig. 4).

Size Measurements

Mean aneurysm dimension measurements (height, maximum and minimum diameters, maximum neck diameters combined) were 4.2 mm on 3D-DSA 4.12 mm on UHR-TOF and 3.89 mm on conventional TOF. Average aneurysm height was 4.34 mm (range 2–11.8 mm) on 3D-DSA, 4.31 mm (range 1.2–11.5 mm) on UHR-TOF and 3.95 mm (range 2.5–11-2 mm) on TOF. Average aneurysm maximum diameters were 4.93 mm (range 1.7–11.4 mm) on 3D-DSA, 4.88 mm (range 1.5–13 mm) on UHR-TOF and 4.66 mm (range 1.7–12.9 mm) on TOF. Average minimum diameters were 4.1 mm (range 1.5–11.4 mm) on 3D-DSA, 3.9 mm (range 1.2–11.4 mm) on UHR-TOF and 3.74 mm (range 1.4–11.4 mm) on TOF. Average maximum neck diameters were 3.39 mm (range 1.3–9 mm) mm on 3D-DSA, 3.37 mm (range 1.3–9.2 mm) on UHR-TOF and 3.18 mm (range 1.3–9.3 mm) on TOF.

Framing Coil Size

A total of 18 aneurysms were judged to be primarily accessible for coiling, 5 for flow-diverter treatment and 2 aneurysms were evaluated to be amenable for both flow-diverter or coiling. Mean diameter of the framing coils decided upon was 3.925 mm (range 2–12 mm) based on 3D-DSA, 3.875 mm (range 2–12 mm) based on UHR-TOF and 4.05 mm (range 2–12 mm) based on conventional TOF.

Flow-diverter Diameter

Mean flow-diverter diameter chosen based on 3D-DSA was 4.39 mm (range 3.5–5 mm) mm, 4.25 mm (range 3–5 mm) mm based on UHR-TOF and 4.22 mm (range 3–5 mm) based on conventional TOF.
Quantitative CNR analysis:
Contrast-to-noise ratio calculations resulted in an average CNR of 1.03 (range 0.06–4.72, SD 1.01) for conventional TOF and an average CNR of 0.32 (range 0.05–0.92, SD 0.26) for UHR-TOF.

Statistical Analysis

Shapiro-Wilk tests revealed a normal distribution of the aneurysm size measurements and a non-normal distribution of framing coil and flow-diverter dimensions. Therefore, a one-way analysis of variance was performed for comparison of the aneurysm dimensions and Wilcoxon-signed rank tests were performed to compare planned framing coil and flow-diverter sizes.
Comparison of aneurysm dimensions showed no statistically significant differences between 3D DSA, UHR-TOF and conventional TOF for largest neck diameter (p = 0.91), aneurysm height (p = 0.78), maximum diameter (p = 0.87) and minimum diameter (p = 0.81).
Comparison of the planned framing coil diameter revealed neither a statistically significant difference between TOF and 3D-DSA (p = 0.19) nor between UHR-TOF and 3D-DSA (p = 0.53).
Comparison of the planned flow-diverter diameter revealed neither a statistically significant difference between TOF and 3D-DSA (p = 0.45) nor between UHR-TOF and 3D-DSA (p = 0.33).
Statistical results are summarized in Table 2.
Table 2
Summary of mean values with standard deviations (in parentheses) of the aneurysm dimensions for conventional TOF, UHR-TOF and 3D-DSA. P-values resulting from the ANOVA (Analysis of Variance) for the comparison of parameters between conventional TOF, UHR-TOF and 3D-DSA are included
 
TOF
UHR TOF
3D DSA
Difference between modalities
Aneurysm height (mm)
3.95 (2.3)
4.31 (2.0)
4.34 (2.14)
P = 0.91
Maximum diameter (mm)
4.66 (2.62)
4.88 (2.31)
4.93 (2.14)
P = 0.78
Minimum diameter (mm)
3.74 (2.4)
3.9 (2.17)
4.1 (2.13)
P = 0.87
Aneurysm neck (mm)
3.18 (2.02)
3.375 (1.76)
3.39 (1.69)
P = 0.81
Comparison of image quality revealed no significant difference between UHR-TOF and conventional TOF (p = 0.17).
Contrast-to-noise ratios of conventional TOF were significantly higher than CNR of UHR-TOF (p = 0.003).

Comparison between Scanners

No significant differences between aneurysm dimension measured with conventional TOF and URH TOF were found in either patients scanned with different MR scanners (p = 0.52–0.75) or with the same scanner (p = 0.39–0.77).

Discussion

In this study, we evaluated the quantitative accuracy and preinterventional diagnostic performance of ultra-high-resolution TOF MR angiography with compressed sensing reconstruction compared to conventional TOF and 3D-DSA in a series of small aneurysms (average maximum diameter 5 mm). The main finding of the study was an equivalent sensitivity of UHR-TOF for intracranial aneurysms and accurate visualization of aneurysm features (irregularities, small vessels originating at the base) compared to 3D DSA. Conventional TOF failed to visualize two aneurysms (small aneurysm adjacent to another and recurrent aneurysm) and showed less anatomical aneurysm features than UHR-TOF.
The sizing of intrasaccular and extrasaccular devices was equally accurate with UHR-TOF and conventional TOF.
To our knowledge, no studies exist comparing the diagnostic performance of UHR-TOF and conventional TOF with DSA for preinterventional assessment of intracranial aneurysms.
In the present study, we were able to decrease the unreconstructed voxel size for the UHR-TOF roughly 5‑fold compared to conventional TOF at 3T. The optimized TOF sequence with compressed sensing reconstruction was described in a previous publication [11]. We showed that the increased spatial resolution of UHR-TOF MRA while keeping scan time in the range of conventional TOF enables visualization of aneurysm features that were not detected with conventional TOF at 3T.
The substantial increase in spatial resolution was achieved at the cost of a significantly decreased CNR. Interestingly, the higher spatial resolution and decreased CNR of UHR-TOF did not result in inferior image quality compared to conventional TOF.
The increased spatial resolution resulted in visualization of one small aneurysm, which, verified by 3D-DSA, changed the endovascular treatment approach from balloon-assisted coiling to placement of a flow-diverter (Fig. 2). In another case, a recurrent aneurysm next to coils and a microsurgical clip that was not detected on conventional TOF could clearly be visualized on UHR-TOF (Fig. 3). The improved visualization of aneurysms adjacent to foreign material introducing magnetic susceptibility is most likely due to less intravoxel dephasing in UHR-TOF because of the small voxel size.
On UHR-TOF, all cases of small arteries originating close to the aneurysm neck were correctly identified. This was superior to conventional TOF, where roughly two thirds of these vessels were visualized (Fig. 4). This feature aiding correct identification of small arteries originating at the aneurysm neck is important to assess the periprocedural risk of occluding these branches. Aneurysm irregularities were slightly underrecognized with conventional TOF and minimally overestimated with UHR-TOF.
No significant differences for aneurysm dimensions were found between modalities despite the smaller mean differences between UHR-TOF and DSA compared to conventional TOF and DSA. Accordingly, comparison between the size of intrasaccular coils and extrasaccular flow-diverters did not result in significant differences either. Although endovascular device sizing is performed based on invasive procedural imaging, potential sizes can be narrowed down based on both preinterventional ultra-high-resolution UHR-TOF but also conventional TOF.
Compressed sensing (CS) reconstruction has been successfully applied to TOF MR angiography in different neurovascular pathologies [4, 68, 10, 1214]. Several publications have focused on the comparison of CS-TOF and conventional TOF MRA [68, 14, 15]. The general conclusions of these publications were a comparable image quality and clinical feasibility of CS-TOF compared to conventional TOF; however, no study is available that utilized compressed sensing reconstruction to heavily increase spatial resolution in order to evaluate its use in the evaluation of intracranial aneurysms.
The main limitation of the present study is that 7 out of 17 conventional TOF examinations were performed at external sites with different scanning parameters; however, all external conventional TOF examinations were performed on state of the art 3T-MRI scanners. Furthermore, the interval from conventional TOF to UHR-TOF was 90 days on average. Potentially, aneurysms might show small changes in size in this period.
In conclusion, ultra-high-resolution TOF MRA showed an equivalent performance in visualization of aneurysm characteristics compared to DSA in our study. Furthermore, all aneurysms were accurately diagnosed; therefore, UHR TOF has the potential to serve as a reliable and noninvasive alternative to preinterventional diagnostic angiography and outperformed conventional TOF in the present study; however, the results of this pilot study need to be confirmed in larger series.

Funding

No specific funding

Declarations

Conflict of interest

T. Schubert, H.S. Husain, P. Thurner, M. Piccirelli, J. Madjidyar, I. Barnaure and Z. Kulcsar declare that they have no competing interests. M. Klarhöfer, M. Schmidt, P. Speier and C. Forman are employees of Siemens Healthineers, no conflict of interest reported.

Ethical standards

This study was conducted according to the 1964 Declaration of Helsinki and its amendments and was approved by the regional ethics committee. Informed consent: all patients signed a general informed consent for retrospective research at the Zurich University Hospital.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Med Neurologie & Psychiatrie

Kombi-Abonnement

Mit e.Med Neurologie & Psychiatrie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

e.Med Neurologie

Kombi-Abonnement

Mit e.Med Neurologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes, den Premium-Inhalten der neurologischen Fachzeitschriften, inklusive einer gedruckten Neurologie-Zeitschrift Ihrer Wahl.

Weitere Produktempfehlungen anzeigen
Literatur
6.
Zurück zum Zitat Fushimi Y, Okada T, Kikuchi T, Yamamoto A, Okada T, Yamamoto T, Schmidt M, Yoshida K, Miyamoto S, Togashi K. Clinical evaluation of time-of-flight MR angiography with sparse undersampling and iterative reconstruction for cerebral aneurysms. NMR Biomed. 2017; https://doi.org/10.1002/nbm.3774.CrossRefPubMed Fushimi Y, Okada T, Kikuchi T, Yamamoto A, Okada T, Yamamoto T, Schmidt M, Yoshida K, Miyamoto S, Togashi K. Clinical evaluation of time-of-flight MR angiography with sparse undersampling and iterative reconstruction for cerebral aneurysms. NMR Biomed. 2017; https://​doi.​org/​10.​1002/​nbm.​3774.CrossRefPubMed
7.
Zurück zum Zitat Lin Z, Zhang X, Guo L, Wang K, Jiang Y, Hu X, Huang Y, Wei J, Ma S, Liu Y, Zhu L, Zhuo Z, Liu J, Wang X. Clinical feasibility study of 3D intracranial magnetic resonance angiography using compressed sensing. J Magn Reson Imaging. 2019;50:1843–51. https://doi.org/10.1002/jmri.26752.CrossRefPubMed Lin Z, Zhang X, Guo L, Wang K, Jiang Y, Hu X, Huang Y, Wei J, Ma S, Liu Y, Zhu L, Zhuo Z, Liu J, Wang X. Clinical feasibility study of 3D intracranial magnetic resonance angiography using compressed sensing. J Magn Reson Imaging. 2019;50:1843–51. https://​doi.​org/​10.​1002/​jmri.​26752.CrossRefPubMed
9.
10.
Zurück zum Zitat Stadler A. Accelerating TOF MRA in clinical practice using sparse MRI with variable poisson density sampling. Intl Soc Mag Reson Med. 2015;23. Stadler A. Accelerating TOF MRA in clinical practice using sparse MRI with variable poisson density sampling. Intl Soc Mag Reson Med. 2015;23.
13.
15.
Zurück zum Zitat Sakata A, Fushimi Y, Okada T, Nakajima S, Hinoda T, Speier P, Schmidt M, Forman C, Yoshida K, Kataoka H, Miyamoto S, Nakamoto Y. Evaluation of cerebral arteriovenous shunts: a comparison of parallel imaging time-of-flight magnetic resonance angiography (TOF-MRA) and compressed sensing TOF-MRA to digital subtraction angiography. Neuroradiology. 2021;63:879–87. https://doi.org/10.1007/s00234-020-02581-y.CrossRefPubMed Sakata A, Fushimi Y, Okada T, Nakajima S, Hinoda T, Speier P, Schmidt M, Forman C, Yoshida K, Kataoka H, Miyamoto S, Nakamoto Y. Evaluation of cerebral arteriovenous shunts: a comparison of parallel imaging time-of-flight magnetic resonance angiography (TOF-MRA) and compressed sensing TOF-MRA to digital subtraction angiography. Neuroradiology. 2021;63:879–87. https://​doi.​org/​10.​1007/​s00234-020-02581-y.CrossRefPubMed
Metadaten
Titel
Ultra-High-Resolution Time-of-Flight MR-Angiography for the Noninvasive Assessment of Intracranial Aneurysms, Alternative to Preinterventional DSA?
verfasst von
Tilman Schubert
Hakim Shakir Husain
Patrick Thurner
Jawid Madjidyar
Isabelle Barnaure
Marco Piccirelli
Markus Klarhöfer
Michaela Schmidt
Peter Speier
Christoph Forman
Zsolt Kulcsar
Publikationsdatum
04.07.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Neuroradiology / Ausgabe 4/2023
Print ISSN: 1869-1439
Elektronische ISSN: 1869-1447
DOI
https://doi.org/10.1007/s00062-023-01320-z

Weitere Artikel der Ausgabe 4/2023

Clinical Neuroradiology 4/2023 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.