Commissural misalignment independently predicts leaflet thrombosis after transcatheter aortic valve implantation

verfasst von: Susanne Jung, Fabian Ammon, Silvia Smolka, Maximilian Moshage, Mohamed Marwan, Stephan Achenbach

Erschienen in: Clinical Research in Cardiology | Ausgabe 1/2024

Abstract

Aims

Transcatheter aortic valve implantation (TAVI) has become a minimally invasive alternative to surgical aortic valve replacement. Hypo-attenuated leaflet thickening (HALT)—a marker of subclinical leaflet thrombosis commonly detected by cardiac computed tomography (CT) after TAVI—may influence valve durability and function. The purpose of this study was to compare commissural alignment of the native and prosthetic aortic valves in cardiac CT in subjects with and without HALT and thereby identify commissural misalignment as potential predictor for leaflet thrombosis after TAVI.

Methods and results

In 170 subjects, 85 with and 85 without HALT in post-TAVI CT, commissural orientation of the prosthesis was determined comparing native and prosthetic aortic valve orientation in cardiac CT by measuring the commissural angle relative to the right coronary ostium in the aortic valve plane. For the prosthetic valve, any deviation ≤ 15° compared to the native valve was classified as “aligned”; 16–30° as “mild”, 31–45° as “moderate” and ≥ 45° as “severe” misalignment.
Among subjects with HALT, median angular deviation was higher (36°, IQR 31°) than in the control group (29°, IQR 29°, p = 0.042). “Severe” misalignment was more frequent in subjects who developed HALT (n = 31, 37%) compared to the control group (n = 17, 20%, p = 0.013). In logistic regression analysis, more severe deviation (p = 0.015, OR = 1.02 per 1° deviation) and “severe” misalignment (p = 0.018, OR = 2.2) represented independent predictors for the occurrence of HALT after TAVI.

Conclusion

Subclinical leaflet thrombosis after TAVI is associated with commissural misalignment. Potential clinical advantages of obtaining commissural alignment remain to be systematically assessed.

Graphical abstract

Association of HALT with commissural misalignment after TAVI. HALT hypo-attenuated leaflet thickening, IQR interquartile range, TAVI transfemoral aortic valve replacement
Abkürzungen
BMI
Body mass index
CT
Computed tomography
HALT
Hypo-attenuated leaflet thickening
IQR
Interquartile range
LCC
Left coronary cusp
LV-EF
Left ventricular ejection fraction
NCC
Non-coronary cusp
p max
Peak pressure gradient
p mean
Mean pressure gradient
RCA
Right coronary artery
RCC
Right coronary cusp
SAVR
Surgical aortic valve replacement
STS
Society of Thoracic Surgeons
TAVI
Transfemoral aortic valve implantation
v max
Peak velocity

Introduction

Transcatheter aortic valve implantation (TAVI) has become a minimally invasive alternative to surgical aortic valve replacement (SAVR) in subjects with high or intermediate surgical risk [1]. Increasingly, TAVI emerges as a valid treatment option for severe symptomatic aortic stenosis even in subjects with lower surgical risk [24]. With the indication of TAVI expanding towards subjects with lower risk profile and longer life expectancy, issues with respect to long-term durability and function of the implanted TAVI prosthesis gain increasing importance.
In this context, the influence of subclinical leaflet thrombosis after TAVI on valve durability and hemodynamics is currently under debate [57]. Hypo-attenuated leaflet thickening (HALT)—a marker of subclinical leaflet thrombosis commonly detected by cardiac computed tomography (CT) after TAVI—can be found in 10–20% of subjects undergoing TAVI or SAVR [812]. Subclinical leaflet thrombosis and reduced leaflet motion in bioprosthetic valves have been described to be associated with a higher incidence of transient ischemic attack and other thromboembolic complications [9, 13]. A recent publication by Garcia et al. even suggests an increased mortality in subjects with HALT compared to those without HALT after TAVI [10]. The elucidation of potential pathomechanisms favoring HALT after TAVI is, therefore, of major importance.
In contrast to SAVR, during which visual inspection allows the alignment of the bioprosthetic valve commissures with those of the native aortic valves, positioning of the bioprosthesis during TAVI usually occurs randomly and may lead to commissural misalignment [14]. Besides potentially causing overlap between the neocommissural posts and the coronary arteries, commissural misalignment has been shown to contribute to aortic regurgitation and leaflet stress [1416].
We hypothesize that commissural misalignment, through altered flow patterns in the aortic root, may furthermore contribute to the occurrence of subclinical leaflet thrombosis. Therefore, the purpose of this study was to compare the commissural alignment of the native and prosthetic aortic valves in cardiac CT in subjects with and without HALT and thereby identify potential predictors for subclinical leaflet thrombosis after TAVI.

Methods

Study population

We present a retrospective analysis of a total of 170 patients—85 patients with HALT following TAVI and 85 controls. Study subjects were chosen from a cohort of 1200 patients who underwent transfemoral TAVI at our center between 2010 and 2020. Of these subjects, those who did not undergo cardiac CT after TAVI or who were lost to regular follow-up were excluded from further analysis. Of the remaining patients, those with HALT were chosen for our analysis and compared to a group of subjects who had also undergone CT after TAVI but had not demonstrated HALT. All of these patients suffered from severe symptomatic aortic valve stenosis and underwent cardiac CT before and after transfemoral TAVI at the Departments of Cardiology and Cardiac Surgery, University Hospital Erlangen, Germany. Subjects who underwent valve-in-valve TAVI were not included in this analysis. Severe aortic stenosis was defined according to 2021 European Society of Cardiology Guidelines for the Management of Valvular Heart Disease [1]. The indication for TAVI was established within the Heart Team. Written informed consent was obtained from each subject for the anonymous use of their data. Approval for retrospective evaluation of patient data for this study was obtained from the local ethics committee and the study was conducted according to the tenets of the Declaration of Helsinki and the principles of good clinical practice guidelines.

Clinical parameters

Clinical data were assessed in each subject, including demographic data, medical history as well as type and size of the implanted prosthesis. For each subject, the STS (Society of Thoracic Surgeons) Score and logistic EuroSCORE (European System for Cardiac Operative Risk Evaluation) were determined. In addition, transthoracic echocardiography was performed in each subject before and after TAVI as well as during follow-up including the assessment of peak and mean pressure gradients (pmax and pmean) as well as peak velocity (vmax) across the aortic valve.

TAVI procedure

All subjects included in the analysis underwent transfemoral TAVI. The procedure was performed in a hybrid operating room under fluoroscopic guidance by a multidisciplinary team consisting of an anesthesiologist, cardiologists and cardiac surgeons. The procedure was either performed in general anesthesia or sedation, depending on the patient’s pre-procedural medical condition and operative risk. Selection of prosthesis type and size was based aortic root anatomy as determined by pre-procedural contrast-enhanced ECG-gated multidetector computed tomography.

CT acquisition and assessment of commissural orientation in cardiac CT

All subjects underwent dual-source CT scanning (Somatom Force, Siemens Healthineers, Erlangen, Germany) twice—up to 3 months before and within 12 months after TAVI. The subjects included in the analysis were non-consecutively selected from a group of patients, who, according to physician decision, received routine follow-up cardiac CT examination. Physician decision to perform a follow-up cardiac CT examination depended on the clinical status, symptoms, echocardiographic findings and renal function of the patients after TAVI. CT data acquisition was performed electrocardiogram-gated and contrast-enhanced according to a site-specific protocol for CT imaging in preparation for TAVI [17]. All cardiac CT scans were evaluated by the same experienced cardiologist who was blinded to all patient characteristics. HALT was defined as hypo-attenuated opacity attached to the valve affecting one or more leaflets evaluated in two-dimensional multiplanar reconstructions, as described previously [1820]. Commissural orientation of the prosthesis was determined by comparing native (pre-TAVI scan) and prosthetic (post-TAVI scan) aortic valve orientation in cardiac CT using the software syngo.via (Siemens Healthineers, Erlangen, Germany) according to the method established by Fuchs et al. [14]. For this purpose, the orientations of the valve and prosthetic commissures were determined by measuring their angle relative to the right coronary artery (RCA) ostium in the aortic valve plane. This results in three angles—from the RCA ostium to the commissure between the right (RCC) and left coronary cusp (LCC), from the RCA ostium to the commissure between the LCC and non-coronary cusp (NCC) as well as from the RCA ostium to the commissure between the NCC and RCC. Deviations of these three angles (∆ angle) were determined comparing the angles in cardiac CT images before and after TAVI. Out of these three angular deviations, one mean angular deviation was finally calculated. For the prosthetic valve, a deviation ≤ 15° compared to the native valve was classified as “aligned”, 16°–30° as “mild” commissural misalignment, 31–45° as “moderate” and ≥ 45° as “severe” commissural misalignment (Fig. 1). This method was feasible in 100% of study subjects.

Statistical analysis

All analyses were performed using SPSS software, version 21.0 (IBM Corporation, Chicago, IL, USA). First, distribution of data was assessed by Kolmogorov–Smirnov test and data were expressed as median (interquartile range, IQR) as they were not normally distributed. Not normally distributed metric data were compared using non-parametric tests for unpaired variables. A two-sided p value < 0.05 was considered statistically significant. For comparison of categorical variables, chi-square test or Fisher’s exact test for unpaired variables was used. Furthermore, logistic regression analysis was performed to assess potential predictors for HALT after TAVI.

Results

Clinical characteristics

Table 1 provides the clinical characteristics of the study groups. Median age in both groups was 81 (7) years (p = 0.640). N = 44 subjects (52%) in the HALT group and n = 45 subjects (53%) in the control group were male (p = 1.0). Logistic euroSCORE was 14 (13) in the HALT group and 12 (13) in the control group (p = 0.248), STS-Score was 2 (2) in both groups (p = 0.437).
Table 1
Clinical characteristics, intra- and post-procedural data of the study population
 
HALT group (n = 85)
Control group (n = 85)
p value
Clinical characteristics
 Age [years]
81 (7)
81 (7)
0.640
 Gender [male/female]
44/41
45/40
1.0
 BMI [kg/m2]
28 (5)
27 (5)
0.063
 Log. euroSCORE [%]
14 (13)
12 (13)
0.248
 STS-Score [%]
2 (2)
2 (2)
0.437
 LV-EF [%]
55 (9)
60 (5)
0.192
 Oral anticoagulation
11 (13%)
17 (20%)
0.301
Intraprocedural data
   
 Category of prosthesis n (%)
   
  Balloon-expandable
40 (47%)
47 (55%)
0.357
  Self-expandable
45 (53%)
38 (45%)
 
 Size of prosthesis n (%)
   
  23 mm
9 (10.6%)
7 (8.2%)
0.599
  25 mm
8 (9.4%)
11 (12.9%)
0.465
  26 mm
15 (17.6%)
21 (24.7%)
0.260
  27 mm
23 (27.1%)
15 (17.6%)
0.141
  29 mm
30 (35.3%)
31 (36.5%)
0.873
Postprocedural data
 pmax [mmHg]
16 (12)
16 (10)
0.469
 pmean [mmHg]
8 (7)
9 (5)
0.343
 vmax [m/s]
2 (1)
2 (1)
0.373
Follow-up
 pmax [mmHg]
18 (14)
15 (10)
0.081
 pmean [mmHg]
9 (8)
8 (5)
0.067
 vmax [m/s]
2 (1)
2 (1)
0.126
Data are given as median (interquartile range, IQR) or n (%)
BMI body mass index, HALT hypo-attenuated leaflet thickening, LV-EF left ventricular ejection fraction, pmax peak pressure gradient, pmean mean pressure gradient, STS Society of Thoracic Surgeons,, vmax peak velocity
N = 11 (13%) subjects from the HALT group and n = 17 (20%) subjects from the control group were on oral anticoagulation (p = 0.301). All other subjects received antiplatelet therapy. The number of subjects under anticoagulation or antiplatelet therapy did not differ significantly between subjects with and without HALT after TAVI (p = 0.301). Left ventricular ejection fraction (LV-EF) was 55 (9) % in the HALT group and 60 (5) % in the control group (p = 0.192).
In the HALT group, 40 subjects (47%) received a balloon-expandable and 45 (53%) a self-expandable valve. In the control group, 47 (55%) subjects received a balloon-expandable and 38 (45%) a self-expandable valve (p = 0.179). Distribution of valve size did not differ significantly between the two groups (Table 1).
Postprocedural pmax and pmean were 16 (12) and 8 (7) mmHg in the HALT group and 16 (10) and 9 (5) mmHg in the control group (p = 0.469 and 0.343, respectively). Vmax was 2 (1) m/s in both groups (p = 0.373). Similarly, pmax and pmean in the follow-up echocardiography did not differ significantly between the two groups (HALT group: 18 (14) and 9 (8) mmHg, control group: 15 (10) and 8 (5) mmHg, p = 0.081 and 0.067, respectively). Vmax in follow-up echocardiography was 2 (1) m/s (p = 0.126).

Commissural orientation

Median angular deviation was significantly higher in the HALT group (36 (31)°) compared to the control group (29 (29)°, p = 0.042). In this context, the angular deviation of the RCA to the RCC/LCC commissure was significantly higher in subjects with HALT (38 (29)°) in comparison to those without HALT (28 (26)°, p = 0.012). Angular deviation of the RCA to the LCC/NCC commissure was numerically higher in the HALT group (37 (42)°) than in the control group (30 (36)°), but did not reach statistical significance (p = 0.108). Same was true for the angular deviation of the RCA to the NCC/RCC commissure (HALT group: 31 (35)°), control group: 29 (21)°), p = 0.203).
Commissural alignment could be found significantly less often in subjects with HALT (n = 13, 15%) than in subjects from the control group (n = 23, 27%, p = 0.045). However, there was no significant difference with respect to “mild” (HALT group: n = 20, 24%; control group: n = 23, 27%; p = 0.362) or “moderate” commissural misalignment (both groups: n = 22, 26%, p = 0.569) in subjects with and without HALT. However, “severe” commissural misalignment could be detected more frequently in subjects with HALT (n = 31, 37%) than in the control group (n = 17, 29%, p = 0.013). A comparison of commissural orientation between subjects with and without HALT after TAVI is given in Table 2.
Table 2
Comparison of commissural alignment between subjects with and without HALT after TAVI
 
HALT (n = 85)
No HALT (n = 85)
p value
∆ Angles [°]
 Median ∆ angle
36 (31)
29 (29)
0.042
 ∆ Angle RCA–commissure RCC/LCC
38 (29)
28 (26)
0.012
 ∆ Angle RCA–commissure LCC/NCC
37 (42)
30 (36)
0.108
 ∆ Angle RCA–commissure NCC/RCC
31 (35)
29 (21)
0.203
Commissural (mis-)alignment
 Alignment
13 (15%)
23 (27%)
0.045
 Mild commissural misalignment
20 (24%)
23 (27%)
0.362
 Moderate commissural misalignment
22 (26%)
22 (26%)
0.569
 Severe commissural misalignment
31 (37%)
17 (29%)
0.013
Data are given as median (interquartile range, IQR) or n (%)
∆ angle angular deviation, HALT hypo-attenuated leaflet thickening, LCC left coronary cusp, NCC non-coronary cusp, RCA right coronary artery, RCC right coronary cusp, TAVI transfemoral aortic valve implantation

Predictors for HALT after TAVI

Logistic regression analysis was performed in order to identify potential predictors for HALT after TAVI (see Table 3). Univariate regression analysis revealed that echocardiographic parameters during follow-up, such as pmax (p = 0.008, OR 1.05 per mmHg), pmean (p = 0.004, OR 1.11 per mmHg) and vmax (p = 0.031, OR 2.3 per m/s) were significantly associated with leaflet thrombosis. Additionally, more severe median angular deviation (p = 0.015, OR 1.02 per 1°), more severe deviation between the RCA and the RCC/LCC commissure (p = 0.009, OR 1.02 per 1°) as well as between the RCA and the LCC/NCC commissure (p = 0.044, OR 1.01 per 1°) could be identified as relevant predictors for HALT after TAVI. In this context, another relevant predictor was the occurrence of severe commissural misalignment (p = 0.018, OR 2.3). In multivariate regression analysis, pmean (p = 0.009, OR 1.12) and severe commissural misalignment (p = 0.010, OR 3.14) remained significant predictors for HALT after TAVI. In contrast, other parameters like size and type of implanted prosthesis (p = 0.935 and p = 0.283), body mass index (BMI, p = 0.132) or anticoagulation regimen (p = 0.851) were no significant predictors for HALT after TAVI.
Table 3
Predictors for the occurrence of HALT after TAVI
 
p value
OR
Univariate regression analysis
 Median ∆ angle
0.015
1.021
 ∆ Angle RCA–commissure RCC/LCC
0.009
1.023
 ∆ Angle RCA–commissure LCC/NCC
0.044
1.013
 Severe CMA
0.018
2.296
 pmax follow-up
0.008
1.050
 pmean follow-up
0.004
1.109
 vmax follow-up
0.031
2.275
Multivariate regression analysis
 pmean follow-up
0.009
1.123
 Severe commissural misalignment
0.010
3.140
∆ angle angular deviation, HALT hypo-attenuated leaflet thickening, LCC left coronary cusp, NCC non-coronary cusp, OR odds ratio, pmax peak pressure gradient, pmean mean pressure gradient, RCA right coronary artery, RCC right coronary cusp, TAVI transfemoral aortic valve implantation, vmax peak velocity

Discussion

In the current analysis, we compared the commissural alignment of the native and prosthetic aortic valves in cardiac CT in subjects with and without HALT after TAVI. There are three key findings: first, median angular deviation was higher in the HALT group compared to the control group, as was angular deviation of the RCA ostium to the RCC/LCC commissure. Second, “severe” commissural misalignment (defined as angular deviation between 45 and 60°) could be detected more frequently in subjects with HALT than in the control group, whereas commissural alignment was found less often in the HALT group. Third, high angular deviation and “severe” commissural misalignment were independent predictors for HALT after TAVI. Therefore, our results suggest that commissural misalignment may contribute to the occurrence of subclinical leaflet thrombosis after TAVI.
The findings of the current analysis are in accordance with those of a study by Khan et al. investigating anatomical characteristics associated with leaflet thrombosis after TAVI in a total of 167 subjects, from which 26 (15.6%) had HALT. There was a higher numerical incidence of commissural misalignment in subjects with HALT compared to those without HALT, but this did not reach statistical significance, possibly due to the low number of subjects with HALT [21]. Our analysis—including a higher number of subjects with HALT—was able to demonstrate a significantly higher incidence of commissural misalignment in subjects with HALT, thereby extending the results by Khan et al.
The findings of the current analysis can be interpreted in the context of Virchow’s triad. According to this theory, there are three factors contributing to the occurrence of thrombosis—blood constitution, endothelial dysfunction and alterations in blood flow patterns. Especially, the last two factors may explain the association between commissural misalignment and HALT described in the current analysis. With respect to endothelial dysfunction, Jilaihawi et al. suggested various factors contributing to delayed re-endothelialization, consecutively increasing the risk of leaflet thrombosis. These factors were stent frame expansion and fracture, implantation depth and symmetry as well as orientation of the native commissural and bioprosthetic leaflet [22]. Another factor that has been discussed to contribute to HALT after TAVI is overexpansion of the bioprosthetic stent frame, leading to endothelial injury and thereby increasing the risk of thrombus formation [23]. Accordingly, factors such as high angular deviation and severe commissural misalignment of the native and prosthetic valve may contribute to endothelial damage and delayed re-endothelialization, thereby increasing the risk of leaflet thrombosis. Regarding alterations in blood flow, data from cardiac magnetic resonance imaging after aortic valve sparing surgery for aortic regurgitation revealed that the formation of neo-sinuses between the native and bioprosthetic valve leaflets after TAVI contributes to increased blood stasis and thereby thrombus formation [24, 25]. Similarly, severe commissural misalignment may lead to altered flow patterns within the aortic root, resulting in increased blood stasis and thereby contributing to thrombus formation and HALT.
The current analysis offers a detailed examination of potential predictors for HALT after TAVI in cardiac CT. In the literature, several clinical and procedural contributors to HALT after TAVI have been described, such as the absence of oral anticoagulation, the use of a large prosthesis, moderate-to-severe paravalvular leakage or balloon-expandable valves [2629]. The current analysis extends the knowledge about potential predictors for HALT after TAVI. In addition to clinical predictors for HALT after TAVI, such as pmax, pmean and vmax during follow-up examination, our analysis revealed severe commissural misalignment as an important independent predictor for HALT after TAVI in cardiac CT.
Currently, the development of feasible and safe TAVI implantation techniques in order to reach commissural alignment is an important topic in the literature [30]. Whereas in subjects undergoing SAVR, visual inspection allows the alignment of the bioprosthetic valve commissures with those of the native aortic valves, positioning of the prosthesis during TAVI occurs randomly and may lead to commissural misalignment [14]. Considering the results of the current analysis, an optimized implantation technique in order to allow commissural alignment might potentially contribute to the prevention of HALT after TAVI and thereby positively influence valve durability and functionality. In this context, a patient- and valve-specific, fluoroscopy-based implantation technique to obtain commissural alignment by rotation of the prosthesis at the level of the aortic valve has recently been published [30].
The current analysis has some limitations. First, its retrospective, single-center, cross-sectional design with a small sample size may impair statistical comparisons between the two groups. Second, the definition of four categories of commissural (mis-)alignment represents an arbitrary approach. However, this technique allows a comparable classification of commissural (mis-)alignment throughout the study population and has been applied successfully in other studies [14, 30]. Third, the results of the current study might possibly suggest, but do not prove an association between commissural misalignment and altered flow patterns in the aortic root, leading to HALT after TAVI. However, the current analysis was not designed in order to explore the pathophysiological, blood-flow related mechanisms behind the development of HALT, but to examine the role of commissural alignment in the development of leaflet thrombosis after TAVI. In this context, further exploration and comparison of aortic flow patterns in subjects with and without commissural misalignment after TAVI might offer interesting new insights into the pathophysiological mechanisms behind the occurrence of leaflet thrombosis. Similarly, the current analysis does not provide any information on clinical consequences of HALT, e.g. thromboembolic complications or clinical advantages of obtaining commissural alignment in subjects undergoing TAVI.
In conclusion, subclinical leaflet thrombosis after TAVI is associated with commissural misalignment. Pathophysiological, blood-flow-related mechanisms behind the development of HALT as well as a potential clinical benefit of attempting commissural alignment remain to be assessed.

Acknowledgements

We gratefully acknowledge the expert technical assistance of Gerd Muschiol and Nadine Hoffart (Department of Cardiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg) in cardiac CT data acquisition.

Declarations

Conflict of interest

Mohamed Marwan received speaker honoraria from Siemens Healthineers, Edwards Lifescience and Amgen. There is a consulting agreement with Edwards Lifescience. All the other authors report that they have no conflicts of interest to declare.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Unsere Produktempfehlungen

Neuer Inhalt

Print-Titel

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

Literatur
1.
Zurück zum Zitat Vahanian A, Beyersdorf F, Praz F, Milojevic M, Baldus S, Bauersachs J et al (2022) 2021 ESC/EACTS Guidelines for the management of valvular heart disease. Eur Heart J 43(7):561–632CrossRefPubMed Vahanian A, Beyersdorf F, Praz F, Milojevic M, Baldus S, Bauersachs J et al (2022) 2021 ESC/EACTS Guidelines for the management of valvular heart disease. Eur Heart J 43(7):561–632CrossRefPubMed
2.
Zurück zum Zitat Bekeredjian R, Szabo G, Balaban U, Bleiziffer S, Bauer T, Ensminger S et al (2019) Patients at low surgical risk as defined by the Society of Thoracic Surgeons Score undergoing isolated interventional or surgical aortic valve implantation: in-hospital data and 1-year results from the German Aortic Valve Registry (GARY). Eur Heart J 40(17):1323–1330CrossRefPubMed Bekeredjian R, Szabo G, Balaban U, Bleiziffer S, Bauer T, Ensminger S et al (2019) Patients at low surgical risk as defined by the Society of Thoracic Surgeons Score undergoing isolated interventional or surgical aortic valve implantation: in-hospital data and 1-year results from the German Aortic Valve Registry (GARY). Eur Heart J 40(17):1323–1330CrossRefPubMed
3.
Zurück zum Zitat Mack MJ, Leon MB, Thourani VH, Makkar R, Kodali SK, Russo M et al (2019) Transcatheter aortic-valve replacement with a balloon-expandable valve in low-risk patients. N Engl J Med 380(18):1695–1705CrossRefPubMed Mack MJ, Leon MB, Thourani VH, Makkar R, Kodali SK, Russo M et al (2019) Transcatheter aortic-valve replacement with a balloon-expandable valve in low-risk patients. N Engl J Med 380(18):1695–1705CrossRefPubMed
4.
Zurück zum Zitat Popma JJ, Deeb GM, Yakubov SJ, Mumtaz M, Gada H, O’Hair D et al (2019) Transcatheter aortic-valve replacement with a self-expanding valve in low-risk patients. N Engl J Med 380(18):1706–1715CrossRefPubMed Popma JJ, Deeb GM, Yakubov SJ, Mumtaz M, Gada H, O’Hair D et al (2019) Transcatheter aortic-valve replacement with a self-expanding valve in low-risk patients. N Engl J Med 380(18):1706–1715CrossRefPubMed
5.
Zurück zum Zitat Rashid HN, Michail M, Ramnarain J, Nasis A, Nicholls SJ, Cameron JD et al (2022) The impact of hypo-attenuated leaflet thickening on haemodynamic valve deterioration following transcatheter aortic valve replacement. J Cardiovasc Comput Tomogr 16(2):168–173CrossRefPubMed Rashid HN, Michail M, Ramnarain J, Nasis A, Nicholls SJ, Cameron JD et al (2022) The impact of hypo-attenuated leaflet thickening on haemodynamic valve deterioration following transcatheter aortic valve replacement. J Cardiovasc Comput Tomogr 16(2):168–173CrossRefPubMed
6.
Zurück zum Zitat Cartlidge TRG, Doris MK, Sellers SL, Pawade TA, White AC, Pessotto R et al (2019) Detection and prediction of bioprosthetic aortic valve degeneration. J Am Coll Cardiol 73(10):1107–1119CrossRefPubMedPubMedCentral Cartlidge TRG, Doris MK, Sellers SL, Pawade TA, White AC, Pessotto R et al (2019) Detection and prediction of bioprosthetic aortic valve degeneration. J Am Coll Cardiol 73(10):1107–1119CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Holmes DR, Mack MJ (2015) Uncertainty and possible subclinical valve leaflet thrombosis. N Engl J Med 373(21):2080–2082CrossRefPubMed Holmes DR, Mack MJ (2015) Uncertainty and possible subclinical valve leaflet thrombosis. N Engl J Med 373(21):2080–2082CrossRefPubMed
8.
Zurück zum Zitat Blanke P, Leipsic JA, Popma JJ, Yakubov SJ, Deeb GM, Gada H et al (2020) Bioprosthetic aortic valve leaflet thickening in the evolut low risk sub-study. J Am Coll Cardiol 75(19):2430–2442CrossRefPubMed Blanke P, Leipsic JA, Popma JJ, Yakubov SJ, Deeb GM, Gada H et al (2020) Bioprosthetic aortic valve leaflet thickening in the evolut low risk sub-study. J Am Coll Cardiol 75(19):2430–2442CrossRefPubMed
9.
Zurück zum Zitat Chakravarty T, Sondergaard L, Friedman J, De Backer O, Berman D, Kofoed KF et al (2017) Subclinical leaflet thrombosis in surgical and transcatheter bioprosthetic aortic valves: an observational study. Lancet 389(10087):2383–2392CrossRefPubMed Chakravarty T, Sondergaard L, Friedman J, De Backer O, Berman D, Kofoed KF et al (2017) Subclinical leaflet thrombosis in surgical and transcatheter bioprosthetic aortic valves: an observational study. Lancet 389(10087):2383–2392CrossRefPubMed
11.
Zurück zum Zitat Makkar RR, Blanke P, Leipsic J, Thourani V, Chakravarty T, Brown D et al (2020) Subclinical leaflet thrombosis in transcatheter and surgical bioprosthetic valves: PARTNER 3 cardiac computed tomography substudy. J Am Coll Cardiol 75(24):3003–3015CrossRefPubMed Makkar RR, Blanke P, Leipsic J, Thourani V, Chakravarty T, Brown D et al (2020) Subclinical leaflet thrombosis in transcatheter and surgical bioprosthetic valves: PARTNER 3 cardiac computed tomography substudy. J Am Coll Cardiol 75(24):3003–3015CrossRefPubMed
12.
Zurück zum Zitat Sondergaard L, De Backer O, Kofoed KF, Jilaihawi H, Fuchs A, Chakravarty T et al (2017) Natural history of subclinical leaflet thrombosis affecting motion in bioprosthetic aortic valves. Eur Heart J 38(28):2201–2207CrossRefPubMed Sondergaard L, De Backer O, Kofoed KF, Jilaihawi H, Fuchs A, Chakravarty T et al (2017) Natural history of subclinical leaflet thrombosis affecting motion in bioprosthetic aortic valves. Eur Heart J 38(28):2201–2207CrossRefPubMed
13.
Zurück zum Zitat Makkar RR, Fontana G, Jilaihawi H, Chakravarty T, Kofoed KF, De Backer O et al (2015) Possible subclinical leaflet thrombosis in bioprosthetic aortic valves. N Engl J Med 373(21):2015–2024CrossRefPubMed Makkar RR, Fontana G, Jilaihawi H, Chakravarty T, Kofoed KF, De Backer O et al (2015) Possible subclinical leaflet thrombosis in bioprosthetic aortic valves. N Engl J Med 373(21):2015–2024CrossRefPubMed
14.
Zurück zum Zitat Fuchs A, Kofoed KF, Yoon SH, Schaffner Y, Bieliauskas G, Thyregod HG et al (2018) Commissural alignment of bioprosthetic aortic valve and native aortic valve following surgical and transcatheter aortic valve replacement and its impact on valvular function and coronary filling. JACC Cardiovasc Interv 11(17):1733–1743CrossRefPubMed Fuchs A, Kofoed KF, Yoon SH, Schaffner Y, Bieliauskas G, Thyregod HG et al (2018) Commissural alignment of bioprosthetic aortic valve and native aortic valve following surgical and transcatheter aortic valve replacement and its impact on valvular function and coronary filling. JACC Cardiovasc Interv 11(17):1733–1743CrossRefPubMed
15.
Zurück zum Zitat Buzzatti N, Romano V, De Backer O, Soendergaard L, Rosseel L, Maurovich-Horvat P et al (2020) Coronary access after repeated transcatheter aortic valve implantation: a glimpse into the future. JACC Cardiovasc Imaging 13(2 Pt 1):508–515CrossRefPubMed Buzzatti N, Romano V, De Backer O, Soendergaard L, Rosseel L, Maurovich-Horvat P et al (2020) Coronary access after repeated transcatheter aortic valve implantation: a glimpse into the future. JACC Cardiovasc Imaging 13(2 Pt 1):508–515CrossRefPubMed
16.
Zurück zum Zitat Tang GHL, Zaid S, Ahmad H, Undemir C, Lansman SL (2018) Transcatheter valve neo-commissural overlap with coronary orifices after transcatheter aortic valve replacement. Circ Cardiovasc Interv 11(10):e007263CrossRefPubMed Tang GHL, Zaid S, Ahmad H, Undemir C, Lansman SL (2018) Transcatheter valve neo-commissural overlap with coronary orifices after transcatheter aortic valve replacement. Circ Cardiovasc Interv 11(10):e007263CrossRefPubMed
17.
Zurück zum Zitat Wuest W, Anders K, Schuhbaeck A, May MS, Gauss S, Marwan M et al (2012) Dual source multidetector CT-angiography before Transcatheter Aortic Valve Implantation (TAVI) using a high-pitch spiral acquisition mode. Eur Radiol 22(1):51–58CrossRefPubMed Wuest W, Anders K, Schuhbaeck A, May MS, Gauss S, Marwan M et al (2012) Dual source multidetector CT-angiography before Transcatheter Aortic Valve Implantation (TAVI) using a high-pitch spiral acquisition mode. Eur Radiol 22(1):51–58CrossRefPubMed
18.
Zurück zum Zitat Blanke P, Weir-McCall JR, Achenbach S, Delgado V, Hausleiter J, Jilaihawi H et al (2019) Computed tomography imaging in the context of transcatheter aortic valve implantation (TAVI) / transcatheter aortic valve replacement (TAVR): An expert consensus document of the Society of Cardiovascular Computed Tomography. J Cardiovasc Comput Tomogr 13(1):1–20CrossRefPubMed Blanke P, Weir-McCall JR, Achenbach S, Delgado V, Hausleiter J, Jilaihawi H et al (2019) Computed tomography imaging in the context of transcatheter aortic valve implantation (TAVI) / transcatheter aortic valve replacement (TAVR): An expert consensus document of the Society of Cardiovascular Computed Tomography. J Cardiovasc Comput Tomogr 13(1):1–20CrossRefPubMed
19.
Zurück zum Zitat Hansson NC, Grove EL, Andersen HR, Leipsic J, Mathiassen ON, Jensen JM et al (2016) Transcatheter aortic valve thrombosis: incidence, predisposing factors, and clinical implications. J Am Coll Cardiol 68(19):2059–2069CrossRefPubMed Hansson NC, Grove EL, Andersen HR, Leipsic J, Mathiassen ON, Jensen JM et al (2016) Transcatheter aortic valve thrombosis: incidence, predisposing factors, and clinical implications. J Am Coll Cardiol 68(19):2059–2069CrossRefPubMed
20.
Zurück zum Zitat Vollema EM, Kong WKF, Katsanos S, Kamperidis V, van Rosendael PJ, van der Kley F et al (2017) Transcatheter aortic valve thrombosis: the relation between hypo-attenuated leaflet thickening, abnormal valve haemodynamics, and stroke. Eur Heart J 38(16):1207–1217CrossRefPubMed Vollema EM, Kong WKF, Katsanos S, Kamperidis V, van Rosendael PJ, van der Kley F et al (2017) Transcatheter aortic valve thrombosis: the relation between hypo-attenuated leaflet thickening, abnormal valve haemodynamics, and stroke. Eur Heart J 38(16):1207–1217CrossRefPubMed
21.
Zurück zum Zitat Khan JM, Rogers T, Weissman G, Torguson R, Rodriguez-Weisson FJ, Chezar-Azerrad C et al (2021) Anatomical characteristics associated with hypoattenuated leaflet thickening in low-risk patients undergoing transcatheter aortic valve replacement. Cardiovasc Revasc Med 27:1–6CrossRefPubMed Khan JM, Rogers T, Weissman G, Torguson R, Rodriguez-Weisson FJ, Chezar-Azerrad C et al (2021) Anatomical characteristics associated with hypoattenuated leaflet thickening in low-risk patients undergoing transcatheter aortic valve replacement. Cardiovasc Revasc Med 27:1–6CrossRefPubMed
22.
Zurück zum Zitat Jilaihawi H, Asch FM, Manasse E, Ruiz CE, Jelnin V, Kashif M et al (2017) Systematic CT methodology for the evaluation of subclinical leaflet thrombosis. JACC Cardiovasc Imaging 10(4):461–470CrossRefPubMed Jilaihawi H, Asch FM, Manasse E, Ruiz CE, Jelnin V, Kashif M et al (2017) Systematic CT methodology for the evaluation of subclinical leaflet thrombosis. JACC Cardiovasc Imaging 10(4):461–470CrossRefPubMed
23.
Zurück zum Zitat Midha PA, Raghav V, Sharma R, Condado JF, Okafor IU, Rami T et al (2017) The fluid mechanics of transcatheter heart valve leaflet thrombosis in the neosinus. Circulation 136(17):1598–1609CrossRefPubMed Midha PA, Raghav V, Sharma R, Condado JF, Okafor IU, Rami T et al (2017) The fluid mechanics of transcatheter heart valve leaflet thrombosis in the neosinus. Circulation 136(17):1598–1609CrossRefPubMed
24.
Zurück zum Zitat Jahren SE, Heinisch PP, Hasler D, Winkler BM, Stortecky S, Pilgrim T et al (2018) Can bioprosthetic valve thrombosis be promoted by aortic root morphology? An in vitro study. Interact Cardiovasc Thorac Surg 27(1):108–115CrossRefPubMed Jahren SE, Heinisch PP, Hasler D, Winkler BM, Stortecky S, Pilgrim T et al (2018) Can bioprosthetic valve thrombosis be promoted by aortic root morphology? An in vitro study. Interact Cardiovasc Thorac Surg 27(1):108–115CrossRefPubMed
25.
Zurück zum Zitat Vahidkhah K, Azadani AN (2017) Supra-annular valve-in-valve implantation reduces blood stasis on the transcatheter aortic valve leaflets. J Biomech 58:114–122CrossRefPubMed Vahidkhah K, Azadani AN (2017) Supra-annular valve-in-valve implantation reduces blood stasis on the transcatheter aortic valve leaflets. J Biomech 58:114–122CrossRefPubMed
26.
Zurück zum Zitat Rashid HN, Michail M, Ihdayhid AR, Dowling C, Khav N, Tan S et al (2021) Clinical predictors and sequelae of computed tomography defined leaflet thrombosis following transcatheter aortic valve replacement at medium-term follow-up. Heart Vessels 36(9):1374–1383CrossRefPubMed Rashid HN, Michail M, Ihdayhid AR, Dowling C, Khav N, Tan S et al (2021) Clinical predictors and sequelae of computed tomography defined leaflet thrombosis following transcatheter aortic valve replacement at medium-term follow-up. Heart Vessels 36(9):1374–1383CrossRefPubMed
27.
Zurück zum Zitat Yanagisawa R, Tanaka M, Yashima F, Arai T, Jinzaki M, Shimizu H et al (2019) Early and late leaflet thrombosis after transcatheter aortic valve replacement. Circ Cardiovasc Interv 12(2):e007349CrossRefPubMed Yanagisawa R, Tanaka M, Yashima F, Arai T, Jinzaki M, Shimizu H et al (2019) Early and late leaflet thrombosis after transcatheter aortic valve replacement. Circ Cardiovasc Interv 12(2):e007349CrossRefPubMed
28.
Zurück zum Zitat Jose J, Sulimov DS, El-Mawardy M, Sato T, Allali A, Holy EW et al (2017) Clinical bioprosthetic heart valve thrombosis after transcatheter aortic valve replacement: incidence, characteristics, and treatment outcomes. JACC Cardiovasc Interv 10(7):686–697CrossRefPubMed Jose J, Sulimov DS, El-Mawardy M, Sato T, Allali A, Holy EW et al (2017) Clinical bioprosthetic heart valve thrombosis after transcatheter aortic valve replacement: incidence, characteristics, and treatment outcomes. JACC Cardiovasc Interv 10(7):686–697CrossRefPubMed
29.
Zurück zum Zitat Reardon MJ, Feldman TE, Meduri CU, Makkar RR, O’Hair D, Linke A et al (2019) Two-year outcomes after transcatheter aortic valve replacement with mechanical vs self-expanding valves: the REPRISE III randomized clinical trial. JAMA Cardiol 4(3):223–229CrossRefPubMedPubMedCentral Reardon MJ, Feldman TE, Meduri CU, Makkar RR, O’Hair D, Linke A et al (2019) Two-year outcomes after transcatheter aortic valve replacement with mechanical vs self-expanding valves: the REPRISE III randomized clinical trial. JAMA Cardiol 4(3):223–229CrossRefPubMedPubMedCentral
30.
Zurück zum Zitat Bieliauskas G, Wong I, Bajoras V, Wang X, Kofoed KF, De Backer O et al (2021) Patient-specific implantation technique to obtain neo-commissural alignment with self-expanding transcatheter aortic valves. JACC Cardiovasc Interv 14(19):2097–2108CrossRefPubMed Bieliauskas G, Wong I, Bajoras V, Wang X, Kofoed KF, De Backer O et al (2021) Patient-specific implantation technique to obtain neo-commissural alignment with self-expanding transcatheter aortic valves. JACC Cardiovasc Interv 14(19):2097–2108CrossRefPubMed
Metadaten
Titel
Commissural misalignment independently predicts leaflet thrombosis after transcatheter aortic valve implantation
verfasst von
Susanne Jung
Fabian Ammon
Silvia Smolka
Maximilian Moshage
Mohamed Marwan
Stephan Achenbach
Publikationsdatum
06.04.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Research in Cardiology / Ausgabe 1/2024
Print ISSN: 1861-0684
Elektronische ISSN: 1861-0692
DOI
https://doi.org/10.1007/s00392-023-02192-6

Weitere Artikel der Ausgabe 1/2024

Clinical Research in Cardiology 1/2024 Zur Ausgabe

Letter to the Editors

Response

Nach Herzinfarkt mit Typ-1-Diabetes schlechtere Karten als mit Typ 2?

29.05.2024 Herzinfarkt Nachrichten

Bei Menschen mit Typ-2-Diabetes sind die Chancen, einen Myokardinfarkt zu überleben, in den letzten 15 Jahren deutlich gestiegen – nicht jedoch bei Betroffenen mit Typ 1.

Erhöhtes Risiko fürs Herz unter Checkpointhemmer-Therapie

28.05.2024 Nebenwirkungen der Krebstherapie Nachrichten

Kardiotoxische Nebenwirkungen einer Therapie mit Immuncheckpointhemmern mögen selten sein – wenn sie aber auftreten, wird es für Patienten oft lebensgefährlich. Voruntersuchung und Monitoring sind daher obligat.

GLP-1-Agonisten können Fortschreiten diabetischer Retinopathie begünstigen

24.05.2024 Diabetische Retinopathie Nachrichten

Möglicherweise hängt es von der Art der Diabetesmedikamente ab, wie hoch das Risiko der Betroffenen ist, dass sich sehkraftgefährdende Komplikationen verschlimmern.

TAVI versus Klappenchirurgie: Neue Vergleichsstudie sorgt für Erstaunen

21.05.2024 TAVI Nachrichten

Bei schwerer Aortenstenose und obstruktiver KHK empfehlen die Leitlinien derzeit eine chirurgische Kombi-Behandlung aus Klappenersatz plus Bypass-OP. Diese Empfehlung wird allerdings jetzt durch eine aktuelle Studie infrage gestellt – mit überraschender Deutlichkeit.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.