Skip to main content
Erschienen in: Oral Radiology 1/2023

Open Access 09.04.2022 | Review Article

Magnetic resonance imaging for preoperative diagnosis in third molar surgery: a systematic review

verfasst von: Adib Al-Haj Husain, Bernd Stadlinger, Sebastian Winklhofer, Marco Piccirelli, Silvio Valdec

Erschienen in: Oral Radiology | Ausgabe 1/2023

Abstract

In recent years, magnetic resonance imaging (MRI) has made great strides through various technical improvements and new sequences, which have made it one of the most promising and leading imaging techniques in the head and neck region. As modern imaging techniques in dentistry aim to reduce radiation exposure, this systematic review evaluated the possibilities, advantages, and disadvantages of advanced imaging diagnostics using dental MRI and its evidence for clinical indications and limitations relevant to mandibular third molar (MTM) surgery. Two reviewers performed multiple database searches (PubMed MEDLINE, EMBASE, Biosis, and Cochrane databases) following the PICOS search strategy using medical subject headings (MeSH) terms, keywords, and their combinations. Ten studies were included in this systematic review. By providing high spatial resolution and excellent soft tissue contrast, black bone MRI sequences such as 3D Double Echo Steady State (DESS) and 3D Short Tau Inversion Recovery (STIR) imaging protocols have the potential to become a valuable alternative to cone-beam computed tomography (CBCT) in future dental clinical routines. Overall, radiation-free MRI represents another step toward personalized dentistry and improved decision-making that avoids ineffectiveness and minimizes risks in oral surgery by taking into account additional patient-side factors such as comorbidity, anatomical norm variations, and imaging biomarkers.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

The extraction of partially erupted or impacted third molars is one of the most frequent oral surgical procedures in dental practice. Thanks to time-efficient, minimally invasive approaches aiming for hard tissue preservation and also novel techniques such as piezosurgery or laser therapy, surgical extraction of third molars has become a routine medical treatment that commonly proceeds without complications [1, 2]. However, due to the close positional relationship of the mandibular third molar (MTM) to the peripheral extracranial branches of the mandibular nerve, in particular the inferior alveolar nerve (IAN) and the lingual nerve (LN), advanced preoperative radiological assessment is of fundamental value for proper surgical case management, assessment of potential perioperative risks, and prevention of unpleasant postoperative complications.
Regarding the occurrence of transient iatrogenic IAN and LN damage associated with MTM surgery, there are heterogeneous datasets in the literature, estimating IAN injuries at approximately 4% (0.4–8.4%) [35] and LN injuries at 0.01–2% [6]. In this context, specific radiographic risk factors such as a close positional relationship between the MTM and the mandibular canal (MC), diversion of the MC, darkening of the MTM roots, and incomplete integrity of the osseous boundaries of the MC may indicate an increased risk for IAN injury [79]. Fortunately, in most cases, spontaneous healing occurs within the first 3–6 months after surgery [10]. However, if no signs of improvement are observed within this period, the risk of permanent neurosensory dysfunction increases, potentially resulting in a relevant compromise of the quality of life of the affected patients [11].
A conventional two-dimensional panoramic radiograph (PAN) is obtained in the daily clinical routine prior to MTM surgery. This overview image captures the entire oral cavity in a single image with relatively short radiation time and exposure (4–30 μSv) and provides anatomical information about the angulation and position of the MTM and the relative position of the MTM roots to the MC. In cases showing the aforementioned radiographic signs of risk, an overlap of anatomic structures, or displacement of the MC, two-dimensional imaging is not sufficient, and three-dimensional radiographic imaging modalities such as computed tomography (CT) or cone-beam computed tomography (CBCT) is indicated [12]. Despite the standardized grayscale value and insufficient soft tissue resolution, CBCT is considered the imaging modality of choice in dentomaxillofacial radiology due to greater accessibility, short examination time, lower radiation dose, and lower cost compared to conventional CT [13]. CBCT generates three-dimensional cross-sectional images of the examined craniofacial and dental structures with high spatial resolution at a radiation dose of approximately 18–200 µS and provides accurate information about relevant anatomical structures, such as the number and shape of MTM roots, the buccolingual location of the MC, and its cortical integrity [14]. However, generating this additional information, which may be elementary for the surgeon performing a high-risk MTM procedure, is associated with an increased lifetime risk for radiation-induced cancers given the increase in CBCT usage in various dental specialties [15, 16]. Considering the efforts to minimize or even eliminate radiation exposure in medical imaging according to the As Low As Reasonably Achievable (ALARA) principle and the proposed paradigm shift towards the As Low As Diagnostically Acceptable (ALADA) principle [17], several promising modalities such as low-dose CBCT or magnetic resonance imaging (MRI) have recently been integrated into dental imaging workflows, resulting in improved virtual surgical planning in oral and maxillofacial surgery with predictable intraoperative outcomes. Reports in the literature have shown that low-dose CBCT imaging protocols can provide confidential diagnostic information with an improved benefit–risk ratio compared to standard-dose CBCT in various clinical situations, e.g., implantology and detection of bony lesions in the jaw [1820]. However, given the large number of manufacturer-specific scan parameters, modification and especially standardization of low-dose imaging protocols are still subject to research.
MRI is a non-invasive imaging technique that is one of the most promising and leading imaging modalities for the diagnosis of diseases and other conditions in the head and neck region [21]. A major advantage of MRI over conventional X-ray imaging is the high soft tissue contrast, which allows much better visualization of specific anatomical structures (e.g., nerves, blood vessels) using magnetic fields without exposing patients to ionizing radiation [22]. Despite the limitations in hard tissue imaging, MRI has advanced rapidly over the past two decades with various technical innovations and advanced imaging protocols, offering a wide range of new diagnostic capabilities in dentistry [23]. The challenges in visualizing dentition with MRI are due to the dense organic constitution, only allowing a limited molecular movement of hydrogen nuclei. This is further affected by the rapid signal decay after radiofrequency excitation so that the transverse relaxation time (T2*) of more than 1 ms, which is necessary to digitize MRI signals for visualization of mineralized dental tissue, cannot be obtained [24, 25]. Recently introduced “black bone” MRI sequences such as 3D double-echo steady-state (DESS) and 3D short tau inversion recovery (STIR) have contributed significantly to an optimized simultaneous visualization of soft and hard tissue, allowing the visualization of the IAN within the MC near the MTM roots without significant diagnostic limitations compared to conventional CBCT [26, 27]. In addition, continuous visualization of the IAN and LN and their finest ramifications are possible [28, 29], improving the diagnostic capabilities in the third molar region and positively impacting preoperative planning in MTM surgery [30, 31]. The residual limitations of MRI in the oral cavity are susceptibility to motion artifacts, complex anatomic courses of small-sized blood vessels and nerves, and image distortion and artifacts due to magnetic field inhomogeneities caused by metallic dental restorations [32].
However, the growing number of MRI studies in dentistry underlines the relevance and the perspectives opened by the targeted use of specific instruments such as mandibular dental coils [33], which represent a tremendous resource for imaging the mandibular third molar region in the oral cavity. Nevertheless, evidence-based guidelines for case-specific indications and limitations are still lacking. Therefore, the aim of this systematic review was to evaluate the indications and limitations of different MRI protocols in terms of diagnostic quality and potential benefits in MTM surgery and to provide guidelines based on daily clinical questions by examining the following PICO question: does preoperative MRI provide accurate and feasible preoperative diagnostic information in healthy volunteers and patients undergoing MTM surgery?

Materials and methods

Search strategy, information sources, and eligibility criteria

This systematic review was performed in accordance with the preferred reporting items for systematic reviews and meta-analysis (PRISMA) guidelines. A search for articles on the use of dental MRI in MTM surgery was performed using the Pubmed MEDLINE, EMBASE, BIOSIS, and Cochrane electronic databases. All English and German articles from 1989 until December 2021 were considered. Data searches were performed using the following MeSH terms, keywords, and their combinations with Boolean operators: third molar extraction, magnetic resonance imaging, nuclear magnetic resonance imaging, diffusion tensor imaging, ultrashort echo-time, neurography, visualization, inferior alveolar nerve, lingual nerve, and mandibular nerve. Thereby the following PICO question was assessed: P-population: human studies with participants undergoing MRI prior to MTM surgery, or healthy patients undergoing MRI to assess the anatomical and spatial relationship of the MTMs in the third molar region; I-intervention: magnetic resonance imaging; C-control: not applicable; O-outcome: feasibility and accuracy of preoperative radiological assessment of the mandibular third molar region; and S-study designs: clinical trials (Table 1). This systematic review was not registered in the International Prospective Register of Systematic Reviews (PROSPERO) platform (no protocol number available).
Table 1
Search strategy according to the focused question (PICO)
Focused question
(PICO)
Does preoperative MRI provide accurate and feasible diagnostic information in healthy volunteers and patients undergoing MTM surgery?
Search strategy
Population
Human studies (patients and/or healthy subjects), aged older than 12 years undergoing MRI prior to MTM surgery
#1—((third molar extraction [MeSH]) OR (inferior alveolar nerve [MeSH]) OR (lingual nerve [MeSH]) OR (mandibular nerve [MeSH]) OR (trigeminal nerve [MeSH]))
 
Intervention
Magnetic resonance imaging
#2—((magnetic resonance imaging [MeSH]) OR (MRI) OR (nuclear magnetic resonance imaging [MeSH]) OR (NMR) OR (diffusion tensor imaging [MeSH]) OR (DTI) OR (ultra-short echo-time [MeSH]) OR (UTE) OR
(maxillofacial imaging))
#3—((visualization) OR (neurography))
 
Comparison
Conventional preoperative radiological assessment
#4—((computed tomography [MeSH]) OR (cone-beam computed tomography [MeSH])
#5—(panoramic radiography [MeSH])
 
Outcome
Feasibility and accuracy of preoperative radiological assessment of MTM region using MRI
#6—((accuracy) OR (feasibility) OR (signal-to-noise-ratio [MeSH]))
 
Search combination(s)
(#1) AND (#2 or #3) OR (#6)
Studies were included based on the following criteria: (1), human studies with participants undergoing MRI prior to MTM surgery, or healthy patients undergoing MRI to assess the anatomical and spatial relationship of the MTMs in the third molar region in randomized or nonrandomized controlled trials and cohort studies; (2), study participants older than 12 years; (3), availability of the full text; and (4), articles in English or German language. Exclusion criteria were: (1), animal, cadaveric, in vitro studies, systematic reviews, narrative reviews, and case reports; (2) patients with other additional pathologies requiring MRI examination of the region of interest.

Data extraction and collection

Two reviewers (A.A.H. and S.V.) conducted the literature searches independently to minimize potential reviewer bias. From all databases, the data were transferred to EndNote (Clarivate, Sydney, Australia) and all duplicates were removed. First, titles and abstracts were screened, and then 55 studies were selected for full-text analysis. If ambiguities occurred, they were clarified through discussion. Finally, ten articles within the scope of this review examining the use of magnetic resonance imaging in MTM surgery were evaluated (Fig. 1).
For each study, the following parameters were recorded by two reviewers (A.A.H and S.V. separately): general information (study design, author, year, country, number of patients treated, age range of patients treated), MRI imaging protocol parameters (MR device, MRI sequence, field strength, acquisition time, type of MRI coil), and outcomes (feasibility and accuracy).

Quality assessment in individual studies

The methodology of the included studies for risk of bias was evaluated based on a modified short version of the Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) according to Edwards et al. [34]. The risk of bias was assessed using 18 criteria from the STROBE statement, whereby a total score of 15 or more out of 18 was considered low risk of bias, a score of 11–14 as medium risk of bias, and studies with a score of 10 or less as high risk of bias.

Results

Study selection

In accordance with the purpose of this systematic review, the literature search identified 723 potentially relevant studies. After duplicates were removed, 515 articles remained. In the first step of study selection, the titles and abstracts were checked, and then 460 articles were excluded (non-human studies, case studies, case reports and reviews, lack of access to the full text of articles). The remaining 55 articles were subjected to full-text screening in a second step. In this process, 45 articles were excluded because they did not meet the inclusion criteria. Finally, ten articles were within the scope of this review.

Study characteristics

In the ten studies included in this systematic review (Miloro et al. [42]; Ferretti et al. [55]; Chau et al. [43]; Kirnbauer et al. [45]; Kang et al. [56]; Burian et al. [30]; Beck et al. [44], Al-Haj Husain et al. [31]; Al-Haj Husain et al. [28]; Valdec et al. [26]), 202 patients underwent an MRI scan preoperatively between 1997 and 2021 at MR field strengths of 1, 1.5, or 3 Tesla using various MRI sequences. The included patients’ age ranged from 12 to 65 years with an acquisition time of 3:30 min to about 20 min (Table 2).
Table 2
Characteristics of included studies
Study number
Author, year, country
Study
Sample size
Age
Study objectives
MRI
sequences
Field strengths
Type of MRI coil
Acquisition time
MR device
Outcome
(feasibility/accuracy)
1
Miloro et al. 1997, USA
Assessment of the lingual nerve in the third molar region using magnetic resonance imaging
10
21–35 years
Determination of the precise in situ location of the lingual nerve in the third molar region by high-resolution MRI
Phase encode time reduction acquisition (PETRA) sequence
1.5 T
surface coil
9:17
min
Signa MRI Unit, General Electric Medical Systems, Milwaukee, Wisconsin, USA
Precise lingual nerve visualization in the third molar region using the PETRA sequence. Assessment of the nerve diameter, the different shapes of the nerve, and the mean vertical and horizontal distances to the lingual crest and the lingual plate of the mandible
2
Ferretti et al. 2009, Italy
Dental magnetic resonance imaging: study of impacted mandibular third molars
29
15–28 years
Evaluation of the relationship between the mandibular canal and impacted mandibular third molars using MRI
T1 weighted (short repetition time (TR)/short echo time (TE)) with spin echo (SE) technique or proton-density (PD)-weighted (long TR/short TE) with turbo spin echo (TSE) technique
1.0 T
radiofrequency head coil
N/A
Magnetom Expert; Siemens Medical Systems, Erlangen, Germany
MRI provided a good depiction of anatomical details of the mandibular canal and its relationship to the MTMs
3
Chau et al. 2012, Hong Kong
Comparison between the use of magnetic resonance imaging and cone-beam computed tomography for mandibular nerve identification
11
N/A
Evaluation of the reliability of the radiographic assessment of the mandibular nerve in MRI or the mandibular canal in CBCT prior to mandibular third molar extraction
T1 weighted Volumetric Interpolated Breath-hold Examination
(VIBE) sequence with fat suppression
3.0 T
12-channel birdcage head and neck coil
4 min
Magnetom Trio Tim syngo, Siemens Medical Solutions, Erlangen, Germany
MRI provided better results in identifying the inferior alveolar nerve/canal and can be used to identify the mandibular nerve in cases that are not accurately visualized on CBCT
4
Kirnbauer et al. 2018,
Austria
Assessment of impacted and partially impacted lower third molars with panoramic radiography compared to magnetic resonance imaging—a proof of principle study
28
13–24 years
Comparison of panoramic radiography and MRI in preoperative radiological assessment considering the positional relationship of impacted and partially impacted mandibular thirds
3D turbo spin echo (TSE) sequence and
3D constructive interference in steady-state (CISS) sequence
3.0 T
8-channel receive-only CPC coil
20:31 min
Magnetom Trio, a TIM System,
Siemens AG, Erlangen, Germany
MRI provided the same information compared with panoramic radiography, with the added advantage of providing three-dimensional information generated without radiation
5
Kang et al. 2018,
China
Investigation of Zero TE MR in preoperative planning in dentistry
22
26–65 years
Comparison of preoperative assessment of the positional relationship between the mandibular third molar and the mandibular canal in mandibular third molar surgery using CBCT and zero echo-time MRI
Ultrashort and zero echo-time (ZTE) MRI
3.0 T
8-channel head neck coil
3:30 min
MR Discovery 750 W, GE Healthcare, Milwaukee,
Wisconsin, USA
ZTE MRI demonstrated the delineation of the teeth and mandible and provided superior visualization of the mandibular canal compared to CBCT
Identical results were obtained in the assessment of the spatial relationship between the mandibular third molar and the mandibular canal
6
Burian et al. 2020,
Germany
MRI of the inferior alveolar nerve and lingual nerve—anatomical variation and morphometric benchmark values of nerve diameters in healthy subjects
30
21–32
years
Imaging of the inferior alveolar nerve and lingual nerve in the mandibular third molar region using black bone MRI sequences
3D Short tau inversion recovery (STIR), 3D Double-echo steady-state (DESS), and 3D T1 fast field echo (FFE)
3.0 T
16-channel Head and Neck Spine Coil
s
5:31 min
Ingenia Elition, Philips Healthcare, Best, the Netherlands
Possibility of accurate visualization of the inferior alveolar nerve and the lingual nerve in the mandibular third molar region. STIR provided the best signal-to-noise ratio. Accurate discrimination of the tissue composition of the mandibular neurovascular bundle was achieved
7
Beck et al. 2021,
Austria
Is MRI a viable alternative to CT/CBCT to identify the course of the inferior alveolar nerve in relation to the roots of the third molars?
53
N/A
Assessment of the spatial relationship between the inferior alveolar and the mandibular third molar using MRI or CT/CBCT images
PD T2 TSE fat saturation (FS) axial, PD TSE FS coronal
3.0 T
64 channel head-and-neck coil
10:44 min
Magnetom Skyra, Siemens Healthcare, Erlangen, Germany
Good inter- and intra-rater agreement in assessing the spatial relationship between IAN and MTM. Additionally, MRI offered advantages in identifying of accessory IAN in comparison to CT/CBCT
8
Al-Haj Husain et al. 2021, Switzerland
Mandibular third molar surgery: intraosseous localization of the inferior alveolar nerve using 3D double-echo steady-state MRI (3D-DESS)
19
18–63 years
Evaluation of the intraosseous position of the inferior alveolar nerve within the mandibular canal’s osseous boundaries prior to third molar extraction treatment using preoperative cone-beam computed tomography and magnetic resonance imaging. Assessment of a conversion factor between both imaging modalities
3D-DESS
with water excitation sequence
3.0 T
64 channel head-and-neck coil
12:24 min
Skyra (release VE11c), Siemens Healthi- neers, Erlangen, Germany
Accurate simultaneous visualization of the nerve tissues within osseous boundaries. A conversion factor from IAC in CBCT and MRI to IAN in MRI was determined
9
Al-Haj Husain et al. 2021, Switzerland
Preoperative visualization of the lingual nerve by 3D double-echo steady-state MRI in surgical third molar extraction treatment
19
18–63 years
Evaluation of the focal and continuous anatomy of the lingual nerve in the surgically relevant region before mandibular third molar surgery
3D-DESS
with water excitation sequence
3.0 T
64 channel head-and-neck coil
12:24 min
Skyra (release VE11c), Siemens Healthi- neers, Erlangen, Germany
Preoperative clarification of the focal and continuous anatomy of the lingual nerve in the third molar region with high reliability
10
Valdec et al. 2021, Switzerland
Comparison of Preoperative Cone-Beam Computed Tomography and 3D-Double Echo Steady-State MRI in Third Molar Surgery
19
18–63 years
Evaluation of the reliability of assessing the positional relationship between the inferior alveolar nerve and the mandibular third molar using CBCT, MRI, and CBCT/MRI image fusion
3D-DESS
with water excitation sequence
3.0 T
64 channel head-and-neck coil
12:24 min
Skyra (release VE11c), Siemens Healthi- neers, Erlangen, Germany
Moderate intra- and inter-rater agreement for the positional relationship was observed, independent of the reader’s experience. MRI provided superior diagnostic benefit regarding early detection of inflammation

Quality assessment in individual studies

The risk of bias assessment in each study was low in 7 and medium in 3 studies. An overview of the percentage responses for each topic is displayed in Table 3.
Table 3
Quality assessment of included studies
Checklist
Miloro et al. [42]
Ferretti et al. [55]
Chau et al. [43]
Kirnbauer et al. [45]
Kang et al. [56]
Burian et al. [30]
Beck et al. [44]
Al-Haj Husain et al. [31]
Al-Haj Husain et al. [28]
Valdec et al. [26]
Objectives: clearly formatted
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
Study design: described in detail
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
Settings: described in terms of location; and relevant dates
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
Participants:eligibility criteria; and methods of selection described
 + ; + 
 + ; + 
−; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
Bias: any efforts to address potential sources of bias described
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
Sample size: explanation of derivation; adequate
−; + 
−; + 
−; + 
−; + 
−; + 
−; + 
 + ; + 
−; + 
−; + 
 + ; + 
Statistical Methods: described; appropriate for data
−; + 
−; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
Participants: described
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
Outcome data: number of outcome events reported
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
Other analysis: any other analyses conducted reported
 + 
 + 
 + 
 + 
 + 
 + 
Limitations: limitations of the study; and any potential bias discussed
−; + 
−; -
−; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
 + ; + 
Interpretation: overall interpretation of results provided
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
 + 
External validity: generalizability of the results discussed
 + 
 + 
 + 
Total out of 18 (percentage)
14 (78%)
12 (67%)
14 (78%)
15 (83%)
15 (83%)
16 (89%)
17 (94%)
16 (89%)
17 (94%)
17 (94%)

Discussion

Magnetic resonance imaging—overview

Dental MRI is a modern, non-invasive cross-sectional imaging technique that provides three-dimensional, detailed anatomical images of the oral cavity and its contents. High-resolution MR images are generated by combining magnetic fields of different strengths, magnetic field gradients, high-frequency magnetic pulses exciting hydrogen protons and magnetic properties of living tissue. Since living tissue in humans is composed of about 70% water, this property is exploited to produce accurate MR images of soft tissues such as internal organs, tendons, ligaments, or muscles. Consequently, in contrast to conventional X-ray, the various image representations in MRI do not reflect the general tissue density but the proton density of the individual tissues [35, 36].
Continuous improvement and optimization of resolution and image quality are based on modifying all three fundamental factors for biomedical imaging—the patient, the imaging device, and the image detector. In this context, the signal-to-noise ratio, which evaluates the ratio between the desired signal intensity and the background noise, and the spatial resolution, which depends on the image voxel size, are the most fundamental parameters for MR image quality [37]. Signal-to-noise ratio can be optimized by increasing the voxel size, the field of view, time to repeat, the slice thickness, and the number of signal acquisitions or decreasing the matrix size and bandwidth by applying specific coils [38]. However, it should be considered that image definition decreases as the signal-to-noise ratio increases. A mandibular dental coil is used to reduce the field of view and increase image resolution without decreasing the signal-to-noise ratio [35]. Depending on the medical issue, especially in oncological imaging or the detection of inflammation, a contrast agent containing gadolinium may be administered intravenously to increase the informative value of MRI. Various changes and advancements in MRI and the introduction of new MRI sequences have steadily improved diagnostic capabilities in recent years, allowing the determination of quantitative parameters such as tissue perfusion, oxygen concentration, and diffusion.
However, it should be noted that there are potential contraindications to MR examinations which include electronic implants, intracorporeal pacemakers, brain and spinal cord stimulators, insulin pumps, cochlear implants, or metallic foreign bodies in soft tissue, severe anaphylactic reactions to MR contrast agents, claustrophobia, and pregnancy. Nowadays, the concept of absolute or relative contraindications has been largely abandoned. Instead, the individual patient situation should be taken into account, including the scan indication, the exact implantation clarification and an individual risk assessment of the examination. Many points that were previously considered as absolute contraindications are now potentially suitable or conditional for MR after clarification and after adjustment of the scan parameters.

Magnetic resonance imaging in MTM surgery

The analysis focused on the impact of imaging protocols and device-specific parameters on the accuracy and feasibility of visualization of the third molar region in MTM surgery performed according to modern surgical concepts. In this context, factors influencing the MR images, such as the MR device, the effect of field strength, type of MRI coil, acquisition time, and MRI sequence on MR images were evaluated. Considering the continuously increasing number of MRI studies in dentistry and the related wide range of applications in preoperative diagnostics, it is necessary to evaluate the evidence of added benefit as well as the indications and limitations of the individual MRI protocols and to make suggestions for optimized individual clinical decision-making in MTM surgery.
The conventional X-ray-based imaging techniques commonly used in MTM surgery, such as PAN or three-dimensional CBCT, can visualize various pathologies and the surgically relevant structures, such as the MTM, the radiolucent MC with the surrounding osseous boundaries, and maxillary sinuses. However, the increasing use of CBCT in MTM surgery is suggested to lead to an additional increase in radiation-induced cancer incidence by 0.46 [39], which is particularly relevant to repeated radiation exposure in radiosensitive, genetically susceptible young adolescents [15]. Other studies suggest that diagnostic radiation exposure from dental radiography may be associated with an increased risk of thyroid cancer and meningioma [40]. Therefore, radiation-free MRI presents a valid alternative that can be used for medical diagnosis, monitoring of treatment, and follow-up care.
Modern surgical concepts in MTM surgery aim for low-risk, minimally invasive, hard tissue preserving approaches, whereby a multimodal approach considering the medical history, intraoperative findings and preoperative radiographic evaluation of the third molar region is crucial to minimize perioperative risks [41]. The depth, angulation and orientation of impaction, number of roots and their morphology, proximity to the IAN and LN and the presence of other pathological processes should always be considered [41]. The results of this systematic review confirm the feasibility and accuracy of preoperative visualization of the MTM region by MRI, with a special focus on the most vulnerable structures, e.g., the IAN and LN. The first MRI studies addressing this issue in MTM surgery were conducted in the late 1990s, whereby phase encode time reduction acquisition (PETRA) sequence, an imaging protocol developed specifically for this purpose, was used to assess the LN visualization and various surgically relevant quantitative and qualitative parameters [42]. The older MRI studies were performed with a field strength of 1 Tesla and the use of conventional, nonspecific MR sequences, resulting in a low signal-to-noise ratio what could be considered insufficient and a limitation from today's perspective. However, it was possible to visualize the neurovascular bundle in the mandibular third molar region as a moderately hyperintense signal, resulting in good contrast due to the lower signal from the bony boundaries of the MC, without distinguishing the nerve tissue from blood vessels.
Current developments are moving toward novel MRI protocols suitable for dental imaging. Special "black bone" MRI sequences such as 3D double-echo steady-state (DESS) and 3D short tau inversion recovery (STIR) sequences provided high-resolution and high-contrast images that allow simultaneous visualization of the inferior alveolar nerve tissue within the osseous boundaries of the mandibular canal, providing a reliable assessment of the positional relationship of MTMs [26, 28, 30, 31]. In these dedicated MRI protocols using the water excitation/fat suppression technique, the IAN and LN appear as a highly hyperintense signal and can be distinguished from the MC due to the myelin layer surrounding the nerves [29].
The 3D DESS sequence allowed focal and continuous visualization of the LN from the foramen ovale to the MTM region with high reproducibility [28] (Fig. 2). This preoperative diagnosis of the exact location of the LN could reduce complications in complex angulated, deeply impacted MTMs and other dentoalveolar surgical procedures performed in anatomic proximity to the LN. In addition, this MRI protocol is considered superior to other black bone MRI protocols in terms of radiographic assessment of quantitative parameters of the LN [28, 30].
However, Burian et al. reconfirmed that, as of today, black bone MRI sequences are best suited for this medical question. The STIR imaging protocol was superior and provided the most promising signal-to-noise ratio and nerve–muscle contrast to noise ratio for the IAN and the LN [30]. Al-Haj Husain et al. demonstrated that the preoperative intraosseous localization of the IAN within the bony borders of the MC is possible and accurate. Thereby, the IAN showed the highest localization within the central MC segments. Additionally, the retention type of the MTM was revealed to influence the intraosseous localization of the IAN, with the nerve being displaced by the MTM in the segments proximal to the contact site [31] (Fig. 3). This phenomenon is described in the literature as the "snake phenomenon". This additional generated information could be of interest to the performing surgeon in high-risk MTM surgeries, leading to improved strategies for perioperative management and thus better outcomes.
In addition to depicting the anatomic course of the IAN and LN, the results obtained in this review demonstrated the utility of MRI in assessing the positional relationship between the IAN and MTM. Regarding the reliability of the assessment of the positional relationship, numerous reports comparing PAN or CBCT with MRI or fused CBCT-MRI images were evaluated. In general, MR imaging using T1 weighted volumetric interpolated breath-hold examination (VIBE) sequence with fat saturation [43], standard investigation MR protocols of the jaw region [44], or blackbone MRI sequences modified specifically for this purpose [26] offered similar results compared with CBCT and was even superior in cases where the MC could not be accurately visualized on CBCT. Compared to PAN, MRI using 3D turbo spin echo (TSE) and 3D constructive interference in steady-state (CISS) sequence provided the same information, with the added advantage of providing three-dimensional information of the region of interest without radiation exposure [45]. In general, it can be concluded that different MRI protocols of the included studies, with their respective advantages and disadvantages, are suitable for determining the relative position of the MTM to the IAN and provide superior visualization of the mandibular canal compared with CBCT. Using CBCT, it is only possible to indirectly visualize the location of the IAN via its cortical bone boundaries, which can be problematic as the MC is difficult or impossible to delineate in CBCT in approximately 20–40% of cases due to low or absent corticalization [46]. In CBCT, an apparent underestimation of the shape and volume of the neurovascular bundle of 1.5 to 5 mm thickness can be observed, thus the use of MRI can contribute significantly to a safer surgical approach [47]. In addition to the now qualitatively and quantitatively good and radiation-free visualization capabilities of bone tissue, the internal structure and microarchitecture of bone could also be assessed preoperatively, providing a new option for preoperative evaluation of the bone quality and allowing the selection of the most appropriate treatment option [48]. MRI scans also enable added diagnostic value compared to CBCT and CT due to their superior soft tissue contrast [26]. Thereby, oral mucosa or gingiva, neurovascular structures such as the IAN or LN in particular, or the dental pulp can be depicted directly [49, 50]. Compared with CBCT or MR imaging alone, the use of fused CBCT/MRI images could offer advantages in preoperative radiographic position assessment in borderline cases that are difficult to evaluate (Fig. 4). This might be especially relevant for borderline cases in which CBCT did not show the continuous osseous boundaries of the MC, while MRI nevertheless revealed a non-contact situation between the roots of the MTM and the IAN [26].
Various modifications and advancements of MRI, such as the introduction of functional MRI (fMRI) and the combination with positron emission tomography (PET), and the implementation of new MRI sequences, have steadily improved the diagnostic capabilities. The drive towards using 3 Tesla MRI is fueled by the advantages of improved signal-to-noise ratio, contrast-to-noise ratio, and high image resolution for dental applications. In many cases, these advantages will allow higher spatial resolution than lower field strength MRI. Alternatively, 1.5 Tesla MRI devices can be used with mandibular dental coils [51] or intraoral coils [52], optimizing image quality and shortening acquisition times. The previously very long scan times of up to 30 min and the low image quality with insufficient resolution, which were considered unsuitable for clinical routine, could, therefore, be overcome and currently be reduced to the range of about 3 min [53]. Furthermore, initial efforts are underway to offer out-of-hospital MRI examinations using recently developed bedside MRI scanners and additional specific equipment, which is an attractive option to make dental MRI scans more time and cost efficient. Nevertheless, further enhancements regarding standardization are still needed to enable the targeted use of MRI scans in routine oral surgical procedures. Considering the artifacts in MRI, motion artifacts, field inhomogeneity, and artifact-causing metallic dental restorations are the major challenges in MR imaging of the oral cavity. It must be taken into account that radiation-based techniques such as CBCT or CT can also cause pronounced artifacts. However, specific applications for artifact suppression are already established and should be further optimized [49, 54].
According to the results of this systematic review, various MRI protocols were suitable for preoperative imaging of the MTM region. However, the articles cited in this review show that MRI-based radiological assessment achieves comparable results to CBCT-based surgical planning without ionizing radiation. The additional information obtained may positively impact various preoperative planning in oral surgery and allow better prediction of surgical difficulties before surgery, leading to a safer surgical approach and a reduction in postoperative discomfort due to nerve damage. Although there is much heterogeneity in the literature related to scanning parameters, MRI currently represents the only promising imaging modality that provides non-invasive direct imaging of the neurovascular bundle, capable of discriminating the neural tissue from the blood vessels. More studies, including randomized control trials, should be performed investigating the benefit of each MRI protocol in MTM surgery to provide an evidence-based understanding of their use and more information about their impact on clinical outcomes. With improved cost-effectiveness and considering the improved risk–benefit ratio, the use of black bone MRI sequences such as DESS and STIR can present a valid alternative that can be used for medical diagnosis, monitoring of treatment, and follow-up care.

Conclusion

From a clinical perspective, the targeted use of dental MRI in indicated MTM sugical cases could support or even replace radiation-based CBCT imaging in preoperative radiological assessment. As shown by the results of these studies, the use of "black-bone" MRI sequences, which have recently been increasingly investigated and modified, provides highly confidential and reliable additional preoperative diagnostic information, improved soft tissue resolution, and high sensitivity in detecting pathologic changes. Consequently, dental MRI could provide the performing surgeon with beneficial information preoperatively, enabling improved personalized therapeutic approaches in high-risk MTM surgery.

Declarations

Conflict of interest

The authors declare that they have no conflict of interests related to this study.

Ethical approval

All procedures followed were in accordance with the ethical standards of the responsible committee on human experimentation (institutional and national) and with the Helsinki Declaration of 1975, as revised in 2008. Informed consent was obtained from all patients for being included in the study .
Informed consent was obtained from each study participant.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Dent – Das Online-Abo der Zahnmedizin

Online-Abonnement

Mit e.Dent erhalten Sie Zugang zu allen zahnmedizinischen Fortbildungen und unseren zahnmedizinischen und ausgesuchten medizinischen Zeitschriften.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Bailey E, Kashbour W, Shah N, Worthington HV, Renton TF, Coulthard P. Surgical techniques for the removal of mandibular wisdom teeth. Cochrane Database Syst Rev. 2020;7:CD004345. Bailey E, Kashbour W, Shah N, Worthington HV, Renton TF, Coulthard P. Surgical techniques for the removal of mandibular wisdom teeth. Cochrane Database Syst Rev. 2020;7:CD004345.
2.
Zurück zum Zitat Brasileiro BF, de Bragança RM, Van Sickels JE. An evaluation of patients’ knowledge about perioperative information for third molar removal. J Oral Maxillofac Surg. 2012;70(1):12–8.CrossRef Brasileiro BF, de Bragança RM, Van Sickels JE. An evaluation of patients’ knowledge about perioperative information for third molar removal. J Oral Maxillofac Surg. 2012;70(1):12–8.CrossRef
3.
Zurück zum Zitat Jerjes W, Swinson B, Moles DR, El-Maaytah M, Banu B, Upile T, et al. Permanent sensory nerve impairment following third molar surgery: a prospective study. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2006;102(4):e1-7.CrossRef Jerjes W, Swinson B, Moles DR, El-Maaytah M, Banu B, Upile T, et al. Permanent sensory nerve impairment following third molar surgery: a prospective study. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2006;102(4):e1-7.CrossRef
4.
Zurück zum Zitat Sigron GR, Pourmand PP, Mache B, Stadlinger B, Locher MC. The most common complications after wisdom-tooth removal: part 1: a retrospective study of 1,199 cases in the mandible. Swiss Dent J. 2014;124(10):1042–6. Sigron GR, Pourmand PP, Mache B, Stadlinger B, Locher MC. The most common complications after wisdom-tooth removal: part 1: a retrospective study of 1,199 cases in the mandible. Swiss Dent J. 2014;124(10):1042–6.
5.
Zurück zum Zitat Qi W, Lei J, Liu YN, Li JN, Pan J, Yu GY. Evaluating the risk of post-extraction inferior alveolar nerve injury through the relative position of the lower third molar root and inferior alveolar canal. Int J Oral Maxillofac Surg. 2019;48(12):1577–83.CrossRef Qi W, Lei J, Liu YN, Li JN, Pan J, Yu GY. Evaluating the risk of post-extraction inferior alveolar nerve injury through the relative position of the lower third molar root and inferior alveolar canal. Int J Oral Maxillofac Surg. 2019;48(12):1577–83.CrossRef
6.
Zurück zum Zitat Pippi R, Spota A, Santoro M. Prevention of lingual nerve injury in third molar surgery: literature review. J Oral Maxillofac Surg. 2017;75(5):890–900.CrossRef Pippi R, Spota A, Santoro M. Prevention of lingual nerve injury in third molar surgery: literature review. J Oral Maxillofac Surg. 2017;75(5):890–900.CrossRef
7.
Zurück zum Zitat Palma-Carrió C, García-Mira B, Larrazabal-Morón C, Peñarrocha-Diago M. Radiographic signs associated with inferior alveolar nerve damage following lower third molar extraction. Med Oral Patol Oral Cir Bucal. 2010;15(6):e886–90.CrossRef Palma-Carrió C, García-Mira B, Larrazabal-Morón C, Peñarrocha-Diago M. Radiographic signs associated with inferior alveolar nerve damage following lower third molar extraction. Med Oral Patol Oral Cir Bucal. 2010;15(6):e886–90.CrossRef
8.
Zurück zum Zitat Park W, Choi JW, Kim JY, Kim BC, Kim HJ, Lee SH. Cortical integrity of the inferior alveolar canal as a predictor of paresthesia after third-molar extraction. J Am Dent Assoc. 2010;141(3):271–8.CrossRef Park W, Choi JW, Kim JY, Kim BC, Kim HJ, Lee SH. Cortical integrity of the inferior alveolar canal as a predictor of paresthesia after third-molar extraction. J Am Dent Assoc. 2010;141(3):271–8.CrossRef
9.
Zurück zum Zitat Blaeser BF, August MA, Donoff RB, Kaban LB, Dodson TB. Panoramic radiographic risk factors for inferior alveolar nerve injury after third molar extraction. J Oral Maxillofac Surg. 2003;61(4):417–21.CrossRef Blaeser BF, August MA, Donoff RB, Kaban LB, Dodson TB. Panoramic radiographic risk factors for inferior alveolar nerve injury after third molar extraction. J Oral Maxillofac Surg. 2003;61(4):417–21.CrossRef
10.
Zurück zum Zitat Coulthard P, Kushnerev E, Yates JM, Walsh T, Patel N, Bailey E, et al. Interventions for iatrogenic inferior alveolar and lingual nerve injury. Cochrane Database Syst Rev. 2014;4:CD005293. Coulthard P, Kushnerev E, Yates JM, Walsh T, Patel N, Bailey E, et al. Interventions for iatrogenic inferior alveolar and lingual nerve injury. Cochrane Database Syst Rev. 2014;4:CD005293.
11.
Zurück zum Zitat Lam NP, Donoff RB, Kaban LB, Dodson TB. Patient satisfaction after trigeminal nerve repair. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003;95(5):538–43.CrossRef Lam NP, Donoff RB, Kaban LB, Dodson TB. Patient satisfaction after trigeminal nerve repair. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003;95(5):538–43.CrossRef
12.
Zurück zum Zitat Dula K, Bornstein MM, Buser D, Dagassan-Berndt D, Ettlin DA, Filippi A, et al. SADMFR guidelines for the use of cone-beam computed tomography/ digital volume tomography. Swiss Dent J. 2014;124(11):1169–83. Dula K, Bornstein MM, Buser D, Dagassan-Berndt D, Ettlin DA, Filippi A, et al. SADMFR guidelines for the use of cone-beam computed tomography/ digital volume tomography. Swiss Dent J. 2014;124(11):1169–83.
13.
Zurück zum Zitat Nakamura T. Dental MRI: a road beyond CBCT. Eur Radiol. 2020;30(12):6389–91.CrossRef Nakamura T. Dental MRI: a road beyond CBCT. Eur Radiol. 2020;30(12):6389–91.CrossRef
14.
Zurück zum Zitat Ohman A, Kivijärvi K, Blombäck U, Flygare L. Pre-operative radiographic evaluation of lower third molars with computed tomography. Dentomaxillofac Radiol. 2006;35(1):30–5.CrossRef Ohman A, Kivijärvi K, Blombäck U, Flygare L. Pre-operative radiographic evaluation of lower third molars with computed tomography. Dentomaxillofac Radiol. 2006;35(1):30–5.CrossRef
15.
Zurück zum Zitat Tsapaki V. Radiation protection in dental radiology - Recent advances and future directions. Phys Med. 2017;44:222–6.CrossRef Tsapaki V. Radiation protection in dental radiology - Recent advances and future directions. Phys Med. 2017;44:222–6.CrossRef
16.
Zurück zum Zitat Sodickson A, Baeyens PF, Andriole KP, Prevedello LM, Nawfel RD, Hanson R, et al. Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults. Radiology. 2009;251(1):175–84.CrossRef Sodickson A, Baeyens PF, Andriole KP, Prevedello LM, Nawfel RD, Hanson R, et al. Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults. Radiology. 2009;251(1):175–84.CrossRef
17.
Zurück zum Zitat Jaju PP, Jaju SP. Cone-beam computed tomography: Time to move from ALARA to ALADA. Imaging Sci Dent. 2015;45(4):263–5.CrossRef Jaju PP, Jaju SP. Cone-beam computed tomography: Time to move from ALARA to ALADA. Imaging Sci Dent. 2015;45(4):263–5.CrossRef
18.
Zurück zum Zitat de Castro HS, Kehrwald R, Matheus RA, Gomes AF, Queiroz PM. Influence of low-dose protocols of CBCT on dental implant planning. Int J Oral Maxillofac Implants. 2021;36(2):307–12.CrossRef de Castro HS, Kehrwald R, Matheus RA, Gomes AF, Queiroz PM. Influence of low-dose protocols of CBCT on dental implant planning. Int J Oral Maxillofac Implants. 2021;36(2):307–12.CrossRef
19.
Zurück zum Zitat Döbelin Q, Stadlinger B, Wiedemeier DB, Bichsel D, Rücker M, Valdec S. Detectability of osseous lesions with a pre-programmed low-dose protocol for cone-beam computed tomography. Appl Sci. 2020;10:4961.CrossRef Döbelin Q, Stadlinger B, Wiedemeier DB, Bichsel D, Rücker M, Valdec S. Detectability of osseous lesions with a pre-programmed low-dose protocol for cone-beam computed tomography. Appl Sci. 2020;10:4961.CrossRef
20.
Zurück zum Zitat Al-Haj Husain A, Döbelin Q, Giacomelli-Hiestand B, Wiedemeier DB, Stadlinger B, Valdec S. Diagnostic accuracy of cystic lesions using a pre-programmed low-dose and standard-dose dental cone-beam computed tomography protocol: an ex vivo comparison study. Sensors (Basel). 2021;21:7402.CrossRef Al-Haj Husain A, Döbelin Q, Giacomelli-Hiestand B, Wiedemeier DB, Stadlinger B, Valdec S. Diagnostic accuracy of cystic lesions using a pre-programmed low-dose and standard-dose dental cone-beam computed tomography protocol: an ex vivo comparison study. Sensors (Basel). 2021;21:7402.CrossRef
21.
Zurück zum Zitat Wippold FJ. Head and neck imaging: the role of CT and MRI. J Magn Reson Imaging. 2007;25(3):453–65.CrossRef Wippold FJ. Head and neck imaging: the role of CT and MRI. J Magn Reson Imaging. 2007;25(3):453–65.CrossRef
22.
Zurück zum Zitat Gahleitner A, Solar P, Nasel C, Homolka P, Youssefzadeh S, Ertl L, et al. Magnetic resonance tomography in dental radiology (dental MRI). Radiologe. 1999;39(12):1044–50.CrossRef Gahleitner A, Solar P, Nasel C, Homolka P, Youssefzadeh S, Ertl L, et al. Magnetic resonance tomography in dental radiology (dental MRI). Radiologe. 1999;39(12):1044–50.CrossRef
23.
Zurück zum Zitat Reda R, Zanza A, Mazzoni A, Cicconetti A, Testarelli L, Di Nardo D. An update of the possible applications of magnetic resonance imaging (MRI) in dentistry: a literature review. J Imaging. 2021;7:75.CrossRef Reda R, Zanza A, Mazzoni A, Cicconetti A, Testarelli L, Di Nardo D. An update of the possible applications of magnetic resonance imaging (MRI) in dentistry: a literature review. J Imaging. 2021;7:75.CrossRef
24.
Zurück zum Zitat Olt S, Jakob PM. Contrast-enhanced dental MRI for visualization of the teeth and jaw. Magn Reson Med. 2004;52(1):174–6.CrossRef Olt S, Jakob PM. Contrast-enhanced dental MRI for visualization of the teeth and jaw. Magn Reson Med. 2004;52(1):174–6.CrossRef
25.
Zurück zum Zitat Lockhart PB, Kim S, Lund NL. Magnetic resonance imaging of human teeth. J Endod. 1992;18(5):237–44.CrossRef Lockhart PB, Kim S, Lund NL. Magnetic resonance imaging of human teeth. J Endod. 1992;18(5):237–44.CrossRef
26.
Zurück zum Zitat Valdec S, Al-Haj Husain A, Winklhofer S, Müller M, Piccirelli M, Stadlinger B. Comparison of preoperative cone-beam computed tomography and 3D-double echo steady-state MRI in third molar surgery. J Clin Med. 2021;10:4768.CrossRef Valdec S, Al-Haj Husain A, Winklhofer S, Müller M, Piccirelli M, Stadlinger B. Comparison of preoperative cone-beam computed tomography and 3D-double echo steady-state MRI in third molar surgery. J Clin Med. 2021;10:4768.CrossRef
27.
Zurück zum Zitat Al-Haj Husain A, Solomons M, Stadlinger B, Pejicic R, Winklhofer S, Piccirelli M, et al. Visualization of the inferior alveolar nerve and lingual nerve using MRI in oral and maxillofacial surgery: a systematic review. Diagnostics (Basel). 2021;11:1657.CrossRef Al-Haj Husain A, Solomons M, Stadlinger B, Pejicic R, Winklhofer S, Piccirelli M, et al. Visualization of the inferior alveolar nerve and lingual nerve using MRI in oral and maxillofacial surgery: a systematic review. Diagnostics (Basel). 2021;11:1657.CrossRef
28.
Zurück zum Zitat Al-Haj Husain A, Valdec S, Stadlinger B, Rücker M, Piccirelli M, Winklhofer S. Preoperative visualization of the lingual nerve by 3D double-echo steady-state MRI in surgical third molar extraction treatment. Clin Oral Investig. 2022;26:2043–53.CrossRef Al-Haj Husain A, Valdec S, Stadlinger B, Rücker M, Piccirelli M, Winklhofer S. Preoperative visualization of the lingual nerve by 3D double-echo steady-state MRI in surgical third molar extraction treatment. Clin Oral Investig. 2022;26:2043–53.CrossRef
29.
Zurück zum Zitat Fujii H, Fujita A, Yang A, Kanazawa H, Buch K, Sakai O, et al. Visualization of the peripheral branches of the mandibular division of the trigeminal nerve on 3D double-echo steady-state with water excitation sequence. AJNR Am J Neuroradiol. 2015;36(7):1333–7.CrossRef Fujii H, Fujita A, Yang A, Kanazawa H, Buch K, Sakai O, et al. Visualization of the peripheral branches of the mandibular division of the trigeminal nerve on 3D double-echo steady-state with water excitation sequence. AJNR Am J Neuroradiol. 2015;36(7):1333–7.CrossRef
30.
Zurück zum Zitat Burian E, Probst FA, Weidlich D, Cornelius CP, Maier L, Robl T, et al. MRI of the inferior alveolar nerve and lingual nerve-anatomical variation and morphometric benchmark values of nerve diameters in healthy subjects. Clin Oral Investig. 2020;24(8):2625–34.CrossRef Burian E, Probst FA, Weidlich D, Cornelius CP, Maier L, Robl T, et al. MRI of the inferior alveolar nerve and lingual nerve-anatomical variation and morphometric benchmark values of nerve diameters in healthy subjects. Clin Oral Investig. 2020;24(8):2625–34.CrossRef
31.
Zurück zum Zitat Al-Haj Husain A, Stadlinger B, Winklhofer S, Müller M, Piccirelli M, Valdec S. Mandibular third molar surgery: intraosseous localization of the inferior alveolar nerve using 3D double-echo steady-state MRI (3D-DESS). Diagnostics (Basel). 2021;11:1245.CrossRef Al-Haj Husain A, Stadlinger B, Winklhofer S, Müller M, Piccirelli M, Valdec S. Mandibular third molar surgery: intraosseous localization of the inferior alveolar nerve using 3D double-echo steady-state MRI (3D-DESS). Diagnostics (Basel). 2021;11:1245.CrossRef
32.
Zurück zum Zitat Gray CF, Redpath TW, Smith FW, Staff RT. Advanced imaging: Magnetic resonance imaging in implant dentistry. Clin Oral Implants Res. 2003;14(1):18–27.CrossRef Gray CF, Redpath TW, Smith FW, Staff RT. Advanced imaging: Magnetic resonance imaging in implant dentistry. Clin Oral Implants Res. 2003;14(1):18–27.CrossRef
33.
Zurück zum Zitat Prager M, Heiland S, Gareis D, Hilgenfeld T, Bendszus M, Gaudino C. Dental MRI using a dedicated RF-coil at 3 Tesla. J Craniomaxillofac Surg. 2015;43(10):2175–82.CrossRef Prager M, Heiland S, Gareis D, Hilgenfeld T, Bendszus M, Gaudino C. Dental MRI using a dedicated RF-coil at 3 Tesla. J Craniomaxillofac Surg. 2015;43(10):2175–82.CrossRef
34.
Zurück zum Zitat Edwards R, Altalibi M, Flores-Mir C. The frequency and nature of incidental findings in cone-beam computed tomographic scans of the head and neck region: a systematic review. J Am Dent Assoc. 2013;144(2):161–70.CrossRef Edwards R, Altalibi M, Flores-Mir C. The frequency and nature of incidental findings in cone-beam computed tomographic scans of the head and neck region: a systematic review. J Am Dent Assoc. 2013;144(2):161–70.CrossRef
35.
Zurück zum Zitat Idiyatullin D, Corum C, Moeller S, Prasad HS, Garwood M, Nixdorf DR. Dental magnetic resonance imaging: making the invisible visible. J Endod. 2011;37(6):745–52.CrossRef Idiyatullin D, Corum C, Moeller S, Prasad HS, Garwood M, Nixdorf DR. Dental magnetic resonance imaging: making the invisible visible. J Endod. 2011;37(6):745–52.CrossRef
36.
Zurück zum Zitat Demirturk Kocasarac H, Geha H, Gaalaas LR, Nixdorf DR. MRI for dental applications. Dent Clin North Am. 2018;62(3):467–80.CrossRef Demirturk Kocasarac H, Geha H, Gaalaas LR, Nixdorf DR. MRI for dental applications. Dent Clin North Am. 2018;62(3):467–80.CrossRef
37.
Zurück zum Zitat Mayerhoefer ME, Szomolanyi P, Jirak D, Materka A, Trattnig S. Effects of MRI acquisition parameter variations and protocol heterogeneity on the results of texture analysis and pattern discrimination: an application-oriented study. Med Phys. 2009;36(4):1236–43.CrossRef Mayerhoefer ME, Szomolanyi P, Jirak D, Materka A, Trattnig S. Effects of MRI acquisition parameter variations and protocol heterogeneity on the results of texture analysis and pattern discrimination: an application-oriented study. Med Phys. 2009;36(4):1236–43.CrossRef
38.
Zurück zum Zitat Di Nardo D, Gambarini G, Capuani S, Testarelli L. Nuclear magnetic resonance imaging in endodontics: a review. J Endod. 2018;44(4):536–42.CrossRef Di Nardo D, Gambarini G, Capuani S, Testarelli L. Nuclear magnetic resonance imaging in endodontics: a review. J Endod. 2018;44(4):536–42.CrossRef
39.
Zurück zum Zitat Petersen LB, Olsen KR, Matzen LH, Vaeth M, Wenzel A. Economic and health implications of routine CBCT examination before surgical removal of the mandibular third molar in the Danish population. Dentomaxillofac Radiol. 2015;44(6):20140406.CrossRef Petersen LB, Olsen KR, Matzen LH, Vaeth M, Wenzel A. Economic and health implications of routine CBCT examination before surgical removal of the mandibular third molar in the Danish population. Dentomaxillofac Radiol. 2015;44(6):20140406.CrossRef
40.
Zurück zum Zitat Memon A, Rogers I, Paudyal P, Sundin J. Dental X-Rays and the risk of thyroid cancer and meningioma: a systematic review and meta-analysis of current epidemiological evidence. Thyroid. 2019;29(11):1572–93.CrossRef Memon A, Rogers I, Paudyal P, Sundin J. Dental X-Rays and the risk of thyroid cancer and meningioma: a systematic review and meta-analysis of current epidemiological evidence. Thyroid. 2019;29(11):1572–93.CrossRef
41.
Zurück zum Zitat Carvalho RW, do Egito Vasconcelos BC. Assessment of factors associated with surgical difficulty during removal of impacted lower third molars. J Oral Maxillofac Surg. 2011;69(11):2714–21.CrossRef Carvalho RW, do Egito Vasconcelos BC. Assessment of factors associated with surgical difficulty during removal of impacted lower third molars. J Oral Maxillofac Surg. 2011;69(11):2714–21.CrossRef
42.
Zurück zum Zitat Miloro M, Halkias LE, Slone HW, Chakeres DW. Assessment of the lingual nerve in the third molar region using magnetic resonance imaging. J Oral Maxillofac Surg. 1997;55(2):134–7.CrossRef Miloro M, Halkias LE, Slone HW, Chakeres DW. Assessment of the lingual nerve in the third molar region using magnetic resonance imaging. J Oral Maxillofac Surg. 1997;55(2):134–7.CrossRef
43.
Zurück zum Zitat Chau A. Comparison between the use of magnetic resonance imaging and conebeam computed tomography for mandibular nerve identification. Clin Oral Implants Res. 2012;23(2):253–6.CrossRef Chau A. Comparison between the use of magnetic resonance imaging and conebeam computed tomography for mandibular nerve identification. Clin Oral Implants Res. 2012;23(2):253–6.CrossRef
44.
Zurück zum Zitat Beck F, Austermann S, Bertl K, Ulm C, Lettner S, Toelly A, et al. Is MRI a viable alternative to CT/CBCT to identify the course of the inferior alveolar nerve in relation to the roots of the third molars? Clin Oral Investig. 2021;25(6):3861–71.CrossRef Beck F, Austermann S, Bertl K, Ulm C, Lettner S, Toelly A, et al. Is MRI a viable alternative to CT/CBCT to identify the course of the inferior alveolar nerve in relation to the roots of the third molars? Clin Oral Investig. 2021;25(6):3861–71.CrossRef
45.
Zurück zum Zitat Kirnbauer B, Jakse N, Rugani P, Schwaiger M, Magyar M. Assessment of impacted and partially impacted lower third molars with panoramic radiography compared to MRI-a proof of principle study. Dentomaxillofac Radiol. 2018;47(4):20170371.CrossRef Kirnbauer B, Jakse N, Rugani P, Schwaiger M, Magyar M. Assessment of impacted and partially impacted lower third molars with panoramic radiography compared to MRI-a proof of principle study. Dentomaxillofac Radiol. 2018;47(4):20170371.CrossRef
46.
Zurück zum Zitat de Oliveira-Santos C, Souza PH, de Azambuja B-C, Stinkens L, Moyaert K, Rubira-Bullen IR, et al. Assessment of variations of the mandibular canal through cone beam computed tomography. Clin Oral Investig. 2012;16(2):387–93.CrossRef de Oliveira-Santos C, Souza PH, de Azambuja B-C, Stinkens L, Moyaert K, Rubira-Bullen IR, et al. Assessment of variations of the mandibular canal through cone beam computed tomography. Clin Oral Investig. 2012;16(2):387–93.CrossRef
47.
Zurück zum Zitat Agbaje JO, de Casteele EV, Salem AS, Anumendem D, Lambrichts I, Politis C. Tracking of the inferior alveolar nerve: its implication in surgical planning. Clin Oral Investig. 2017;21(7):2213–20.CrossRef Agbaje JO, de Casteele EV, Salem AS, Anumendem D, Lambrichts I, Politis C. Tracking of the inferior alveolar nerve: its implication in surgical planning. Clin Oral Investig. 2017;21(7):2213–20.CrossRef
48.
Zurück zum Zitat Techawiboonwong A, Song HK, Leonard MB, Wehrli FW. Cortical bone water: in vivo quantification with ultrashort echo-time MR imaging. Radiology. 2008;248(3):824–33.CrossRef Techawiboonwong A, Song HK, Leonard MB, Wehrli FW. Cortical bone water: in vivo quantification with ultrashort echo-time MR imaging. Radiology. 2008;248(3):824–33.CrossRef
49.
Zurück zum Zitat Probst M, Richter V, Weitz J, Kirschke JS, Ganter C, Troeltzsch M, et al. Magnetic resonance imaging of the inferior alveolar nerve with special regard to metal artifact reduction. J Craniomaxillofac Surg. 2017;45(4):558–69.CrossRef Probst M, Richter V, Weitz J, Kirschke JS, Ganter C, Troeltzsch M, et al. Magnetic resonance imaging of the inferior alveolar nerve with special regard to metal artifact reduction. J Craniomaxillofac Surg. 2017;45(4):558–69.CrossRef
50.
Zurück zum Zitat Probst M, Burian E, Robl T, Weidlich D, Karampinos D, Brunner T, et al. Magnetic Resonance Imaging as a Diagnostic Tool for Periodontal Disease: a prospective study with correlation to standard clinical findings - is there added value? J Clin Periodontol. 2021;48:929–48.CrossRef Probst M, Burian E, Robl T, Weidlich D, Karampinos D, Brunner T, et al. Magnetic Resonance Imaging as a Diagnostic Tool for Periodontal Disease: a prospective study with correlation to standard clinical findings - is there added value? J Clin Periodontol. 2021;48:929–48.CrossRef
51.
Zurück zum Zitat Sedlacik J, Kutzner D, Khokale A, Schulze D, Fiehler J, Celik T, et al. Optimized 14 + 1 receive coil array and position system for 3D high-resolution MRI of dental and maxillomandibular structures. Dentomaxillofac Radiol. 2016;45(1):20150177.CrossRef Sedlacik J, Kutzner D, Khokale A, Schulze D, Fiehler J, Celik T, et al. Optimized 14 + 1 receive coil array and position system for 3D high-resolution MRI of dental and maxillomandibular structures. Dentomaxillofac Radiol. 2016;45(1):20150177.CrossRef
52.
Zurück zum Zitat Ludwig U, Eisenbeiss AK, Scheifele C, Nelson K, Bock M, Hennig J, et al. Dental MRI using wireless intraoral coils. Sci Rep. 2016;6:23301.CrossRef Ludwig U, Eisenbeiss AK, Scheifele C, Nelson K, Bock M, Hennig J, et al. Dental MRI using wireless intraoral coils. Sci Rep. 2016;6:23301.CrossRef
53.
Zurück zum Zitat Probst FA, Schweiger J, Stumbaum MJ, Karampinos D, Burian E, Probst M. Magnetic resonance imaging based computer-guided dental implant surgery-A clinical pilot study. Clin Implant Dent Relat Res. 2020;22(5):612–21.CrossRef Probst FA, Schweiger J, Stumbaum MJ, Karampinos D, Burian E, Probst M. Magnetic resonance imaging based computer-guided dental implant surgery-A clinical pilot study. Clin Implant Dent Relat Res. 2020;22(5):612–21.CrossRef
54.
Zurück zum Zitat Hilgenfeld T, Prager M, Heil A, Schwindling FS, Nittka M, Grodzki D, et al. PETRA, MSVAT-SPACE and SEMAC sequences for metal artefact reduction in dental MR imaging. Eur Radiol. 2017;27(12):5104–12.CrossRef Hilgenfeld T, Prager M, Heil A, Schwindling FS, Nittka M, Grodzki D, et al. PETRA, MSVAT-SPACE and SEMAC sequences for metal artefact reduction in dental MR imaging. Eur Radiol. 2017;27(12):5104–12.CrossRef
55.
Zurück zum Zitat Ferretti F, Malventi M, Malasoma R. Dental magnetic resonance imaging: study of impacted mandibular third molars. Dentomaxillofac Radiol. 2009;38(6):387–92.CrossRef Ferretti F, Malventi M, Malasoma R. Dental magnetic resonance imaging: study of impacted mandibular third molars. Dentomaxillofac Radiol. 2009;38(6):387–92.CrossRef
56.
Zurück zum Zitat Kang Y, Hua C, Wu B, Zhou Z, Xu Y, Wu P, et al. Investigation of zero TE MR in preoperative planning in dentistry. Magn Reson Imaging. 2018;54:77–83.CrossRef Kang Y, Hua C, Wu B, Zhou Z, Xu Y, Wu P, et al. Investigation of zero TE MR in preoperative planning in dentistry. Magn Reson Imaging. 2018;54:77–83.CrossRef
Metadaten
Titel
Magnetic resonance imaging for preoperative diagnosis in third molar surgery: a systematic review
verfasst von
Adib Al-Haj Husain
Bernd Stadlinger
Sebastian Winklhofer
Marco Piccirelli
Silvio Valdec
Publikationsdatum
09.04.2022
Verlag
Springer Nature Singapore
Erschienen in
Oral Radiology / Ausgabe 1/2023
Print ISSN: 0911-6028
Elektronische ISSN: 1613-9674
DOI
https://doi.org/10.1007/s11282-022-00611-4

Weitere Artikel der Ausgabe 1/2023

Oral Radiology 1/2023 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update Zahnmedizin

Bestellen Sie unseren kostenlosen Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.