The aim of this study was to compare the dosimetric differences of the four setup-field plans and the suitable arc modes of VMAT plans for the Halcyon in bilateral breast radiotherapy and to analyse the plan quality and delivery efficiency by comparing the dosimetric differences between the Halcyon and the Trilogy to guide the clinical application in bilateral breast radiotherapy.
Materials and methods
Patient selection and volume delineation
From September 2006 to December 2018, CT image datasets of 10 patients diagnosed with bilateral breast cancer and who received bilateral breast radiotherapy at Shandong Cancer Hospital were selected. The clinical target volume (CTV) included all bilateral breast tissue, excluding local lymph node region. The planning target volume (PTV) was generated by expanding a 5-mm margin from the CTV and was shrunk to 5 mm below the skin on the skin side. The organs at risk (OARs) include the total lung, heart, left ventricle (LV), left anterior descending artery (LAD) and liver. The skin is defined as the 3-mm region below the body outside of the PTV. The normal structures were defined as the body minus the PTV (B-P).
Treatment planning
With the Halcyon version 1.0, all imaging setup fields are taken using digital megavoltage imaging panels. When designing a Halcyon plan, four different setup fields can be selected. For the 10 patients, we designed four VMAT plans with different setup fields on the Halcyon: high‐quality MV CBCT (the gantry rotates clockwise from 260° to 100°, delivering 10 MUs, simply called CBCT-H); low‐dose MV CBCT (delivering 5 MUs in a clockwise gantry rotation from 260° to 100°, simply called CBCT-L); high‐quality orthogonal MV radiograph pair (images acquired with 0° and 90° and delivering 2 MUs for each field, simply called MV-H) and low‐dose orthogonal MV radiograph pair (images acquired with 0° and 90° and delivering 1 MU for each field, simply called MV-L). For the four plans, two anticlockwise 160°–200°and two clockwise 200°–160° rotation arcs were used.
On the Trilogy, whole and partial arc plans, referred to as T-4arc and T-8arc, respectively, were generated. The whole-arc plan consisted of two anticlockwise 160°–200° and two clockwise 200°–160° rotation arcs. The partial-arc plan consisted of total 8 partial arcs. For unilateral breast, four 100° arcs like a bowknot were generated. In the two plans, the medial x-jaw was set to the minimum site (− 2 cm) to minimize the irradiated volume of the lungs and heart. A 6 MV X-ray was used, and the dose rate was set to 600 MU/min. Two whole and partial arc plans, referred to as H-4arc and H-8arc, were designed on the Halcyon with the same arc angle as T-4arc and T-8arc mentioned above. In the Halcyon plans, a 6 MV FFF X-ray was used at the maximum dose rate of 800 MU/min. Low-dose MV CBCT was selected for image guidance as Flores-Martinez et al. [
6] suggested.
The prescription dose was 50 Gy in 2-Gy fractions. All plans were designed with the Eclipse version 15.5 treatment planning system (Varian Medical Systems, Palo Alto, CA, USA) using an analytic anisotropic algorithm (AAA). For all VMAT plans, the PTV was extended to 5 mm outside the skin, named PTVop, a 10-mm bolus was used on the skin outside the PTV, and the dose of PTVop was optimized. In the final dose calculation, the bolus was deleted. The dose normalization was 95% volume of the PTV received 100% prescription dose. All plans used the same optimization parameter settings, with the goal of minimizing the doses to the lungs, heart and LAD while ensuring PTV dose coverage. No dose constraints were applied to the skin and LV during the optimization.
Dosimetric evaluation
The dose statistics of the plans were based on dose-volume histogram (DVH) analysis. For PTV, the dose of 2% and 98% volume (D
2, D
98), the volume receiving 107% and 110% of the prescribed dose (V
107 and V
110) and the mean dose (D
mean) were analysed. The conformity index (CI) and the homogeneity index (HI) of the PTV were calculated according to the following formula:
$$CI=\frac{{{TV}_{PV}}^{2}}{TV\times PV}$$
TV
PV represents the volume of the PTV wrapped by the prescription dose, the TV represents the volume of the PTV, and the PV represents the total volume wrapped by the prescription dose. Larger CI values indicate the better conformity of the target [
13].
$$HI=\frac{{D}_{2}-{D}_{98}}{{D}_{p}}\times 100{\%}$$
D
p represents the prescribed dose. Lower HI values indicate the better uniformity of the target [
14].
For OARs, the VX and mean doses were analysed. VX represents the irradiated volume of X Gy dose.
The number of MUs was analysed for all plans. The delivery times of T-4arc, T-8arc, H-4arc and H-8arc were recorded. The delivery time was recorded from the first field beam on to the last field beam off, excluding the positioning time.
Statistical analysis
All data were statistically analysed using the Statistical Package for Social Sciences v20.0 software (SPSS Inc., Chicago, IL, USA). First, a one-way analysis of variance (ANOVA) using Bonferroni’s multiple comparisons test was applied to compare the four different setup-field plans on the Halcyon. Second, whole and partial arc plans on the same LINAC were compared to determine the most suitable field mode for the machine. The Mann–Whitney U test was used. The better Halcyon plan was selected for further comparisons. Third, a statistical comparison of the better Halcyon plan and the Trilogy plan was implemented to analyse the dosimetric differences between the two machines using the Mann–Whitney U test. The differences were considered statistically significant when p < 0.05.
Discussion
In radiotherapy for breast cancer, high position accuracy is essential to prevent under-dose in the target and excessive irradiation to OARs. Image guidance must be taken before each patient is treated on the Halcyon. Compared to two-dimensional (2D) position verification, three-dimensional (3D) position verification like CBCT can more accurately measure 3D vector changes and observe the body position rotation [
15,
16]. For breast tissue, CBCT is very useful because it can provide 3D soft tissue and bony anatomy information and can be compared with the planning CT to assess the accuracy of the setting [
16,
17]. Rossi et al. [
18] reported that CBCT matching is recommended when breast/chest wall patients were treated with the VMAT technique. However, CBCT may be associated with more additional dose and time-consuming than orthogonal planar images. On Halcyon, MV CBCT imaging process take only 15 s. It's beneficial for patients. In C-arm accelerators such as the Trilogy, kilovoltage (KV) CBCT can be used to correct the patient’s position if 3D position verification is necessary. In Halcyon 1.0, MV CBCT was used for image-guided radiotherapy (IGRT). Compared to KV imaging, MV imaging has some advantages, such as identical isocenter as the treatment beam and no metal artifacts. Compared with KV imaging, the main disadvantages of MV imaging are higher dose and lower image quality. About Halcyon’s MV CBCT, Malajovich et al. [
19] reported that the highest tissue dose of MV CBCT ranges from 2 to 7 cGy per fraction in different treatment sites, which is equivalent to the fractional dose of KV CBCT during breast and pelvic IGRT application, and MV CBCT images of Halcyon is able to identify different soft tissues and lack of metal-induced artifacts. On Halcyon, two different dose image mode could be selected to apply the CBCT. Compared to the low-dose mode, the high-quality imaging mode does not provide material advantages [
19]. We found that high-quality CBCT plans increased the OARs doses compared to low-dose CBCT plans. Therefore, we inferred that low‐dose MV CBCT was the optimal setup-field mode. Flores‐Martinez et al. [
6] compared four different setup-field plans for unilateral breast cancer, and in their opinion, low-dose MV CBCT was the most suitable technique for patients treated on the Halcyon. Their results agreed with ours. On Halcyon, the dose of the setup field is incorporated in the calculation of the planned dose, and the irradiated doses to the target and the OARs are also reflected in the total plan dose, which is more intuitive.
Partial-arc plans on the Trilogy showed more dosimetric advantage, especially in low-dose volumes of the heart, left ventricle, and lungs and the mean doses of the heart, lungs, and liver. This is because while designing a partial-arc plan, it is possible to artificially choose the arc degree that irradiates less OAR volume. Rotating the collimator angle and fixing the jaw can further reduce the influence of the leakage between the MLC on the dose of the OARs, which can minimize the dose of the OARs. Boman et al. [
11] compared the dosimetric differences between the whole and partial arc VMAT plans of unilateral breast cancer, including regional lymph node irradiation, and the results showed that partial-arc plans significantly reduced the dose of the ipsilateral lung and the V
5 of the heart but increased the V
5 of the contralateral breast. The result is similar to ours, but the cases in our study are bilateral breast cancer, which does not involve the dose of the contralateral breast. Comparing the two plans of the Halcyon, the results were contrary to the Trilogy’s, and the whole-arc plans showed better dosimetry. For the PTV, in addition to the mean dose, the whole-arc plan was better than partial-arc plans in terms of the maximum dose, minimum dose, conformity and uniformity. For OARs, partial-arc plans increased the doses to the heart, LV, LAD, and lungs. The results showed that partial-arc plans have no advantage for the Halcyon, which may be related to the jawless setting and the fewer arc degrees. According to the results described above, when designing a treatment plan for bilateral breast cancer, we can choose a more suitable arc setting according to the corresponding LINAC.
Based on the results described above, we mainly compared the dosimetric differences between the Halcyon's whole-arc plan and the Trilogy's two plans. All plans met clinical requirements. For the PTV, apart from the mean dose, there were no significant differences in other dosimetric parameters between the two whole-arc plans on the Halcyon and the Trilogy. Compared with the T-8arc plan, H-4arc showed worse Dmean, D2 and V107, and better D98 and CI of the PTV. In short, the plans of the two machines were comparable in terms of target dose.
Darby et al. [
20] found a linear relationship between the mean dose of the heart and the incidence of ischaemic heart disease, and the incidence increased by 7.4% for each 1-Gy increase in the mean dose. Therefore, the mean dose of the heart is often used as a reference for cardiac toxicity. However, the dose to the cardiac substructure also needs to be considered in radiotherapy. Some studies believe that the LAD and LV are important parts of the heart in the context of radiation-induced heart disease [
21‐
23]. In this study, for low-dose irradiated volumes of the heart and LV, partial-arc plans on the Trilogy showed the lowest values, while for high-dose irradiated volumes, the whole-arc plans on the Halcyon showed the lowest values. For the mean dose of the heart, the partial-arc plans on the Trilogy showed the lowest value, and the whole-arc plans on the Halcyon showed the highest values. This may be related to the additional radiation dose to the heart from each MV CBCT scan. For all dosimetric parameters of the LAD, there is no significant difference between whole-arc plans on the Halcyon and the two plans on the Trilogy.
For lungs, the partial-arc plans on the Trilogy reduced the low-dose irradiated volume (V
5, V
10) and the mean dose but increased the high-dose irradiated volume (V
40). There was no significant difference in the comparison of the V
20 and V
30 in the lungs between all plans of the two LINACs. For all dosimetric parameters of the lungs, there was no significant difference between the two whole-arc plans on the two LINACs. Fiorentino et al. [
24] retrospectively analysed the VMAT plans of 16 patients with bilateral breast cancer. For lungs, the average values of the D
mean, V
5 and V
20 were 11.8 ± 2.3 Gy, 78.9 ± 15.3% and 15.7 ± 5%, respectively. No acute and late complications above grade 2 were observed during the 24 months of follow-up. In our study, the mean dose, V
5 and V
20 of the lungs in all plans were lower than in their study.
The plans on the Halcyon increased the skin’s dose compared to the two plans on the Trilogy. O’Grady et al. [
25] found that 6 X FFF fields on the Halcyon increased the superficial dose compared to FF fields for breast cancer radiotherapy, as demonstrated by in vivo measurements, phantom measurements, and planning comparisons. Their results were the same as ours. Because the rays are softened after the flattening filter is removed, the 6 MV X-rays in FFF mode are equivalent to lower-energy rays, resulting in a shallower depth of the dose build-up area, thereby increasing the superficial dose. Studies on the Monte Carlo (MC) have shown that the influence of contamination electrons in the FFF mode is greater, which further leads to an increase in surface dose [
26,
27]. Barsky et al. [
5] retrospectively analysed 34 breast cancer cases treated on the Halcyon, and the results showed that breast cancer cases were well tolerated on Halcyon; additionally, the acute toxicity was comparable to the published reports using conventional LINACs. The difference from our study is that they used tangential fields instead of the VMAT in our study. On the Halcyon, the effect of VMAT on the skin dose needs more prospective clinical studies to be proved.
In contrast to Halcyon, KV CBCT could be used for image-guided on Trilogy. Although the dose of KV CBCT is less than that of MV CBCT, it is not negligible. The difference with Halcyon is that the imaging dose of Trilogy is not calculated into the entire plan. So, in fact, the dose difference between the two machines may vary slightly. The whole-arc plans on Halcyon reduced the number of MUs compared to whole-arc plans on Trilogy and showed similar MUs compared to partial-arc plans on Trilogy. On Halcyon, the imaging dose is integrated into plans and thus it is contributing to the PTV and OAR dose. However, in Trilogy the imaging dose is not accounted for plans, what might explain the lower MUs required on Halcyon.
Previous studies have demonstrated that the Halcyon could reduce the treatment time significantly compared with conventional LINACs [
3‐
7]. Our study proved the same results in radiotherapy for bilateral breast cancer. This may be related to the Halcyon's faster gantry and MLC speed and higher dose rate. The shortening of the treatment time can reduce intra-fraction movement, improve patient comfort, and increase machine throughput.
An issue that needs to be addressed is the correction of rotation errors. On C-arm accelerators like Trilogy, we can use the six degree of freedom (6-DoF) couch to correct the rotation errors. The combination of 6-DoF couch and CBCT can correct the translational and rotational setup errors and improve the positioning accuracy obviously [
28,
29]. However, on O-ring accelerators like Halcyon, the installed 3-DoF couch allows only for correction of translational shifts. On the Halcyon, if large rotation errors are found, time-consuming repositioning may be required to improve the positioning accuracy. This drawback might significantly compromise the advantage of faster treatment plan delivery with Halcyon compared to Trilogy. Therefore, in order to reduce the positioning time and improve the positioning accuracy, it is necessary to improve the immobilization equipment or adopt online 3D IGRT equipment, such as surface imaging. 3D surface imaging system is a quick and non-invasive method to assist the patient in setting up, which could improve the accuracy and speed of patient positioning in breast cancer radiotherapy [
30,
31]. On Halcyon, 3D surface imaging is especially suitable, which could correct rotation errors in patient positioning and monitor patient movement during beam delivery [
32]. The combination of 3D surface imaging and daily CBCT may greatly improve treatment accuracy and reduce positioning time, which needs more research to prove in the future.
Many studies [
33‐
35] have shown that deep inspiration breath hold (DIBH) can significantly reduce the radiation dose to the heart and lungs for breast cancer radiotherapy. However, the technique requires patients to hold their breath for a certain amount of time. Minimizing the treatment time is beneficial to patients with DIBH. The combination of Halcyon and VMAT can further reduce the treatment time and expand the application range of DIBH. The robustness of VMAT plans under respiratory motion or other influences which might change form or position of PTV needs to be considered. Field-in-field and fixed-field intensity-modulated radiation therapy can reduce the influence of respiratory movement and skin deformation or swelling during treatment by opening the MLC outside the skin or adding skin flash. When designing VMAT plans, we could increase the robustness of the plan by adding a virtual bolus on the skin [
11,
36,
37].
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.